← Product Code [ITI](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI) · K991658

# PERMOBIL POWERED WHEELCHAIR 1280 (K991658)

_Permobil AB · ITI · Oct 8, 1999 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI/K991658

## Device Facts

- **Applicant:** Permobil AB
- **Product Code:** [ITI](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI.md)
- **Decision Date:** Oct 8, 1999
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The intended use of Permobil powered wheelchair 1280 is to provide indoor and outdoor mobility to persons restricted to a sitting position.

## Device Story

Front-wheel-driven, battery-powered wheelchair; provides indoor/outdoor mobility for users restricted to sitting position. Input: user manipulation of joystick integrated into Penny & Giles Pilot+ controller. Operation: two 12V 73Ah batteries power drive unit (motor/gear/brakes) mounted on welded steel base frame. Output: mechanical movement of wheelchair. Used in various environments; operated by patient. Healthcare provider uses device to restore mobility; output affects patient independence and quality of life.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Welded steel base frame; front-wheel drive; two 12V 73Ah batteries; Penny & Giles Pilot+ controller; joystick interface; pressure die-cast aluminum controller enclosure.

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Permobil Chairman L/CS 8 ([K960951](/device/K960951.md))
- G-424 power wheelchair ([K983677](/device/K983677.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "permobil" in a bold, sans-serif font, with the "o" in "mobil" stylized as a circle. The word is enclosed in a thin, black oval. The text and oval are black against a white background. 

- ブ

#### OCT 8 1999

1252

## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

# K991658

## PERMOBIL POWERED WHEELCHAIR 1280

| Submitter                                                                                                                                                                                                           | Permobil AB<br>Box 120<br>S-861 23 Timrå<br>Sweden                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone<br>Fax                                                                                                                                                                                                        | +46 60 59 59 00<br>+46 60 57 52 50                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Contact person<br>E-mail address                                                                                                                                                                                    | Bengt Persson<br>bengt.persson@permobil.se                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Date Prepared                                                                                                                                                                                                       | May 10 1999                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Device name                                                                                                                                                                                                         | Powered wheelchair 1280                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification name                                                                                                                                                                                                 | Powered wheelchair                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Identification of predicate devices                                                                                                                                                                                 | Permobil Chairman L/CS 8 (K960951)<br>G-424 power wheelchair (K983677)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Intended use                                                                                                                                                                                                        | The intended use of Permobil powered wheelchair 1280 is to provide indoor and<br>outdoor mobility to persons restricted to a sitting position.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Description of the device                                                                                                                                                                                           | The Permobil powered wheelchair 1280 is front wheel driven, battery powered,<br>motor driven and is controlled by the Penny & Giles compact powered wheelchair<br>controller Pilot +. The joystick is integrated in the controller.<br>The wheelchair is powered by two 12V 73 Ah batteries.<br>Theoretical driving range on the fully charged batteries is 35 km.<br>The wheelchair consists of two basic sub-sections. These are the base with motor<br>controller and the body supporting system. In this submission, only the design<br>changes made to base is included. The base is of welded steel construction and<br>includes the base frame, front drive wheels with drive unit (motor/gear/brakes),<br>batteries and pivoting rear casters. The motor controller is mounted to the left or<br>right armrest, depending on user needs.<br>The motor controller, including the joystick, is housed in a pressure die cast<br>aluminum enclosure. |
| Substantial equivalence<br>The Permobil powered wheelchair 1280 is substantial equivalent with respect to,<br>intended use, energy source and materials, because those are identical to the<br>predicate device(s). |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

The key changes are frame suspension and use of the Pilot + controller from Penny & Giles instead of the PG8 controller from Penny & Giles.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "permobil" in a bold, sans-serif font, enclosed in an oval shape. The word is written in lowercase letters, with the "o" slightly smaller and raised compared to the other letters. The oval shape is a simple black line that surrounds the word, providing a border for the logo.

## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

## PERMOBIL POWERED WHEELCHAIR 1280

Jos
1699/1658

### Safety and effectiveness

The Permobil powered wheelchair 1280 has in substantials the same technological characteristics and the same safety and effectiveness as the predicate device(s) and the minor design changes declared in this submission do not raise new questions of safety and effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a bird-like figure, possibly representing an eagle or other national symbol. The bird is depicted with flowing lines, giving it a sense of movement and freedom.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 8 1999

Mr. Bengt Persson Quality Manager Permobil A.B. Engelbrekts Vag 101 S-191 62 Sollentuna, Sweden

Re: K991658

> Trade Name: Permobil Powered Wheelchair 1280 Regulatory Class: II Product Code: ITI Dated: July 7, 1999 Received: July 12, 1999

Dear Mr. Persson:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

#### Page 2 - Mr. Bengt Persson

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely vours.

Mark N Milhusan

Celia M. Witten, Ph.D., M.D. Director

Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "permobil" in a bold, sans-serif font, with the "o" in "mobil" being smaller and slightly raised. The word is enclosed in a black oval. The background is white, providing a clear contrast to the black text and oval.

## PERMOBIL POWERED WHEELCHAIR 1280

### Indication for Use

The intended use is to provide indoor and outdoor mobility to persons restricted to a sitting position.

#### 510(k) number

Not assigned at the writing of this submission

#### Device name

Not decided at writing of this submission Working name, also use in this submission, 1280

#### Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription use (per 21 CFR 801.109) □

or

Over the counter use of

Mark M. Mellikerson

---

(Division Sign Off)

(Division Sign-Off) Division of Ggeneral Restorative Devices

510(k) number

(k) number K99/65Y

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI/K991658](https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI/K991658)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
