← Product Code [IOR](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR) · K051369

# SYRYKER SORANO WHEELCHAIR (K051369)

_Stryker Canada · IOR · Sep 26, 2005 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR/K051369

## Device Facts

- **Applicant:** Stryker Canada
- **Product Code:** [IOR](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR.md)
- **Decision Date:** Sep 26, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3850
- **Device Class:** Class 1
- **Review Panel:** Physical Medicine

## Indications for Use

The Stryker Sorano Wheelchair is a mechanical wheelchair intended to provide mobility to persons restricted to a seated position.

## Device Story

Stryker Sorano is a manual, user-propelled mechanical wheelchair for indoor/outdoor use. Frame constructed of titanium or chrome-moly steel; features rear wheels with handrims and front pivoting casters. Includes fiber glass composite suspension system connecting frame to rear axle. Optional lever drive system provides two forward gear ratios and disc braking, eliminating need to grip handrims. Modular design allows user adjustment of seat height/angle, backrest angle/height, and footrest position. Footrest module is removable; optional folding backrest available for compact storage. Operated by the user to provide mobility; intended to improve patient independence.

## Clinical Evidence

Bench testing only. Device meets performance standards specified in ANSI/RESNA WC/Vol. 1-1998 requirements and test methods for wheelchairs.

## Technological Characteristics

Materials: Titanium or chrome-moly steel frame, fiber glass composite suspension. Rigid, non-folding frame. Propulsion: Manual handrims or optional lever drive with two gear ratios and disc brakes. Adjustability: Modular seat, backrest, and footrest components. Standards: ANSI/RESNA WC/Vol. 1-1998.

## Regulatory Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Invacare Top End Terminator Titanium Manual Wheelchair ([K012167](/device/K012167.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

SEP 26 2005

struker

K05/369

**Canada**

## 510K Summary: Stryker Sorano Wheelchair

Trade Name: Stryker Sorano

Common Name: Manual Wheelchair

Classification Name: Wheelchair, Mechanical 8911OR

Device Sponsor:

Stryker Canada 45 Innovation Drive Hamilton, ON, Canada L9H 7L8

Regulatory Class: Class I 21 CFR 890.3850 Mechanical Wheelchair

Contact Person:

Kathryn Ronalds, Senior Regulatory Affairs Representative

Predicate Device:

The Stryker Sorano Wheelchair is substantially equivalent to the Invacare Top End Terminator Titanium Manual Wheelchair (K012167, Aug 1/01).

Intended Use:

The Stryker Sorano Wheelchair is a mechanical wheelchair intended to provide mobility to persons restricted to a seated position.

Device Description:

The Stryker Sorano Wheelchair is a manually operated, user propelled mechanical wheelchair. It is intended to provide mobility to persons restricted to a seated position.

The product consists of a titanium frame (also available in chrome-moly steel), large rear wheels with handrims for propelling the chair, and smaller front pivoting casters for stability. The wheelchair is a lightweight manual chair designed for everyday use, both indoors and outdoors. The wheelchair is a rigid, non-folding type wheelchair.

{1}------------------------------------------------

The Stryker Sorano offers a suspension system made of a fibre glass composite. The two suspension elements are part of the frame construction as they connect the main frame to the rear wheel axle. As it is part of the frame, the suspension is a standard feature.

The Stryker Sorano offers an additional propulsion method to conventional handrims. A lever drive option may be added that provides two forward gear ratios an incorporates a disc braking system. Handrim use is still possible as handrims are standard on all units. The key benefits to the drive system are that it eliminates the need to grip the handrim, provides mechanical advantage and provides an enhanced braking system.

The Stryker Sorano also offers modularity and adjustability features. The seat height, angle, rear seat-to-floor height, and front seat-to-floor height may be adjusted by the end user. The footrest module with caster wheels may be adjusted forward and backward, and the footrest height may be adjusted. The backrest angle and height may also be adjusted. In addition, the footrest module is removable and may be replaced with optional sports or other special purpose footrests. The seat/backrest module and the footrest module may be removed by the user, allowing the entire wheelchair to be packed in a compact space for travel. The optional folding backrest enhances this feature.

Performance Data:

The Stryker Sorano Wheelchair will meet the applicable performance standards specified in Rehabilitation Engineering Sociey of North American (RESNA) Standard ANSI/ RESNA WC/Vol. 1-1998 "Requirements and Test Methods for Wheelchairs."

Substantial Equivalence (SE) Rational:

The Stryker Sorano Wheelchair is equivalent in intended use, safety, and effectiveness to existing devices being marketed by Invacare.

Safety and Effectiveness:

The Stryker Sorano Wheelchair does not raise any new safety and efficacy concerns when compared to a similar device already legally marketed. Therefore, the Stryker Sorano Wheelchair is substantially equivalent to this existing device.

Signed: Kathryn Ronalds Dated: May, 20 2005

Kathryn Ronalds Senior Regulatory Affairs Representative

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services, USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, facing to the left.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 2 6 2005

Ms. Sheryl Bagalio Regulatory and Quality Affairs Supervisor Stryker Canada 45 Innovation Drive Hamilton, ON, Canada L9H 7L8

Re: K051369

Trade/Device Name: Stryker Sorano Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical wheelchair Regulatory Class: I Product Code: IOR Dated: August 18, 2005 Received: August 19, 2005

Dear Ms. Bagalio:

We have reviewed your Section 510(k) premarket notification of intent to market the device WC nave rowed your be received the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for association to May 28, 1976, the enactment date of the Medical Device Amendments, or to conniner of the May 20, 2017 11:11 in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic For (10) has the device, subject to the general controls provisions of the Act. The r ou may, diere, eroyisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your de rise to such additional controls. Existing major regulations affecting your device can may or sations of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Sheryl Bagalio

This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin manketing your antial equivalence of your device to a legally premarket nother device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your as 11-07-20120. Also, please note the regulation entitled, Contact the Office of Compuners and (21 CFR Part 807.97). You may obtain Misolanding by reference to premail.org/w/index.php?title=Pivision of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

2

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Statement of Indications for Use

| 510K Number:         | TBD                                                                                                                               |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Device Number:       | Stryker Sorano Wheelchair                                                                                                         |
| Indications for Use: | The Stryker Sorano Wheelchair is a mechanical wheelchair intended to provide mobility to persons restricted to a seated position. |

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ AND/OR Prescription Use _ (Part 21 CFR 801.109)

(Optional Format 1-2-96)

## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

2

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number_KOS 1369 369

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR/K051369](https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR/K051369)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
