K120526 · Jiangsu Yuyue Medical Equipment & Supply Co., Ltd. · IOR · Mar 8, 2012 · Physical Medicine
Device Facts
Record ID
K120526
Device Name
YUYUE K2 WHEELCHAIR, YUYUE K4 WHEELCHAIR
Applicant
Jiangsu Yuyue Medical Equipment & Supply Co., Ltd.
Product Code
IOR · Physical Medicine
Decision Date
Mar 8, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Attributes
3rd-Party Reviewed
Indications for Use
The YUYUE K2/K4 wheelchair is indicated for providing mobility to persons limited to a sitting position.
Device Story
YUYUE K2/K4 is a mechanical, manual wheelchair providing mobility for individuals limited to a sitting position. Device features a rigid steel frame, nylon upholstery, two 24-inch rear wheels, and two 8-inch front casters. Operated by the user or an assistant; intended for indoor and outdoor use on smooth surfaces (flooring, concrete, asphalt, pocked dirt) with inclines under 9 degrees. Provides mobility support; enhances patient independence. No electronic or software components.
Clinical Evidence
Bench testing only. Device performance verified against ISO 7176 standards (stability, braking, dimensions, strength, and ignition resistance) and ISO 10993-5 for in vitro cytotoxicity. No clinical data provided.
Technological Characteristics
Mechanical wheelchair; rigid steel frame; nylon upholstery; 24" rear wheels; 8" front casters. Materials meet ISO 7176-16 (ignition resistance) and ISO 10993-5 (cytotoxicity). Complies with ISO 7176-1 (stability), 7176-3 (brakes), 7176-5 (dimensions), 7176-7 (seating), 7176-8 (strength), and 7176-15 (labeling). Flame retardance tested per California Technical Bulletin 117.
Indications for Use
Indicated for persons limited to a sitting position requiring mobility assistance. Suitable for indoor and outdoor use on smooth surfaces free of large obstacles and inclines not exceeding 9 degrees.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
K080114 — HUAYI H035 BASIC WHEELCHAIR · Danyang Huayi Medical Supply & Equipment Co, Ltd. · Feb 15, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
MAR - 8 2012
K120526
PAGE 1.0F3
## SECTION10.0 SUMMARY OF SAFETY AND EFFECTIVENESS
.
l
| 510(K) SUMMARY | |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 10.1 | ADMINISTRATION INFORMATION |
| 10.11 | SPONSOR IDENTIFICATION<br>JiangSu YuYue Medical Equipment & Supply Co., Ltd.<br>Zhenxin Road, Yunyang Economical Development Zone,<br>Danyang City, Jiangsu Province, CHINA, 212300.<br>Contact Person: Mr. James Lin<br>Tel: (86511)86900991<br>Fax:(86511)86900991<br>email: james.lin@yuyue.com.cn |
| 10.12 | ESTABLISHMENT REGISTRATION NUMBER: Pending |
| 10.13 | OFFICIAL CONTACT PERSON<br>Mr. Jiuguang Song<br>Managing Director<br>Strategic Advantage Consulting.Inc. Jiuguang Song<br>Managing Director<br>STRATEGIC ADVANTAGE CONSULTING.INC<br>3119 Brick Lane, Decatur. GA30033<br>Tel: 404-294-0743<br>Fax:831-300-3572<br>songjiuguang@yahoo.com |
| 10.14 | DATE OF PREPARATION OF THIS SUMMARY: August 10, 2011 |
| 10.15 | PROPRIETARY (TRADE) NAME: YUYUE K2 wheelchair, YUYUE K4 wheelchair |
| 10.16 | COMMON NAME: Wheelchair |
| 10.17 | CLASSIFICATION NAME: Wheelchair, Mechanical |
| 10.18 | REGULATION NUMBER: 21 CFR 890.3850 |
| 10.19 | PROPOSED REGULATORY CLASS: Class I |
| 10.20 | MEDICAL SPECIFICATIONS: Physical medicine |
| 10.21 | DEVICE PRODUCT CODES: 89 IOR |
| 10.3 | DESCRIPTION OF DEVICE:<br>The YUYUE K2/K4 wheelchair is a wheelchair that provides mobility to |
| | persons limited to a sitting position. It consists of rigid, mechanical, steel<br>frame and nylon upholstery back and seat that meet ISO 7176-16:<br>Resistance to ignition of Upholstered parts. It has two 24" rear wheels and<br>two 8" front casters for turning and maneuverability. The YUYUE K2/K4<br>wheelchair is intended for the use in indoors and outdoors, over smooth<br>surface (all standard indoor flooring surfaces, concrete, asphalt and<br>pocked dirt) that free of large obstacles and inclines greater than 9<br>degrees. |
| 10.4 | INDICATION FOR USE<br>The YUYUE K2/K4 wheelchair is indicated for providing mobility to<br>persons limited to a sitting position. |
| 10.5 | COMPARISON OF TECHNICAL CHARACTERISTICS<br>YUYUE K2/K4 wheelchair and Danyang Huayi K2 wheelchair (K080207)<br>have the same technological characteristics. |
| | TECHNICAL CHARACTERISTICS SUMMARY<br>YUYUE K2/K4 wheelchair production meets the following standards:<br>• ISO 7176-1 Wheelchair: Determination of static stability<br>• ISO 7176-3 Wheelchair: Determination of efficiency of brakes.<br>• ISO 7176-5: Determination of overall dimension, mass and turning<br>space.<br>• ISO 7176-7: Measurement of seating and wheel dimensions.<br>• ISO 7176-8 Wheelchair: Requirements and test methods for static,<br>impact and fatigue strengths.<br>• ISO 7176-15 Wheelchair: Requirements for information disclosure,<br>documentation and labeling. |
| 10.6 | • ISO 7176-16 Wheelchair: Resistance to ignition of upholstered parts –<br>requirements and test methods.<br>• ISO 10993-5-2009: Biological Evaluation of Medical Devices – Part5:<br>Testes for in vitro Cytotoxicity<br>• ASTM D751:2006 Section 11-Procedure A- Tensile Strength<br>• Section A, Part I and Section D, Part II of California Technical Bulletin<br>117 Requirement: Test Procedure and Apparatus for Testing the Flame<br>Retardance of Resilient Filling Materials Used in Upholstered Furniture.<br>• PVC leather/Nylon fabric/liner in accordance with: Section E, Part I of<br>California Technical Bulletin 117 Requirement: Test Procedure and<br>Apparatus for Testing the Flame Retardance of Resilient Filling<br>Materials Used in Upholstered Furniture. |
508 ·
{1}------------------------------------------------
252
ไ
2 OF 3
509
.
:
.
-
{2}------------------------------------------------
SUBSTANTIAL EQUIVALENCE: In the domestic market sales, YUYUE K2/K4 wheelchair has good quality and stability. Market information feedback to the product quality and reliable. YUYUE K2/K4 wheelchair and Danyang Huayi K2 wheelchair 10.8 (K080207) have the same structure and effective consistent performance, and customers responded well. YUYUE K2/K4 wheelchair and Danyang Huayi K2 wheelchair (K080207) are substantially equivalent products in all areas impacting safety and effectiveness. CONCLUSION: YUYUE K2/K4 wheelchair confirm fully to the standards which are mentioned in Section as well as applicable 21 CFR references, and meet pinhole FDA requirements, biocompatibility requirements and 10.9 labeling claims required by these standards. There are no safety/efficiency issues or claims that differ from the predicate devices cited.
4120526 PAGE 3 OF 3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a human figure in profile, with multiple overlapping lines creating a sense of depth and movement.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JiangSu YuYue Medical Equipment & Supply Co., Ltd. % Underwriters Laboratories Inc. Mr. Ned Devine 333 Pfingsten Road Northbrook, Illinois 60062
MAR 26 2012
Re: K120526
Trade/Device Name: YUYUE K2/K4 wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical wheelchair Regulatory Class: Class I Product Code: IOR Dated: February 14, 2012 Received: February 22, 2012
Dear Mr. Devine:
This letter corrects our substantially equivalent letter of March 8, 2012.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Ned Devine
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark M. Millbern
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# SECTION 4.0 INDICATIONS FOR USE
## Indications for Use
510(k) Number (if known):
Device Name: YUYUE K2/K4 wheelchair
#### Indications For Use:
The YUYUE K2/K4 wheelchair is indicated for providing mobility to persons limited to a sitting position.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Eyaluation (ODE)
Signature
(Division Sign-Off) (Division Sign-cical, Orthopedic, and Restorative Devices
510(k) Number K120526
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.