K222729 · Wu’S Tech Co., Ltd. · INI · Mar 28, 2023 · Physical Medicine
Device Facts
Record ID
K222729
Device Name
Electrical Scooter, WT-T4QP2
Applicant
Wu’S Tech Co., Ltd.
Product Code
INI · Physical Medicine
Decision Date
Mar 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Indications for Use
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Device Story
WT-T4QP2 is a lithium-ion battery-powered, four-wheeled electric scooter for indoor/outdoor use. User operates via turning tiller for steering and control panel for speed and forward/reverse selection. Releasing speed lever automatically engages electromagnetic motor brake. Device provides mobility for individuals restricted to sitting. Performance is limited to 4 mph max speed, 6-degree incline, and 250 lbs weight capacity. User must observe labeling regarding ground clearance and curb climbing limits. Benefits include increased mobility for users with restricted sitting capabilities.
Clinical Evidence
Bench testing only. No clinical data provided. Device performance validated against ISO 7176 series (stability, brakes, energy consumption, dimensions, speed, seating, strength, climatic, obstacle climbing, ignition resistance) and RESNA WC-2 (power/control systems, electromagnetic compatibility). Biocompatibility testing (cytotoxicity, sensitization, irritation) performed per ISO 10993-1:2018.
Technological Characteristics
Four-wheeled electric scooter; aluminum alloy frame; PU solid tires; lithium-ion battery (25.2 VDC/10.35 Ah); electromagnetic braking; Dynamic DR50 electronic controller. Materials: TPV (handle bar grip), PU (seat), PVC (speed lever). Standards: ISO 7176 series, RESNA WC-2:2019, ISO 10993-1:2018.
Indications for Use
Indicated for persons restricted to a sitting position requiring mobility assistance. No specific age or gender limitations provided.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 28, 2023
Wu's Tech Co., Ltd. % Dr. Jen Ke-Min Contact Person Chinese European Industrial Research Society No. 58. Fu-Chiun St Hsin-Chu City, Taiwan 300113 Taiwan
Re: K222729
Trade/Device Name: Electrical Scooter, WT-T4QP2 Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: February 17, 2023 Received: February 27, 2023
## Dear Dr. Jen Ke-Min:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Tushar Bansal -S
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222729
Device Name Electric Scooter, WT-T4QP2
Indications for Use (Describe)
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) number: K222729
# 510(k) Summary (per 21 CFR 807.92)
| Submitter's Name: | Wu's Tech Co., Ltd.<br>No.225, Yuan Peir St., Hsinchu, Taiwan ROC 30093 |
|------------------------|-------------------------------------------------------------------------|
| Date summary prepared: | August 28, 2022 |
| Proprietary Name: | Electrical Scooter, model WT-T4QP2 |
| Common or Usual Name: | Electrical Scooter |
| Classification Name | Motorized Three-Wheeled Vehicle,<br>Class II, 21 CFR 890.3800 |
| Product Code: | INI |
| Contact Person | Dr. KE-MIN JEN (email: ceirs.jen@msa.hinet.net)<br>TEL: 886-3-5208829 |
| Predicate Device | PUC Perfect Union Co., Ltd. |
| Device name | i3 Foldable Mobility Scooter |
| 510(k) Number | K191356 |
#### ● Indications for Use:
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
#### ● Device Description:
WT-T4QP2 is for indoor or outdoor uses, and it is Lithium-ion battery powered and configured with four solid wheels, a seat, a turning tiller, and a control panel.
WT-T4QP2 is driven by two PU solid rear wheels (7.28" x 1.96") and steered by two PU solid castors (6.49" x 1.96"). The user can control the turning tiller to control the castors to steer the scooter's direction. Control panel has 3 functions of battery level indication, speed lever and forward/reverse selector. The user can pull the speed lever to drive
{4}------------------------------------------------
forward or drive backward. Release the speed lever completely and the electromagnetic brake in the motor will be activated automatically, and the scooter will stop.
The maximum weight capacity of WT-T4QP2 is 250 lbs. (115 kg). Its maximum forward speed is 4 mph (6.4 km/h), and its maximum safe climbing incline angle is 6 degrees.
The following surfaces are recommended NOT to operate on:
- � Sand surface
- Wet or icy surface �
- Road maintenance hole metal cover �
- � Avoid going up multiple steps.
- � Avoid using escalators. Use the elevator.
- Too steep incline over 6 degrees.
- Ground clearance to battery over 1.1"/28 mm
- Curb climbing ability over 1.58"/40 mm
#### Performance Testing: ●
- ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability, 1999.
- ISO 7176-2:2017 Wheelchairs Part 2: Determination of dynamic stability of ● electric wheelchairs. .
- ISO 7176-3: 2012 Wheelchairs - Part 3: Determination of effectiveness of brakes.
- ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs for determination of theoretical distance range,
- ISO 7176-5:2008 Wheelchairs - Part 5: Determination of overall dimensions, mass and maneuvering space,
- ISO 7176-6:2001 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs,
- ISO 7176-7:1998 Wheelchairs - Part 7: Determination of seating dimensions -Definitions and measuring method,
- ISO 7176-8:1998 Wheelchairs - Part 8: Static, impact and fatigue strength for manual wheelchairs.
- ISO 7176-9:2009 Wheelchairs Part 9: Climatic tests for electric wheelchairs, ●
{5}------------------------------------------------
- ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs,
- ISO 7176-16: 2012 Wheelchairs – Part 16: Resistance to ignition of postural support devices
- RESNA WC-2:2019 Section 14 Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods
- RESNA WC-2:2019 Section 21 Requirements and test method electromagnetic compatibility of powered wheelchairs and motorized scooters.
| Patient-contacting<br>Parts | Materials | Rational according<br>to<br>Wu's Tech expertise | Tests conducted | Verdict |
|-----------------------------|----------------------------|-------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|---------|
| Handle bar grip | DIOSHY<br>TPV | No customer<br>claim with<br>adverse<br>conditions for<br>more than 300<br>thousand units in<br>the market. | Cytotoxicity,<br>Sensitization<br>and Irritation<br>tests | Pass |
| Seat | PU | No customer<br>claim with<br>adverse<br>conditions for<br>more than 300<br>thousand units in<br>the market. | Cytotoxicity,<br>Sensitization<br>and Irritation<br>tests | Pass |
| Speed lever | PVC<br>(NAN YA<br>FR-PA66) | No customer<br>claim with<br>adverse<br>conditions for<br>more than 250<br>thousand units in<br>the market. | Cytotoxicity,<br>Sensitization<br>and Irritation<br>tests | Pass |
#### Biocompatibility evaluation per ISO 10993-1:2018 ●
{6}------------------------------------------------
#### Comparison table ●
| | Devices<br>Specifications | Subject device | Predicate device | Comparison<br>verdict |
|-------------------------------------|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|-----------------------|
| | Manufacturer | Wu's Tech Co., Ltd. | PUC Perfect Union Co.,<br>Ltd. | -- |
| | Proprietary name | Electrical Scooter | Foldable Mobility<br>Scooter | -- |
| | Model name | WT-T4QP2 | i3 | -- |
| | 510(k) No. | TBA | K191356 | -- |
| | Indication for use | The device is intended<br>for medical purposes<br>to provide mobility to<br>persons restricted to a<br>sitting position. | The device is intended<br>for medical purposes to<br>provide mobility to<br>persons restricted to a<br>sitting position. | Same |
| | Overall Dimensions<br>(LxWxH) | 36.2" x 16.9" x 35.8" | 39.9" x 18.4"x 38.6" | Similar |
| | Manual folding | Yes | Yes | Same |
| | Operating<br>environments | Indoor/ outdoor | Indoor/outdoor | Same |
| | Number of Wheels | 4 | 4 | Same |
| | Frame | Fixed / Aluminum<br>alloy | Fixed / Aluminum alloy | Same |
| | Front wheel size &<br>type | 6.49"x 1.96"x2<br>(PU solid tire) | 7" x 2"x 2<br>(PU Solid tire) | Similar |
| | Rear wheel size & type | 7.28"x 1.96"x2<br>(PU solid tire ) | 8" x 2"x2<br>(PU solid tire) | Similar |
| | Battery | One Lithium-ion<br>battery pack<br>25.2 VDC /10.35 Ah | Two Lithium-ion<br>batteries<br>25.4 DCV/ 11.6 Ah | Similar |
| | Off-board charger | Off-board charger | Same | |
| Charger type & rating | Input AC100-240V,<br>Output DC 29.4V, 2A | Input: AC100-240V<br>Output: DC 29.4 V | | |
| Battery weight | 4.4 lbs. (1.98 kg) | 3.8 lbs. (1.72 kg) | Similar | |
| Armrest | No | No | Same | |
| Motor | DC24V / 180W | DC 24V /270W | Similar | |
| Driving System | Rear wheel drive | Rear wheel drive | Same | |
| Brake System | Electromagnetic brake | Electromagnetic brake | Same | |
| Anti-tipper | Yes | Yes | Same | |
| Max Forward Speed | 4 mph /6.4 km/h | 3.7 mph /6.0 km/h | Similar | |
| Biocompatibility | ISO 10993-1:2018<br>ISO 10993-5:2009<br>ISO 10993-10:2010 | ISO 10993-1:2009<br>ISO 10993-5:2009<br>ISO 10993-10:2010 | Same | |
| Height adjustable tiller | No | Yes | Different | |
| Back upholstery | PU Foam | Fabric | Different | |
| Maximum safe<br>operational incline | 6 degrees | 8 degrees | Different | |
| Weight, w/ Battery | 45.8 lbs. (20.8 kg) | 59.5 lbs. (26.99 kg) | Different | |
| w/o Battery | 41.4 lbs.(18.82 kg) | 55.6 lbs. (25.24 kg) | Different | |
| Cruising Range | 9.06 miles/14.5 km | 5.67 miles /9.13 km | Different | |
| Max Loading | 250 lbs./115 kg | 300 lbs. / 137 kg | Different | |
| Turning Radius | 55.9" / 1420 mm | 59" / 1500 mm | Different | |
| Electronics controller | Dynamic DR50 | Penny & Giles S-Drive | Different | |
| Ground clearance | 1.1"/ 28 mm | 1.7"/ 43 mm | Different | |
| Curb climbing | 1.58"/ 40 mm | 2"/50 mm | Different | |
{7}------------------------------------------------
{8}------------------------------------------------
#### ● Substantial equivalent discussion
The same or similar items will not be discussed and only the different items will be discussed.
The height of turning tiller for the subject device cannot be adjusted and it can be adjusted for the predicate device. A height-adjustable turning tiller means the user can change the height of turning tiller according to the sizes of different users, thus bring more convenience to the users. A height adjustable tiller also has a risk of loose of fixing knob, and this harm will lead to serious injury of users in some cases. Though, a fixed-height tiller has no this kind of convenience, it can keep safety well. There are no any new safety or effectiveness concerns raised by the subject device due to a fixed height tiller.
The material of back upholstery for the subject device is PU Foam and it is Fabric for the predicate device. The PU Foam material has soft feeling to users, and fabric material has some kinds of hardness feeling to users. The safety concerns for back upholstery focus on the ignition resistance testing, and the subject device passed the ISO 7176-16:2012 Wheelchairs – Part 16: Resistance to ignition of postural support devices. Thus, there are no any new safety or effectiveness concerns raised by the subject device for the different back upholstery materials used.
The Maximum safe operational incline for the subject device is 6 degrees and it is 8 degrees for the predicate device. The difference of 2 degrees incline means the predicate device can climb steeper incline than the subject device. But we have informed the users of not to operate the subject device on the incline higher than 6 degrees in the labeling. So, there are no any new safety and effectiveness concerns raised by subject device for a 6-degree maximum safe operational incline.
The Scooter weight with battery is 45.8 lbs. for subject device and it is 59.5 lbs. for predicate device. The predicate device is heavier than the subject device by 13.7 lbs. The 13.7 lbs. difference mainly comes from the larger overall dimensions of frame for predicate device. Larger scooter weight will lead to a smaller cruising range according to the principle of conservation of energy. The difference between these two scooters' weight will not raise any new safety and effectiveness concerns for the subject device.
{9}------------------------------------------------
The Cruising range is 9.06 miles for the subject device and it is 5.67 miles for the predicate device. The difference of cruising range just comes from different designs of the scooters, not related to safety or effectiveness concerns. Moreover, we provide the 9.06 miles cruising range of WT-T4QP2 in the relevant labeling to inform the users of this cruising range. Normally, scooter with larger cruising range leads to lower frequency of charging battery, thus bring about more convenience to the user. There are no any new safety and effectiveness concerns raised by the subject device due to the larger cruising range.
The Max loading is 250 pounds for the subject device and it is 300 pounds for the predicate device. We provide the max loading data in the WT-T4QP2 labeling to inform users of this data. The users have the duty of noticing and observing the labeling, thus, there are no any new safety and effectiveness concerns raised by the subject device due to the different max loading.
The Turning radius is 55.9 inches for subject device and it is 59 inches for the predicate device. The larger turning radius will have more limitations of turning around space to the scooter with larger radius, i.e., predicate device. So, there are no any new safety and effectiveness concerns raised by the subject device.
The Electronics controller is Dynamic DR50 for subject device and it is PG S-Drive for predicate device. The subject device which contains the electronic controller DR50 together passes the RESNA WC-2:2019 Section 14, thus safety of electronic controller is ensured, and there are no any new safety and effectiveness concerns raised by the subject device due to using different electronic controller.
The Ground clearance is 1.1 inches for subject device and it is 1.7 inches for predicate device. The users should notice this data in labeling when operate the scooter to pass the obstacle, otherwise there is damage to the chassis of scooter. If the users observe the labeling, there are no any new safety and effectiveness concerns raised by the subject device due to the smaller ground clearance.
The Curb climbing is 1.58 inches for the subject device and it is 2 inches for predicate device. The users should notice this data in labeling when operate the scooter to meet the curb, otherwise the scooter will not pass over the curb and will have impact on the scooter. If the users observe the curb labeling, there are no any new safety and effectiveness concerns raised by the subject device due to the smaller climbing curb.
{10}------------------------------------------------
#### ● Conclusions:
The subject device, WT-T4QP2, is as safe and effective as, and functions in a manner equivalent to the predicate device - i3 Foldable Mobility Scooter. The conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device identified in the submission. Thus, the subject device is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.