The power wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
Power wheelchair (models D03, D05, D06, D09, D11, D16, D22, D23, D26, D36) is a foldable, battery-operated, 2-wheel drive mobility device. Inputs include user commands via joystick or Bluetooth Low Energy (BLE) remote controller. The control system processes inputs to drive brushless motors and actuate electromagnetic brakes for speed/direction control and stopping. The device is intended for indoor/outdoor use by disabled or elderly individuals. The joystick has priority over remote control for safety. The device provides mobility, allowing users to navigate environments while seated. Healthcare providers or caregivers do not operate the device during standard use; it is patient-operated. Benefits include improved mobility and independence for users with limited physical capacity.
Clinical Evidence
No clinical or animal studies were conducted. Substantial equivalence is supported by bench testing, including static/dynamic stability, brake performance, energy consumption, obstacle climbing, electromagnetic compatibility (IEC 60601-1-2), wireless coexistence (ANSI C63.27), and biocompatibility (ISO 10993).
Technological Characteristics
Foldable aluminum alloy frame; 2-wheel drive with brushless motors; 24VDC Li-ion battery; electromagnetic braking system; joystick and BLE remote control interface (2.402-2.480 GHz). Dimensions vary by model. Complies with ISO 7176 series for powered wheelchairs, IEC 62133-2 for batteries, and IEC 60601-1-2 for EMC. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for disabled or elderly persons limited to a seated position requiring indoor and outdoor mobility assistance.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Predicate Devices
Portable Folding Electric Wheelchair (Model: DC01) (K212092)
Reference Devices
Electric Wheelchair (Model: BBR-LY-01-01) (K223393)
Related Devices
K242252 — Power Wheelchair · Anhui Jbh Medical Apparatus Co., Ltd. · Dec 27, 2024
K242468 — Power Wheelchair (D10, D12, D15, D17, D20, D37) · Anhui Jbh Medical Apparatus Co., Ltd. · Dec 31, 2024
K251159 — Power wheelchair · Zhenjiang Assure Medical Equipment Co., Ltd. · Dec 22, 2025
K251606 — Power Wheelchair (D26) · Anhui Jbh Medical Apparatus Co., Ltd. · Jun 25, 2025
K170787 — Powered Wheelchair DYW30A(D09) · Nanjing Jin Bai HE Medical Apparatus Co., Ltd. · Aug 9, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
January 2, 2025
Anhui JBH Medical Apparatus Co., Ltd. % Ivy Wang Technical Manager Shanghai SUNGO Management Consulting Co., Ltd. Room 1 401, Dongfang Building, 1500# Century Ave. Shanghai. 200122 China
Re: K242448
Trade/Device Name: Power Wheelchair (D03, D05, D06, D09, D11, D16, D22, D23, D26, D36) Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: December 3, 2024 Received: December 3, 2024
Dear Ivy Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Tushar Bansal -S
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
## K242448
## Device Name
Power Wheelchair (D03, D05, D06, D09, D11, D16, D22, D23, D26, D36)
Indications for Use (Describe)
The power wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
## K242448
#### SUBMITTER l.
Name: Anhui JBH Medical Apparatus Co., Ltd.
Address: No. 116 Qicang Road, Industrial District, Ming Guang City, Chuzhou, Anhui, China
Name of contact person: Wei Yuanwei
Telephone: +86-0550-8108866
Fax: +86-0550-8106688
Email: outreach@jbhmedical.com
Date prepared: 2024-12-31
#### II. Device
Device trade name: Power wheelchair
Models: D03, D05, D06, D09, D11, D16, D22, D23, D26, D36
Classification name: Powered wheelchair
Regulation class: 2
Regulation number: 21CFR 890.3860
Panel: Physical Medicine
Product code: ITI
## III. Predicate device
K212092
Portable Folding Electric Wheelchair (Model: DC01)
Anhui JBH Medical Apparatus Co., Ltd
#### Reference device
K223393
Electric Wheelchair (Model: BBR-LY-01-01)
Shanghai BangBang Robotics Co., Ltd.
{5}------------------------------------------------
#### IV. Device description
The power wheelchair, model name: D03, D05, D06, D09, D11, D16, D22, D23, D26, D36, is an indoor/outdoor, foldable, battery-operated 2-wheel drive) powered wheelchair. It consists of two parts: the electrical part and the wheelchair main body. The electric part includes brushless motor, electromagnetic brake system, controller, rechargeable Lithium-lon battery and off-board battery charger. The main parts of the wheelchair include front wheels, rear wheels, frame, armrest, backrest and seat cushion.
The control system, including the controller handle (joystick), is equipped on the control pad that attaches to the one of the arm rests. When the joystick is released, the electromagnetic brakes will be actuated, and the power wheelchair is slowed to a stop.
The power wheelchair, model name: D03, D05, D06, D09, D11, D16, D22, D23, D26, D36, can be controlled by joystick or remote controller via Bluetooth Low Energy (BLE) wireless communication interface. The user can lock and unlock the device remotely via the BLE interface using the remote controller. For safety, controller rocker control is priority over the remote control by design.
The power wheelchair can be folded/ expanded manually. The Left and right handrails (joystick) can be interchange.
#### V. Indication for use
The power wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
{6}------------------------------------------------
## VI. Comparison of technological characteristics with the predicate device
Table 1 General Comparison
| Attribute | Subject device | Predicate device | Reference device | Discussion/Conclusion |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Manufacturer | Anhui JBH Medical Apparatus<br>Co., Ltd. | Anhui JBH Medical Apparatus<br>Co., Ltd. | Shanghai BangBang Robotics<br>Co., Ltd. | / |
| Proprietary<br>name, model | Power Wheelchair<br>D03, D05, D06, D09, D11, D16,<br>D22, D23, D26, D36 | power wheelchair, DC01 | Electric Wheelchair<br>(Model: BBR-LY-01-01) | / |
| 510(k) number | TBD | K212092 | K223393 | / |
| Device<br>classification<br>name | Class II | Class II | Class II | Same |
| Classification<br>regulations | 21 CFR 890.3860 | 21 CFR 890.3860 | 21 CFR 890.3860 | Same |
| Product code | ITI | ITI | ITI | Same |
| Indication for<br>use | The Power wheelchair is a motor<br>driven, indoor and outdoor<br>transportation vehicle with the<br>intended use to provide mobility<br>to a disabled or elderly person<br>limited to a seated position. | The wheelchair (Model: DC01) is<br>a motor driven, indoor and<br>outdoor transportation vehicle<br>with the intended use to provide<br>mobility to a disabled or elderly<br>person limited to a seated<br>position. | The intended use of the<br>Electric Wheelchair (Model:<br>BBR-LY-01-01) is to provide<br>outdoor and indoor mobility to<br>persons limited to a seated<br>position that are capable of<br>operating a powered<br>wheelchair. | Same |
| Intended user | disabled or elderly person<br>limited to a seated position | disabled or elderly person<br>limited to a seated position | disabled or elderly person<br>limited to a seated position | Same |
{7}------------------------------------------------
| Use condition | indoor and outdoor use | indoor and outdoor use | indoor and outdoor use | Same |
|---------------|------------------------|------------------------|------------------------|------|
| Type of use | OTC | OTC | OTC | Same |
## Table 2 Basic Parameters Comparison
| Attribute | Subject device | Predicate device | Reference device | Discussion/Conclusion |
|------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device length | 930mm – 1150mm | Not publicly available | 1075mm | Minor difference on dimension do not affect safety and effectiveness. |
| Device Width | 530mm – 658mm | Not publicly available | 628mm | |
| Stowage Length | 530mm – 945mm | Not publicly available | 895mm | |
| Stowage width | 280mm – 610mm | Not publicly available | 628mm | All safety and performance have been validated with the maximum rated weight dummy. |
| Stowage height | 740mm -830mm | Not publicly available | 395mm | |
| Ground clearance | 25mm - 120mm | Not publicly available | Not publicly available | Minor difference on ground clearance do not affect effectiveness and safety performance. All safety and performance have been validated with the maximum rated weight dummy. |
| Number of wheels | 4 | 4 | 4 | Same |
| Front wheel size/type | 7" /PU Solid tire | 7" /PU Solid tire | 10 in | larger size of driven wheel will provide better stability |
| Rear<br>wheel<br>size/type | 8" /PU Solid tire for D03, D05,<br>D06, D22, D23<br>12" /PU Solid tire for D09,<br>D11, D16, D26<br>11.5"/ PU solid tire for D36 | 8" /PU Solid tire | 10 in | and safety performance |
| Function<br>of<br>wheels | Front wheels: driven wheels<br>suitable for rotation,<br>acceleration, retrograde<br>Rear wheels: driving wheels<br>to control the speed and<br>direction | Front wheels: driven wheels<br>suitable for rotation,<br>acceleration, retrograde<br>Rear wheels: driving wheels to<br>control the speed and direction | Front wheels: driven wheels<br>suitable for rotation,<br>acceleration, retrograde<br>Rear wheels: driving wheels to<br>control the speed and direction | Same |
| Frame design | The frame of the wheelchair<br>is type capable of front and<br>rear close. The main part of<br>the frame can be folded for<br>saving space and convenient<br>storage and transportation. | The frame of the wheelchair is<br>type capable of front and rear<br>close. The main part of the<br>frame can be folded for saving<br>space and convenient storage<br>and transportation. | The frame of the wheelchair is<br>type capable of front and rear<br>close. | Same |
| Folding<br>mechanism | Manually fold/ expand f | Manually fold/ expand | Automatically fold/<br>expand drove by motor | Same |
| Movement<br>control method | By Joystick control and<br>Bluetooth remote controller | By Joystick control | By Joystick control and<br>Bluetooth smartphone app<br>control | Same with reference<br>device, has one more<br>control method through<br>Bluetooth. |
| Driving system | Direct drive on the rear<br>wheels | Direct drive on the rear wheels | Direct drive on the rear wheels | Same |
| Brake system | Intelligent electromagnetic<br>brake system | Intelligent regenerative<br>Electromagnetic brake | Electromagnetic | Same |
| Max speed<br>forward | Up to 6 km/h (3.75 mph),<br>continuously adjustable | Up to 6 km/h (3.75 mph),<br>variable | 6km/h | Same |
| Main frame<br>material | Aluminum alloy | Carbon fiber | Not applicable | Different material used for<br>frame, that such<br>difference will not impact<br>the safety and<br>effectiveness of the<br>subject device as the<br>performance tests are<br>conducted according to<br>ISO 7176 series. |
| Armrest cushion | Polyurethane (PU) | Polyurethane (PU) | Not applicable | Same |
| Speed settings | 5 | 5 | 5 | Same |
| Differences | | | | |
| Battery | li-ion battery; rechargeable,<br>24 VDC 10Ah/ 6Ah*2 | Li-ion, Rechargeable; 24 VDC<br>20Ah | 1 rechargeable lithium-ion<br>battery<br>Ratings: 24 V 20Ah | Same rated voltage with<br>different power capacity,<br>both batteries are tested<br>according to IEC62133-2,<br>this difference will not<br>affect the safety and<br>performance of the<br>subject device. |
| Maximum<br>distance of travel<br>on the fully<br>charged battery | 23 km for model D36<br>20 km for model D03, D05,<br>D06, D09, D11, D16, D22,<br>D23, D26 | 20 km | 20.6 km | Minor difference on travel<br>distance will not cause<br>different performance.<br>This difference will not<br>raise any new safety and<br>effectiveness concerns. |
| Battery charger | Off-board charger<br>Input: 100-240Vac, 50/60Hz,<br>1.5A<br>Output: 24 Vdc, 2A | Off-board charger<br>Input: 100-240V, 50/60Hz, 1.5A<br>Output: 24 Vdc, 2A | Input: 100-240VAC<br>50-60Hz 1.9A<br>Output: 24V 4A | More wide range of input<br>voltage in the device<br>which will not cause new<br>safety and effectiveness<br>concerns raised. |
| Minimum braking<br>distance from<br>maximum speed | Forward: <1m | Forward:0.5m | 1.2m | Minor difference on braking<br>distance will not raise any<br>new safety and<br>effectiveness concerns. |
| Max loading<br>weight | 120kg for models D03, D05,<br>D06, D09, D11, D16, D22,<br>D23, D26<br>150kg for D36 | 120kg | 120kg | Minor difference on<br>loading weight will not<br>cause different<br>performance. more<br>loading weight provide<br>more convenient and<br>stable performance for the<br>transportation. |
| Maximum safe<br>operational<br>incline degree | 8 ° | 8 ° | 6 ° | Same |
| | | | | |
| Motor | Brush motor; 24VDC; 250W/;<br>2pcs for D09, D11, D16, D22, D26, D36<br>Brush motor; 24VDC; 180W/;<br>2pcs for D03, D05, D06, D23 | Brushless DC motor; 24 VDC;<br>180 W; 2 pcs | Not applicable | minor difference on motor power will not cause different performance.<br>larger power will provide more driving force, no safety and effectiveness concerns raised. In addition, although brush motor and brushless motor has some difference, while all safety and performance tests are performed, it will not affect the difference on clinical use. |
| Electronic<br>controller | Dual Drive Controller for<br>Brushless Motor | Dual Drive Controller for<br>Brushless Motor | Not applicable | Same |
| Remote<br>controller | Via Bluetooth | N/A | Via Bluetooth | Same with reference device |
| Wireless RF frequency range | 2.402 GHz to 2.480 GHz | N/A | 2.400GHz ~ 2.4835GHz | Minor difference on the frequency range, however subject device complies with FCC 47 CFR 15.247 and RF exposure requirements, and ISO 7176-12:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs, these differences do not affect safety and effectiveness. |
| Wireless operation range | 10m | N/A | 10m | Same with reference device |
| Turning Radius | D05, D22 – 450mm<br>D03, D06, D09, D26– 750mm<br>D11 - 770mm<br>D16 - 938mm<br>D23 – 675mm<br>D36 - 850mm | 800 mm | 760mm | The minor difference in the turning radius is caused by different size of wheelchair and may cause a little bit inconvenience when it turns in a narrow space while the minor difference will not raise any new safety and effectiveness concerns. |
| Maximum<br>obstacle climbing | | 20 mm | | |
| | 20 mm for models D05, D06,<br>D22<br>25mm for D03, D09, D11,<br>D16, D23, D26<br>40mm for D36 | 20 mm | 45mm | Minor difference in the<br>obstacle climbing will not<br>impact the safety and<br>effectiveness of the<br>subject device. |
| seat cushion/<br>back cushion | mesh fabric + sponge | linen cloth filled with PU<br>foam | Not applicable | different material used for<br>parts in contact with user,<br>which such differences<br>will not impact the safety<br>and effectiveness of the<br>subject device as<br>biocompatibility tests are<br>carried out according to<br>ISO 10993 series. |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
## Table 3 Safety comparison
| Attribute | Subject device | Predicate device | Reference device | Discussion/Conclusion |
|------------------|------------------------------------------------------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------|--------------------------------------------------------------|
| Biocompatibility | All user directly contacting materials are compliance with ISO 10993-1 | All user directly contacting materials are compliance with ISO 10993-1 | All user directly contacting materials are compliance with ISO 10993-1 | Same |
| EMC | ISO 7176-21 & IEC60601-1-2<br>ANSI C63.27 | ISO 7176-21 & IEC60601-1-2 | ISO 7176-21 | Extra wireless coexistence testing was conducted and passed. |
| Performance | ISO 7176 series<br>Battery safety testing: IEC 62133-2 | ISO 7176 series<br>Battery safety testing: IEC 62133-2 | ISO 7176 series | Same |
| Labeling | Conforms to FDA Regulatory | Conforms to FDA Regulatory | Conforms to FDA Regulatory | Same |
{14}------------------------------------------------
#### VII. Summary of substantial equivalence discussion
The proposed device and predicate device are complying to the same ISO standards, ISO 7176-1:2014, ISO 7176-2:2017, ISO 7176-3:2012, ISO 7176-4:2008, ISO 7176-5:2008, ISO 7176-6:2018, ISO 7176-7:1998, ISO 7176-8:2014, ISO 7176-9:2009, ISO 7176-10:2008, ISO 7176-11:2008, ISO 7176-13:1989, ISO 7176-14:2022, ISO 7176-15:1996, ISO 16840-10:2021, ISO 7176-21:2009, ISO 7176-22:2014, ISO 7176-25:2013, IEC 62133-2:2017, IEC 60601-1-2: 2020, ANSI C63.27:2021, AAMI TIR69:2017, and FDA guidance document for wheelchairs. The proposed device performs in a similar manner to the predicate device. All these tests have corresponding requirements/ control criteria following above mentioned standards. And the test results show that the subject product is substantially equivalent to the predicate device in performance.
The performance testing demonstrates that the subject device is substantially equivalent to the predicate devices regarding Static ability, The Dynamic stability, Brake performance, Theoretical distance range, Dimension and weight, Maximum speed, Dimension of wheel Static, impact and fatigue strengths, Climatic tests, Obstacle-climbing ability, Dummy, friction of test surfaces, Power and control systems, Documentation and labeling, Resistance to iqnition, Electromagnetic Compatibility and Electrical Safety, Batteries and chargers.
The biocompatibility of the Predicate device and Subject device meet the requirements of the ISO 10993-5:2009 & ISO10993-10:2010/ ISO 10993-23:2021. Although the standard version updated, test methods for the subject device and the predicate device are same.
The proposed device has additional Bluetooth remote control way to operate the wheelchair which is different with the predicate device, while is substantially equivalent to the reference device. Although the remote-control tool is different, the Bluetooth skill is the same. The cybersecurity has been validated to be effective and do not raise new issues of safety or effectiveness.
The non-clinical laboratory data support the safety and performance of the subject device and demonstrate that the subject device should perform as intended in the specified use conditions.
11
{15}------------------------------------------------
### VIII. Summary of non-clinical testing
#### >> Performance testing-bench
The following performance data were provided to verify that the subject device met all design
specifications and provided support of the substantial equivalence determination.
- Risk Analysis developed in accordance with ISO 14971: 2019.
- . Software validation
- . ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability
- ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs
- . ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
- . ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range
- . ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space
- ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs
- ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
- . ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strength
- . ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs
- . ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs
- . ISO 7176-11:2012 Wheelchairs -- Part 11: Test dummies
- . ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces.
- ISO 7176-14:2022 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods
- . ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling.
{16}------------------------------------------------
- . ISO 16840-10:2021 Wheelchair seating - Part 10: Resistance to ignition of postural support devices - Requirements and test method.
- . ISO 7176-21:2009 Wheelchairs - Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters
- . ISO 7176-25:2013 Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs
- . Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2020.IEC 62133-2:2017 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
- . Wireless coexistence testing in accordance with ANSI C63.27:2021 (American National Standard for Evaluation of Wireless Coexistence) and AAMI TIR69:2017 (Risk management of radio-frequency wireless coexistence for medical devices and systems)
#### > Biocompatibility of patient-contacting material
Biocompatibility Tests are carried out in accordance with ISO 10993-1: 2018, including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2021) and irritation (ISO 10993-23: 2021). Additionally, the low-biocompatibility-risk materials policy in Attachment G of FDA's 2023 Guidance on "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" was invoked for eligible tissue-contacting materials.
#### IX. Summary of clinical testing
No animal study and clinical studies are available for our device. Clinical testing was not required to demonstrate the substantial equivalence of the power wheelchair to its predicate device.
#### X. Conclusions
The conclusion drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device K212092.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.