ATTO Mobility Scooter (ATTO);ATTO Mobility Scooter (ATTO SPORT);ATTO Mobility Scooter (ATTO SPORT MAX)

K222703 · Moving Life, Ltd. · INI · May 19, 2023 · Physical Medicine

Device Facts

Record IDK222703
Device NameATTO Mobility Scooter (ATTO);ATTO Mobility Scooter (ATTO SPORT);ATTO Mobility Scooter (ATTO SPORT MAX)
ApplicantMoving Life, Ltd.
Product CodeINI · Physical Medicine
Decision DateMay 19, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3800
Device ClassClass 2

Indications for Use

The Moving Life ATTO Mobility Scooter is an indoor/outdoor scooter that provides transportation for a disabled or elderly person.

Device Story

The ATTO Mobility Scooter is a battery-powered, three-wheeled, collapsible electric scooter designed for elderly or disabled individuals. It is operated by the user via a thumb throttle and tiller. The device features a front-wheel brushless DC hub motor, spring-applied electronic-release brakes, and a regenerative braking system. Models ATTO SPORT and ATTO SPORT MAX include additional manual front-wheel disc brakes and software-based speed control. The device is used in indoor and outdoor settings. The user controls speed and direction; the microprocessor-based controller monitors system performance. The device provides mobility, allowing users to navigate environments independently. It is collapsible for transport and storage.

Clinical Evidence

No clinical data. Bench testing only. Compliance with ISO 7176 series (sections 2-6, 9-11, 14, 16, 21, 25) and EN 12184:2014 standards confirmed performance and safety.

Technological Characteristics

Three-wheeled electric scooter; 48V DC brushless hub motor; Li-ion battery (48V, 5.2Ah); microprocessor-based controller; spring-applied electronic-release brakes; regenerative braking; manual front disc brake (SPORT models); airless NPT tires; collapsible frame. Designed per ISO 7176 and EN 12184:2014.

Indications for Use

Indicated for disabled or elderly persons requiring indoor/outdoor transportation assistance.

Regulatory Classification

Identification

A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable. May 19, 2023 Moving Life Ltd Igor Naroditsky RA Adviser Havat Alenby Netzer Sereni, 7039500 Israel Re: K222703 Trade/Device Name: ATTO Mobility Scooter (ATTO):ATTO Mobility Scooter (ATTO SPORT);ATTO Mobility Scooter (ATTO SPORT MAX) Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: April 17, 2023 Received: April 17, 2023 Dear Igor Naroditsky: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Tushar Bansal -S Tushar Bansal, PhD Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K222703 Device Name ATTO Mobility Scooter (ATTO) ATTO Mobility Scooter (ATTO SPORT) ATTO Mobility Scooter (ATTO SPORT MAX) Indications for Use (Describe) ATTO Mobility Scooter is an indoor/outdoor scooter that provides transportation for a disabled or elderly person. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) SUMMARY (K222703) ## ATTO Mobility Scooter #### A. Contact Details Applicant Name: Moving Life Ltd Havat Allenby, Netzer Sereni, 7039500 Israel Contact Name: Igor Naroditsky, Regulatory Adviser Phone: + 972 54 4384386 Date Prepared: Aug 15, 2022 #### B. Device Name: Trade Name: ATTO Mobility Scooter Models: ATTO, ATTO SPORT, ATTO SPORT MAX Common Name: Vehicle, Motorized 3-Wheeled Classification Name: Motorized three-wheeled vehicle; INI; 890.3800 #### C. Legally Marketed Predicate Device(s) | 510(k) Number | Product Code | Trade Name | Manufacturer | |---------------|--------------|-----------------------|-----------------| | K160909 | INI | ATTO Mobility Scooter | Moving Life Ltd | #### D. Device Description ATTO Mobility Scooter is an electrical battery powered, three-wheeled scooter intended to provide mobility for elderly or disabled individuals in a variety of indoor and outdoor settings. The ATTO is a collapsible electric scooter which has a front-wheel brushless DC hub motor. The scooter can be collapsed in 12 seconds and can also be disassembled with ease. The scooter controlled through a thumb throttle and protected by an electronic-release brake system, has a driving range up to 20 km between charges. It is capable of carrying a driver weighing up to 136 (for ATTO MAX) kg. It moves both directions and handles a 6° slope. It can come to a full stop within 1.1m when on a horizontal plane and within 2.2 m when on a 6° slope. The differences between the ATTO Mobility Scooter and its predicates are: - । Introduction of the two additional models, ATTO SPORT, and ATTO SPORT MAX with - । Additional Manual Front Wheal brake for extra safety {4}------------------------------------------------ - Software Speed control feature (in ATTO SPORT, and ATTO SPORT MAX only) । - -Enhanced LCD display - -Extra wide rear wheel - -Rear light handle - Powerful LED Headlights - - -Shock-absorbing airless NPT tires - -Regenerative braking control system - Modified controller - - -Modified charger and battery #### ய் Intended Use/indications for use The Moving Life ATTO Mobility Scooter is an indoor/outdoor scooter that provides transportation for a disabled or elderly person. #### F. Substantial Equivalence Comparison The proposed ATTO Mobility Scooter is as safe and effective as the ATTO Mobility Scooter (K160909) by Moving Life. The Scooter has the same intended uses and same indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Moving Life ATTO Mobility Scooter and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the Moving Life ATTO Mobility Scooter is as safe and effective as the ATTO Mobility Scooter (K160909) by Moving Life. Thus, the proposed ATTO Mobility Scooter is substantially equivalent. | Model | Predicate ATTO Mobility<br>Scooter (K160909) | Proposed ATTO Mobility Scooter<br>(K222703) | |--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------| | Indication for Use | The Moving Life ATTO Mobility<br>Scooter is an indoor/outdoor<br>scooter that provides<br>transportation for a disabled or<br>elderly person. | The Moving Life ATTO Mobility<br>Scooter is an indoor/outdoor<br>scooter that provides<br>transportation for a disabled or<br>elderly person. | | Front wheel size | 203 X 27 mm | 8 in | | Rear wheel size | 228 X 57 mm | 9 in | | Ground clearance | 100 mm | 100 mm | | Number of wheels | 3 | 3 | | Length | 1200 mm | 1200 mm | | Width | 560 mm | 560 mm | | Max load | 100 kg | 120 kg for ATTO and ATTO Sport.<br>136 kg for ATTO SPORT MAX | | Required width of<br>angled corridor | 1000 mm | 1000 mm | #### G. Substantial Equivalence Chart {5}------------------------------------------------ | Model | Predicate ATTO Mobility<br>Scooter (K160909) | Proposed ATTO Mobility Scooter<br>(K222703) | |----------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Turn radius | 1350 mm | 1350 mm | | Battery | Li-Ion battery 48V 5.2 ah | Li-ion Battery 5.2 Ah; DC 48 V;<br>249.6 Wh | | Charger | AC input.100-240 volt,<br>50/60Hz, DC output 54.6 volt.<br>2 Amp for lithium-ion battery | AC input. 100-240 volt,<br>50/60Hz, DC output 54.6 volt. 2<br>Amp for lithium-ion battery | | Type of controller | Microprocessor based,<br>proportional and integral<br>controller which continually<br>monitors the ATTO systems. | Microprocessor based,<br>proportional and integral controller<br>which continually monitors the<br>ATTO systems. | | Maximum speed<br>(forward) | 6.4 km/h | 6.4 km/h for ATTO and ATTO<br>Max. 9.6 km/h for Max SPORT | | Maximum speed<br>(reverse) | 4 km/h | 4 km/h | | Travel distance | 15 km | 20/17 km | | Electrical System: | 48-volt DC | 48-volt DC | | Drive system | Front wheel, direct drive. | Front wheel, direct drive. | | Motor | 48-volt DC, Brushless electric<br>motor | 48-volt DC, Brushless electric motor | | Brake | Spring Applied Electrical<br>Released Brakes with<br>freewheel mode | Spring Applied Electrical<br>Released Brakes with freewheel<br>mode with regenerative braking<br>control system. Additional front<br>disk brake for extra safety | | Tiller | Adjustable locking for driving<br>comfort | Adjustable locking for driving<br>comfort | | Braking Time | Not publicly available | .5s | | Braking Distance | Not publicly available | 1.1m | | Speed Control | Velocity and current control | Velocity and current control | | Seat | Folding seat | Folding seat | #### H. Non-clinical Testing Moving Life ATTO Mobility Scooter is designed in accordance with the ISO7176 series including all relevant sections i.e., Section 2, Section 3, Section 4, Section 5, Section 6, Section 9, Section 10, Section 11, Section 14, Section 14, Section 16, Section 21 and Section 25, etc. ATTO performance and safety testing is conducted according to EN 12184:2014. All Performance testing met the predetermined acceptance values. The Moving Life ATTO Mobility Scooter functioned as intended and its functionality observed was as expected. #### l. Clinical Testing No clinical testing is included in this submission. #### J. Technological Characteristics ATTO Mobility Scooter is a same as the commercially available ATTO Mobility Scooter (K160909) by Moving Life and based on the same fundamental technology as the predicate devices. Both are {6}------------------------------------------------ electric scooters that are battery operated and have automatic braking systems. Batteries and battery chargers are provided with each scooter. Performance parameters of the devices are very similar. There are some minor technology differences between the proposed ATTO Mobility Scooter and the commercially available ATTO Mobility Scooter (K160909) by Moving Life. None of these differences raises new issues of safety and effectiveness. #### K. Conclusion The safety and effectiveness of the ATTO Mobility Scooter were demonstrated by the testing in compliance with national and international standards. The intended use, basic technology, and the features of the ATTO Mobility Scooter are similar to the predicate device. No new issues of safety and effectiveness is raised by the differences between the proposed ATTO Mobility Scooter and its predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...