K963382 · Quickie Designs, Inc. · ITI · Apr 28, 1997 · Physical Medicine
Device Facts
Record ID
K963382
Device Name
QUICKIE P100 SERIES
Applicant
Quickie Designs, Inc.
Product Code
ITI · Physical Medicine
Decision Date
Apr 28, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Quickie Powered Wheelchairs empower physically challenged persons by providing a means of mobility.
Device Story
Motorized wheelchair; features include push handles, adjustable armrests, backrest, seat frame, cushion, footrests, casters, joystick controller, motors, brakes, batteries, and drive wheels. Device modification involves replacing Dynamics controller with Penny and Giles (P&G) controller. Controller utilizes joystick input to manage motor speed, acceleration, deceleration, and braking. Used by physically challenged individuals for mobility; operated via joystick. Output is physical movement of wheelchair. Benefits include restored or enhanced mobility for users.
Clinical Evidence
Bench testing only. Comparative testing of speed, acceleration, and braking distance performed against predicate device with Dynamics controller. Electromagnetic compatibility testing (20 V/m EMI) performed. Software validation, including requirements, design, verification, validation, and hazard analysis, completed.
Indicated for physically challenged persons requiring powered mobility assistance.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Predicate Devices
P100
P110
P190
P500
P200
P210
P320
P300
Everest and Jennings Lancer 2000
Everest and Jennings Sabre
Everest and Jennings Sabre ES
Prode Health Care Inc. Jazzy
Reference Devices
Hoveround Personal Mobility Vehicles
Related Devices
K963381 — QUICKIE P300 SERIES · Quickie Designs, Inc. · Apr 28, 1997
K232121 — Power wheelchair (Model:BC-EA8000) · Ningbo Baichen Medical Devices Co., Ltd. · Sep 15, 2023
K963385 — QUICKIE POWER WHEELCHAIRS P200 SERIES · Quickie Designs, Inc. · Apr 28, 1997
K253980 — Electrically Powered Wheelchair (Model YC-01) · Yurob Rehabilitation Medical Co., Ltd. · Mar 27, 2026
K231508 — Power Wheelchair, W5521 · Zhejiang Innuovo Rehabilitation Devices Co.,Ltd · Nov 9, 2023
Submission Summary (Full Text)
{0}
CONFIDENTIAL
APR 28 1997
K963382
# Appendix F
Summary of Safety and Effectiveness for Quickie Powered Wheelchair with Penny and Giles Controller
August 22, 1996
Quickie Designs Inc. Powered Wheelchair
F
{1}
Submitter Quickie Designs Inc.
2842 Business Park Ave.
Fresno, CA 93727
209-292-2171 Fax-209-292-2741
Dave Counts
Date summary was prepared 8/21/1996
Name(s) of the device
P100 series
Identification of predicate device(s)
P100, P110, P190, P500, P500, P200, P210, P320, P300
Everest and Jennings Lancer 2000, Sabre, Sabre ES
Prode Health Care Inc. Jazzy
Description of the device
Quickie powered wheelchairs consist of typical features found on any wheelchair, such as push handles, adjustable armrests, backrest, seat frame and cushion, footrests, and casters. Because these are motorized wheelchairs, they also consist of joy stick controller, motors, brakes, batteries and drive wheels.
Intended Use
Quickie Powered Wheelchairs empower physically challenged persons by providing a means of mobility.
Comparison of device characteristics to predicate
This 510(k) is for a change in the wheel chair controller. The power wheelchair specifications have not been changed due to this modification. The New controller subject to this 510(k) is the P&G Controller that has been available on other lines of power wheelchairs, including Pride Health Care’s Jazzy, Everest and Jennings Lancer 2000 and Sabre, and Hoveround Personal Mobility Vehicles. Everest and Jennings recently made the same switch in controllers as Quickie; replacing the Dynamics Controller with a P&G 8 controller.
The P&G controller is based on Hitachi’s H8/532 microprocessor while Dynamics is based on Motorola’s 68HC705C8. The power wheelchair specifications have not been changed due to this modification.
Non clinical testing
{2}
Testing of all programmable parameters has been performed. These parameters include the following: maximum and minimum forward and reverse speeds, maximum and minimum turn speeds, maximum and minimum acceleration and deceleration, and maximum and minimum turn acceleration and deceleration. Comparative testing regarding speed, acceleration, brake distance demonstrate that the wheelchair performs as the predicate device with the Dynamics controller. Results demonstrate the wheelchairs perform according to specification. Electromagnetic compatibility testing was also performed on Quickie Power Wheelchairs, and results demonstrate that the wheelchairs pass the 20 V/m EMI test. Software validation information includes the software requirements, design, development and verification and validation of the controller as well as a hazards analysis and mitigation associated with the safety of the controller.
**510(k) Number** K963382
Not Known at the writing of this summary.
## Conclusion
The Quickie Power Wheelchair is Substantially Equivalent to the previous versions of the Quickie Wheelchairs which utilize the Dynamics controller as well as other power wheelchair models which already incorporate the P&G 8 controller.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches