← Product Code [IOR](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR) · K223527

# Little Wave Clik, Rogue XP, Little Wave XP, Spark (K223527)

_Ki Mobility, LLC · IOR · Aug 31, 2023 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR/K223527

## Device Facts

- **Applicant:** Ki Mobility, LLC
- **Product Code:** [IOR](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR.md)
- **Decision Date:** Aug 31, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3850
- **Device Class:** Class 1
- **Review Panel:** Physical Medicine
- **Attributes:** Pediatric

## Indications for Use

The Ki Mobilty Little Wave Clik, Rogue XP, and Spark manual wheelchairs are intended to provide mobility to pediatrics limited to a seating position.

## Device Story

Manual wheelchairs (Little Wave Clik, Rogue XP, Little Wave XP, Spark) provide mobility to pediatric users restricted to sitting position; self-propelled via hand-rims or attendant-pushed via handles. Devices feature high-strength aluminum frames (rigid or folding); adjustable components (seat depth, backrest angle, center of gravity, caster position) to accommodate pediatric growth. Used in clinical or home settings. Healthcare providers/caregivers configure device settings to optimize patient positioning and support. Output is the physical mobility platform itself; enhances patient independence and mobility. Conforms to ISO 7176-19:2022 and ANSI/RESNA WC-4 Section 19:2017 for use as motor vehicle seats.

## Clinical Evidence

Bench testing only. Devices were tested against recognized consensus standards including ISO 7176-1 (static stability), ISO 7176-3 (brakes), ISO 7176-5 (dimensions/mass), ISO 7176-8 (static/impact/fatigue strength), and ISO 7176-19 (motor vehicle use). Results confirmed performance equivalent to predicate devices.

## Technological Characteristics

Materials: High-strength aluminum tubular frames. Configuration: Rigid (Clik, Rogue XP, Little Wave XP) or folding (Spark) frames. Adjustability: Seat depth, backrest angle, center of gravity, caster housing. Connectivity: None. Sterilization: Not applicable. Software: None.

## Regulatory Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Zippie GS ([K983639](/device/K983639.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 31, 2023

Ki Mobility LLC Mark Murphy Vice President of Operations 5201 Woodward Drive Stevens Point, Wisconsin 54481

Re: K223527

Trade/Device Name: Little Wave Clik, Rogue XP, Little Wave XP, Spark Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I, reserved Product Code: IOR Dated: August 29, 2023 Received: August 30, 2023

Dear Mark Murphy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Tushar Bansal -S

for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K223527

Device Name

Little Wave Clik; Rogue XP; Little Wave XP; Spark;

Indications for Use (Describe)

The Ki Mobility Little Wave Clik, Rogue XP, and Spark manual wheelchairs are intended to provide mobility to pediatrics limited to a seating position.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| 510(k) #:                          | K223527                                                  | 510(k) Summary | Prepared on: 2023-01-11 |
|------------------------------------|----------------------------------------------------------|----------------|-------------------------|
| Contact Details                    |                                                          |                | 21 CFR 807.92(a)(1)     |
| Applicant Name                     | Ki Mobility LLC                                          |                |                         |
| Applicant Address                  | 5201 Woodward Drive Stevens Point WI 54481 United States |                |                         |
| Applicant Contact Telephone        | (715) 303-6447                                           |                |                         |
| Applicant Contact                  | Mr. Mark Murphy                                          |                |                         |
| Applicant Contact Email            | mmurphy@kimobility.com                                   |                |                         |
| Device Name                        |                                                          |                | 21 CFR 807.92(a)(2)     |
| Device Trade Name                  | Little Wave Clik; Rogue XP; Little Wave XP; Spark        |                |                         |
| Common Name                        | Mechanical wheelchair                                    |                |                         |
| Classification Name                | Wheelchair, Mechanical                                   |                |                         |
| Regulation Number                  | 890.3850                                                 |                |                         |
| Product Code                       | IOR                                                      |                |                         |
| Legally Marketed Predicate Devices |                                                          |                | 21 CFR 807.92(a)(3)     |
| Predicate #                        | Predicate Trade Name (Primary Predicate is listed first) | Product Code   |                         |
| K983639                            | Zippie GS                                                | IOR            |                         |
| Device Description Summary         |                                                          |                | 21 CFR 807.92(a)(4)     |

The Ki Mobility Little Wave Clik manual wheelchair is intended to provide mobility to pediatrics restricted to a sitting position. The Little Wave Clik manual wheelchair can be self-propelled by the occupant with access to hand-rims on the rear wheels or moved by an attendant with access to push handles. It is a rigid frame wheelchair wight capacity for all models of 165 pounds/75 kg. The Little Wave Clik wheelchair has a welded tubular high strength 1.125″ aluminum frame with bend/trim configurations that vary the seat angle, knee angle, front and rear seat heights, seat depth, and caster wing offsets provide more room for the user's legs without increasing the frame length or reducing seat support.

The seat frame is dimpled to create a consistent array for frame mounted components in 0.25" increments. An adjustable clamp is used to set the rear wheel center of gravity position relative to the backrest. A set of clamps and an aluminum cross tube comprise an adjustable cross strut to seat width. The strut position may be adjusted for optimal positioning for rigidity for rigidity for rigidity for rigidity for rigi any given seat depth. The front caster mount also contains a button to mate with dimples along the front tubes for vertical caster adjustment.

The Little Wave Clik has several components intended to grow with the pediatric user either with built in adjustment, or growth kits of separate parts. The Clik cross tubes come in growth kits with multiple accommodated chair widths. Growth of the chair is also accommodated from footplates that can handle multiple widths and are adjustable to fit footrest length.

The Little Wave Clik manual wheelchair has a folding backrest frame for ease of storage and transport. It utilizes discrete holes to adjust the backrest angle relative to the seat tube. It also utilizes an adjustable position backrest to adjust the seat depth.

The tubular aluminum frame accommodates both seat slings and removable depth adjustable aluminum seat pans that have overlapping portions that can grow in depth. The pans and slings are with wheechair seat cushions. The backrest frame accommodates either back upholstery or planar/contoured wheelchair backrests. For adjustable back 

{4}------------------------------------------------

upholstery has straps that accommodate multiple seat widths. Seat pans and slings have overlapping portions that can grow in depth. The frame is also designed to accommodate aftermarket seating.

The seat cushion is a separate medical device adapted for use to the Little primary contact surface to the occupant along with the backrest upholstery. Little Wave Clik components such as amrests will also have contact to the occupant. The chair may be used as a seat in a motor vehicle and conforms to ISO 7176-19:2022 as well as ANSI/RESNA WC-4 Section 19: 2017 requirements. Two sets of transit tie down loops are bolted to the frame.

The Clik XP and Clik XPe order forms are order form variations of the same device to meet different market's requirements for pricing and reimbursement. All can be configured identically.

The Ki Mobility Rogue XP manual wheelchair is intended to provide mobility to pediatics restricted to a sitting position. The Rogue XP manual wheelchair an be self-propelled by the occupant with access to hand-rims on the rear wheels or moved by an attendant with access to push handles. It is a rigid frame wheelchair with a standard weight capacity for all models of 200 pounds/91 kg. The Rogue XP wheelchair has a welded tubular high strength 1.375" aluminum frame with bend/trim configurations that vary the seat angle, knee angle, front and rear seat heights, seat depth. Taper frames provide additional positioning of user's legs without the use of other components or accessories.

The Roque XP frame has an index of holes for mounting the backrest. Additional hole can be provided behind the backrest to accommodate adjustable clamp is used to set the rear wheel center of gravity position relative to the backrest. The chair utilizes a welded cross strut to maximize the rigidity of the chair. The Rogue XP has a welded caster wing for stiffness with a knuckle housing. An internal pinion allows adjustment of caster and backwards. The caster mount takes multiple fork and caster types, including a +3/4" stem option.

The Rogue XP has several components intended to grow with the pediatric user either with built in adjustment, or growth kits of separate parts. The cross tube weldment is split in the can grow the chair up to 3 inches in width the rigidizer bar on the backrest has a similar split with insert to grow the width of the chair is also accommodated from footplates that can handle multiple widths and are adjustable to fit footrest length.

The Rogue XP manual wheelchair has a folding backrest frame for ease of storage and transport. It utilizes discrete holes the backrest angle relative to the seat tube.

The tubular aluminum frame accommodates both seat slings and removable aluminum seat pans. The pans and slings are available for use with wheelchair seat cushions by adherence view AB tape. The backrest frame accommodates either back upholstery or planar/ contoured wheelchair backrests. For adjustable back upholstery has straps that accommodate multiple seat widths. The frame is also designed to accommodate aftermarket seating.

The seat cushion is a separate medical device adapted for use to the Rogue XP and is the primary contact surface to the occupant along with the backrest upholstery. Rogue XP components such as armrests will also have contact to the occupant.

The chair may be used as a seat in a motor vehicle and conforms to ISO 7176-19:2022 as well as ANSJ/RESNA WC-4 Section 19: 2017 requirements. Two sets of transit tie down loops are bolted to the frame.

The Rogue XP, Rogue XPe and Rogue XP TTL order form variations of the same device to meet different market's requirements for pricing and reimbursement. All can be configured identically.

The Ki Mobility Little Wave XP manual wheelchair is intended to provide mobility to pediatics restricted to a sitting position. The Little Wave XP manual wheelchair can be self-propelled by the occupant with access to hand-rims on the rear wheels or moved by an attendant with access to push handles. It is a rigid frame wheelchair with a standard weight capacity for all models of 165 pounds/75 kg. The Little Wave XP wheelchair has a welded tubular high strength 1.375″ aluminum frame with bend/trim configurations that vary the seat angle, knee angle, front and rear seat heights, seat depth, and front frame provide additional positioning of user's legs without the use of other components or accessories.

The Little Wave XP frame has an index of holes for mounting the backrest. Additional holes can be provided behind the backrest to accommodate adjustable depth. An adjustable clamp is used to set the rear wheel center of gravity position relative to the backrest. The chair utilizes a welded cross strut to main . The front caster mount is a two-part system with a clamp embedded that allows adjustment of caster angle front to back. The caster mount takes multiple fork and caster types, including a +3/4" stem option.

The Little Wave XP has several components intended to grow with the pediatric user either with built in adjustment, or growth kits of separate parts. The cross-tube weldment is split in the center with an insert that up to 3 inches in width. The rigidizer bar on the backrest has a similar split with insert to grow the chair. Growth of the chair is also accommodated from footplates that can handle multiple widths and are adjustable to fit footrest length.

The Little Wave XP manual wheelchair has a folding backrest frame for ease of storage and transport. It utilizes discrete holes to adjust the backrest angle relative to the seat tube.

The tubular aluminum frame accommodates both seat slings and removable aluminum seat pars. The pans and shor use with wheelchair seat cushions by adherence via AB tape. The backrest frame accommodates either back upholstery or planar/ contoured wheelchair backrests. For adjustment, the tension adjustable back upholstery has straps that accommodate multiple seat widths. The frame is also designed to accommodate aftermarket seating.

The seat cushion is a separate medical device adapted for use to the primary contact surface to the occupant along with the backrest upholstery. Little Wave XP components such as armrests will also have contact to the occupant.

{5}------------------------------------------------

The chair may be used as a seat in a motor vehicle and conforms to ISO 7176-19: 2022 as well as ANSI/RESNA WC-4 Section 19: 2017 requirements. Two sets of transit tie down loops are bolted to the frame. All available options and accessories are listed on the order form in the labeling section of this submission. The Ki Mobility Spark manual wheelchair is intended to provide mobility to pediatics restricted to a sitting position. The Spark manual wheelchair can be self-propelled by the occupant with access to handrims on the rear wheels or moved by an attendant with access to push handles. It is a folding frame wheelchair with a standard weight capacity for all models of 165 pounds/75 kg.

The Spark wheelchair has a welded tubular high strength aluminum frame with configurations that vary the front and rear seat heights, seat depth, rear wheel position and caster housing position. A width adjustable cross brace connects the wheelchair side frames and allows for folding of the frame for storage.

The Spark has several components intended to grow with built in adjustment within a range of its initial configuration. Arm rests offer multiple options, including a pediatric t-arm with integrated side guard, standard thatable locking extendable flip up arm rests and tubular flip up arm rests. Footplates can handle multiple widths and are adjustable to fit footrest length.

The tubular aluminum frame accommodates aluminum seat pans for use with wheelchair seat cushions. The frame is designed to accommodate aftermarket seating systems.

The Spark manual wheelchair has a folding backrest frame for ease of storage and transport. It utilizes discrete holes the backrest angle relative to the seat tube and to adjust the seat depth.

Seat cushions and backrests are separate medical devices and are primary contact surfaces to the occupants such as armrests and footrests will also have contact to the occupant.

The chair may be used as a seat in a motor vehicle and conforms to ISO 7176-19:2022 as well as ANSI/RESNA WC-4 Section 19: 2017 requirements. Two sets of transit tie down loops are bolted to the frame.

# Intended Use/Indications for Use

The Ki Mobilty Little Wave Clik, Rogue XP, and Spark manual wheelchairs are intended to provide mobility to pediatrics limited to a seating position.

# Indications for Use Comparison

The Ki Mobility Little Wave Clik, Rogue XP, Little Wave XP and Spark manual wheelchairs are intended to provide mobility to persons limited to a sitting position. The primary use is by pediatrics in need of manual wheeled mobility. This is consistent with the intended use of the Zippie GS.

# Technological Comparison

The Ki Mobilty Little Wave Clik manual wheelchair, in comparison to predicate device, shares intenced use, similarity of configuration of components, wheels and accessories. The seat depth adjustment paradigms of both devices are similar, moving the backrest relative to the front frame. The Little Wave Clik utilizes an array of dimples to adjust the backrest, whereas the Zippie GS utilizes discrete throughbolts. The center of gravity is adjustable on both products, with the Little Wave Clik utilizing an array of dimples, while the Zippie GS utilizes discrete through holes. The Ki Mobility Little Wave Clik uses a nonfolding welder rigidity to the backrest. The Zippie GS utilizes a dual seat rail system to provide two attachment points for each back cane to provide rigidity. Footrests are separate from side frames. The Zippie GS offers both folding frames, while the Little Wave Clik only offers a nonfolding frame with integrated footrest tubes. Additionally, the Zippie GS offers a reversible frame, putting the large wheels at the front of the chair and the casters at the rear, while the Little Wave Clik only offers the conventional orientation with the large wheels in the rear. Different seat heights can be achieved on both products by interchanging wheels and casters, as well as by adjusting the height of the rear wheels on its support, or the caster forks. The backrest angle of both products is adjustable by use of an array of holes to set discrete seat to backrests of both products fold down for stowage. The backrests of both products are either fixed height single tubes or a set of telescoping tubes to ackrest heights desired. The Zippie GS provides a folding mechanism to fold the chair for stowage, while the Little Wave Clik does not fold laterally. However, the Little Wave Clik aluminum frame does not present a significant change to the safety or effectiveness of the wheelchair function and has been qualified through testing per the applicable standards. The Little Wave Clik components and accessories do not pose new risks, being as safe and effective as the predicate device. The Little Wave Clik is substantially equivalent to the predicate device in intended use, design, materials, and operating principles with no new issues of safety or effectiveness.

The Ki Mobility Roque XP manual wheelchair, in comparison to predicate device, shares intended use, similarity of configuration of

## 21 CFR 807.92(a)(6)

# 21 CFR 807.92(a)(5)

### 21 CFR 807.92(a)(5)

{6}------------------------------------------------

| IOR PREDICATE COMPARISON TABLE   |                                                                                                                                    |                                                                                                                                      |                                                                                                                                             |                                                                                                                                               |                                                                                                                           |  |  |  |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|--|--|--|
| Device                           | Ki Mobility Spark                                                                                                                  | Ki Mobility Rogue XP                                                                                                                 | Ki Mobility Little Wave XP                                                                                                                  | Ki Mobility Little Wave Clik                                                                                                                  | Zippie GS                                                                                                                 |  |  |  |
| 510(k) Number<br>Intended Use    | K223527<br>The Ki Mobility Spark manual wheelchair is<br>intended to provide mobility to persons limited<br>to a sitting position. | K223527<br>The Ki Mobility Rogue XP manual wheelchair is<br>intended to provide mobility to persons limited<br>to a sitting position | K223527<br>The Ki Mobility Little Wave XP manual<br>wheelchair is intended to provide mobility to<br>persons limited to a sitting position. | K223527<br>The Ki Mobility Little Wave Clik manual<br>wheelchair is intended to provide mobility to<br>persons limited to a sitting position. | K983639<br>The Zippie GS manual wheelchair is intended to<br>provide mobility to persons limited to a sitting<br>position |  |  |  |
| Indications for Use              | The Ki Mobility Spark manual wheelchair is<br>intended to provide mobility to pediatrics<br>limited to a seating position          | The Ki Mobility Rogue XP manual wheelchair is<br>intended to provide mobility to pediatrics<br>limited to a seating position         | The Ki Mobility Little Wave XP manual<br>wheelchair is intended to provide mobility to<br>pediatrics limited to a seating position          | The Ki Mobility Little Wave Clik manual<br>wheelchair is intended to provide mobility to<br>pediatrics limited to a seating position.         | The Zippie GS manual wheelchair is intended to<br>provide mobility to pediatric users limited to a<br>sitting position    |  |  |  |
| Design<br>Weight Limit           | Ki Mobility Spark<br>165lb/75kg (with or without transit)                                                                          | Ki Mobility Rogue XP<br>200lb/91kg (with or without transit)                                                                         | Ki Mobility Little Wave XP<br>165lb/75kg (with or without transit)                                                                          | Ki Mobility Little Wave Clik<br>165lb/75kg (with or without transit)                                                                          | Zippie GS<br>165lb/75kg (with or without transit)                                                                         |  |  |  |
| Frame Material                   | Aluminum                                                                                                                           | Aluminum                                                                                                                             | Aluminum                                                                                                                                    | Aluminum                                                                                                                                      | Aluminum                                                                                                                  |  |  |  |
| Seat Width<br>Seat Depth         | 10"-16"<br>12"-18                                                                                                                  | 10" - 18"<br>10" - 20'                                                                                                               | 10" - 16"<br>10" - 16"                                                                                                                      | 8" - 16<br>8" - 16                                                                                                                            | 10" - 18"<br>10" - 18"                                                                                                    |  |  |  |
| Seat Height                      | (13"-21") Front Seat-to-Floor                                                                                                      | 13.5" - 20.5" (Front Seat-to-Floor)                                                                                                  | 11.5" - 20.5" (Front Seat-to-Floor)                                                                                                         | 11.5" - 20.5" (Front Seat-to-Floor)                                                                                                           | 14" - 20" (Front Seat-to-Floor)                                                                                           |  |  |  |
| Stowage Width                    | (11"-18.5") Rear Seat-to-Floor<br>12.6" (Folded Laterally)                                                                         | 11.5 - 20" (Rear Seat-to-Floor<br>23.6"                                                                                              | 11.5 - 20" (Rear Seat-to-Floor)<br>22.6"                                                                                                    | 11.5" - 20" (Rear Seat-to-Floor)<br>21.3"                                                                                                     | 12.5"-20" (Rear Seat-to-Floor<br>11.8" (Lateral Folding Frame Option) 20.5" (Rigid<br>Frame Option)                       |  |  |  |
| Stowage Depth<br>Stowage Height  | 29.5"<br>25.6"                                                                                                                     | 31.5"<br>24.4"                                                                                                                       | 28.9"<br>23.6"                                                                                                                              | 28.9"<br>26.6"                                                                                                                                | 29.5"<br>24.0"                                                                                                            |  |  |  |
| Transit approved?                | Yes (optional)                                                                                                                     | Yes (optional)                                                                                                                       | Yes (optional)                                                                                                                              | Yes (optional)                                                                                                                                | Yes (optional)                                                                                                            |  |  |  |
| Product weight<br>Turning Radius | 28.7lb<br>28.9"                                                                                                                    | 261b<br>26.4"                                                                                                                        | 16Ib<br>23.8"                                                                                                                               | 311b<br>20.7"                                                                                                                                 | 29Ib<br>32.5"                                                                                                             |  |  |  |
| Frame Options:                   | Width Adjustable Folding<br>Depth Adjustable<br>Footrest Swing Away<br>Footrest Super Low - Swing Away                             | Straight frame - non-folding<br>Tapered frame - non-folding                                                                          | Straight frame - non-folding<br>Tapered frame non-folding                                                                                   | Standard - non folding<br>Offset frame - non-folding                                                                                          | Folding<br>Non-folding                                                                                                    |  |  |  |
| Back Angle Range                 | 85°-115°, 5° increments                                                                                                            | 80° = 100°, 2° increments                                                                                                            | 81° - 102°, 3° increments                                                                                                                   | 80° - 100°, 2° increments                                                                                                                     | 85° = 120°, 5° increments                                                                                                 |  |  |  |
| Backrest Styles                  | Stroller Handle Back Post<br>Height Adjustable Straight with Push Handle<br>Dynamic Rocker Back                                    | Height Adjustable<br>Fixed Height<br>Fixed Height Offset                                                                             | Height Adjustable<br>Fixed Height<br>Fixed Height Offset                                                                                    | Height Adjustable<br>Fixed Height<br>Fixed Height Offset                                                                                      | Height adjustable<br>Fixed Height                                                                                         |  |  |  |
|                                  | Stroller Handle: 20"-24"                                                                                                           | Adjustable height: 9"-18"                                                                                                            | Adjustable height: 11"-18"                                                                                                                  | Height Adjustable                                                                                                                             | Fixed Height Stroller: 19", 22", 25"                                                                                      |  |  |  |
| Back Heights                     | Height Adjustable: 13"-24"                                                                                                         | Fixed Height: 9"-18"<br>Fixed Height Offset: 11"-17"                                                                                 | Fixed Height: 9" - 18"                                                                                                                      | Fixed Height                                                                                                                                  | Height Adj (Rigid): 16"-22" or 19"-25"<br>Height Adj (Folding): 17"-23" or 20"-26"                                        |  |  |  |
|                                  | Std Push Handle fixed on back canes                                                                                                | Std Push Handle fixed on back canes                                                                                                  | Fixed Height Offset: 11"-17"<br>Std Push Handle fixed on back canes                                                                         | Fixed Height Offset<br>Std Push Handle fixed on back canes                                                                                    | Fixed push handles on back canes                                                                                          |  |  |  |
|                                  | Removeable Stroller Back Handle                                                                                                    | Low Profile Push Handle fixed on back canes                                                                                          | Low Profile Push Handle fixed on back canes                                                                                                 | Low Profile Push Handle fixed on back canes                                                                                                   |                                                                                                                           |  |  |  |
|                                  |                                                                                                                                    |                                                                                                                                      |                                                                                                                                             |                                                                                                                                               |                                                                                                                           |  |  |  |
| Push Handles                     |                                                                                                                                    | Folding push Handle                                                                                                                  | Folding push Handle                                                                                                                         | Folding push handle                                                                                                                           | Angle adjustable push handle (non-folding only)                                                                           |  |  |  |
|                                  |                                                                                                                                    | Bolt-On Push Handle<br>Ergo Stroller Handle                                                                                          | Bolt-On Push Handle<br>Ergo Stroller Handle                                                                                                 | Bolt-on Push Handle<br>Ergo Stroller Handle                                                                                                   |                                                                                                                           |  |  |  |
|                                  |                                                                                                                                    | Single Stroller Handle                                                                                                               | Single Stroller Handle                                                                                                                      | Single Stroller Handle                                                                                                                        |                                                                                                                           |  |  |  |
|                                  | None                                                                                                                               | Double Stroller Handle<br>Non-adjustable Height Rigidizer Bar                                                                        | Double Stroller Handle<br>Non-adjustable Height Rigidizer Bar                                                                               | Double Stroller Handle                                                                                                                        | None                                                                                                                      |  |  |  |
| Rigidizer bars                   |                                                                                                                                    | Adjustable Height Rigidizer Bar                                                                                                      | Adjustable Height Rigidizer Bar                                                                                                             | Adjustable Height Rigidizer Bar                                                                                                               |                                                                                                                           |  |  |  |
|                                  | Single Post Height Adjustable                                                                                                      | Height Adjustable T-Arm (Tall, Std, low)                                                                                             | Height Adj. T-Arm (Std, low, tall and pediatric)                                                                                            | Height Adjustable T-Arm (Tall, Std, low and<br>pediatric)                                                                                     | Single Post (std and low)                                                                                                 |  |  |  |
|                                  | Height Adjustable T-Arm (Std, Tall, Low)                                                                                           | Angle Adj. Locking Extendable Flip Up Armrest                                                                                        | Tubular Flip Up Armrest                                                                                                                     | Tubular Flip Back Armrest                                                                                                                     | T-Post w/foam pad (regular, tall)                                                                                         |  |  |  |
|                                  |                                                                                                                                    |                                                                                                                                      |                                                                                                                                             |                                                                                                                                               |                                                                                                                           |  |  |  |
| Armrests                         | Angle Adjustable Locking Extendable Flip Up                                                                                        | Swing away armrest (short and tall)                                                                                                  | Locking Extendable Flip Back armrest                                                                                                        | Angle Adj. Locking Extendable Flip Back Armrest   Flip Up                                                                                     |                                                                                                                           |  |  |  |
|                                  | Pediatric Height Adjustable T-Arm                                                                                                  | Desk and Full Arm Pad - Standard and Waterfall                                                                                       | Swing away armrest                                                                                                                          | Swing away armrest (short and tall)                                                                                                           | Adjustable Locking Flip up                                                                                                |  |  |  |
|                                  | Tubular Flip up                                                                                                                    | styles<br>Foam Grip                                                                                                                  | Desk and Full length arm pads                                                                                                               | Desk and Full length arm pads                                                                                                                 | Length Adjustable locking flip up                                                                                         |  |  |  |
|                                  | Arm Pads (Std, Waterfall, Foam Grip)                                                                                               |                                                                                                                                      | Foam grip type arm pads                                                                                                                     | Foam Grip                                                                                                                                     |                                                                                                                           |  |  |  |
|                                  | Push to lock<br>Pull to Lock<br>Push to Lock Flush Mount                                                                           | Push to Lock<br>Pull to Lock<br>Push to Lock Flush Mount                                                                             | Push to Lock<br>Pull to Lock                                                                                                                | Push to Lock<br>Pull to lock<br>Push to lock Flush Mount                                                                                      | Push to lock<br>Pull to lock                                                                                              |  |  |  |
| Wheel Lock                       | Hemi-Wheel Locks                                                                                                                   | Low Profile Scissor Lock<br>Short Thro Scissor Lock<br>Under Seat Scissor Lock                                                       | Low Profile Scissor lock<br>Short Thro Scissor lock                                                                                         | Low Profile Scissor lock<br>Short Thro Scissor lock                                                                                           |                                                                                                                           |  |  |  |
|                                  |                                                                                                                                    | Drum Brake                                                                                                                           | Drum brake                                                                                                                                  | Drum brake                                                                                                                                    | Foot Lock                                                                                                                 |  |  |  |
|                                  | Extension handles (push and pull)                                                                                                  | Dual Drum Brake w/push or pull<br>Extension handles (push and pull)                                                                  | Dual Drum brake w/push or pull<br>Extension handles (push and pull)                                                                         | Dual Drum brake w/push or pull<br>Extension handles (push and pull)                                                                           | Hub Lock (12 and 16" wheels)<br>Extension handles (push and pull)                                                         |  |  |  |
|                                  | Grade Aids                                                                                                                         | Grade Aids                                                                                                                           | Grade aids                                                                                                                                  | Grade aids                                                                                                                                    |                                                                                                                           |  |  |  |
| Anti-Tips                        | Removable/swing-up rear Anti-tips                                                                                                  | Removable/swing-up rear Anti-tips                                                                                                    | Removable/swing-up rear Anti-tips                                                                                                           | Removable/swing-up rear Anti-tips                                                                                                             | Removable swing up rear anti-tips                                                                                         |  |  |  |
|                                  |                                                                                                                                    | Stainless Steel Spoke (18Ct.): 18", 20", 22", 24",                                                                                   |                                                                                                                                             |                                                                                                                                               | Front Anti-tips                                                                                                           |  |  |  |
|                                  | Stainless Steel Spoke (18Ct.): 18", 20", 22", 24"                                                                                  | 25"                                                                                                                                  | Stainless Steel Spoke (18 Ct.): 18", 20", 22", 24"                                                                                          | Stainless Steel Spoke (18 Ct.): 18", 20", 22", 24",                                                                                           | 18", 20", 22" and 24" Spoke                                                                                               |  |  |  |
| Rear Wheels                      | Mag: 20", 22", 24",                                                                                                                | Mag: 20", 22", 24"                                                                                                                   | Mag: 20", 22", 24"                                                                                                                          | Mag: 20", 22", 24"                                                                                                                            | 12, 16, 20", 22" and 24" mag wheels                                                                                       |  |  |  |
|                                  | Composite Fiber Spoke (18Ct.): 22" 24"                                                                                             | Composite Fiber Spoke (18Ct.): 20" 22", 24", 25" Composite Fiber Spoke (18 Ct.): 20", 22", 24"                                       |                                                                                                                                             | Composite Fiber Spoke (18 Ct.): 20", 22", 24"                                                                                                 |                                                                                                                           |  |  |  |
|                                  | Pneumatic                                                                                                                          | Composite Fiber Spoke (12Ct.): 22", 24", 25"<br>Pneumatic                                                                            | Composite Fiber Spoke (12 Ct.): 22", 24"<br>Pneumatic                                                                                       | Composite Fiber Spoke (12 Ct.): 22", 24"<br>Pneumatic                                                                                         | Pneumatic                                                                                                                 |  |  |  |
| Tires                            | Pneumatic with Airless Insert                                                                                                      | Pneumatic with Airless Insert                                                                                                        | Pneumatic with Airless Insert                                                                                                               | Pneumatic with Airless Insert                                                                                                                 | Pneumatic with Airless Insert                                                                                             |  |  |  |
|                                  | Pneumatic with puncture protection<br>Solid Polyurethane                                                                           | Pneumatic with puncture protection<br>Solid Polyurethane                                                                             | Pneumatic with Puncture Protection<br>Solid Polyurethane                                                                                    | Pneumatic with Puncture Protection<br>Solid Polyurethane                                                                                      | Solid Polyurethane                                                                                                        |  |  |  |
|                                  | Aluminum Anodized                                                                                                                  | Aluminum Anodized                                                                                                                    | Aluminum Anodized                                                                                                                           | Aluminum Anodized                                                                                                                             | Aluminum                                                                                                                  |  |  |  |
|                                  | Aluminum with Plastic Coating<br>Aluminum with Non-Slip Tape                                                                       | Aluminum with Plastic Coating<br>Aluminum with Non-Slip Tape                                                                         | Aluminum with Plastic Coating<br>Aluminum with Non-Slip Tape                                                                                | Aluminum with Plastic Coating<br>Aluminum with Non-Slip Tape                                                                                  | Aluminum Plastic Coated                                                                                                   |  |  |  |
| Handrims                         | Projection Knob with Plastic Coating                                                                                               | Projection Knob with Plastic Coating                                                                                                 | Projection Knob with Plastic Coating                                                                                                        | Projection Knob with Plastic Coating                                                                                                          | Projection Knob                                                                                                           |  |  |  |
|                                  | Ergonomic Thumb Grip: Flexible Polymer                                                                                             | Ergonomic Thumb Grip: Flexible Polymer                                                                                               | Ergonomic Thumb Grip: Flexible Polymer<br>Ergonomic Thumb Grip: Aluminum with Grip                                                          | Ergonomic Thumb Grip: Flexible Polymer                                                                                                        |                                                                                                                           |  |  |  |
|                                  | Ergonomic Thumb Grip: Alum with Grip Coating                                                                                       | Ergonomic Thumb Grip: Alum with Grip Coating                                                                                         | Coating                                                                                                                                     | Ergonomic Thumb Grip: Alum with Grip Coating                                                                                                  |                                                                                                                           |  |  |  |
| Camber<br>Caster Wheels          | 0°, 2°, 4°<br>.75" Roller Blade (3", 4")                                                                                           | 0°, 2°, 3° [drive ready], 4°, 6°, 8°<br>0.75" x 3", 4"- Rollerblade                                                                  | 0°, 2°, 3° [drive ready], 4°, 6°, 8°<br>0.75" x 3" . 4" - Roller Blade                                                                      | 0°, 2°, 3° [drive ready], 4°, 6°, 8°<br>0.75" x 3", 4" - Roller Blade                                                                         | 0<br>3″, 4″ and 5″ lighted rollerblade                                                                                    |  |  |  |
| (W x Ø)                          | .75" Lighted Roller Blade (3",4",5")                                                                                               | 0.75" x 3", 4", 5" - Lighted Rollerblade                                                                                             | 0.75" x 3", 4", 5" - Lighted Roller Blade                                                                                                   | 0.75" x 3", 4", 5" - Lighted Roller Blade                                                                                                     | 5" low poly, 5x1.5 semi-pneumatic                                                                                         |  |  |  |
|                                  | 1" Polyurethane (4", 5",6", 7",8")<br>1" Polyurethane Aluminum (4", 5")                                                            | 1" x 4", 5", 6" - Polyurethane<br>1" x 4", 5" - Polyurethane Aluminum                                                                | 1" x 4", 5", 6" - Polyurethane<br>1″ x 4″, 5″ - Polyurethane Aluminum                                                                       | 1" x 4", 5", 6" = Polyurethane<br>1" x 4", 5" - Polyurethane Aluminum                                                                         | 6" poly, 6" pneu, 6x1.5 semi-pneumatic<br>8" poly, 8" pneumatic, 8x1.5 Semi-pneu.                                         |  |  |  |
|                                  | 1" Pneumatic (6",8")                                                                                                               | 1" x 6" - Pneumatic                                                                                                                  | 1" x 6" - Pneumatic                                                                                                                         | 1" x 6" - Pneumatic                                                                                                                           | 8x2 pneu, 8x2 pneu/airless                                                                                                |  |  |  |
|                                  | 1.5" Polyurethane (4", 5",6")                                                                                                      | 1.5" x 4", 5", 6" = Polyurethane                                                                                                     | 1.5" x 4", 5", 6" - Polyurethane                                                                                                            | 1.5" x 4", 5", 6" - Polyurethane<br>1.5" x 4", 5", 6" =  Soft Roll Aluminum                                                                   |                                                                                                                           |  |  |  |
|                                  | 1.5" Soft Roll Aluminum (3", 4", 5", 6")<br>2″ Polyurethane (6″)                                                                   | 1.5" x 4", 5", 6" - Soft Roll Aluminum                                                                                               | 1.5" x 4", 5", 6" - Soft Roll Aluminum                                                                                                      |                                                                                                                                               |                                                                                                                           |  |  |  |
|                                  | 2" Pneumatic w/ Foam Insert (8")                                                                                                   |                                                                                                                                      |                                                                                                                                             |                                                                                                                                               |                                                                                                                           |  |  |  |
|                                  | Aluminum Fork<br>Fork Stem (Std, 3/" Longer, 1 ½" Longer)                                                                          | Aluminum Fork<br>Single Sided Fork                                                                                                   | Aluminum Fork<br>Single Sided Fork                                                                                                          | Aluminum Fork<br>Single Sided Fork                                                                                                            | Aluminum Fork                                                                                                             |  |  |  |
| Caster Forks                     | Shock Absorbing (Aluminum, Composite)                                                                                              | Shock Absorbing Fork: (Aluminum, Composite)                                                                                          | Shock Absorbing Fork: (Aluminum, Composite)                                                                                                 | Shock Absorbing Fork: (Aluminum, Composite)                                                                                                   |                                                                                                                           |  |  |  |
|                                  | Caster Pin Locks                                                                                                                   |                                                                                                                                      |                                                                                                                                             |                                                                                                                                               |                                                                                                                           |  |  |  |
|                                  | Housing (Std, Performance)                                                                                                         |                                                                                                                                      |                                                                                                                                             |                                                                                                                                               |                                                                                                                           |  |  |  |
|                                  | Swing Away, Footrest, Extension Mount                                                                                              | Integral to frame                                                                                                                    | Integral to frame                                                                                                                           | Integral to frame                                                                                                                             | Swing Away Footrest, Extension mount                                                                                      |  |  |  |
| Footrest Hangers                 | Swing Away, Footrest, Front Mount<br>Elevating Leg Rest (PRO, PRO Pediatric)                                                       |                                                                                                                                      |                                                                                                                                             |                                                                                                                                               | Swing Away Footrest, Front Mount<br>Elevating Footrest (ext and front mount)                                              |  |  |  |
|                                  | Residual Limb Support                                                                                                              |                                                                                                                                      |                                                                                                                                             |                                                                                                                                               | Articulating Footrest (ext. mount)                                                                                        |  |  |  |
|                                  | Hanger Release (Classic, 4 Way In)<br>Composite                                                                                    | Tubular w/o Footplate (Open)                                                                                                         | Tubular without Footplate (Open)                                                                                                            |                                                                                                                                               |                                                                                                                           |  |  |  |
|                                  | Composite Angle Adjustable                                                                                                         | Tubular with Plastic Footplate                                                                                                       | Tubular with Plastic Footplate                                                                                                              |                                                                                                                                               | Composite                                                                                                                 |  |  |  |
| Footplates                       | Aluminum Angle Adjustable<br>Aluminum Locking Multi-Angle Adjustable                                                               | Aluminum Angle Adj.<br>Aluminum Angle Adj. with Riser                                                                                | Aluminum Angle Adjustable<br>Aluminum Angle Adjustable with Riser                                                                           | Aluminum Angle Adjustable with Riser                                                                                                          | Aluminum Angle Adjustable<br>Aluminum Footplate                                                                           |  |  |  |
|                                  | One Piece Flip Up Adjustable                                                                                                       | High Mount Aluminum Angle Adj                                                                                                        | High Mount Aluminum Angle Adjustable                                                                                                        | High Mount Aluminum Angle Adjustable                                                                                                          |                                                                                                                           |  |  |  |

{7}------------------------------------------------

|                            | Swing Away Extensions (Short, Med, Long)    | Angle Adj. Flip-Under                       | Angle Adjustable Flip Under                 | Angle Adjustable Flip Under                 | Footplate flip-up                   |
|----------------------------|---------------------------------------------|---------------------------------------------|---------------------------------------------|---------------------------------------------|-------------------------------------|
|                            | Front Mount Extension                       | High Mount Angle Adj. Flip-Under            | High Mount Angle Adjustable Flip Under      | High Mount Angle Adjustable Flip Under      |                                     |
| Seating and<br>positioning | Removable Aluminum Seat Pan                 | Solid Seat pan                              | Solid Seat pan                              | Depth Adjustable Seat Pan                   | Solid Seat/Back                     |
|                            |                                             | Standard Seat Sling                         | Standard Seat Sling                         | Standard Seat Sling                         |                                     |
|                            | Seat Cushions                               | Seat Cushions                               | Seat Cushions                               | Seat Cushions                               | Jay Cushions                        |
|                            | Backrest                                    | Backrests                                   | Backrests                                   | Backrest                                    | Jay Back                            |
|                            | Lumbar pads                                 | Lumbar pads                                 | Lumbar pads                                 | Lumbar pads                                 |                                     |
|                            | Shoulder guides                             | Shoulder guides                             | Shoulder guides                             | Shoulder guides                             |                                     |
|                            | Headrest                                    | Headrests                                   | Headrests                                   | Headrests                                   |                                     |
|                            | Chest supports                              | Chest supports                              | Chest supports                              | Chest supports                              |                                     |
|                            | Supports (Chest and Trunk)                  | Supports (Chest and Trunk)                  | Supports (Chest and Trunk)                  | Supports (Chest and Trunk)                  |                                     |
|                            | Laterals                                    | Laterals                                    | Laterals                                    | Laterals                                    | Toe Loop                            |
|                            | Positioning belts                           | Pelvic Positioning belts                    | Pelvic Positioning belts                    | Pelvic Positioning belts                    | Positioning belts                   |
|                            | Calf straps                                 | Calf straps                                 | Calf straps                                 | Calf straps                                 | Leg Strap                           |
|                            | Heel loops                                  |                                             |                                             |                                             | Heel loops                          |
| Accessories                | Ki Mobility Spark                           | Ki Mobility Rogue XP                        | Ki Mobility Little Wave XP                  | Ki Mobility Little Wave Clik                | Zippie GS                           |
| Sideguards                 | Composite Side guards (adult and pediatric) | Composite Side guards (adult and pediatric) | Composite Side guards (adult and pediatric) | Composite Side guards (adult and pediatric) | Composite Side guards (Std and Low) |
|                            |                                             | Aluminum Side guards (adult and pediatric)  | Aluminum Side guards (adult and pediatric)  | Aluminum Side guards (adult and pediatric)  |                                     |
| Backpack/Pouch             | Yes                                         | Yes                                         | Yes                                         | Yes                                         | Yes                                 |
| Spoke Guards               | Yes                                         | Yes                                         | Yes                                         | Yes                                         | Yes                                 |
| Cane/Crutch Holder         | Yes                                         | Yes                                         | Yes                                         | Yes                                         | No                                  |
| ELR Gel Pads               | Yes                                         | No                                          | No                                          | No                                          | No                                  |
| Impact guards              | Yes                                         | Yes                                         | Yes                                         | Yes                                         | Yes                                 |
| Canopy                     | Yes                                         | No                                          | No                                          | Yes                                         | Yes                                 |
| Tool Kit                   | Yes                                         | Yes                                         | Yes                                         | Yes                                         | Yes                                 |
| One-Arm Drive              | Yes                                         | Yes                                         | No                                          | Yes                                         | Yes                                 |

{8}------------------------------------------------

components, wheels and accessories. The seat depth adjustment paradigms of both devices are similar, moving the backrest relative to the front frame. Both the Roque XP and the Zippie GS utilizes dis…

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR/K223527](https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR/K223527)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
