ELECTRO-LITE ELITE POWERED WHEELCHAIR

K970575 · Damaco · ITI · Aug 5, 1997 · Physical Medicine

Device Facts

Record IDK970575
Device NameELECTRO-LITE ELITE POWERED WHEELCHAIR
ApplicantDamaco
Product CodeITI · Physical Medicine
Decision DateAug 5, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

For the non-ambulatory or mobility impaired individual.

Device Story

Electro-Lite Powered Wheelchair is a mobility aid for non-ambulatory or mobility-impaired individuals. Device provides powered locomotion to assist users with daily mobility. Operated by the user via integrated controls; intended for use in various environments where manual wheelchair use is insufficient. Benefits include increased independence and mobility for patients with physical impairments.

Clinical Evidence

No clinical data provided; substantial equivalence determination based on regulatory review.

Technological Characteristics

Powered wheelchair; class II device (Product Code: ITI). Technical specifications and materials not detailed in provided documentation.

Indications for Use

Indicated for non-ambulatory or mobility impaired individuals requiring powered wheelchair assistance.

Regulatory Classification

Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Related Devices

Submission Summary (Full Text)

{0} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Ron Higginbotham Quality Assurance Damaco 5105 Maureen Lane Moorpark, California 93021 AUG - 5 1997 Re: K970575 Electro-LiteElite Powered Wheelchair Regulatory Class: II Product Code: ITI Dated: June 13, 1997 Received: June 16, 1997 Dear Mr. Higginbotham: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {1} Page 2 - Mr. Ron Higginbotham This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Enclosure {2} 510(k) Number: K970575 Device Name: Electro-Lite™ Powered Wheelchair Indications For Use: For the non-ambulatory or mobility impaired individual. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![img-1.jpeg](img-1.jpeg) Prescription Use ☐ Per 21 CFR 801.109) OR Over-The Counter Use ☑
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...