← Product Code [INI](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/INI) · K070817

# LERADOTECH SCOOTER, MODEL SB (K070817)

_Lerado China Limited · INI · Apr 13, 2007 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/INI/K070817

## Device Facts

- **Applicant:** Lerado China Limited
- **Product Code:** [INI](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/INI.md)
- **Decision Date:** Apr 13, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3800
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine

## Indications for Use

The device is intended for medical purposes to provide mobility to persons restricted to a seated position.

## Device Story

LERADOTECH Scooter, SB is battery-operated indoor/outdoor motorized 3-wheeled vehicle; provides mobility for individuals restricted to seated position. Rider operates device via hand controls on steering column. Device features base with seat; disassembles for transport; includes onboard battery charger. Electronic systems, controller, and switching power supply match predicate. Operator adjusts speed via throttle tiller. Device offers agile dimensions, smaller turning radius, and safe climbing angle compared to predicate. Intended for use by individuals needing mobility support; facilitates independent movement.

## Clinical Evidence

Bench testing only. EMC testing performed per ANSI/RESNA WC/Vol.2-1998, CISPR 11:1990, EN61000-3-2:1995, and IEC61000-3-3:1995. Upholstery passed resistance ignition testing by SGS.

## Technological Characteristics

Battery-operated motorized 3-wheeled vehicle. Features hand-controlled steering column, seat, and onboard charger. Electronic systems, controller, and switching power supply are UL certified. Back upholstery material tested for resistance ignition. Dimensions and weight capacity optimized for agility. Speed continuously adjustable via throttle tiller.

## Regulatory Identification

A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.

## Predicate Devices

- COMFORT Scooter LY-EW415 ([K063032](/device/K063032.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K070817

APR 1 3 2007

Image /page/0/Picture/1 description: The image shows a logo with the word "LERADO" in large, bold letters. A stylized image of a tree is placed between the letters "R" and "D". Below the word "LERADO" is the text "LERADO GROUP" in a smaller font. Below that is some text in a different language.

LERADO CHINA LIMITED

No. 22, Kuang Fu Road, Chia Tai Industrial Tai Pao City, Chia Yi Hsien, Tahwan, ROC. Tel: 886-5-2475629 Fax: 886-5-2379672 E-mail:edwardlaglerado.com tw http:// www.lerado.com

## 510(k) SUMMARY " દ્ધ

Submitter's Name: LERADO CHINA LIMITED

Dong Sheng Road North Zhong Shan, CHINA

Date summary prepared:

Device Name:

March 17, 2007

Proprietary Name: Common or Usual Name: Classification Name:

Electrical Scooter Motorized 3-Wheeled Vehicle, Class II, 21 CFR 890.3800

Indications for Use:

The device is intended for medical purposes to provide mobility to persons restricted to a seated position.

LERADOTECH Scooter, SB

Description of the device:

The LERADOTECH Scooter, SB is an indoor / outdoor Electrical Scooter that is battery operated. It has a base with four-wheeled with a seat. The movement of the scooter is controlled by the rider who uses hand controls located at the top of the The device can be disassembled for transport and is provided with steering column. an onboard battery charger.

Performance Testing:

EMC Report ANSI / RESNA WC/Vol.2-1998, CISPR 11: 1990, EN61000-3-2: 1995, IEC61000-3-3: 1995 (Electrically Electrical Scooters, controller, and their chargers requirements and test methods)

Legally marketed device for substantial equivalence comparison: COMFORT Scooter LY-EW415 (K063032)

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for LERADO Group. The logo is in black and white and features the company name in a stylized font. Above the company name is a triangular shape that resembles a mountain. Below the company name are some Chinese characters.

LERADO CHINA LIMITED No. 22, Kuang Fu Road, Chia Tai Industrial Tai Pao City, Chia Yi Hsien, Taiwan, ROC Tel: 886-3-2475629 · Fax: 886-5-2379672 E-mail edwardl@lerado.com tw http:// www.lerado.com

Summary for substantial equivalence comparison:

The new device SB and the predicate device LY-EW415 are the same intended uses and warranty period, and the electronic systems between the two devices are all passed by the UL certificated. Except the recharger of the electronic systems, the two devices are the same suppliers. Especially, the two devices use the same controller and switching power supplier. Besides, the back upholstery is the same material, and also passed the resistance ignition test by SGS. Thus the same safety level for the two devices is assured.

The major difference existing for the two devices are tires size, weight limit, weight capabilities, and the dimensions for the new device are more agile and smaller than those of the predicate device. Thus the new device can drive smaller turning radius and safe climbing angle. The overall appearance differences are not safety aspect. So the new device is substantially equivalent to the predicate devices in this aspect.

The maximum speed is 6.0 mph for the new device and 5.0 mph for the predicate device that are also under the 6 miles maximum speed limited. The speeds can be continuous adjusted by the throttle tiller. The operators can set the adequate speed according to their feeling and need, and the maximum speed differences do not mean any performance differences. They are substantially equivalent.

The cruising range of the new device is 36 km and 75 km for the predicate device. This is mainly due to the fact that the batteries for the new device are smaller. Certainly the real range depends on the practice environments, i.e., weight, surface, incline, and temperature. For the real life use, the two devices are substantially equivalent.

Based on the above the information and the analysis, we know that the subject device and the predicate devices have the same intended use, the same technological aspects and only minor dimensions or data differences exist. Thus the subject device and the predicate device are substantially equivalent.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Lerado China Limited % Republic of China Chinese-European Industrial Research Society Dr. Ke-Min Jen Official Correspondent No. 58, Fu-Chiun Street Hsin-Chu City China (Taiwan) 30067

APR 1 3 2007

Re: K070817

Trade/Device Name: LERADOTECH Scotter, SB Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized three-wheeled vehicle Regulatory Class: Class II Product Code: INI Dated: March 17, 2007 Received: March 26, 2007

Dear Dr. Jen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

## Page 2 – Dr. Ke-Min Jen

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510 (K) Number ( If Known ): __ K

Device Name: LERADOTECH SCOOTER, SB

Indications for Use:

The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.

(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number 16070817

Prescription Use

(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/INI/K070817](https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/INI/K070817)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
