MECHANICAL FOLDING WHEEL CHAIR

K973491 · Heartway Medical Products Co., Ltd. · IOR · Dec 4, 1997 · Physical Medicine

Device Facts

Record IDK973491
Device NameMECHANICAL FOLDING WHEEL CHAIR
ApplicantHeartway Medical Products Co., Ltd.
Product CodeIOR · Physical Medicine
Decision DateDec 4, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3850
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The design and construction of this device is such that it presents a mobility unit allowing users to be transported by an attendent or to be self propelled. This device is intended to provide mobility to invalids who are otherwise incapable of ambulation with patient ambulatory assist devices. The models to be offered are of different sizes allowing for user preferance because of physical needs. Options for users include removeable armrests & detachable footrests to accomodate user requirements.

Device Story

Mechanical folding wheelchair; provides mobility for individuals unable to ambulate. Features include dual rear axle positioning for seat height adjustment, reinforced frame, and various arm/footrest configurations. Operated by user or attendant; used in clinical or home settings. Provides mobility assistance; improves patient independence and accessibility.

Clinical Evidence

No clinical data. Bench testing only.

Technological Characteristics

Mechanical folding wheelchair; materials include aluminum footplates, molded wheels, and leatherette upholstery. Features dual axle bushings for seat height adjustment (17-3/4" to 19-3/4"), reinforced top inner rails, and snap-on solid tires. Manual operation; no electronic or software components.

Indications for Use

Indicated for invalids incapable of ambulation with patient ambulatory assist devices, requiring a mobility unit for transport by an attendant or self-propulsion.

Regulatory Classification

Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a logo for "HEARTWAY". The logo consists of the word "HEARTWAY" in bold, sans-serif font, with a stylized graphic above it. The graphic is composed of several horizontal lines that are arranged in a semi-circular or arched shape, resembling a series of stacked, curved bars. The overall design is simple and bold, with a focus on the text and the geometric shape above it. ## HEARTWAY MEDICAL PRODUCTS CO NO. 4, ROAD 5, TAICHUNG INDUSTRIAL PARK, HUNG. TAIWAN FI. O. C 80357 (Sales Den DEC - 4 1997 Internet: http://www.webtree.net.tw/heartway/ E-mail: heartway@webtree.net.tw K973491 Office of Device Evaluation 510(k) Document Mail Center (HFZ-401) Food & Drug Administration 1390 Piccard Drive Rockville, Maryland 20850 Dear Sir or Madam, In accordance with the requirements of section 510(k) of the Federal Food and Drug Cosmetic Act we hereby request to register for commercial distribution for the following device. | | Device Name ------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name ------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | | | Heartway Models--------------------------------------------------------------------------------------------------------------------------------------------------------------- | | Proprietary Name ------------------------------------------------------------------------------------------------------------------------------------------------------------- | | | Established Registration No. ------------------------------------------903564 | | | FDA Classification ----------------------------------------------------------------------------------------------------------------------------------------------------------- | | ### Summary The design and construction of this device is such that it presents a mobility unit allowing users to be transported by an attendent or to be self propelled. This device is intended to provide mobility to invalids who are otherwise incapable of ambulation with patient ambulatory assist devices. The wheelchairs listed above are substantially equivalent to many products being Marketed at this time. Wheelchairs of this type compare with the Invacare Tracer Plus Series (See Exhibit II) and the Everest & Jennings Traveler Series (see Exhibit III) {1}------------------------------------------------ # New Traveler Image /page/1/Picture/1 description: The image shows a close-up of a wheelchair. The wheelchair is black and silver. The wheels are large and have a silver rim. The seat is not visible in the image. TRAVELER QUALITY AND DURABILITY. . . Ever evolving to meet the needs of our customers, the "New" Traveler enhances its reputation as the rental wheelchair of choice. Now available with dual rear axle positioning for seat heights from 17-3/4" to 19-3/4", seat widths from 16" to 22" wide and a weight capacity of up to 300 pounds. The Traveler has earned its reputation for enduring the rigors of all types of use. . . designed to survive the journey. - · Dual axle positioning ... seat heights between 17-3/4" and 19-3/4". - · Front slide post design provides added rigidity and easy folding. - · Reinforced top inner rail for extra strength. - · Strong upholstery inner liner tested at 350 lbs. per inch, lasts twice as long as comparable models. - · Low-maintenance, reinforced molded wheels with snap-on solid tires. - · Desk length or full length arms available for easy transfer and access. - · Superior E&J cam-action footrest design combined with quality manufacturing provide a snug fit and a true swing-away motion. - · 300 lb. capacity on 20" and 22" wide models. - · 2 year limited warranty on frame and crossbraces. Image /page/1/Picture/12 description: The image shows the logo for Everest Jennings. The text "Everest Jennings" is in a bold, serif font. Below the company name is the tagline "Value through Innovation" in a smaller, italicized font. The logo is simple and professional. {2}------------------------------------------------ # TM Image /page/2/Picture/1 description: The image is a black and white photograph of a metal structure. The structure appears to be a framework of some kind, with vertical and diagonal beams. The beams are connected by smaller pieces of metal. The image is dark and grainy, and the structure is not clearly defined. DUAL AXLE BUSHINGS Allows for easy seat conversion of seat heights from 17-3/4" to 19-3/4" through use of different wheels and caster combinations. ### FEATURES - ·Padded, double-embossed burgundy leatherette upholstery and padded armrests. - •Durable 24" x 1" black molded wheel with snapon rubber tire. - •8″ x 1″ black molded front casters. - · Rugged cam-action swing-away detachable footrests or elevating legrests. - •Durable aluminum footplate. - ·Choice of 3 frame styles: std. fixed arm, Std/Hemi/Extra wide detachable arm and Std/Hemi recliners. - · Available with a variety of optional arm styles. All specifications subject to change without notice. ### "New" Traveler™ Dimensions and Weights (inches and pounds) | "New" Traveler<br>Models | SEAT | | ARMS | | BACK | OVERALL | | | EXTENSIONS | | | | WT. | | |------------------------------------------|------|----|---------|---------|---------|---------|---------|---------|------------|---------|----------|----|------|----| | | | | | | | | | | FOOTRESTS | | LEGRESTS | | LBS. | | | Fixed Arm, Narrow Adult | 16 | 16 | $141/4$ | $101/8$ | $161/2$ | 30 | 36 | $221/2$ | $11$ | $161/2$ | 22 | 16 | 23 | 35 | | Fixed Arm, Adult | 18 | 16 | $161/4$ | $101/8$ | $161/2$ | 30 | 36 | $241/2$ | $11$ | $161/2$ | 22 | 16 | 23 | 36 | | Detachable Arm,<br>Narrow Adult, Hemi | 15 | 16 | $173/4$ | $93/4$ | $161/2$ | 30 | 36 | $241/4$ | $11$ | $161/2$ | 20 | 16 | 23 | 40 | | Detachable Arm, Adult, Hemi | 18 | 16 | $173/4$ | $93/4$ | $161/2$ | 30 | 36 | $261/4$ | $11$ | $161/2$ | 20 | 16 | 23 | 41 | | Detachable Arm, Narrow Adult | 16 | 16 | $193/4$ | $93/4$ | $161/2$ | 30 | 36 | $241/4$ | $11$ | $161/2$ | 22 | 16 | 23 | 40 | | Detachable Arm, Adult | 18 | 16 | $193/4$ | $93/4$ | $161/2$ | 30 | 36 | $261/4$ | $11$ | $161/2$ | 22 | 16 | 23 | 41 | | Detachable Arm<br>Narrow Adult Recliner | 16 | 17 | $193/4$ | $93/4$ | $233/8$ | 36 | $531/8$ | $241/4$ | 13 | $161/2$ | 22 | 15 | 21 | 55 | | Detachable Arm, Adult Recliner | 18 | 17 | $193/4$ | $93/4$ | $233/8$ | 36 | $531/8$ | $261/4$ | 13 | $161/2$ | 22 | 15 | 21 | 55 | | Detachable Arm<br>Adult-Wide (20") | 20 | 16 | $193/4$ | $93/4$ | $161/2$ | 30 | 36 | $281/4$ | $113/4$ | $161/2$ | 22 | 16 | 23 | 47 | | Detachable Arm<br>Adult-Extra Wide (22") | 22 | 16 | $193/4$ | $93/4$ | $161/2$ | 30 | 36 | $301/4$ | $113/4$ | $161/2$ | 22 | 16 | 23 | 48 | "Without front rigging. All dimensions are + or - 1/4" NOTE: Everest & Jennings maintains the policy of continual product improvement, therefore we reserve the right to make changes without notice. **Everest & Jennings**® Value through Innovation United States 4203 Earth City Exprswy St. Louis, MO 63045 203 Earth City Exprswy St. Louis, MO 63045 For additional information call 1-800-235-4661 Canada LAK 218 For additional information call 1-905-669-2381 111 Snidecroft Road Concord, Ontario, Canada Mexico Calle 3 No. 631 Zona Industrial Codigo Postal 44940 Guadalajara, Jalisco Mexico For additional information call (011) 52-36-12-12-34 PN# 02003023 7/96 Printed in the USA {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and strength. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC - 4 1997 Mr. Kenny Ho ·President Heartway Medical Products Company, Limited Number 4, Road 5, Taichung Industrial Park Taichung, Taiwan R.O.C. K973491 Re: Mechanical Folding Wheelchair Requlatory Class: I Product Code: IOR Dated: October 27, 1997 -October 31, 1997 Received: Dear Mr. Ho: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. Please note: concerning your device in the Federal Register. this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ #### Page 2 - Mr. Kenny Ho This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Colin M. Whitten, Ph.D. ia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page of 510(k) Number (if known): K973491 STANDARD FOLDING WHEELCHAIR Device Name: Indications For Use: The design and construction of this device is such that it presents a mobility unit allowing users to be transported by an attendent or to be self propelled. This device is intended to provide mobility to invalids who are otherwise incapable of ambulation with patient ambulatory assist devices. The models to be offered are of different sizes allowing for user preferance because of physical needs. Options for users include removeable armrests & detachable footrests to accomodate user requirements. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) to see Division of General Restorative Dev 510(k) Number Prescription Use ._ (Per 21 CFR 801 109) OR Over-the-Counter Use. \ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...