← Product Code [ITI](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI) · K070433

# LERADO, POWER WHEELCHAIR, PB (K070433)

_Lerado China Limited · ITI · Mar 16, 2007 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI/K070433

## Device Facts

- **Applicant:** Lerado China Limited
- **Product Code:** [ITI](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI.md)
- **Decision Date:** Mar 16, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The device is intended for medical purposes to provide mobility to persons restricted to a seated position.

## Device Story

Battery-operated indoor/outdoor powered wheelchair; four-wheeled base with seat. Rider operates device via hand controls on steering column. Disassemblable for transport; includes onboard battery charger. Electronic system utilizes UL-certified components (motor, batteries, charger, controller). Provides mobility for users restricted to seated position. Differences from predicate include increased incline capability (10 degrees vs 5 degrees) and extended cruising range (22.5 miles vs 15 miles).

## Clinical Evidence

Bench testing only. Compliance with ANSI/RESNA WC/Vol.2-1998, CISPR 11:1990, EN61000-3-2:1995, and IEC61000-3-3:1995 for electromagnetic compatibility and electrical safety.

## Technological Characteristics

Battery-operated powered wheelchair; four-wheeled base. Electronic controller (Dynamic DA series). Fabric back upholstery. Meets ANSI/RESNA WC/Vol.2-1998 standards. UL-certified electrical components.

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- WU'S Powered Wheelchair, MAMBO 36X ([K050010](/device/K050010.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K070433

Image /page/0/Picture/1 description: The image shows the logo for Lerado Group. The logo is in black and white and features the word "LERADO" in large, bold letters. Above the word "LERADO" is a triangular shape that resembles a mountain. Below the word "LERADO" is the word "GROUP" in smaller letters. Below the word "GROUP" are some Chinese characters.

LERADO CHINA LIMITED No. 22, Kuang Fu Road, Chia Tai Industrial Tai Pao City, Chia Yi Hsien, Taiwan, ROC Tel: 886-5-2475629 Fax: 886-5-2379672 E-mail:edwardl@lerado.com.rw http:// www.lerado.com

MAR 16 2007

February 5, 2007

## 510(k) SUMMARY "

#### Submitter's Name: Lerado China Limited

No. 22, Kuang Fu Road, Chia Tai Industrial,Tai Pao City, Chia Yi Hsien, 612, Taiwan, ROC

Date summary prepared:

Device Name:

Proprietary Name: LERADQ Power Wheelchair, PB Common or Usual Name: Powered Wheelchair Classification Name: Powered Wheelchair, Class II, 21 CFR 890.3860

Indications for Use:

The device is intended for medical purposes to provide mobility to persons restricted to a seated position.

Description of the device:

The LERADO Power Wheelchair, PB is an indoor / outdoor Powered Wheelchair that is battery operated. It has a base with four-wheeled with a seat. The movement of the Wheelchair is controlled by the rider who uses hand controls located at the top of the steering column. The device can be disassembled for transport and is provided with an onboard battery charger.

Performance Testing:

EMC Report ANSI / RESNA WC/Vol.2-1998, CISPR 11: 1990, EN61000-3-2: 1995, IEC61000-3-3: 1995 (Electrically Powered Wheelchairs, controller, and their chargers - requirements and test methods)

Legally marketed device for substantial equivalence comparison: WU'S Powered Wheelchair, MAMBO 36X (K050010)

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for LERADO GROUP. The logo is in black and white and features the company name in large, bold letters. Above the word "GROUP" is a triangular shape that resembles a mountain. Below the company name are some Chinese characters.

LERADO CHINA LIMITED No. 22, Kuang Fu Road, Chia Tai Industrial Tai Pao City, Chia Yi Hsien, Taiwan, ROC Tel: 886-5-2475629 Fax: 886-5-2379672 E-mail:edwardl@lerado.com.tw http:// www.lerado.com

#### C.2 COMPARISON SUMMARY

### ( We place the related information for the predicate device in the following pages. )

We can know from the above table that the intended use between the two devices is Mainframes of two devices are not foldable and all meet the strength and the same. fatigue tests, thus they are similar for the material aspects. The overall dimensions are similar. The weight capabilities, maximum speed, suspension of cross brace, footplates, armrest, and the warranty are all the same. Back upholstery material is also the same fabric.

Especially the electronic systems between two devices are the same suppliers, and all passed by the UL certificated, for instance the motor, batteries, and recharge. Though for the electronic controller, the predicate used by Penny & Giles and the subject device used by Dynamic DA series that are also passed by the UL certificated. Thus the same safety level for the two devices is assured.

There is different incline capability for the two devices. The new device can drive under 10 degrees and the predicate device claim 5 degrees slope. We place the specification limit of 10 degrees in the page 2 of the Owner's Manual. Thus, the user is not allowed to operate the device on the incline angle higher than the specified angle. The safety levels of the two devices are the same when operating the devices on the different inclines. They are substantially equivalent.

The cruising range per charge is different, the new device is 22.5 miles and 15 miles for the predicate device. Certainly the real range depends on the practical environments, i.e., weight, surface, incline, and temperature. For the real life use, the two devices are substantially equivalent.

To sum up, the major differences existing are the overall dimensions, cruising range, and the incline degrees are differences between the two devices. The overall So the new device is substantially appearance differences are not safety aspect. equivalent to the predicate devices in this aspect.

Based on the above the information and the analysis, we know that the subject device and the predicate device have the same intended use the same technological aspects and only minor dimensions and material differences exist. We believe that FDA can decide the subject device and the predicate device are substantially equivalent.

{2}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three strands instead of the usual one or two.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Lerado China Limited % Dr. Ke-Min Jen Official Correspondent No. 58, Fu-Chiun St. Hsin-Chu Citv. 30067 Taiwan, Republic of China

MAR 1 6 2007

Re: K070433

Trade/Device Name: Lerado Power Wheelchair, PB Regulation Number: 21 CFR 890.3860 Regulation Name: Powered wheelchair Regulatory Class: Class II Product Code: ITI Dated: February 5, 2007 Received: February 14, 2007

Dear Dr. Jen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 – Dr. Ke-Min Jen

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510 (K) Number ( If Known ):__________________________________________________________________________________________________________________________________________________

Device Name: LERADO, Power Wheelchair, PB

Indications for Use:

The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use _ V

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices

| Page          | 1 of 1  |
|---------------|---------|
| 510(k) Number | k070433 |

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI/K070433](https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/ITI/K070433)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
