← Product Code [IOR](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR) · K213790

# Aluminum back folding wheelchair SYIV 100-LQXAL2482-407 (K213790)

_Mobility Source Medical Technology Co., Ltd. · IOR · Mar 14, 2022 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR/K213790

## Device Facts

- **Applicant:** Mobility Source Medical Technology Co., Ltd.
- **Product Code:** [IOR](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR.md)
- **Decision Date:** Mar 14, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3850
- **Device Class:** Class 1
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Device Story

Manual wheelchair; propelled by occupant via hand-rims or pushed by caregiver via rear handles. Designed for mobility assistance for individuals with illness, injury, or disability. Composed of painted aluminum alloy frame, fireproof/breathable seat/backrest, armrests, detachable leg rests, 8-inch front casters, 24-inch rear wheels, brake assembly, and anti-overturn device. Used in various settings (home, clinic, etc.) by patients or caregivers. Provides mobility, facilitating daily activities and independence. Output is physical transport of the user.

## Clinical Evidence

Bench testing only. No clinical data. Performance verified against ISO 7176 series (stability, brakes, dimensions, strength, ignition resistance) and ISO 10993 series (biocompatibility). Usability study with 15 participants confirmed the user manual is understandable and the device is operable by users.

## Technological Characteristics

Frame: Aluminum alloy 6061-T6 (Ø 25.4mm, 2.0mm wall). Armrests: Aluminum alloy 6061-T6 (Ø 19mm, 1.5mm wall). Tires: PU solid. Seat/backrest: Flame-resistant fabric. Mechanical operation. Foldable frame. Anti-overturn device. Standards: ISO 7176-1, -3, -5, -7, -8, -11, -13, -15, -16, -22; ISO 10993-1, -5, -10.

## Regulatory Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- AST Model MA012 and MS019 Rehab Wheelchair ([K181795](/device/K181795.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 14, 2022

Mobility Source Medical Technology Co., Ltd. % Shanfeng Jiang Regulation Control Manager Guangzhou Junyi Information Technology Co., Ltd. Room 304, Building A, 62 Nanyun2 Road Huangpu District, Science City, Guangdong 510663 China

Re: K213790

Trade/Device Name: Manual wheelchair SYIV 100-LQXAL2482-407 Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I, reserved Product Code: IOR Dated: December 6, 2021 Received: December 6, 2021

Dear Shanfeng Jiang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K213790

Device Name Manual Wheelchair SYIV1OO-LQXAL2482-407

Indications for Use (Describe)

The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

Date of Summary Preparation: Nov,08, 2021

## 1. Submitter's Identifications

Submitter's Name: Mobility Source Medical Technology Co., Ltd.

- 1) Address: 160 Heheng Road, Yanghe Town, Gaoming District, Foshan City, Guangdong Province, China, 528513
2 ) Contact Person: Evan Wong Contact Title: QA Manager Contact E-mail Address: evan@lqxmedical.com Telephone: +86- 757-88807188 Fax: +86-757-88809768 3 ) Manufacturer: Guangdong Shunde LQX Health Technology Inc. Address: Gujian Industrial Zone, Liangle Road Daliang Town, Shunde District Foshan City, Guangdong Province, China, 528000 Tel: +86-757-28091601 Fax: +86-757-28091602 Email: evan@lqxmedical.com

### 2. Correspondent's Identifications

Correspondent's Name: Guangzhou Junyi Information Technology Co., Ltd. Address: Room 304, building A, 62 Nanyun2 Road, Science City, Huangpu District, Guangzhou China ZIP Code: 510663 Contact Person: Shanfeng Jiang Contact Title: Regulation Control Manager Contact E-mail Address:jiang13620586569@126.com Telephone: +86-20-82329549 Fax: +86-20-82329549

## 3. Name of the Device

Device Classification Name: Wheelchair, Mechanical Product Name: Manual wheelchair SYIV100-LQXAL2482-407 Trade Name: Aluminum back folding wheelchair Model: SYIV100-LQXAL2482-407 Classification Panel: Physical Medicine Regulation Number: 21 CFR 890.3850 Product Code: IOR Device Classification: Class I

## 4. The Predicate Devices

{4}------------------------------------------------

#### AST Model MA012 and MS019 Rehab Wheelchair K181795

## 5. Device Description

## Operating principle

A mechanical wheelchair is a chair with wheels, designed to be a replacement for walking, where it is propelled by the seated occupant turning the rear wheels by hand. There are also handles behind the seat for someone else to do the pushing. Wheelchairs are used by people for whom walking is difficult or impossible due to illness, injury, or disability.

#### 【Structure composition

The device is mainly composed of painted frame, fireproof and breathable seat cushion, armrest, backrest, hand push tube, detachable leg, 8-inch front wheel, 24-inch rear wheel, brake group and anti-overturn device.

## 【Main materials

Frame: high strength aluminum alloy 6061 - T6, external diameter of Ø 25.4 mm, wall thickness of 2.0 mm.

Armrest: high strength aluminum alloy 6061 - T6, external diameter of Ø 19 mm, wall thickness of 1.5 mm

Brake assembly: aluminum alloy

Seat and backrest: flame resistant fabrics

Tire: PU

## 6. Intended Use of Device

The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.

#### 7. Summary of Substantial Equivalence

#### Table 1 Comparison to Predicate Device

|                     | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Comparison                                                                                                                                                                                                                                                                                                                                                  |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k Number         | K213790                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K181795                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | ------                                                                                                                                                                                                                                                                                                                                                      |
| Product Code        | IOR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | IOR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                        |
| Proprietary Name    | Manual<br>Wheelchair<br>SYIV100-LQXAL2482<br>-407                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | AST Model MA012 and<br>MS019 Rehab Wheelchair                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | ------                                                                                                                                                                                                                                                                                                                                                      |
| Model               | SYIV100-LQXAL2482<br>-407                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | MA012 and MS019                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | ------                                                                                                                                                                                                                                                                                                                                                      |
| Manufacturer        | Guangdong Shunde LQX<br>Health Technology Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Sichuan AST Medical<br>Equipment Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | ------                                                                                                                                                                                                                                                                                                                                                      |
| Indications for Use | The device is intended<br>for medical purposes to<br>provide mobility to<br>persons restricted to a                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | The AST Model MA012 and<br>MS019 Rehab Wheelchairs<br>are to provide mobility to<br>persons limited to a sitting                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                        |
|                     | sitting position.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | position.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                             |
| Type of Use         | Over-the-counter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Over-the-counter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                        |
| Basic Design        | 1. The wheelchair<br>adopts aluminum alloy<br>tube frame and<br>ergonomic design, the<br>surface of the piano<br>paint treatment, to<br>achieve high strength,<br>high load bearing, light<br>weight, small volume,<br>and at any time folding<br>structure, do not take up<br>space, easy to carry;<br>2. Seat depth and<br>backrest angle can be<br>adjusted according to<br>human comfort.<br>3. Backrest height can<br>be adjusted.<br>4. The armrest of the<br>wheelchair can be lifted<br>back, the height of the<br>armrest can be adjusted,<br>the front and back can<br>slide, the footplate can<br>rotate, and the angle can<br>be adjusted.<br>5. Front and rear wheels<br>adopt PU solid tire,<br>fireproof and breathable<br>seat cushion and<br>backrest.<br>6. Backrest can be back<br>folding.<br>7. The device can be<br>equipped with drum<br>brake, easy for the<br>caregiver control and<br>down the steep slope<br>control flexibility and<br>stability. | The AST Model MA012 and<br>MS019 Rehab Wheelchair are<br>manual wheelchairs.<br>They have adjustable<br>armrests, and multiple axle<br>position. The casters are<br>6"/7"/8" PU wheels with<br>height adjustable forks and<br>the rear wheels are<br>20"/22"/24"*1-3/8"<br>polyurethane (MA012 and<br>MS019).<br>Aluminum frame with liquid<br>coated and folding backrest<br>Flip-up and detachable<br>armrest PU pad<br>Detachable footrest<br>Front casters PU tires and<br>quick release rear wheel PU<br>tires<br>Nylon cushion | Similar<br>The structure<br>and materials<br>of main<br>frame of the<br>proposed<br>device and<br>predicate<br>device are<br>same, only<br>individual<br>materials are<br>different,<br>such as back<br>upholstery<br>and seat<br>upholstery,<br>etc.<br>The<br>differences<br>do not raise<br>safety and<br>effectiveness<br>of the<br>proposed<br>device. |

{5}------------------------------------------------

{6}------------------------------------------------

|           | Materials    |        | with aluminum alloy<br>brake assembly.<br>9. There is Equipped<br>with anti-overturn<br>device.<br>Frame: high strength<br>aluminum alloy 6061 -<br>T6, external diameter of<br>$Ø$ 25.4 mm, wall<br>thickness of 2.0 mm.<br>Armrest: high strength<br>aluminum alloy 6061 -<br>T6, external diameter of<br>$Ø$ 19 mm, wall<br>thickness of 1.5 mm<br>Brake assembly:<br>aluminum alloy<br>Seat and backrest: flame<br>resistant fabrics<br>Tire: PU | AST Model MA012 Rehab<br>Wheelchair: Aluminum frame<br>with liquid coated and folding<br>backrest<br>Flip-up and detachable<br>armrest PU pad<br>Detachable footrest<br>Front casters PU tires and<br>quick release rear wheel PU<br>tires<br>Nylon cushion | Similar<br>The Main<br>structural<br>materials and<br>wheels<br>materials are<br>same.<br>The<br>differences<br>do not raise<br>safety and<br>effectiveness<br>of the<br>proposed<br>device. |
|-----------|--------------|--------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|           | Components   |        | The device is mainly<br>composed of painted<br>frame, fireproof and<br>breathable seat cushion,<br>armrest, backrest, hand<br>push tube, detachable<br>leg, 8-inch front wheel,<br>24-inch rear wheel,<br>brake group and<br>anti-overturn device.                                                                                                                                                                                                   | AST Model MA012 Rehab<br>Wheelchair: Main frame,<br>handle sleeve, back<br>upholstery, seat upholstery,<br>armrest pad, armrest, side<br>panel, front casters, rear<br>wheel, legrest, footplate,<br>brake, front fork, cross-<br>brace.                    | Same<br>Although the<br>expressions<br>are different,<br>the main<br>components<br>are same.                                                                                                 |
|           | Control Mode |        | Mechanical                                                                                                                                                                                                                                                                                                                                                                                                                                           | Mechanical                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                         |
|           |              | Length | 1070mm                                                                                                                                                                                                                                                                                                                                                                                                                                               | Model MA012:<br>1173mm( $\pm$ 1mm)                                                                                                                                                                                                                          | Similar<br>The two<br>physical                                                                                                                                                               |
| Dimension |              | Width  | 625mm                                                                                                                                                                                                                                                                                                                                                                                                                                                | Model MA012:<br>645mm ( $\pm$ 1mm)                                                                                                                                                                                                                          | dimensions<br>are different.<br>The                                                                                                                                                          |
|           |              |        |                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                             |                                                                                                                                                                                              |
|           |              | Height | 1035mm                                                                                                                                                                                                                                                                                                                                                                                                                                               | Model MA012:<br>892mm ( $\pm$ 1mm)                                                                                                                                                                                                                          | difference<br>does not<br>affect the<br>effectiveness<br>and safety                                                                                                                          |

{7}------------------------------------------------

| Net weight                   | 38.3 lbs/17.4kg                      | Model MA012:<br>17.2kg/38lbs                                                                                                                                    | Similar<br>The<br>proposed<br>device<br>is<br>heavier than<br>the predicate<br>device.                                                                          |                                                                                                             |
|------------------------------|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Weight capacity              | 300lbs/136kg                         | Model MA012:<br>300lbs/136kg                                                                                                                                    | Same                                                                                                                                                            |                                                                                                             |
| Seat width                   | 446mm(17.6")                         | Model MA012:<br>16"(406mm)<br>18"(457mm)<br>20"(508mm)                                                                                                          | Similar<br>Different<br>seat sizes do<br>not raises the<br>safety<br>and<br>effectiveness<br>of<br>the<br>device.                                               |                                                                                                             |
| Seat height                  | 518mm(20.4")                         | Model MA012:<br>19.7"(500mm)                                                                                                                                    |                                                                                                                                                                 |                                                                                                             |
| Seat depth                   | 495mm(19.5")                         | Model MA012:<br>16"-20"<br>(406mm-508mm)                                                                                                                        |                                                                                                                                                                 |                                                                                                             |
| Frame type                   | foldable                             | foldable                                                                                                                                                        | Same                                                                                                                                                            |                                                                                                             |
| Cross-brace<br>configuration | 18"                                  | 14", 16", 18", 20" or 22"                                                                                                                                       | Same<br>Cross-brace<br>sizes<br>of<br>predicate<br>device<br>include with<br>that<br>of<br>proposed<br>device.                                                  |                                                                                                             |
| Back style                   | Fixed                                | Model MS019: Fixed<br>Model MA012: Adjustable                                                                                                                   | Same                                                                                                                                                            |                                                                                                             |
| Anti-tippers                 | Optional                             | Optional                                                                                                                                                        | Same                                                                                                                                                            |                                                                                                             |
| Wheel<br>construction        | Quick release                        | Quick release                                                                                                                                                   | Same<br>Wheel<br>construction<br>of proposed<br>device<br>is<br>same<br>as<br>predicate<br>device model<br>SIVFH2A10<br>2.                                      |                                                                                                             |
| Tires                        | Front:200mm (8")<br>Rear: 600mm(24") | Front: 6",7",8"<br>Rear: 20",22",24"                                                                                                                            | Same<br>The<br>tires                                                                                                                                            |                                                                                                             |
|                              |                                      |                                                                                                                                                                 | sizes of<br>proposed<br>device are<br>included to<br>those of<br>predicate<br>device.                                                                           |                                                                                                             |
|                              | Armrest                              | Flip back armrest                                                                                                                                               | Model MA012: Height<br>Adjustable desk length<br>armrest, Flip back<br>Model MS019: Fixed or<br>adjustable height; desk or full<br>length; removable            | Similar<br>The height of<br>armest of<br>proposed<br>device can<br>not be<br>adjustable.                    |
|                              | Foot rest                            | Optional/ swing away                                                                                                                                            | Optional/ swing away                                                                                                                                            | Same                                                                                                        |
|                              | Rear Axle Position                   | Single                                                                                                                                                          | Multiple                                                                                                                                                        | Similar<br>The rear axle<br>of proposed<br>device has<br>only one<br>size.                                  |
|                              | Frame Material                       | high strength aluminum<br>alloy                                                                                                                                 | Model MA012: Aluminum<br>Model MS019: Steel                                                                                                                     | Same<br>The frame<br>material of<br>proposed<br>device is<br>same as<br>predicate<br>device model<br>MA012. |
|                              | Safety Feature                       | Manual Wheel Lock                                                                                                                                               | Manual Wheel Lock                                                                                                                                               | Same                                                                                                        |
|                              | Standard                             | ISO7176-1<br>ISO7176-3<br>ISO7176-5<br>ISO7176-7<br>ISO7176-8<br>ISO7176-11<br>ISO7176-13<br>ISO7176-15<br>ISO7176-16<br>ISO7176-22<br>ISO10993-1<br>ISO10993-5 | ISO7176-1<br>ISO7176-3<br>ISO7176-5<br>ISO7176-7<br>ISO7176-8<br>ISO7176-11<br>ISO7176-13<br>ISO7176-15<br>ISO7176-16<br>ISO7176-22<br>ISO10993-1<br>ISO10993-5 | Same                                                                                                        |
|                              | ISO10993-10                          | ISO10993-10                                                                                                                                                     |                                                                                                                                                                 |                                                                                                             |
|                              |                                      |                                                                                                                                                                 |                                                                                                                                                                 |                                                                                                             |

{8}------------------------------------------------

{9}------------------------------------------------

#### 8. Substantial Equivalence discussion:

Our device and the predicate device are almost identical in terms of all areas described in the above table (Table 1). There are some minor differences with the predicate device don't affect the safety or effectiveness of the subject device.

The following table (Table 2) shows similarities and differences of the performance between our device and the predicate device. Tests were conducted following the recommended procedures outlined in the respective consensus standards, and results for "Manual Wheelchair SYIV100-LQXAL2482-407" met all relevant requirements in the test standards, our internal specifications, and are comparable to the predicate device.

| Description                                              | Proposed Device        | Predicate device       |
|----------------------------------------------------------|------------------------|------------------------|
| Static stability                                         | Meets ISO 7176-1:2014  | Meets ISO 7176-1:2014  |
| Effectiveness of brakes                                  | Meets ISO 7176-3:2012  | Meets ISO 7176-3:2012  |
| Dimensions, mass and<br>maneuvering space                | Meets ISO 7176-5:2008  | Meets ISO 7176-5:2008  |
| Seating and wheel<br>dimensions                          | Meets ISO 7176-7:1998  | Meets ISO 7176-7:1998  |
| Static, impact, and fatigue<br>strengths                 | Meets ISO 7176-8:2014  | Meets ISO 7176-8:2014  |
| Information disclosure,<br>documentation and<br>Labeling | Meets ISO 7176-15:1996 | Meets ISO 7176-15:1996 |
| Resistance to ignition                                   | Meets ISO 7176-16:2012 | Meets ISO 7176-16:2012 |

#### Table 2: Comparison of Performance Testing

The tests were performed following the general requirements outlined in ISO 7176-11:2012, ISO 7176-13:1989, ISO 7176-22:2014.

A brief discussion of the non-clinical testing data that was submitted, referenced or relied on to demonstrate that the Proposed Device is safe and effective, and whose performance meets the requirements of its user-defined acceptance criteria and intended uses:

"Mechanical Wheelchair SIVFH2A102" meets performance requirements per ISO 7176-1:2014, ISO 7176-3: 2012, ISO 7176-5:2008, ISO 7176-7:1998, ISO 7176-8:2014, ISO 7176-11:2012, 7176-13: 1989, ISO 7176-15: 1996, ISO 7176-16:2012 and ISO 7176-22:2014. It is safe and effective, and its performances meet the requirements of the pre-defined acceptance criteria and intended use.

A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence.

According to the non-clinical test results, the proposed devices are as safe, as effective and perform as well as the predicate device. So the proposed device is 

{10}------------------------------------------------

Substantially Equivalent (SE) to the predicate device which is US legally market device.

## 9. Non-Clinical Tests Performed:

following testing was performed on the "Mechanical Wheelchair The SYIV100-LQXAL2482-407" in accordance with the requirements of the design control regulations and established quality assurance procedures.

Safety and performance:

ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability

ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes

ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space

ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions

ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths

ISO 7176-11:2012 Wheelchairs - Part 11: Test dummies

ISO 7176-13:1989 Wheelchairs – Part 13: Determination of friction of test surface

ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling

ISO 7176-16:2012 Wheelchairs - Part 16: Resistance to ignition of postural support devices

ISO 7176-22:2014 Wheelchairs - Part 22: Set-up procedures

Biocompatibility:

ISO 10993-1:2009 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

ISO 10993-5:2009 Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity

ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and delayed-type hypersensitivity

## 10. Usability Study:

The 15 participants read the User Manual before initial use of the device. They can understand the questions and answer them. The 15 participants can operate the device referring to the User Manual.

Based on the verification & validation testing records provided, we concluded that the Manual Wheelchair, Model SYIV100-LQXAL2482-407 is considered to meet the usability requirement as defined in the verification & validation test report.

These non-clinical tests demonstrate that the labeling for the Manual Wheelchair, SYIV100-LQXAL2482-407, Model by Guangdong Shunde LQX Health Technology Inc. and bv US-Distributors is comprehensive, was well understood by a variety of "patients", and provided sufficient information for safe and proper use of the device.

{11}------------------------------------------------

# 11. Conclusion:

Based on the comparison of the proposed device of SYIV100-LQXAL2482-407 is determined to be Substantially Equivalent (SE) to the predicate device of AST Model MA012 and MS019 Rehab Wheelchair, in respect of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR/K213790](https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR/K213790)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
