TACAMO MANUAL WHEELCHAIR

K101719 · Dalton Instrument Corp. · IOR · Aug 25, 2010 · Physical Medicine

Device Facts

Record IDK101719
Device NameTACAMO MANUAL WHEELCHAIR
ApplicantDalton Instrument Corp.
Product CodeIOR · Physical Medicine
Decision DateAug 25, 2010
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 890.3850
Device ClassClass 1

Indications for Use

The intended use of the TACAMO Manual Wheelchair is to provide mobility to persons limited to a sitting position.

Device Story

TACAMO is a rigid, non-folding, manually operated mechanical wheelchair. Input: user or attendant physical force for propulsion and steering. Operation: frame constructed of 1.25" and 1" OD steel tubing; 24" rear wheels with hand rims; 8" front casters. Features a manually operated 'Tilt in Space' mechanism for patient stability, comfort, and positioning; activated by rear release levers that disengage gear rack segments from notched actuator rods via flexible cables. Provides 30 degrees of tilt and 30 degrees of recline. Used in clinics, nursing homes, or home settings. Output: physical mobility and postural adjustment. Benefits: provides independent or assisted mobility and facilitates patient care tasks (e.g., feeding) through tilt/recline functionality.

Clinical Evidence

Bench testing only. Device performance verified against ANSI/RESNA WC/Vol.1-1998 (Requirements and Test Methods for Wheelchairs) and DIS ISO 7176 (1992) Part 8 (Static, Impact, and Fatigue Strengths).

Technological Characteristics

Rigid, non-folding manual wheelchair. Materials: 1.25" and 1" OD steel tubing. Mechanical actuation: manual 'Tilt in Space' via release levers, flexible cables, gear rack segments, and notched actuator rods. Dimensions: 24" rear wheels, 8" front casters. Standards: ANSI/RESNA WC/Vol.1-1998, DIS ISO 7176 Part 8.

Indications for Use

Indicated for persons limited to a sitting position requiring mobility assistance. Suitable for use by individuals or as an attendant-propelled transport device in clinical or home environments.

Regulatory Classification

Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ May 13, 2010 # - TH DMC Image /page/0/Picture/2 description: The image shows the logo for Dalt Instruments. The logo features the letters 'DI' in a stylized font, with an oval shape surrounding the 'I'. Above the logo, the text 'K101719' is written in a handwritten style. The overall image appears to be a scan or photocopy, with some imperfections and variations in contrast. JUN. 1.8 2010 ### Received # 510(k) SUMMARY DALTON INSTRUMENT CORPORATION'S Model TACAMO Manual Wheelchair AUG 2 5 2010 # Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared. Dalton Instrument Corp. 3788 Arapaho Road Addison, TX 75001 Tel: 469-522-1200 Fax: 469-522-1202 Contact Person: President, Mei Lein Date Prepared: May 13, 2010 # Name of Device and Name/Address of Sponsor Model TACAMO Manual Wheelchair Dalton Instrument Corp. 3788 Arapaho Road Addison, TX 75001 Tel: 469-522-1200 Fax: 469-522-1202 Common or Usual Name: Manual Wheelchair Classification Name: Wheelchair, Mechanical 89IOR #### Predicate Devices The Dalton Instrument Corp. Model TACAMO substantially equivalent to the Invacare Manual Wheelchair Model Action AT II (K984447, 1/8/1999) and Sunrise Medical/Quickie Design Model "Quickie TSR" Manual Wheelchair (K952641, 6/9/1995) Intended Use: the intended use of the Model TACAMO Manual Wheelchair is to provide mobility to persons limited to a sitting position. {1}------------------------------------------------ ## Technological Characteristics and Substantial Equivalence #### A. Device Description The DALTON INSTRUMENT CORPORATION Model TACAMO is a manually operated, self propelled, manual, mechanical wheelchair. The primary design function and use is to provide mobility to persons who may be restricted to a sitting position. This product may also be used as an attendant propelled patient transport device in a health care environment such as a hospital, nursing home or long term care facilities. The product has standard manual wheelchair features. It consists of the frame, 24" rear wheels with hand rims for propelling the chair, and 8'' front casters for steering and turning. The wheelchair frame is constructed from both 1.25" and 1" outside diameter(OD) round, mechanical, steel tubing. The side frames are of welded construction and are secured to the remaining frame members using screws and bolts. This device is a rigid, "non folding" type wheel chair that incorporates a solid seating surface. TACAMO also comes with wheelchair cushion and seating system. TACAMO also includes a "Tilt in Space" feature which allows the seat and back of the wheelchair to be tilted. This feature is used for those patients who require a tilt feature for stability, comfort or head control. It also serves as an attendant aid in those situations where a patient need to be tilted to be fed, or attended to in some situations. The "Tilt in Space" feature is manually operated and is activated by two release levers located at the rear of the wheelchair. Adjusting the tilt is achieved by squeezing the two release lever handles inward simultaneously. When the levers are squeezed, flexible cables attached to each release lever disengage gear rack segments from independent actuator rods which are notched to match the teeth of the gear segments. Once the desired tilt angle has been obtained, the handles are released and the chair will remain at the angle chosen. The chair has a tilt angle range of 30 degrees and 30 degrees of reclining angle. ### B. Substantial Equivalence The DALTON INSTRUMENT CORPORATION Model TACAMO substantially equivalent to numerous manual, mechanical wheelchairs currently in market place, {2}------------------------------------------------ for which FDA has granted marketing clearance through the 510(k) premarket notification process. More specifically the DALTON INSTRUMENT CORPORATION Model TACAMO is substantially equivalent to the Sunrise Medical/Quickie Design Model "Quickie TSR" Manual Wheelchair(K952641 6/9/95) and the Invacare Model AT II Manual Wheelchair(K984447 1/8/1999) Each of the products is a manual operated, attendant or user propelled, manual mechanical wheelchair with the same intended function and use which is to provide mobility to persons that may be limited to a seated position. All products consist basically of a mechanical frame to support the wheelchair, 24" rear wheels for propelling the wheelchair, and 8" casters for turning and steering. Additionally, each of these wheelchairs incorporates a manually operated "tilt in space" feature for patient comfort and positioning. #### Performance Data The DALTON INSTRUMENT CORPORATION wheelchair TACAMO meets the applicable performance requirements specified in: - . Rehabilitation Engineering Society of North America (RESNA) Standard ANSI/RESNA WC/Vol.1-1998 "Requirements and Test Methods for Wheelchairs(including Scooters)" - DIS ISO 7176 (1992) Part 8: Requirements and Test Methods for Static, ● Impact and Fatigue Strengths {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three stripes forming its body and wings. The eagle is facing right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Dalton Instrument Corporation % Mei Lein President 3788 Arapaho Road Addison, Texas 75001 AUG 2 5 2010 Re: K101719 Trade/Device Name: TACAMO Manual Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical wheelchair Regulatory Class: Class I Product Code: IOR Dated: August 9, 2010 Received: August 16, 2010 Dear Mei Lein: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ #### Page 2 - Mei Lein device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Marki McMullerson Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use K101719 510(K) Number (if known): K101719 Device Name: TACAMO Manual Wheelchair Indications for Use: The intended use of the TACAMO Manual Wheelchair is to provide mobility to persons limited to a sitting position. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _X (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) . Page 1 of : Sim-Off sion of Surgical. Orthopedic, i Kestorative Devices 510(k) Number K101719
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...