← Product Code [ITI](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI) · K143014

# F5 (K143014)

_Permobil AB · ITI · Mar 25, 2015 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI/K143014

## Device Facts

- **Applicant:** Permobil AB
- **Product Code:** [ITI](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI.md)
- **Decision Date:** Mar 25, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine

## Indications for Use

The intended use of the F5 powered wheelchair is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair.

## Device Story

F5 Powered Wheelchair provides indoor/outdoor mobility for users with limited mobility; battery-powered, front-wheel motor-driven chassis. User operates via joystick interface connected to R-net 120 amp controller. Joystick actuation signals controller to release solenoid electromechanical brakes; chair moves based on joystick input. Releasing joystick triggers automatic deceleration and brake re-engagement. Features include tilt, recline, and elevation functions. Used in home or community settings by the patient. Output is physical mobility; healthcare provider uses device to improve patient independence and mobility. Benefits include enhanced obstacle climbing compared to central-driven chairs.

## Clinical Evidence

Bench testing only. No clinical data provided. Compliance with ISO 7176 series standards (1-6, 8-11, 14-16, 19, 21) demonstrated static/dynamic stability, braking efficiency, energy consumption, maneuverability, speed/acceleration, structural strength, climatic resistance, obstacle climbing, and electromagnetic compatibility.

## Technological Characteristics

Steel frame construction; front-wheel drive; two 12VDC 73Ah Group M24 batteries; R-net 120 amp controller; solenoid electromechanical brakes. Joystick user interface. Connectivity: R-net controller system. Complies with ISO 7176 standards for stability, strength, and EMC.

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- M300 & M400 ([K103477](/device/K103477.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three human profiles facing right, layered on top of each other, with a wing-like shape above them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2015

Permobil AB Jan Åström Director Quality & Environment Per Uddens VAG 20, Timra, SE 86123 Vasternorrland SWEDEN

Re: K143014

Trade/Device Name: F5 Powered Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: February 23, 2015 Received: February 23, 2015

Dear Mr. Åström:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Felipe Aquel -S

Carlos L. Peña, PhD, MS for

Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K143014

Device Name F5 Powered Wheelchair

Indications for Use (Describe)

The intended use of the F5 powered wheelchair is to provide indoor mobility to persons limited to a seating position that are capable of operating a powered wheelchair.

Type of Use (Select one or both, as applicable)

|  | For Routine Use (Per 49 CFR 395.8) - Driver        |
|--|----------------------------------------------------|
|  | For Time Sensitive Use (Per 49 CFR 395.8) - Driver |

|X | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# ORIGINAL, TRADITIONAL 510(K) NOTIFICATION PERMOBIL POWERED WHEELCHAIR: F5

Image /page/3/Picture/1 description: The image shows the Permobil logo. The word "permobil" is written in a bold, sans-serif font in blue color. Below the word "permobil" is the text "THE POWER OF MOBILITY" in a smaller, sans-serif font in black color.

# 510(k) Summarv

Submitter Permobil AB Box 120 S-861 23 Timrå Sweden

Phone: +46 60 595900 Facsimile: +46 60 575250

Contact Person: Jan Åström e-mail address: jan.astrom@permobil.se

Date Prepared: February, 2015

Trade name: F5

Common or Usual Name: Powered Wheelchair

Classification Name: Powered wheelchair

Product Code: ITI

Predicate Devices: M300 & M400 (K103477) manufactured by Permobil AB.

### Intended use:

The intended use of the F5 powered wheelchair is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair.

#### Description of device:

F5 Powered Wheelchair is battery powered, front wheel motor driven and is controlled by the R-net 120 amp controller. The user interface is a joystick.

The F5 is powered by two 12VDC 73Ah. Group M24. approximate driving range on fully charged batteries is up to 25km (15.5 miles), depending on use and the terrain the chair is driven on. The chair frame is a steel construction and includes two front drive units (motor, gear and brake), two batteries and two rear pivoting casters. Depending on the user's needs, the joystick motor control is mounted to the left or right armrest.

When the user activates the joystick, the controller receives a signal to release the brakes. With the brakes released, the chair is allowed to move in the joystick is actuated. When the user releases the joystick, the chair slows to a stop and the brakes are automatically reengaged. The solenoid electromechanical brakes allow the user to stop by letting go of the joystick.

#### Comparison to Predicate Devices:

The F5 is substantially equivalent to the M300 & M400 (#K103477). The F5 has the same intended uses and similar indications, technological characteristics and principles of operation. F5 has slightly more power than compared predicated device but no changes in speed occur. F5 has the same option in tilt, recline and elevation functions as the predicated device, see below table. These functions working in the same technological characteristics as the predicated M300 & M400.

{4}------------------------------------------------

# ORIGINAL, TRADITIONAL 510(K) NOTIFICATION PERMOBIL POWERED WHEELCHAIR: F5

Image /page/4/Picture/1 description: The image shows the Permobil logo. The word "permobil" is written in a blue, sans-serif font. Below the word "permobil" is the phrase "THE POWER OF MOBILITY" in a smaller, black, sans-serif font.

| Functions | M300 & M400<br>Predicated device | F5<br>Submitted device |
|-----------|----------------------------------|------------------------|
| Tilt      | X                                | X                      |
| Recline   | X                                | X                      |
| Elevation | X                                | X                      |

The submitted device differs from the predicated device on its position of the drive wheel. The predicated device have a shorter turning radius and are less comfortable compared to a front driven chair because the placement of the tires on the chassis. A front wheel driven chair have a better obstacle climbing than a central driven chair. The submitted device are tested and having same or improved results as the already predicated device. These technological differences do not raise any new issues in safety and effectiveness.

Other specific differences between the F5 and the M300 & M400 (K103477) are:

- * F5 has slightly larger pivoting caster wheels than the predicated device.
- * Slightly specific dimensions such as height, length, weight and turning radius are different.

These minor technological differences between the F5 and its predicate device M300 & M400 raise no new issues of safety or effectiveness. Performance data demonstrates that the F5 is as safe and effective as the M300 & M400. Thus, the F5 is substantially equivalent.

### Non-Clinical Testing:

The F5 complies to the below standards:

| Standard    | Name                                                                                                                       | FDA Recognition number |
|-------------|----------------------------------------------------------------------------------------------------------------------------|------------------------|
| ISO 7176-1  | Determination of static stability                                                                                          | 16-158                 |
| ISO 7176-2  | Determination of dynamic stability of electric wheelchairs                                                                 | 16-159                 |
| ISO 7176-3  | Determination of efficiency of brakes                                                                                      | 16-192                 |
| ISO 7176-4  | Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range                    | 16-162                 |
| ISO 7176-5  | Determination of dimensions, mass and maneuverings space                                                                   | 16-163                 |
| ISO 7176-6  | Determination of maximum speed, acceleration and deceleration of electric wheelchairs                                      | 16-29                  |
| ISO 7176-8  | Requirements and test methods for static, impact and fatigue strengths                                                     | -                      |
| ISO 7176-9  | Climatic tests for electric wheelchairs                                                                                    | 16-167                 |
| ISO 7176-10 | Determination of obstacle-climbing ability of electrically powered wheelchairs                                             | 16-164                 |
| ISO 7176-11 | Test dummies                                                                                                               | 16-190                 |
| ISO 7176-14 | Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods                | 16-165                 |
| ISO 7176-15 | Requirements for information disclosure, documentation and labelling                                                       | 16-27                  |
| ISO 7176-16 | Resistance to ignition of postural support devices                                                                         | 16-191                 |
| ISO 7176-19 | Wheeled mobility devices for use in motor                                                                                  | -                      |
| ISO 7176-21 | Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and motorized scooters | 16-166                 |

{5}------------------------------------------------

# ORIGINAL, TRADITIONAL 510(K) NOTIFICATION PERMOBIL POWERED WHEELCHAIR: F5

Image /page/5/Picture/1 description: The image shows the Permobil logo. The word "permobil" is written in a blue, sans-serif font. Below the word "permobil" is the phrase "THE POWER OF MOBILITY" in a smaller, black, sans-serif font. The logo is simple and modern.

Clinical testing is not applicable.

### Conclusions:

The F5 and the predicated device M300 & M400 are substantial equivalence. F5 has the same general intended use and similar indications, principles of operation, and similar technological characteristics as the previously cleared M300 & M400. The differences between the devices are minor and do not raise any new issues of safety and effectiveness because both devices have passed all necessary testing and are considered safe and effectives for use.

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI/K143014](https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI/K143014)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
