K242628 · Lichtmega Technology Co., Ltd. · INI · Dec 20, 2024 · Physical Medicine
Device Facts
Record ID
K242628
Device Name
Scooter (PS30)
Applicant
Lichtmega Technology Co., Ltd.
Product Code
INI · Physical Medicine
Decision Date
Dec 20, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Indications for Use
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
The PS30 Scooter is a motor-driven, three-wheeled mobility vehicle for disabled or elderly users. It features an aluminum alloy frame, seat, tiller console, electric motor, electromagnetic brake, and a rechargeable lithium battery. The rider controls movement via a throttle lever and speed control dial on the tiller. The electromagnetic brake engages automatically when the device is not in use. Designed for operation on flat surfaces, the scooter provides mobility assistance to users in a seated position. It is intended for both indoor and outdoor use. The device does not involve complex software or AI; it functions as a standard motorized mobility aid.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including performance, stability, braking, energy consumption, and electromagnetic compatibility testing in accordance with ISO 7176 series standards and IEC 60601-1-2:2020.
Technological Characteristics
Aluminum alloy frame; electric motor; electromagnetic brake; rechargeable lithium battery; off-board charger. Complies with ISO 7176-1 through 7176-15, 7176-21, 7176-22, 7176-25, ISO 16840-10, and IEC 60601-1-2:2020. Mechanical control system via tiller-mounted throttle and speed dial.
Indications for Use
Indicated for disabled or elderly persons with limited mobility who require a seated position for indoor and outdoor transportation.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
K081980 — KYMCO FORU, MODEL EQ35 · Besteam Technology, Inc. · Aug 4, 2008
K083303 — LANDLEX, MODEL: 402X · Besteam Technology, Inc. · Jan 26, 2009
K253631 — Mobility Scooter (CL-Q3, CL-Q4) · Zhejiang Wei Ling New Energy Technology Co., Ltd. · Feb 23, 2026
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
December 20, 2024
Lichtmega Technology Co., Ltd. % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th Floor, Dongfang Building, 1500# Century Ave Shanghai. 200122 China
Re: K242628/S001
Trade/Device Name: Scooter (PS30) Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: December 3, 2024 Received: December 3, 2024
Dear Ivy Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Tushar Bansal -S
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
#### K242628
Device Name
Scooter (PS30)
Indications for Use (Describe)
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: K242628 | | | 510(k) Summary | | Prepared on: 2024-12-03 |
|------------------------------------|----------------------------------------------------------|-----------------------------------------------------------------------------|---------------------------------------------------------------------------|--|-------------------------|
| Contact Details | | | | | 21 CFR 807.92(a)(1) |
| Applicant Name | | Lichtmega Technology CO., Ltd. | | | |
| Applicant Address | | #101, No. 200, Cangjing Road, Songjiang District Shanghai Shanghai<br>China | | | |
| Applicant Contact Telephone | | +86-021-578986 | | | |
| Applicant Contact | | Mr. Hyman Zhu | | | |
| Applicant Contact Email | | Hyman@lichtmega.com | | | |
| Correspondent Name | | | Shanghai Sungo Management Consulting Company Limited | | |
| Correspondent Address | | | 14th Floor, Dongfang Building, 1500# Century Ave Shanghai 200122<br>China | | |
| Correspondent Contact Telephone | | | +86-58817802 | | |
| Correspondent Contact | | | Ms. Ivy Wang | | |
| Correspondent Contact Email | | | haiyu.wang@sungoglobal.com | | |
| Device Name | | | | | 21 CFR 807.92(a)(2) |
| Device Trade Name | | Scooter (PS30) | | | |
| Common Name | | Motorized three-wheeled vehicle | | | |
| Classification Name | | Vehicle, Motorized 3-Wheeled | | | |
| Regulation Number | | 890.3800 | | | |
| Product Code(s) | | INI | | | |
| Legally Marketed Predicate Devices | | | | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | | | Product Code |
| K182471 | Scooter | | | | INI |
| Device Description Summary | | | | | 21 CFR 807.92(a)(4) |
{5}------------------------------------------------
This scooter is a motor driven, indoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
The PS30 Scooter has a base with aluminum allov frame, two rear wheels, two anti-tip wheels, a seat, a tiller console, an electric motor, an electromagnetic brake, 1 rechargeable lithium Battery with an off-board charger. The movement of the scooter is controlled by the rider who operates the throttle lever, speed control dial and handle on the device is installed with an electromagnetic brake that will engage automatically when the scooter is not in use and the brake cannot be used manually. The Scooter only can be operated on the flat road.
The scooter is provided in four colors - cherry Red, Neptune Blue, Sunburst Orange, Frozen Grey Mat, for the outer shell of wheels and consoler panel.
## Intended Use/Indications for Use
lt is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
## Indications for Use Comparison
The proposed device has same indications for use to the predicate device.
# Technological Comparison
The design and technological characteristics of the proposed device scooter is identical to the predicate device. There are minor differences between the device is the dimension, frame material, travel distance, turning radius, ground clearance, controller, tire dimension, slope grade ability, maximum speed and loading weight. According to ISO7176 series standards and the test records support its safety and effectiveness. There is no deleterious effectiveness due to the minor differences do not influence the intended use of the device.
Therefore, the proposed Scooter is substantially equivalent (SE) to The Scooter (K182471).
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
## 21 CFR 807.92(a)(5)
# 21 CFR 807.92(a)(5)
# 21 CFR 807.92(a)(6)
{6}------------------------------------------------
Non-clinical tests were conducted to verify that the subject device met all design specifications and provided support of the substantial equivalence determination. The test results demonstrated that the proposed device complies with the following standards: ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes ISO 7176-42008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatique strength ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs ISO 7176-11:2012 Wheelchairs -- Part 11: Test dummies ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces. ISO 7176-14:2022 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling. ISO 16840-10:2021 Wheelchair seating - Part 10: Resistance to ignition of postural support devices – Requirements and test method ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters ISO 7176-22: 2014 Wheelchairs - Part 22: Set-up procedures ISO 7176-25:2013 Wheelchairs - Batteries and chargers for powered wheelchairs Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2020.
The clinical study is not applicable.
The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device K182471.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.