Indicated for physically challenged persons requiring powered mobility assistance.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
K220747 — Power Wheelchair (N5515B) · Zhejiang Innuovo Rehabilitation Devices Co.,Ltd · May 12, 2022
K220827 — Electric Wheelchair · Foshan Dahao Medical Technology Co., Ltd. · Jul 5, 2022
K230964 — Power Wheelchair (N5909) · Zhejiang Innuovo Rehabilitation Devices Co.,Ltd · Dec 13, 2023
K963382 — QUICKIE P100 SERIES · Quickie Designs, Inc. · Apr 28, 1997
Submission Summary (Full Text)
{0}
CONFIDENTIAL
K963381
APR 28 1997
# Appendix F
Summary of Safety and Effectiveness for Quickie Powered Wheelchair with Penny and Giles Controller
August 22, 1996
Quickie Designs Inc. Powered Wheelchair
F
{1}
Submitter Quickie Designs Inc.
2842 Business Park Ave.
Fresno, CA 93727
209-292-2171 Fax-209-292-2741
Dave Counts
Date summary was prepared 8/21/1996
Name(s) of the device
P300
Identification of predicate device(s)
P100, P110, P190, P500, P500, P200, P210, P320, P300
Earnest and Jennings Lancer 2000, Sabre, Sabre ES
Prode Health Care Inc. Jazzy
Description of the device
Quickie powered wheelchairs consist of typical features found on any wheelchair, such as push handles, adjustable armrests, backrest, seat frame and cushion, footrests, and casters. Because these are motorized wheelchairs, they also consist of joy stick controller, motors, brakes, batteries and drive wheels.
Intended Use
Quickie Powered Wheelchairs empower physically challenged persons by providing a means of mobility.
Comparison of device characteristics to predicate
This 510(k) is for a change in the wheel chair controller. The power wheelchair specifications have not been changed due to this modification. The New controller subject to this 510(k) is the P&G Controller that has been available on other lines of power wheelchairs, including Pride Health Care’s Jazzy, Everst and Jennings Lancer 2000 and Sabre, and Hoveround Personal Mobility Vehicles.
August 22, 1996
Quickie Designs Inc. Powered Wheelchair
{2}
Everest and Jennings recently made the same switch in controllers as Quickie; replacing the Dynamics Controller with a P&G 8 controller.
The P&G controller is based on Hitachi’s H8/532 microprocessor while Dynamics is based on Motorola’s 68HC705C8. The power wheelchair specifications have not been changed due to this modification.
## Non clinical testing
Testing of all programmable parameters has been performed. These parameters include the following: maximum and minimum forward and reverse speeds, maximum and minimum turn speeds, maximum and minimum acceleration and deceleration, and maximum and minimum turn acceleration and deceleration. Comparative testing regarding speed, acceleration, brake distance demonstrate that the wheelchair performs as the predicate device with the Dynamics controller. Results demonstrate the wheelchairs perform according to specification. Electromagnetic compatibility testing was also performed on Quickie Power Wheelchairs, and results demonstrate that the wheelchairs pass the 20 V/m EMI test. Software validation information includes the software requirements, design, development and verification and validation of the controller as well as a hazards analysis and mitigation associated with the safety of the controller.
**510(k) Number** K963381
Not Known at the writing of this summary.
## Conclusion
The Quickie Power Wheelchair is Substantially Equivalent to the previous versions of the Quickie Wheelchairs which utilize the Dynamics controller as well as other power wheelchair models which already incorporate the P&G 8 controller.
August 22, 1996
Quickie Designs Inc. Powered Wheelchair
F 2
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.