Cajun Commando 4x4

K152672 · Lil Critter, LLC · ITI · Jan 21, 2016 · Physical Medicine

Device Facts

Record IDK152672
Device NameCajun Commando 4x4
ApplicantLil Critter, LLC
Product CodeITI · Physical Medicine
Decision DateJan 21, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

To provide mobility to people who are unable to walk and are capable of operating a powered wheelchair.

Device Story

Cajun Commando 4x4 is an electric indoor/outdoor powered wheelchair; provides mobility for users unable to walk. Device features a steel frame, four-wheel drive, and a joystick-based controller mounted on the armrest. User operates joystick to control movement; controller manages motor speed and direction. Powered by 24V direct drive gear-in-line motors and AGM batteries. Includes electromagnetic, regenerative braking system. Used in indoor and outdoor environments on firm terrain. Healthcare provider/clinician assesses patient capability to operate the device; patient self-operates during daily use. Benefits include restored independent mobility for users with walking impairments.

Clinical Evidence

Bench testing only. Compliance with ANSI/RESNA WC-1:2009 and WC-2:2009 standards for static/dynamic stability, braking effectiveness, energy consumption, dimensions, maneuvering space, speed/acceleration/deceleration, climatic testing, obstacle climbing, and electromagnetic compatibility.

Technological Characteristics

Steel frame construction; 4-wheel drive; 4x 24V 4-pole direct drive gear-in-line motors; 2x 80 AMP/hour AGM Group 24 12V batteries; Dynamic DX2-PMA90 electronic controller; electromagnetic/regenerative braking system. Dimensions: 57" length, 41.5" height, 38" turning radius. Complies with ANSI/RESNA WC-1:2009 and WC-2:2009 standards.

Indications for Use

Indicated for individuals unable to walk who possess the physical and cognitive capacity to operate a powered wheelchair.

Regulatory Classification

Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular pattern around the caduceus symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 21, 2016 Lil Critter, LLC Brenda Brant President 2701 Oliver Road Monroe, Louisiana 71201 Re: K152672 Trade/Device Name: Cajun Commando 4x4 Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: December 21, 2015 Received: December 24, 2015 Dear Brenda Brant: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the {1}------------------------------------------------ electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Michael J. Hoffmann -A for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use 510(k) Number (if known) K152672 Device Name Cajun Commando 4x4 Indications for Use (Describe) To provide mobility to people who are unable to walk and are capable of operating a powered wheelchair. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review the seller in the solled to reasing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo for Cajun Mobility. The logo features a cartoon lobster on the left side of the image. To the right of the lobster, the word "CAJUN" is written in red, and below that, the word "MOBILITY" is written in black. A Division of Lil Critter, LLC 510(k) Summary Cajun Commando 4x4 Lil Critter, LLC 510(k) K152672 Traditional | Applicant's Name: | Lil Critter, LLC | |-------------------|-----------------------| | Physical Address: | 2701 Oliver Road | | | Monroe, LA 71201 | | Mailing Address: | 2703 Stokes Street | | | Monroe, LA 71201 | | Phone: | 318.366.5772 | | Contact Person: | Brenda Brant | | | President | | Email: | cajunmobility@aol.com | | Date: | January 20, 2016 | | Device Name: | Cajun Commando 4x4 | | Device Name: | Cajun Commando 4x4 | |----------------------|---------------------| | Common Name: | Powered Wheelchair | | Classification Name: | Wheelchair, Powered | | Product Code: | ITI | | Regulatory Class: | II | | Regulation Number: | 21 CFR 890.3860 | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Cajun Mobility, which is "A Division of Lil Critter, LLC". The logo features a cartoon crawfish with googly eyes and a large claw. The words "CAJUN" are in red, and the word "MOBILITY" is in black. Predicated Device: Extreme 4x4-X8 Power Wheelchair K090350 Device Description: The Cajun Commando 4x4 is an electric indoor/outdoor wheelchair. The Cajun Commando is similar in design, composition and function to the Magic Mobility Extreme 4x4-8. The Cajun Commando 4x4 does not introduce any new technology or material that departs from current industry standard or practice. The Cajun Commando 4x4 is a four wheel drive wheelchair with a controller attached to the armrest allowing the rider to control the movement of the chair. Intended Use: To provide mobility to people who are unable to walk and are capable of operating a powered wheelchair. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows a logo for Cajun Mobility, a division of Lil Critter, LLC. The logo features a cartoon crawfish with googly eyes and a large claw. The words "CAJUN MOBILITY" are written in red and black, respectively, next to the crawfish. The text "A Division of Lil Critter, LLC" is written in black below the main logo. ## Substantial Equivalence Discussion The Commando 4x4 is the next generation of the Extreme X8 and has the same intended function and use as the Extreme X8. The Cajun Commando is similar in design, composition and function to the Extreme 4x4-8. The Cajun Commando 4x4 does not introduce any new technology or material that departs from current industry standard or practice. There are no differences in subject device or predicated device that effect safety or effectiveness. {6}------------------------------------------------ # Comparison Chart Substantial Equivalence | | Extreme 4x4- X8 K090350 | Cajun Commando 4x4<br>K140704 | |---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------| | Type | Wheelchair, power | Wheelchair, power | | Indications for Use | To provide mobility to persons<br>limited to a sittingposition, who<br>have the capacity of operating a<br>powered wheelchair. | To provide mobility to people who<br>are unable to walk & are capable<br>of operating a power wheelchair. | | Where Used | Designed to operate indoors and<br>over a variety of outdoor surfaces<br>including soft/rough terrain with<br>inclines up to 1 in 14 | Designed to operate indoors as<br>well as outdoors in firm terrain<br>with inclines up to 1 in 14 | | Design | Compact, with front frame<br>articulation | Rugged steel design with<br>conventional wheelchair steering | | Construction | Steel Frame | Steel Frame | | Dimensions | Overall length: 45.25"<br>Overall height: 37" | Overall length: 57" with rear anti-<br>tip<br>Overall height: 41.5" | | Wheelchair Seat Height | 19" | 19.7" | | Wheel Size | 4x 14" | 4x 14.6" | | Wheelchair Turning Radius | 52" | 38" | | Rolling Base Weight (without seat<br>& batteries) | 145 lbs | 149 lbs | | Range | 15.5 miles | 11.8 miles (max) | | Maximum Speed | 6.2 miles | 6.6 mph | | Maximum Weight capacity | 400 lbs | 350 lbs | | Ground Clearance | 4" | 4.5" | | Controller | Dynamic DX2-PMA70L electronic<br>control and joystick<br>*programmable to meet individual<br>needs *battery charge level gauge<br>*attaches to either side | Dynamic DX2-PMA90 electronic<br>control and joystick<br>*programmable to meet individual<br>needs *battery charge level gauge<br>*attaches to either side | | Motors | 4-24V, 4 pole direct drive gear in<br>line motors/gearbox<br>*Freewheeling lever/motor lock<br>releases & engages rear motor<br>brakes | 4-24V,4 pole direct drive gear in<br>line motors/gearbox<br>*Freewheeling lever/ motor lock<br>release & engages all motor<br>brakes | | Suspension | None | None | | Batteries | 2- 73 AMP/hour AGM Group 24<br>12 volt Weight 55 lbs | 2-80 AMP/hour AGM Group 24<br>12 volt<br>Weight 49.2 lbs | | Battery Charger Input Voltage | 110 volt | 110 volt | | Type of Braking System | Electromagnetic,regenerative<br>brakes with a free-wheeling mode | Electromagnetic, regenerative<br>brakes with a free-wheeling mode | | Seat Width | 16" – 24" | 18" – 22" | | Footrest | 1 or 2 piece, fixed or flip up, angle and height adjustable rigid footplates, and others | 1 or 2 piece, fixed or flip up angle and height adjustable rigid footplates, and others | | Armrest | Height adjustable, removable | Height adjustable, removable | | Safety Features | Electromagnetic brake engaged when chair is stationary. When electromagnetic brakes are disengaged, chair cannot be driven. Drive is inhibited with some seat positioning functions. | Electromagnetic brake engaged when chair is stationary. When electromagnetic brakes are disengaged, chair cannot be driven. | {7}------------------------------------------------ {8}------------------------------------------------ ## Discussion of Non-Clinical Data | WC-1:2009 | American National Standard for Wheelchairs- Volume 1:<br>Requirements and Test Methods for Wheelchairs (including<br>Scooters) Section 1-Determination of Static Stability | |-----------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | WC-2:2009 | American National Standard for Wheelchairs- Volume 2:<br>Additional Requirements for Wheelchairs (including scooters)<br>With Electrical Systems Section 2- Determination of Dynamic<br>Stability of Electrically Powered Wheelchairs | | WC-2:2009 | American National Standard for Wheelchairs- Volume 2:<br>Additional Requirements for Wheelchairs (including scooters)<br>With Electrical Systems Section 3: Determination of<br>Effectiveness of Brakes | | WC-2:2009 | Section 4: Energy Consumption of Electrically Powered<br>Wheelchairs and Scooters for Determination of<br>Theoretical Distance Range | | WC-1:2009 | American National Standard for Wheelchairs- Volume 1:<br>Requirements and Test Methods for Wheelchairs<br>(including scooters) Section 5- Determination of Dimensions,<br>Mass, and Maneuvering Space | | WC-2:2009 | American National Standard for Wheelchairs – Volume 2:<br>Additional Requirements for Wheelchairs (including scooters) | | | With Electrical Systems Section 6: Determination of<br>Maximum Speed, Acceleration and Deceleration of<br>Electrically Powered Wheelchairs | | WC-1:2009 | American National Standard for Wheelchairs – Volume 1:<br>Requirements and Test Methods for Wheelchairs (including<br>Scooters) Section 7: Method of Measurement of Seating and<br>Wheel Dimensions | | WC-1:2009 | American National Standard for Wheelchairs – Volume 1:<br>Requirements and Test Methods for Wheelchairs (including<br>Scooters) Section 8: Requirements and Test Methods for<br>Static, Impact and Fatigue Strengths | | WC-2:2009 | American National Standard for Wheelchairs – Volume 2:<br>Additional Requirements for Wheelchairs (including scooters)<br>With Electrical Systems Section 9: Climatic Tests for<br>Electrically Powered Wheelchairs | | WC-2:2009 | American National Standard for Wheelchairs – Volume 2:<br>Additional Requirements for Wheelchairs (including scooters)<br>With Electrical Systems Section 10: Determination of<br>Obstacle-Climbing Ability of Electrically Powered Wheelchairs | | WC-2:2009 | American National Standard for Wheelchairs – Volume 2:<br>Additional Requirements for Wheelchairs (including scooters)<br>With Electrical Systems Section 14: Power and Control<br>Systems for Electrically Powered Wheelchairs-Requirements<br>And Test Methods | | WC-2:2009 | American National Standard for Wheelchairs – Volume 2:<br>Additional Requirements for Wheelchairs (including scooters)<br>With Electrical Systems Section 21: Requirements and Test<br>Methods for Electromagnetic Compatibility of Electrically<br>Powered Wheelchairs and Motorized Scooters | {9}------------------------------------------------ {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows a logo for "Cajun Mobility, A Division of Lil Critter, LLC". The logo features a cartoon crawfish with googly eyes and a large claw. The words "Cajun Mobility" are written in a large, red font, with "Mobility" stacked below "Cajun". The text "A Division of Lil Critter, LLC" is written in a smaller, black font below the main logo. ## Substantial Equivalence Conclusion The proposed Cajun Commando 4x4 power wheelchair has similar indications for use as the predicate Extreme X8 power wheelchair. The subject and predicate have the same intended function and similar technological characteristics. There are technological differences between the subject and predicate devices, however, the results of performance testing demonstrate that these differences do not raise any questions of safety or effectiveness. The nonclinical data demonstrate that the Commando 4x4 is as safe, as effective, and performs as well as the Extreme X8. Therefore, the Cajun Commando 4x4 is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...