← Product Code [ITI](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI) · K143180

# F3 (K143180)

_Permobil AB · ITI · Mar 27, 2015 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI/K143180

## Device Facts

- **Applicant:** Permobil AB
- **Product Code:** [ITI](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI.md)
- **Decision Date:** Mar 27, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The intended use of the F3 powered wheelchair is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair.

## Device Story

F3 Powered Wheelchair is a battery-powered, front-wheel-drive mobility device. User operates the chair via a joystick interface mounted to an armrest. Input from the joystick is processed by an R-net 120 amp controller, which manages motor drive units, gear, and solenoid electromechanical brakes. Actuating the joystick releases brakes and initiates movement; releasing the joystick slows the chair to a stop and automatically reengages brakes. The device features a steel frame, two front drive wheels, and two rear pivoting casters. It supports optional tilt, recline, and elevation functions. Used in indoor and outdoor environments by patients with limited mobility. The device provides independent mobility, allowing users to navigate terrain and obstacles. Healthcare providers prescribe the device to improve patient mobility and quality of life.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including compliance with ISO 7176 series standards for stability, braking, energy consumption, dimensions, speed, obstacle climbing, and electromagnetic compatibility.

## Technological Characteristics

Steel frame construction; front-wheel drive; R-net 120 amp controller; powered by two 12VDC 60Ah or 73Ah batteries (Group M34 or M24). Features solenoid electromechanical brakes. Complies with ISO 7176-1 through 7176-21 standards for safety, performance, and testing. Connectivity is limited to the integrated controller system.

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- M300 & M400 ([K103477](/device/K103477.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, with flowing lines suggesting movement or connection. The profiles are arranged in a way that creates a sense of depth and unity. Encircling the design is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 27, 2015

Permobil AB Jan Åström Director Quality & Environment Per Uddens VAG 20, Timra, SE 86123 Vasternorrland SWEDEN

Re: K143180

Trade/Device Name: F3 Powered Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: February 10, 2015 Received: February 10, 2015

Dear Mr. Åström:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Felipe Aquel -S

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K143180

Device Name F3 Powered Wheelchair

Indications for Use (Describe)

The intended use of the F3 powered wheelchair is to provide indoor mobility to persons limited to a seating position that are capable of operating a powered wheelchair.

| Type of Use (Select one or both, as applicable) | <table><tr><td></td><td>      Exportation Use (Part 367.2(a) Category Five)     </td></tr><tr><td></td><td>      Own-Use Consumption (Part 367.2(a) Category Six)     </td></tr></table> |  | Exportation Use (Part 367.2(a) Category Five) |  | Own-Use Consumption (Part 367.2(a) Category Six) |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|-----------------------------------------------|--|--------------------------------------------------|
|                                                 | Exportation Use (Part 367.2(a) Category Five)                                                                                                                                            |  |                                               |  |                                                  |
|                                                 | Own-Use Consumption (Part 367.2(a) Category Six)                                                                                                                                         |  |                                               |  |                                                  |

|X | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# ORIGINAL, TRADITIONAL 510(K) NOTIFICATION PERMOBIL POWERED WHEELCHAIR: F3

# 510(k) Summarv

Submitter Permobil AB Box 120 S-861 23 Timrå Sweden

Phone: +46 60 595900 Facsimile: +46 60 575250

Contact Person: Jan Åström e-mail address: jan.astrom@permobil.se

Date Prepared: February, 2015

Trade name: F3

Common or Usual Name: Powered Wheelchair

Classification Name: Powered wheelchair

Product Code: ITI

Predicate Devices: M300 & M400 (K103477) manufactured by Permobil AB.

### Intended use:

The intended use of the F3 powered wheelchair is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair.

#### Description of device:

F3 Powered Wheelchair is battery powered, front wheel motor driven and is controlled by the R-net 120 amp controller. The user interface is a joystick.

The F3 is powered by two 12VDC 60Ah or tow 12VDC 73Ah. Group M34 batteries or Group M24. approximate driving range on fully charged batteries is up to 25km (15.5 miles), depending on use and the terrain the chair is driven on. The chair frame is a steel construction and includes two front drive wheels with drive units (motor, gear and brake), two batteries and two rear pivoting casters. Depending on the user's needs, the joystick motor control is mounted to the left or right armrest.

When the user activates the joystick, the controller receives a signal to release the brakes. With the brakes released, the chair is allowed to move in the joystick is actuated. When the user releases the joystick, the chair slows to a stop and the brakes are automatically reengaged. The solenoid electromechanical brakes allow the user to stop by letting go of the joystick.

#### Comparison to Predicate Devices:

The F3 is substantially equivalent to the M300 & M400 (#K103477). The F3 has the same intended uses and similar indications, technological characteristics and principles of operation. F3 has slightly more power than compared predicated device but no changes in speed occur. F3 has the same option in tilt, recline and elevation functions as the predicated device, see below table. These functions working in the same technological characteristics as the predicated M300 & M400.

{4}------------------------------------------------

# ORIGINAL, TRADITIONAL 510(K) NOTIFICATION PERMOBIL POWERED WHEELCHAIR: F3

Image /page/4/Picture/1 description: The image shows the logo for Permobil. The word "permobil" is written in a bold, sans-serif font in blue color. Below the word "permobil" is the phrase "THE POWER OF MOBILITY" in a smaller, sans-serif font in black color.

| Functions | M300 & M400<br>Predicated device | F3<br>Submitted device |
|-----------|----------------------------------|------------------------|
| Tilt      | X                                | X                      |
| Recline   | X                                | X                      |
| Elevation | X                                | X                      |

The submitted device differs from the predicated device on its position of the drive wheel. The predicated device have a shorter turning radius and are less comfortable compared to a front driven chair because the placement of the tires on the chassis. A front wheel driven chair have a better obstacle climbing than a central driven chair. The submitted device are tested and having same or improved results as the already predicated device. These technological differences do not raise any new issues in safety and effectiveness.

Other specific differences between the F3 and the M300 & M400 (K103477) are:

- * F3 has slightly larger pivoting caster wheels than the predicated device.
- * Slightly specific dimensions such as height, length, weight and turning radius are different.

These minor technological differences between the F3 and its predicate device M300 & M400 raise no new issues of safety or effectiveness. Performance data demonstrates that the F3 is as safe and effective as the M300 & M400. Thus, the F3 is substantially equivalent.

## Non-Clinical Testing:

The F3 complies to the below standards:

| Standard    | Name                                                                                                                                | FDA recognized standards |
|-------------|-------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| ISO 7176-1  | Determination of static stability                                                                                                   | 16-158                   |
| ISO 7176-2  | Determination of dynamic stability of<br>electric wheelchairs                                                                       | 16-159                   |
| ISO 7176-3  | Determination of efficiency of brakes                                                                                               | 16-192                   |
| ISO 7176-4  | Energy consumption of electric<br>wheelchairs and scooters for determination<br>of theoretical distance range                       | 16-162                   |
| ISO 7176-5  | Determination of dimensions, mass and<br>maneuverings space                                                                         | 16-163                   |
| ISO 7176-6  | Determination of maximum speed,<br>acceleration and deceleration of electric<br>wheelchairs                                         | 16-29                    |
| ISO 7176-8  | Requirements and test methods for static,<br>impact and fatigue strengths                                                           | -                        |
| ISO 7176-9  | Climatic tests for electric wheelchairs                                                                                             | 16-167                   |
| ISO 7176-10 | Determination of obstacle-climbing ability<br>of electrically powered wheelchairs                                                   | 16-164                   |
| ISO 7176-11 | Test Dummies                                                                                                                        | 16-190                   |
| ISO 7176-14 | Power and control systems for electrically<br>powered wheelchairs and scooters -<br>Requirements and test methods                   | 16-165                   |
| ISO 7176-15 | Requirements for information disclosure,<br>documentation and labelling                                                             | 16-27                    |
| ISO 7176-16 | Resistance to ignition of postural support<br>devices                                                                               | 16-191                   |
| ISO 7176-19 | Wheeled mobility devices for use in motor                                                                                           | -                        |
| ISO 7176-21 | Requirements and test methods for<br>electromagnetic compatibility of electrically<br>powered wheelchairs and motorized<br>scooters | 16-166                   |

{5}------------------------------------------------

# ORIGINAL, TRADITIONAL 510(K) NOTIFICATION PERMOBIL POWERED WHEELCHAIR: F3

Image /page/5/Picture/1 description: The image shows the Permobil logo. The word "permobil" is written in blue, with the "o" in "mobil" being a circle. Below the word "permobil" is the phrase "THE POWER OF MOBILITY" in a smaller, black font.

### Clinical Testing:

Clinical testing is not applicable.

### Conclusions:

The F3 and the predicated device M300 & M400 are substantial equivalence. F3 has the same general intended use and similar indications, principles of operation, and similar technological characteristics as the previously cleared M300 & M400. The differences between the devices are minor and do not raise any new issues of safety and effectiveness because both devices have passed all necessary testing and are considered safe and effectives for use.

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI/K143180](https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI/K143180)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
