Y.E.S POWERED WHEELCHAIR. T180

K032512 · Yeou Eir Shuen Machinery Co., Ltd. · ITI · Nov 12, 2003 · Physical Medicine

Device Facts

Record IDK032512
Device NameY.E.S POWERED WHEELCHAIR. T180
ApplicantYeou Eir Shuen Machinery Co., Ltd.
Product CodeITI · Physical Medicine
Decision DateNov 12, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Device Story

Y.E.S. Powered Wheelchair T180 provides mobility for individuals restricted to sitting position. Device operates via electric motor and user-controlled interface; allows navigation in indoor/outdoor environments. Intended for use by patients with mobility impairments. Output is physical movement of wheelchair; assists patient in daily activities and independence. Operated by patient.

Clinical Evidence

Bench testing only.

Technological Characteristics

Powered wheelchair; electric motor drive; user-controlled interface; class II device per 21 CFR 890.3860.

Indications for Use

Indicated for persons restricted to a sitting position requiring powered mobility assistance.

Regulatory Classification

Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a bird-like figure. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 1 2 2003 Yeou Eir Shuen Machinery Co., Ltd. C/o Dr. Ke-Min Jen ROC Chinese-European Industrial Research Society No. 58, Fu-Chiun St. Hsin-Chu City, China (Taiwan) 300 Re: K032512 Trade/Device Name: Y.E.S. Powered Wheelchair, T180 Regulation Number: 21 CFR 890.3860 Regulation Name: Powered wheelchair Regulatory Class: II Product Code: ITI Dated: October 3, 2003 Received: October 6, 2003 Dear Dr. Jen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Page 2 - Dr. Ke-Min Jen This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, miriam C. Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510 (K) NUMBER ( IF KNOW ): _ TBA DEVICE NAME: ## Y.E.S. POWERED WHEELCHAIR, T180 INDICATIONS FOR USE: The device is intended for medical purposes to provide mobility to persons restricted to a sitting position. ( PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE C.. ANOTHER PAGE Concurrence of CDRH, office of Device Evaluation (ODE ) OR Prescription Use Over - The - Counter - Use_ ( Per 21 CFR 801.109 ) ( Optional Format 1-2-96 ) Miriam C. Provost eneral. Restorative and Neurological Device F1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...