K203806 · Yamaha Motor Co., Ltd. · ITI · May 21, 2021 · Physical Medicine
Device Facts
Record ID
K203806
Device Name
JWX-1 Plus (Navigo 16/Navigo 24)
Applicant
Yamaha Motor Co., Ltd.
Product Code
ITI · Physical Medicine
Decision Date
May 21, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device JWX-1 PLUS( NAVIGO 24" ) is a Powered Wheelchair Conversion Kit and suitable for a disabled person who can not walk or have restrictions on walking. It is intended for medical purposes to provide a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and flexibility.
Device Story
Add-on drive kit converting manual wheelchairs to powered wheelchairs; consists of left/right drive units, controller, battery, and charger. User operates via joystick to control direction and motor start/stop; toggle switch used for power. Microprocessors control electrical current from rechargeable batteries (Ni-MH or Li-ion) to wheel-in motors based on joystick input. Provides regenerative braking and downhill speed regulation. Used by disabled individuals to increase mobility and flexibility. Safety features include manual parking brake and electromotive/electromechanical brakes. Tested for stability, braking effectiveness, energy consumption, and electromagnetic compatibility.
Clinical Evidence
Bench testing only. Device performance validated against ISO 7176 standards for static/dynamic stability, braking effectiveness, energy consumption, maximum speed/acceleration, climatic testing, and electromagnetic compatibility (IEC 60601-1-2). Battery safety verified per IEC 62133-1/2 and IEC 60335-2-29. Software lifecycle processes verified per IEC 62304.
Technological Characteristics
Powered wheelchair conversion kit; DC brushless motors (DC36V 110W x 2); Ni-MH or Li-ion battery power; joystick-based control interface; regenerative and electromechanical braking; pneumatic tires; quick-release axles. Complies with ISO 7176 series for powered wheelchairs and EN 12184:2014. Software follows IEC 62304 standards.
Indications for Use
Indicated for disabled persons unable to walk or with walking restrictions requiring powered wheelchair propulsion for increased mobility.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
K252101 — JWG-1 · Yamaha Motor Co., Ltd. · Mar 19, 2026
K232121 — Power wheelchair (Model:BC-EA8000) · Ningbo Baichen Medical Devices Co., Ltd. · Sep 15, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 21, 2021
Yamaha Motor CO., LTD. Mikio Saitou Official Correspondent 2500 Shingai Iwata, Shizuoka 438-8501 Japan
Re: K203806
Trade/Device Name: JWX-1 Plus (Navigo 16"/Navigo 24") Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: March 1, 2021 Received: March 1, 2021
Dear Mikio Saitou:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director, Acute Injury Device Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203806
Device Name JWX-1 PLUS( NAVIGO 16"/NAVIGO 24" )
Indications for Use (Describe)
The device JWX-1 PLUS( NAVIGO 24" ) is a Powered Wheelchair Conversion Kit and suitable for a disabled person who can not walk or have restrictions on walking. It is intended for medical purposes to provide a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and flexibility.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| 区 Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
- 1. Submitter information
| Manufacture Name: | YAMAHA MOTOR CO., LTD.<br>2500 Shingai, Iwata<br>Shizuoka 438-8501, Japan |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Michiyo YAMAZAKI, General Manager<br>JW Wheelchairs Division, Smart Power Vehicle Business Unit<br>YAMAHA MOTOR CO., LTD.<br>Tel: +81-538-32-2106<br>Fax: +81-538-38-9217<br>E-mail: yamazakim@yamaha-motor.co.jp |
| 2. Date Prepared | December 1, 2020 |
| 3. Device | |
| Type in common name: | Electric add-on drive for wheelchairs |
| Name: | JWX-1 PLUS( NAVIGO 16"/NAVIGO 24" ) |
| Classification Name: | Wheelchair, Powered (21 CFR 890.3860), Class II |
| Product Code: | ITI |
| Device Structure: | Right Drive Unit<br>Left Drive Unit<br>Controller<br>Battery<br>Battery Charger |
| 4. Predicate Device: | Alber GmbH e-fix E35/E36 (K161241)<br>YAMAHA MOTOR CO., LTD. JWX-2 (K140204) |
| 5. Device Description | The JWX-1 PLUS( NAVIGO 16"/NAVIGO 24" ) is an add-on drive for wheelchairs. The JWX-1 PLUS turns a<br>manually propelled wheelchair into a powered wheelchair. It is intended for medical purposes to prov<br>a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and<br>flexibility |
The device consists of the left/ right drive units, controller, battery and the battery charger. The drive units replace the original wheels of the manual wheelchair.
purposes to provide
Yamaha Ni-MH battery JWB2 and Yamaha Li-ion be used for the device. They are recharged can be charged also on the wheelchair condition. ESB1 does not supply the driving the charging to prevent the wheelchair to move.
- 6. Indications for use
The device JWX-1 PLUS( NAVIGO 24″ ) is a Powered Wheelchair Conversion Kit and suitable for a disabled person who can't walk or have restrictions on walking. It is intended for medical purposes to provide a means for a disabled person to take over the wheelchair and increase mobility and flexibility.
7. Comparison to Predicate Devices
The device has the similar technological characteristics as the predicate device. The microprocessors in both devices control the electrical current from the rechargeable batteries to the wheel-in motors in response to the user's joystick operations. See COMPARISON TABLE below.
{4}------------------------------------------------
8. Non-Clinical Testing
The device JWX-1 PLUS( NAVIGO 16" /NAVIGO 24" ) was tested with the following standards:
ISO 7176–1·1999 Wheelchairs - Part 1: Determination of static stability
ISO 7176-2:2001 ,Wheelchairs - Part 2:Determination of dynamic stability of electrically powered wheelchairs
ISO 7176-3 Third edition 2012-12-15. Wheelchairs - Part 3: Determination of effectiveness of brakes
ISO 7176-4 Third edition 2008-10-01, Wheelchairs – Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range
ISO 7176-6 Second edition 2001-10-01, Wheelchairs – Part 6: Determination of maximum speed, acceleration of electric wheelchairs
ISO 7176-8:1998, Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths
ISO 7176-9 Third edition 2009-11-15 Wheelchairs - Part 9: Climatic tests for electric wheelchairs
ISO 7176-11 Second edition 2012-12-01, Wheelchairs - Part 11: Test dummies
ISO 7176-14 Second edition 2008-02-15 Wheelchairs - Part 14: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods
ISO 7176-15 First edition 1996-11-15, Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling
ISO 7176-21 Second edition 2009-04-01. Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers
ISO 7176-22:2000, Wheelchairs - Part 22: Set-up procedures
IFC 62304 edition 10 2015-06 Medical device software - Software life cycle processes.
ISO 14971 Third Edtion 2019-12, Medical devices - Application of risk management to medical devices
ISO 10993-1 Fifth edition 2018-10. Biological evaluation of medical devices - Part 1:Evaluation and testing within a risk management process
ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity.
ISO 10993-10 Third edition 2010-08-01. Biological evaluation of medical devices -- Part 10: Tests for iriitation and skin sensitization
IEC 62133-1 Edition 1.0 2017-02,
Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 1: Nickel systems
{5}------------------------------------------------
### IEC 62133-2 Edition 1.0: 2017-02,
Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems
### IEC 60601-1-2 Edition 4.0 2014-02,
medical electrical equipment – part 1–2: general requirements for safety – collateral standard: electromagnetic compatibility – requirements and tests (edition2:2001 with amendment 1:2004) (aam/ansi/iec 60601–1–2:2001 with amendment 1:2004 is the u.s..(General I (QS/RM))
IEC 60335-2-29 Ed. 5.0 :2016 (b) , Household and similar electrical appliances – Safety – Part 2–29:Particular requirements for battery chargers
#### 9. Summary
The device JWX-1 PLUS( NAVIGO 16" ) has the same intended use and similar technological characteristics as the predicate device. The device does not raise any new questions of safety or effectiveness. The device is substantially equivalent to the predicate device.
{6}------------------------------------------------
| Comparison table e-fix E35/E36 versus JWX-2 versus<br>JWX-1 PLUS(NAVIGO 16"/NAVIGO 24")<br>(Same performance are indicated by boldface.) | | | note:<br>The primary predicate device is e-fix E35/E36(K161241).<br>The reason why JWX-2(K140204) as complementary device is used in the Substantial Equivalence<br>Discussion is that NI-MH, LI battery registered in the K140204 device will be used in the subject<br>device. The NI-MH, LI battery are compatible to the subject devices. | |
|------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | primary predicate device | complementary predicate device | Subject Device | Rationale |
| | e-fix E35 | JWX-2 | JWX-1 PLUS<br>(NAVIGO 16"/NAVIGO 24") | |
| | e-fix E36 | K140204 | | |
| | K161241 | | | |
| Intended Use | The e-fix is an add-on drive for<br>wheelchairs. The e-fix E35/E36 turns a<br>manually propelled wheelchair into a<br>powered wheelchair. It is intended for<br>medical purposes to provide a means for a<br>disabled person to take over the<br>propulsion of the wheelchair and increase<br>mobility and flexibility. | The device JWX-2 is a Power Assist<br>Wheelchair Conversion Kit and suitable for<br>the manual wheelchair users who are<br>limited in their field of activities because<br>of their physical conditions. The<br>device can expand their field of activities<br>by assisting their wheelchair operating<br>force. | The device JWX-1 PLUS is a Powered<br>Wheelchair Conversion Kit and suitable for a<br>disabled person who can not walk or have<br>restrictions on walking.<br>It is intended for medical purposes to provide a<br>means for a disabled person to take over the<br>propulsion of the wheelchair and increase<br>mobility and flexibility. | On safety and effectiveness view point, the subject device is as<br>same as the predicate devices. Because it is equal to other<br>ones at the concept of expanding driver's mobility by add-on<br>device. |
| Total Weight | 18.9kg(Li-ion) | 17 kg (Ni-MH)<br>17.7kg (Li-ion) | 20 kg (Ni-MH)/24", 18 kg (Ni-MH)/16"<br>21kg (Li-ion)/24", 19kg (Li-ion)/16" | The Subject Device is similar weight.<br>So it has no negative influence on the safety of the device. |
| Braking<br>technologies | Manual parking brake<br>Electromotive brake<br>Electromechanical brake | Manual parking brake<br>Electromotive brake | Manual parking brake<br>Electromotive brake<br>Electromechanical brake | The subject device has similar breaking means as e-fix E35/E36. So it<br>has no negative influence on the safety of the device.<br><br>Report Page:<br>18-38 to 18-47 10 Propulsion and braking system<br>17-75 8.12 Brakes<br>Performance tests of parking brake, braking distance etc. passed.<br>Test results conform to our intended use. |
| Speed, acceleration, deceleration | Preselected max-speed, microprocessor-controlled speed, acceleration and deceleration. The user controls these parameter with the joystick. | Microprocessor-controlled speed, acceleration and deceleration. The user controls these parameter with the hand-rim. | Preselected max-speed, microprocessor-controlled speed, acceleration and deceleration. The user controls these parameter with the joystick. | The subject device has similar speed control means as e-fix E35/E36. So it has no negative influence on the safety of the device.<br>Report Page:<br>18-21 8.1.4 Maximum Downhill speed<br>18-26 8.1.8 Maximum speed<br>18-42 10.2.1(a) Requirement<br>17-72 5.3. Wheel Chair<br>Performance tests of maximum forward/reverse/ downhill speed etc. passed.<br>Test results conform to our intended use. |
| User interface | Joystick & push buttons | Hand-rim | Joystick & Toggle switch | On the subject device and e-fix E35/E36, joystick controls the direction and motor start/stop. Compare to e-fix E35/E36, the defferences between JWX-1 PLUS (See Owner's Manual 4.2 page 13-55 and 4.3 page 13-59) and E35/36 (User Manual page 13-139) are power on/off and speed setting means. JWX-1 PLUS uses toggle switch that can be operated even by people with weak hands. Toggle switch is easier to operate than push button or dial because it dosn't need complicated move or force.<br>The toggle switch has no influence on the safety of the powered propulsion.<br>Report Page:<br>17-64 TEST REPORT PPP31025:2011A<br>18-51 12 electrical system<br>Performance tests of electrical safety passed. Test results conform to our intended use. |
| User interface 2<br>(For attendant / assistant) | Lever, Dial and Toggle switch | Dial and Push switch | | User interface 2 is the input device for attendant.<br>JWX-1 PLUS's power on/off switch (See Owner's Manual 6. page 13-65) is as same as the push button of E35/E36 (See User Manual Intuitive attendant control page 13-180).<br>Both E35/E36 and JWX-1 PLUS have a speed control dial. On JWX-1 PLUS, there are some distance from glip to dial, so there are less misoperation during operation.<br>E35/E36 has a switch to toggle the direction forward and backward, and a lever to toggle motor start and stop. Whereas JWX-1 PLUS has a button to start or stop the motor for each forward and backward direction.<br>Input means are slight different but functions are almost same. Motor on/off switches of JWX-1 PLUS are separated and different sizes in each direction so there are less confusion of the direction.<br>So it has no negative influence on the safety of the device.<br>Report Page:<br>17-64 TEST REPORT PPP31025:2011A<br>18-51 12 electrical system<br>Performance tests of electrical safety passed. Test results conform to our intended use. |
{7}------------------------------------------------
{8}------------------------------------------------
| Max. User Weight | Speed Range with Power Assist | Max Safe Slope | Max Range per Charge | Type of Motor | Rated Power of Motor |
|----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| 120kg | up to 6Km/h | up to 20% | up to 16km<br>(up to 13miles with the<br>optionally available high<br>capacity battery pack) | DC Brushless Motor | DC36V 110W × 2 |
| 160kg | up to 6Km/h | up to 15% | 40km & above (Li-ion)<br>20km & above (Ni-MH) | AC servomotor<br>(DC Brushless Motor) | DC36V 150W × 2 |
| 130kg | up to 6Km/h | up to 6 degrees (up to 10%) | 16km & above (Ni-MH)/24"<br>30km & above (Li-ion)/24"<br>15km & above (Ni-MH)/16"<br>25km & above (Li-ion)/16" | AC servomotor<br>(DC Brushless Motor) | 30mins rated DC24V 110W x 2 |
| 125kg/24", 100kg/16"<br>The subject device has a lower user weight<br>capacity and has no negative influence on the<br>safety of the device. | Same<br><br>Report Page:<br>18-21 8.1.4 Maximum Downhill speed<br>18-26 8.1.8 Maximum speed<br>18-42 10.2.1(a) Requirement<br>17-72 5.3. Wheel Chair<br>Performance tests of electrical safety passed. Test<br>results conform to our intended use. | up to 6 degrees (up to 10%)<br>The subject device has lower climbing capacity<br>but it does not adversely affect safety or<br>effectiveness of the subject device. | Compare to e-fix E35/E36, The Subject Device has less range.<br>However, it has no negative influence on the safety of the<br>device<br><br>Report Page:<br>18-20 to 18-22 8.1.2 Ability to climb rated slope.<br>Performance tests of Max Safe Slope passed. Test results<br>conform to our intended use. | Compare to e-fix E35/E36, the Subject Device has more range.<br>However, it has no negative influence on the safety of the<br>device.<br><br>Report Page:<br>18-27 8.1.9 Distance range<br>18-61 13.2 q)<br>Performance tests of Max Range per Charge passed. Test<br>results conform to our intended use. | Same<br><br>The differences in motor performance do not affect safety of<br>the subject device. |
{9}------------------------------------------------
| Battery | Battery Charger | Battery Mount | 510(k) Number |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|----------------|
| Type: Li-ion<br>Capacity: 36V 6.0Ah | Input: 100V-240V AC 50-60Hz 1.35A<br>Output: 45V 1.5A DC | Holder fixed to frame with Velcro strap. | K161241 |
| Type: Ni-MH(Dry)<br>Capacity: 24V 6.7Ah(Nom.) | Ni-MH specific charger<br>Input: 100-240V AC 50-60Hz 1.8A<br>Output: 29V 2.6A DC<br>(JWC-2) | Holder is in the bag, and the bag is fixed to frame with Velcro strap. | K140204 |
| Type: Li-ion(Dry)<br>Capacity: 25V11.8Ah(Nom.)<br>*Selectable | Li-ion specific charger<br>Input: 100-240V AC 50-60Hz 2A<br>Output: 29.2V 3A DC<br>(ESC1) | Holder is in the bag, and the bag is fixed to frame with Velcro strap. | Subject Device |
| Type: Ni-MH(Dry)<br>Capacity: 24V 6.7Ah(Nom.) | Ni-MH & Li-ion common charger<br>Input: 100-240V AC 50-60Hz 1.1A<br>Output: 29V 2.6A DC (for Ni-MH)<br>29.2V 3A DC (for Li-ion)<br>(ESC3) | Battery mount way is the same as JWX-2 one that has already recognized at K140204. | |
| Type: Li-ion(Dry)<br>Capacity: 25V11.8Ah(Nom.)<br>*Selectable | JWC-2 is the charger for Ni-MH battery. ESC1 is the charger<br>for Li-ion battery. Both chargers has discontinued so the new<br>ESC3 charger is unified two chargers, (See EMC TEST REPORT<br>Appendix 2 page 17-246).<br>ESC3 can charge both Ni-MH and Li-ion.<br>Charging way of each battery is different. When charging Ni-MH<br>battery, it is put on the receptacle on the body of the charger.<br>When charging Li-ion battery, it is connected by charging cable<br>from the charger. These manners are as same as each previous<br>charger, therefore it is hard to misunderstand the usage of it.<br>So it has no negative influence on the safety of the device. | | |
| The Subject Device's batteries are the same as JWX-2 ones that<br>has already recognized at K140204. | Report Page:<br>18-51 to 18-53 12.1 General requirement<br>17-189 EMC TEST REPORT<br>17-287 TEST RPORT IEC60335-2-29<br>Performance tests of battery charger for electrical safety, EMC<br>passed. Test results conform to our intended use. | | |
| Report Page:<br>10-24 to 10-47 1. INTRODUCTION<br>17-84 to 17-186 TEST REPORT IEC62133-2<br>Performance tests of battery electrical safety passed. Test results<br>conform to our intended use. | | | |
{10}------------------------------------------------
| Wheel Size | Tire | Quick Release Axle | Left/Right wheel<br>Synchronized Control | Regenerative Brake | Down Slope Speed Regulation | Certification |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|--------------------|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| 24"/22" | Puncture proof | QR only | Provided | Provided | Provided | EN12184<br>(Electrically powered wheelchairs, scooters<br>and their chargers – Requirements and test<br>methods) |
| 24" / 22" | Pneumatic | QR only | Provided | Provided | Provided | EN12184:2014<br>(Electrically powered wheelchairs, scooters<br>and their chargers - Requirements and test<br>methods) |
| 24" / 16" | Pneumatic | QR only | Provided | Same<br> | Provided | EN12184:2014<br>(Electrically powered wheelchairs, scooters and<br>their chargers - Requirements and test methods) |
| 16 inch model is for the user who needs the wheelchair<br>with recline or tilt function (See Pre Sale Information<br>Attachment-1 page 13-273).<br>16 inch model has the same platform with 24 inch model,<br>and it has the same safety function and reliability.<br>Toruque, speed, and other parameters have been tuned<br>for 16inch.<br>16 inch unit is proved to meet all regarding standards of<br>EN12184:2014, such as max speed, stopping distance,<br>static and dynamic stability, durability and so on.<br>The wheel size has no influence on the safety of the<br>powered propulsion. | | | | | | |
| Report Page:<br>18-76 TEST REPORT NO.20-C061<br>18-80 TEST REPORT NO.20-C063<br>Performance tests of wheel chair mechanical safety with<br>24/16 inch wheel passed. Test results conform to our<br>intended use. | | | | Same | Same | Same |
| | | | | Same<br>Report Page:<br>18-38 to 18-47 10 Propulsion and braking system<br>17-75 8.12 Brakes<br>Performance tests of braking system etc. passed.<br>Test results conform to our intended use. | Provided | |
| | | | | | Same<br>Report Page:<br>18-21 8.1.4 Maximum Downhill speed<br>18-26 8.1.8 Maximum speed<br>18-42 10.2.1(a) Requirement<br>17-72 5.3. Wheel Chair<br>Performance tests of downhill speed passed.<br>Test results conform to our intended use. | |
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.