K983662 · Warepalmy Enterprise, LLC (Usa) · INI · Dec 18, 1998 · Physical Medicine
Device Facts
Record ID
K983662
Device Name
C.T.M. MOBILITY SCOOTER, MODEL # HS-686
Applicant
Warepalmy Enterprise, LLC (Usa)
Product Code
INI · Physical Medicine
Decision Date
Dec 18, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Indications for Use
The C.T.M. Mobility Scooter HS-686 is an indoor/outdoor scooter which provides transportation for a disabled or elderly person.
Device Story
C.T.M. Mobility Scooter HS-686 is a battery-powered, four-wheeled motorized scooter for elderly or disabled individuals. Device features platform base, adjustable seat, and steering column with hand-operated controls. Rider uses thumb-control levers for forward/reverse movement and steering column for direction. Includes horn, headlight, turn signals, and speed dial. Powered by two 12V sealed lead acid batteries; includes 24V constant current charger. Disassembles into five components without tools for transport. Features automatic electric braking system on rear wheels; brakes engage upon release of thumb controls or system failure. Includes self-diagnostic warning light for battery status and system malfunctions. Free-wheeling mode allows manual maneuvering. Used in indoor/outdoor environments; operated by the rider.
Clinical Evidence
No clinical data or comparative performance testing submitted. Compliance demonstrated via bench testing per FDA Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs, and Motorized Three Wheeled Vehicles (July 1995).
Technological Characteristics
Four-wheeled battery-powered scooter; two 12V sealed lead acid (SLA) batteries; 24V constant current charger. Electric braking system (rear wheels). Hand-operated controls on steering column. Disassembles into five parts; max weight 59.5 lbs per part. 11-inch diameter wheels. Self-diagnostic system with LED flash codes for fault detection (battery, brake, electrical).
Indications for Use
Indicated for transportation of elderly or disabled persons in indoor and outdoor settings.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
{1}------------------------------------------------
BEC 1 8 1998
K983662
# 510(k) SUMMARY
Submitter's Name: Warepalmy Enterprise LLC (USA) 1725 NE Orenco Station Parkway Hillsboro, OR 97124 (503) 693-6516
Date summary prepared:
October 6, 1998
Device name:
Proprietary name: Common or usual name: Classification name:
C.T.M. Mobility Scooter HS-686 Electric scooter. Motorized three-wheeled vehicle, Class II, 21 CFR 890.3800.
Legally marketed device for substantial equivalence comparison: Shoprider The TE889 submitted by Pride Health Care, Inc. and cleared for marketing under 510(k) *K920654.
Description of the device:
The C.T.M. Mobility Scooter HS-686 is an indoor/outdoor four-wheeled scooter which is battery operated. It consists of a platform which connects the four wheels, an adjustable steering column, and a seat for the rider. The rider drives the scooter using hand controls located at the top of the steering column. It can be disassembled into five parts for transport in a car trunk. It comes with a battery charger.
Intended use of device:
The device is an indoor/outdoor scooter that provides transportation for a disabled or elderly person.
Technological characteristics:
The device features and use parameters of the C.T.M. Mobility Scooter HS-686 and Shoprider scooters are very similar. Both are battery operated, have motors less than 1 horsepower, and have automatic braking systems. Batteries and battery chargers are similar and are provided with the scooters. Use parameters are very similar, varying only in minor parameters such as the curb climbability of the respective scooters.
Testing conducted:
Tests listed in the Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs, and Motorized Three Wheeled Vehicles, July 1995, were conducted and the results included in the subject 510(k) submission.
Performance testing:
Comparative performance testing and clinical evaluations were not submitted as part of this 510(k).
{2}------------------------------------------------
# SECTION 3 - INTENDED USE
Intended use of device:
The C.T.M. Mobility Scooter HS-686 is an indoor/outdoor scooter that provides transportation for a disabled or elderly person.
Intended use of predicate device:
The Shoprider-TE889 is a motorized scooter designed to be used indoors and outdoors by an elderly or disabled person.
### Comparison:
The intended uses of the two products are the same. Each provides increased mobility for one individual, who is also the operator. Each device can be used in indoor and outdoor settings.
Labeling:
The intended use of the C.T.M. Mobility Scooter HS-686 is stated on page 4 of the Instructions for Use found in Appendix II. The intended use of the Shoprider can be inferred from the Instructions for Use and the promotional material. These can be found in Appendix IV.
{3}------------------------------------------------
C.T.M. Mobility Scooter HS-686 510(k) Notification Page 6
# SECTION 4 - DEVICE INFORMATION
## Executive Summary
The C.T.M. Mobility Scooter HS-686 is an indoor/outdoor four-wheeled scooter which is battery operated. It has a base which connects the four wheels, an adjustable steering column, and a seat for the rider. It is driven by the rider using hand controls located at the top of the steering column. It can be disassembled into five parts for transport. It is provided with a battery charger.
The C.T.M. Mobility Scooter HS-686 is a new device, not a modification of a previously cleared device. Numerous motorized scooters are currently on the market.
### Device Description
#### Scooter
The C.T.M. Mobility Scooter HS-686 is a battery-powered scooter with four wheels. Its design is similar to many other models on the market. Its main parts are a platform base which connects the front and rear wheels, an adjustable seat, and a steering column with control panel. It also has a detachable front basket. The scooter is designed for one person and can be used indoors or outdoors. Diagrams illustrating the scooter features are included in Appendix I. Diagram 1 shows a side view of the scooter with all major parts labeled. Diagram 2 shows front, rear, and side views of the C.T.M. Mobility Scooter HS-686 with major dimensions given.
The scooter can be disassembled for transportation into five components: the rear drive unit, the front platform unit which includes the steering column, the seat, and the two batteries. The heaviest part weighs 59.5 lbs. The whole scooter with the batteries weighs 196.7 lbs; without batteries 145.9 lbs. Assembly of the scooter does not require any tools. The wheel and tire assemblies are 11" in diameter.
The electrical cables run from the control panel, through the steering column, under the platform, and to the batteries. When assembling the scooter, electrical connections must be made between the front and rear units. Each battery must be connected to the electrical system. There is also a circuit breaker in the line.
All controls for the scooter are located at the top of the steering column and are operated with the hands. Diagram 3 in Appendix I shows the controls on the top of the steering column. Handlebars protrude from either side of the steering column. These keep the hands in the correct position to operate the controls and to steer. The whole steering column is turned in the direction that the operator wants to go. The minimum turning radius of the scooter is 162 cm. Just inside the handlebars are two thumb-control levers. Pushing these levers allows the scooter to move. The right lever moves the scooter forward; the left one moves it backward. As
{4}------------------------------------------------
soon as either thumb control lever is released, the brake engages. On the back of the steering column are the key hole and the charger plug-in. The battery charger plugs directly into the steering column to recharge the batteries.
On the control panel there are buttons for sounding the horn, turning on/off the headlight, and signaling turns. There is also a speed dial with a rabbit indicating higher speed at one end and a turtle representing lower speed at the other end. Maximum forward speed on a flat surface is 5.5 mph. Maximum speed in reverse is 2.2 mph. Turning the speed knob to the turtle picture reduces the speed to 2.2 mph in the forward direction and to 0 in the reverse direction. There is a power reserve indicator which shows the battery reserve. Below this light are red, yellow, and green color zones. When this light is completely lit, the batteries are fully charged. When the light is lit only in the red and yellow zones, the batteries should be recharged as soon as possible.
There is also a self-diagnostic warning light which blinks to indicate selected problems with the scooter. Generally speaking, the scooter can not be driven when the self-diagnostic light is flashing. The number of flashes helps determine the problem. When the light is making only one flash, indicating low batteries, the scooter can still be driven. The self-diagnostic signals are:
| Number of flashes | Problem |
|-------------------|--------------------------------|
| 1 | battery needs recharging |
| 2 | dead battery |
| 3 | battery overcharge |
| 4 | electrical short |
| 5 | brake problem |
| 6-9 | electrical system malfunctions |
Explanation of the problems and solutions are described in the Instructions for Use for the scooter, which can be found in Appendix II.
The braking system is electric. It is on the rear wheels only. The cables for this system run from the steering column to the rear unit. The brakes are automatically "on" except when one of the thumb-control levers is being pushed. Only under these conditions will the brakes disengage. Release of the thumb-control levers engages the brake. Braking time is less than 1 second and braking distance is less than 1 meter. If the electrical brake system fails, the brakes will default to the closed or "brakes on" position, thereby stopping the scooter. A diagram of the brakes is in Diagram 4 in Appendix I.
There is also a free-wheeling device on many scooters. On the C.T.M. Mobility Scooter HS-686 it is located near the base of the seat post. The normal position of the lever is "D". In this position the scooter will not move without turning on the
{5}------------------------------------------------
key and having charged batteries. Moving the lever to the position marked "N" allows the scooter to be rolled freely without the key being turned on. This feature is very useful when maneuvering the scooter for battery charging or storage. The scooter is not intended to be ridden at all in this free-wheeling mode.
## Batteries
The C.T.M. Mobility Scooter HS-686 runs on two 12 volt SLA (sealed lead acid) batteries. These batteries are provided with the scooter. Replacements can be obtained from the distributor or other local sources.
### Battery Charger
The battery charger is also provided with the scooter. It is a 24 volt constant current charger. It plugs into a standard 110 volt wall socket or into a 220 volt socket.
#### Accessories
Optional accessories include: a rear basket, a storage cover, a cane holder, and a mirror.
### Regulatory Status
The C.T.M. Mobility Scooter HS-686 is a new device which has not been previously submitted to the FDA.
{6}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing right, with flowing lines beneath them.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 18 1998
Mr. Robert S. McQuate R.S. McQuate & Associates, Inc. Representing Warepalmy Enterprise LLC (USA) 3636 East Columbine Drive Phoenix, Arizona 85032
Re : K983662 C.T.M. Mobility Scooter HS-686 Trade Name: K983663 C.T.M. Mobility Scooter HS-666 Trade Name: Regulatory Class: II Product Code: INI Dated: November 12, 1998 November 13, 1998 Received:
Dear Mr. McQuate:
We have reviewed your Section 510(k) notifications of intent to market the devices referenced above and we have determined these devices are substantially equivalent (for the indications for use stated in the enclosures) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), they may be subject to such additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ಗ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your devices in the Federal Reqister. Please note: this response to your premarket notification submission does
{7}------------------------------------------------
Page 2 - Mr. Robert S. McQuate
not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic enroagh eadiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in a classification for warrodevices and thus, permits your devices to proceed to the market.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
C. Collins McWhirter, Ph.D.
ia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{8}------------------------------------------------
| | CT M. Mobility Scooter HS-686 | |
|--|-------------------------------|--|
| | 510(k) Notification | |
| | Page 2 | |
.
# Indications for Use Statement
:
510(k) Number ( if known):
| 510(k) Number (if known): | |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name: | C.T.M. Mobility Scooter HS-686 |
| Indications for Use: | The C.T.M. Mobility Scooter HS-686 is an indoor/outdoor scooter which provides transportation for a disabled or elderly person. |
| (Please do not write below this line) | |
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
| Prescription Use<br>(Per 21 CFR 801.109) | OR |
| | Over-The-Counter Use <span style="text-decoration: overline;">X</span> |
| | <div> <img alt="Signature" src="data:image/png;base64,iVBORw0KGgoAAAANSUhEUgAAAAEAAAABCAQAAAC1HAwCAAAAC0lEQVR42mNkYAAAAAYAAjCB0C8AAAAASUVORK5CYII="/> </div> (Division Sign-Off) Division of General Restorative Devices |
| 510(k) Number | K98 3662 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.