K192658 · Jerry Medical Instrument (Shanghai) Co., Ltd. · IOR · Apr 1, 2020 · Physical Medicine
Device Facts
Record ID
K192658
Device Name
Manual Wheelchair
Applicant
Jerry Medical Instrument (Shanghai) Co., Ltd.
Product Code
IOR · Physical Medicine
Decision Date
Apr 1, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Indications for Use
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Device Story
Manual wheelchair for disabled or elderly individuals; provides mobility via manual operation of rear tires by user or pushing by caretaker. Consists of frame, footplate, front castors, rear wheels, anti-tippers, armrests, and brakes. Available in four models (JR201-JR204) with variations in wheel size (6/8-inch front, 22/24-inch rear) and armrest adjustability. Used in indoor and outdoor environments. Provides mobility support to improve patient independence and facilitate caregiving. No electronic or software components.
Clinical Evidence
No clinical data; bench testing only. Device performance verified against ISO 7176 standards (stability, braking, dimensions, fatigue, ignition resistance) and biocompatibility standards (ISO 10993-5, ISO 10993-10).
Technological Characteristics
Manual wheelchair; aluminum alloy frame; PU/PVC/Oxford cloth components. Dimensions vary by model. No energy source; manual operation. Biocompatibility per ISO 10993-1. Testing per ISO 7176-1, 3, 5, 7, 8, 11, 13, 15, 16, 22.
Indications for Use
Indicated for disabled and elderly persons (weight < 100kg) restricted to a sitting position requiring mobility assistance.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 30, 2020
Jerry Medical Instrument (Shanghai) Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608, No.738, Shangcheng Rd., Pudong Shanghai, China 200120
Re: K192658
Trade/Device Name: Manual Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I, reserved Product Code: IOR Dated: January 5, 2020 Received: February 20, 2020
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, Ph.D. Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K192658
Device Name Manual Wheelchair
Indications for Use (Describe)
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
#### 1.0 Submitter's information
Name: JERRY MEDICAL INSTRUMENT (SHANGHAI) CO., LTD. Address: Building 12, No. 615 Fengdeng Rd, Malu Town, Jiading District, Shanghai 201801, China Tel: 86-13817397985 Fax: 86-21-59517526 Contact: Jianguo Chen Date of Preparation: Sep.20, 2019
#### Designated Submission Correspondent
Mr. Boyle Wanq Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn
### 2.0 Device information
Trade name: Manual Wheelchair Common name: Mechanical Wheelchair Classification name: Wheelchair, Mechanical Model(s): JR201, JR202, JR203, JR204
#### 3.0 Classification
Production code: IOR Regulation number: 21 CFR 890.3850 Classification: Class I Panel: Physical Medicine
#### 4.0 Predicate device information
Manufacturer: JIANGYIN NEWRISE MEDICAL EQUIPMENT CO., LTD. Device: Manual Wheelchair 510(k) number: K180852
### 5.0 Indication for Use Statement
{5}------------------------------------------------
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
#### 6.0 Device description
The proposed device, Manual Wheelchair, driven by manual operation of rear tire, or pushed by caretaker.
This product is mainly designed to be used to take care of the disabled or elderly, it is driven by manual operation of rear tire, or pushed by caretaker.
The Manual Wheelchair consists of Frame, Footplate, Front castor, Rear wheel, Anti-tipper, Armrest, Brake, Central fork.
The Manual Wheelchair has four models: JR201, JR202, JR203, JR204.
- JR201: 6 inch front wheel and 22 inch rear tire, without height adjustable armrest, flip-up armrest, detachable footrest, connecting brake
- JR202: 6 inch front wheel and 22 inch rear tire, with height adjustable armrest, flip-up armrest, detachable footrest, connecting brake
- JR203: 8 inch front wheel and 24 inch rear tire,, without height adjustable armrest, flip-up armrest, detachable footrest, connecting brake
- JR204: 8 inch front wheel and 24 inch rear tire, with height adjustable armrest, ● flip-up armrest, detachable footrest, connecting brake
Max. loading can not be over than 100Kgs.
### 7.0 Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
16-195 ISO 7176-1 Third edition 2014-10-01 Wheelchairs - Part 1: Determination of static stability 16-192 ISO 7176-3 Third edition 2012-12-15 Wheelchairs - Part 3: Determination of effectiveness of brakes 16-163 ISO 7176-5 Second edition 2008-06-01 Wheelchairs - Part 5: Determination of overall dimensions, mass and manoeuvring space 16-196 ISO 7176-7 First Edition 1998-05-15 Wheelchairs - Part 7: Measurement of seating and wheel dimensions 16-197 ISO 7176-8 Second editon 2014-12-15 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue
{6}------------------------------------------------
strengths
16-190 ISO 7176-11 Second edition 2012-12-01
Wheelchairs - Part 11: Test dummies
16-25 ISO 7176-13 First edition 1989-08-01
Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces
16-27 ISO 7176-15 First edition 1996-11-15
Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling
16-191 ISO 7176-16 Second edition 2012-12-01
Wheelchairs - Part 16: Resistance to ignition of postural support devices
16-198 ISO 7176-22 Second edition 2014-09-01
Wheelchairs - Part 22: Set-up procedures
ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
## 8.0 Clinical Test Conclusion
No clinical study implemented for the mechanical wheelchair.
## 9.0 Technological Characteristic Comparison Table
| Item | Proposed device | Predicated device | Remark |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------|
| Product Code | IOR | IOR | SE |
| Regulation No. | 21 CFR 890.3850 | 21 CFR 890.3850 | SE |
| Class | I | I | SE |
| Product name | Manual Wheelchair | Manual Wheelchair | - |
| 510(k) No. | | K180852 | - |
| Models | JR201, JR202, JR203,<br>JR204 | XSG106A | - |
| Intended Use | The device is intended for<br>medical purposes to provide<br>mobility to persons<br>restricted to a sitting<br>position | The device is intended for<br>medical purposes to provide<br>mobility to persons<br>restricted to a sitting<br>position. | SE |
| Use<br>environment | Indoor, outdoor | Indoor, outdoor | SE |
| Patient<br>Population | Disabled and elderly people<br>(less than 100kg) | Disabled or an elderly<br>person | * Gap 1 |
| Product<br>structure | Frame, Footplate, Front<br>castor, Rear wheel,<br>Anti-tipper, Armrest, Brake. | Frame, Footplate, Front<br>castor, Rear wheel,<br>Anti-tipper, Armrest, Brake. | SE |
#### Table1-General Comparison
{7}------------------------------------------------
| | Central fork | | |
|---------------------|------------------------------------------------------------------|------------------------------------------------------------------|--------|
| Height adjustable | With height adjustable, or, without height adjustable | No height adjustable | *Gap 2 |
| Driving system | Driven by manual operation of rear tire, or pushed by caretaker. | Driven by manual operation of rear tire, or pushed by caretaker. | SE |
| Footplate | Yes | Yes | SE |
| Number of wheels | 4 | 4 | SE |
| Main frame material | Aluminium alloy | Aluminium alloy | SE |
* Gap analysis:
Gap 1: the two device share same target population, but the proposed device has stricter scope, only for whose weight is less than 100kg, which is defined in the product intended use, and the predicate device does not demonstrate this limit. After clarification in the product intended use, the difference can not raise additional safety concerns.
Gap 2: the models of JR201, JR202, JR203, R204 have difference in armrest type (with height adjustable, or, without height adjustable), the predicate device's armrest is without height adjustable. The difference in height adjustable does not bring additional safety and effectiveness concerns;
{8}------------------------------------------------
| Item | Table2 Performance Comparison<br>Proposed Device | | | | Predicate Device | Remark |
|---------------------------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|--------------------------------------------------|----------|
| | JR201 | JR202 | JR203 | JR204 | | |
| Braking system | Connecting brake | | | | Connecting brake | SE |
| Max. loading<br>(kg) | 100 | | | | 100 | SE |
| Overall<br>dimensions | 1000*530*970<br>mm | 1000*530*970<br>mm | 1015*655*955<br>mm | 1015*655*955<br>mm | 1030*640*930 mm | * Gap 3 |
| Seat dimension | Depth:435 mm<br>Height:445 mm<br>Width:<br>410/450/510 mm | Depth:435 mm<br>Height:445 mm<br>Width:<br>410/450/510 mm | Depth:435 mm<br>Height:500 mm<br>Width:<br>410/450/510 mm | Depth:435 mm<br>Height:500 mm<br>Width:<br>410/450/510 mm | Depth: 460 mm<br>Height: 420 mm<br>Width: 410 mm | * Gap 4 |
| Rear wheel | Size: 22 inch<br>Tire: no<br>pneumatic | Size: 22 inch<br>Tire: no<br>pneumatic | Size: 24 inch<br>Tire: no<br>pneumatic | Size: 24 inch<br>Tire: no<br>pneumatic | Size: 610 mm<br>Tire: no pneumatic | * Gap 5 |
| Wheel lock | Push-to-Lock | Push-to-Lock | Push-to-Lock | Push-to-Lock | Pull-to-Lock | * Gap 6 |
| Ground<br>clearance | 70mm | 70mm | 70mm | 70mm | 150mm | * Gap 7 |
| Armrest | Without height<br>adjustable | Height adjustable | Without height<br>adjustable | Height adjustable | Without height adjustable | * Gap 8 |
| Casters (front<br>wheel) | Size: 6 inch | Size: 6 inch | Size: 8 inch | Size: 8 inch | Size: 200mm | * Gap 9 |
| Static stability<br>(forward, with<br>non-lockable<br>front wheels) | 15° | 15° | 15° | 15° | 17° | * Gap 10 |
| Static stability<br>(rearward, with<br>non-lockable<br>rear wheels) | 15° | 15° | 15° | 15° | 17° | * Gap 11 |
| Static stability<br>(rearward, with<br>lockable rear<br>wheels) | 13.1° | 13.1 ° | 13.1 ° | 13.1 ° | 16° | * Gap 12 |
| Static stability<br>(lateral forward,<br>Left/Right) | 15° | 15° | 15° | 15° | 16° | * Gap 13 |
| Parking brakes<br>(Facing<br>downhill, Facing<br>uphill) | 15° | 15° | 15° | 15° | 16° | * Gap 14 |
| Min. Turning<br>back diameter | 1700mm | 1700mm | 1700mm | 1700mm | 1700mm | SE |
| Net Weight | 14.8 kgs | 15 kgs | 16.3 kgs | 16.5 kgs | 18kg | * Gap 15 |
| Anti-tip wheels | Removable | Removable | Removable | Removable | Removable | SE |
| Suspension | No | No | No | No | No | SE |
# Table2 Performance Comparison
{9}------------------------------------------------
* Gap analysis:
Gap 3-5, the two devices have different dimensions, rear wheel size, this specification is only affects the device, it could not affects the safety and effectiveness of proposed device are meet the design specification and been conducted into performance test, so the subject is as safe, as effective, and performs as well as the legally marketed prodicate device; Gap 6, the two devices have different wheel lock, which only influence the operation method: pull or push, it does not create additional safety and effectiveness concerns;
{10}------------------------------------------------
Gap 7, the two devices have different ground clearance, which impact the trafficability characteristic, but not affects the safety and effectiveness of proposed device, the ground clearance of proposed device is approved by its test report;
Gap 8, the models of JR201, JR202, R204 have difference in armrest adjusting (with height adjustable), the predicate device's armrest is without height adjustable, only lie one more function, does not bring additional safety and effectiveness concerns;
Gap 9, the different specification in caster size is only affects the appearance of the safety and effectiveness of proposed device;
Gap 10-14, the static stability of the proposed device, both of them is enough for usual using condition, the difference does not raise additional safety and effectiveness concerns;
Gap 15, the net weight of proposed device is lower than the predicate device, but is close, the difference does not raise and effectiveness concerns.
{11}------------------------------------------------
#### ltem Proposed Device Predicate Device Remark Performance test ISO 7176-1:2014, ISO 7176-1:2014, SE ISO 7176-3:2012. ISO 7176-3:2012, ISO 7176-5 :2008, ISO 7176-5 :2008, ISO 7176-7:1998, ISO 7176-7:1998, ISO 7176-8:2014, ISO 7176-8:2014, ISO 7176-11:2012, ISO ISO 7176-11:2012, ISO 7176-13:1989, ISO 7176-13:1989, ISO 7176-15:1996, 7176-15:1996, ISO 7176-16:2012, ISO 7176-16:2012, ISO 7176-22:2014. ISO 7176-22:2014. * Gap 16 Main materials Frame, Anti-tip, Central Frame: Aluminium allov: fork: Aluminium alloy; Wheel, Armrest: PU; Armrest: PU; Rear Wheel: PU Solid Casters: Material Casters: PVC Solid PP+30%GF+PU; Material Rear wheel: hub+spoke(36 pcs) +aluminium alloy hand rim: Backrest: oxford cloth; Brake: Steel Materials Armrest: PU; * Gap 16 Armrests, Seat Base, Back Inner cushion: oxford Cover: Artificial leather contacting user cloth Comply with ISO SE Biocompatibility of Comply with ISO 10993-1, materials 10993-1, FDA Guidance, FDA Guidance, Tests contacting user Tests included included Cytotoxicity (ISO Cytotoxicity (ISO 10993-5:2009), 10993-5:2009), Sensitization and Sensitization and Intracutaneous Intracutaneous Reactivity (ISO Reactivity (ISO 10993-10:2010) 10993-10:2010) Conforms to FDA SE Label and Conforms to FDA Labeling Regulatory Requirements Regulatory Requirements
#### Table3 Safety Comparison
* Gap analysis:
Gap 16, the two devices have different materiasl on the Armrests, Seat Base, Back Cover, the two materials has passed the bio-compatibility tested on Cytotoxicity, Sensitization and Intracutaneous Reactivity based on the same test standards: ISO 10993-5:2009, and ISO 10993-10:2010; therefore, no safety and effectiveness concerns raised.
{12}------------------------------------------------
#### 10.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device and raises no new questions of safety or effectiveness.The differences between both devices are insignificant in terms of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.