Disposable Concentric Needle Electrode 0.30x25mm Red (TE/B50700-001); Disposable Concentric Needle Electrode 0.40x25mm Yellow (TE/B50700-002); Disposable Concentric Needle Electrode 0.30x30mm Pink (TE/B50700-006); Disposable Concentric Needle Electrode 0.45x37mm Green (TE/B50700-003); Disposable Concentric Needle Electrode 0.45x50mm Blue (TE/B50700-004); Disposable Concentric Needle Electrode 0.60x75mm Violet (TE/B50700-005)

K253581 · Technomed Europe · IKT · Mar 15, 2026 · Physical Medicine

Device Facts

Record IDK253581
Device NameDisposable Concentric Needle Electrode 0.30x25mm Red (TE/B50700-001); Disposable Concentric Needle Electrode 0.40x25mm Yellow (TE/B50700-002); Disposable Concentric Needle Electrode 0.30x30mm Pink (TE/B50700-006); Disposable Concentric Needle Electrode 0.45x37mm Green (TE/B50700-003); Disposable Concentric Needle Electrode 0.45x50mm Blue (TE/B50700-004); Disposable Concentric Needle Electrode 0.60x75mm Violet (TE/B50700-005)
ApplicantTechnomed Europe
Product CodeIKT · Physical Medicine
Decision DateMar 15, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.1385
Device ClassClass 2

Indications for Use

Disposable Concentric Needle Electrodes are intended for use with recording, monitoring equipment for the recording of biopotential signals including electroencephalography (EEG), electromyography (EMG) and nerve potential signal.

Device Story

Disposable Concentric Needle Electrode (DCNE) is an invasive, single-use recording electrode; interfaces between patient muscle tissue and EMG/EEG monitoring equipment. Device consists of a needle cannula (reference electrode) and an active core wire (active electrode) insulated from each other. During clinical neurophysiological testing, the needle is inserted into muscle tissue; it detects weak bio-electric signals. Signals are transmitted to a differential amplifier via a 5-pole DIN connector; the amplifier subtracts the reference signal from the active signal to eliminate noise. Output is a recorded biopotential signal used by clinicians for diagnostic assessment. Benefits include targeted muscle activity recording, reduced infection risk due to single-use design, and minimal patient discomfort. Used in highly controlled clinical environments by certified electrophysiology professionals.

Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including visual inspection (20x magnification), puncture force/friction testing, electrical impedance (capacitance/resistance), and mechanical push/pull testing of the needle-to-connector interface. Biocompatibility was demonstrated per ISO 10993-1, including cytotoxicity, sensitization, intracutaneous reactivity, and acute systemic toxicity testing.

Technological Characteristics

Invasive needle electrode; cannula and core wire construction. Materials: Stainless steel AISI304 (cannula), Tungsten (core wire), Polyamide PA6 (hub), Polyesterimide (insulation). Energy: Passive signal conduction. Connectivity: 5-pole DIN connector. Sterilization: Ethylene oxide (SAL 10^-6). Tip geometry: Trocar, 15° angle. Dimensions: 0.30-0.60mm diameter, 25-75mm length.

Indications for Use

Indicated for patients requiring clinical neurophysiological testing, including EEG, EMG, and nerve potential signal recording. Used by trained healthcare professionals.

Regulatory Classification

Identification

A diagnostic electromyograph needle electrode is a monopolar or bipolar needle intended to be inserted into muscle or nerve tissue to sense bioelectrical signals. The device is intended for medical purposes for use in connection with electromyography (recording the intrinsic electrical properties of skeletal muscle).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION March 15, 2026 Technomed Europe Alan Yik Senior Regulatory Affairs Specialist Wiebachstraat 25a Kerkrade, 6466 NG Netherlands Re: K253581 Trade/Device Name: Disposable Concentric Needle Electrode 0.30x25mm Red (TE/B50700-001); Disposable Concentric Needle Electrode 0.40x25mm Yellow (TE/B50700-002); Disposable Concentric Needle Electrode 0.30x30mm Pink (TE/B50700-006); Disposable Concentric Needle Electrode 0.45x37mm Green (TE/B50700-003); Disposable Concentric Needle Electrode 0.45x50mm Blue (TE/B50700-004); Disposable Concentric Needle Electrode 0.60x75mm Violet (TE/B50700-005) Regulation Number: 21 CFR 890.1385 Regulation Name: Diagnostic Electromyograph Needle Electrode Regulatory Class: Class II Product Code: IKT, GXZ Dated: November 17, 2025 Received: November 17, 2025 Dear Alan Yik: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. {1} K253581 - Alan Yik Page 2 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. {2} K253581 - Alan Yik Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Patrick Antkowiak -S for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253581 | ? | | Please provide the device trade name(s). | | ? | | Disposable Concentric Needle Electrode 0.30x25mm Red (TE/B50700-001); Disposable Concentric Needle Electrode 0.40x25mm Yellow (TE/B50700-002); Disposable Concentric Needle Electrode 0.30x30mm Pink (TE/B50700-006); Disposable Concentric Needle Electrode 0.45x37mm Green (TE/B50700-003); Disposable Concentric Needle Electrode 0.45x50mm Blue (TE/B50700-004); Disposable Concentric Needle Electrode 0.60x75mm Violet (TE/B50700-005) | | | | Please provide your Indications for Use below. | | ? | | Disposable Concentric Needle Electrodes are intended for use with recording, monitoring equipment for the recording of biopotential signals including electroencephalograph (EEG), electromyography (EMG) and nerve potential signal. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} Technomed medical accessories Taking you further. Step by step. # Disposable Concentric Needle Electrode (DCNE) 510(k) # 510(k) Summary # K253581 A. Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Technomed Europe Wiebachstraat 25a, Kerkrade, 6466 NG, Netherlands Phone: +44 (0) 1994 240798 Contact Person: Alan Yik, Senior Regulatory Affairs Specialist (ayik@welcony.com) Date Prepared: 22nd Jan 2026 B. Device Name | Device Trade Name | Disposable Concentric Needle Electrode 0.30x25mm Red (TE/B50700-001); | | --- | --- | | | Disposable Concentric Needle Electrode 0.40x25mm Yellow (TE/B50700-002); | | | Disposable Concentric Needle Electrode 0.30x30mm Pink (TE/B50700-006); | | | Disposable Concentric Needle Electrode 0.45x37mm Green (TE/B50700-003); | | | Disposable Concentric Needle Electrode 0.45x50mm Blue (TE/B50700-004); | | | Disposable Concentric Needle Electrode 0.60x75mm Violet (TE/B50700-005) | Common Name Needle Electrode Classification Name Electrode, Needle Regulation Number 890.1385 Product Code(s) IKT, GXZ C. Legally Marketed Predicate Devices Primary Predicate Device: Technomed EEG / EMG needle electrodes (K990015) Product Code: 21 CFR § 882.1350 GXZ Technomed Europe Wiebachstraat 25a 6466 NG Kerkrade The Netherlands T +31 (43) 408 6868 infotechnomed@welcony.com www.technomedmedical.com Chamber of Commerce Reg.nr. 14049781 IBAN nr.: NL80ABNA0441485987 Vat-nr.: NL803379936B01 {5} Technomed medical accessories Taking you further. Step by step. # Disposable Concentric Needle Electrode (DCNE) 510(k) ## Reference Devices 1. Disposable hypodermic EMG needle electrode (K062437) Product Code: 21 CFR § 890.1385 IKT 2. Disposable Monopolar and Subdermal Needle Electrodes (K130136) Product Code: 21 CFR § 890.1385 IKT Subsequent Product Code: 21 CFR § 882.1350 GXZ Product Code: 21 CFR § 890.1385 IKT ## D. Device Description Summary The Disposable Concentric Needle Electrode (DCNE) is a recording electrode used for recording of bio-electric, electromyography (EMG) or electroencephalography (EEG) signals in patients undergoing clinical neurophysiological testing. It is inserted into muscle tissue of patient, acting as an interface between the EMG equipment and the patient. It is connected to the EMG cable (not in this submission), which is connected to the EMG equipment to detect bio-electric signals distally. The subject devices are invasive, single-use and supplied sterile. It is used by healthcare professionals, specifically trained and certified in electrophysiology and recording/monitoring techniques. The subject devices consist of 6 variants. The differences between the variants are the diameter, length of the needle and the color of the handle. ## E. Intended Use/ Indications for Use Disposable Concentric Needle Electrodes are intended for use with recording, monitoring equipment for the recording of biopotential signals including electroencephalography (EEG), electromyography (EMG) and nerve potential signal. ## F. Technological Comparison ### Technological Characteristics Technomed Europe Wiebachstraat 25a 6466 NG Kerkrade The Netherlands T +31 (43) 408 6868 infotechnomed@welcony.com www.technomedmedical.com Chamber of Commerce Reg.nr. 14049781 IBAN nr.: NL80ABNA0441485987 Vat-nr.: NL803379936B01 {6} Technomed media taking you further. Step by step. # Disposable Concentric Needle Electrode (DCNE) 510(k) Technologically the subject device DCNE is identical to the predicate devices as covered by K990015. Modifications have been made to the joining method from needle to the connector cage, changing from a soldering connection between needle and connector cage to a crimped connection which is glued to the hub. This hub also has a different inner shape compared to the predicate device. Also, the plating finish of the brass connector (Tin vs gold (predicate device) and the core wire material are changed from stainless steel to tungsten to achieve a better performance. Finally, the tip geometry has been modified by adjusting the grinding angles to improve puncturing and sliding behavior. As demonstrated by the non-clinical testing submitted with this submission, the proposed modification does not raise new or different questions of safety or effectiveness. Non-clinical test was performed and attached in this submission. The basic design of the subject device DCNE in this submission is very similar to the design of the predicate device as covered by K990015. Both devices are the interface between the diagnostic or monitoring device and the patient. They transmit electrical signals from the human body to the monitoring device. Whilst the joining method of the needle to the connector cage, the core wire material, finished plating of the connector, tip geometry and internal shape of the hub are different, the functionality of the product to transmit electrical signals in both the predicate and the subject device is unchanged. The same crimping system is used in the reference device DHNE which its functionality is proven to be safe and effective in the previous submission K062437. Refer to Attachment 3.07.a - Transit and Aging Rationale for the detail comparison. In addition, there is new materials used in the subject devices which has been proven safe by compare with the reference device DMNE cleared under K130136. Refer to Attachment 3.05.06 - Biocompatibility for detail for the demonstration of biocompatibility of subject device. To evaluate the difference between the two products, detailed information can be found in the Substantial Equivalence Summary Table (See Table 1). In summary, the following can be concluded: - Compatibility and connections: Both the predicate and subject device can be connected to a reusable cable connected with a 5-pole DIN connector; and therefore, are compatible with any intraoperative neuromonitoring equipment that is equipped with 5-pole DIN connector. - Sizes: Hub diameter and length, Needle length and diameter are identical - Tip geometry: The tip geometry is change due to a change in the grinding angles to improve puncturing and sliding behavior Technomed Europe Wiebachstraat 25a 6466 NG Kerkrade The Netherlands T +31 (43) 408 6868 infotechnomed@welcony.com www.technomedmedical.com Chamber of Commerce Reg.nr. 14049781 IBAN nr.: NL80ABNA0441485987 Vat-nr.: NL803379936B01 {7} Technomed media taking you further. Step by step. # Disposable Concentric Needle Electrode (DCNE) 510(k) - Materials: Materials used in the product are identical for the outside needle and the hub. The differences are the internal joining method of the needle to the connector cage, changing from solder to crimp bushing with glue; the connector plating, which does not have patient contact; and the core wire material, which has changed from stainless steel to tungsten. - Other information: sterility and sterilization, shelf life and packaging are identical for the predicate and subject device. For the predicate device, tensile strengths between needle and hub were not specified in the original submission (K990015), however have been compared in the non-clinical performance testing. Both the predicate device and the device that is subject of this submission conform to the set specifications and testing requirements. Thus, the subject device DCNE is substantial equivalence to the predicate devices in terms of the technological principles. # Principles of Operations The subject device Disposable Concentric Needle Electrodes (DCNE) has the same principles of operation as with the predicate device cleared under K990015. Both the subject and predicate devices need to be inserted into muscle tissue. They are the interface between the EMG equipment and the patient. They transmit signals from the human body to the EMG equipment. These signals are extremely weak. Therefore, amplification and elimination of unwanted noise is essential. Since the human body acts like a radio antenna, it is picking up interference from broadcast radio transmissions or even from the lighting and electrical wiring in the laboratory. Special amplifiers, called differential amplifiers, are used to help eliminate the unwanted interference signals and to provide the high gain necessary (up to 1V) before the signal can be adequately recorded. This is usually obtained from a surface electrode on a body part away from the muscle under investigation, and away from any contracting musculature (especially the heart), i.e., a so-called reference electrode. This electrode will pick up the unwanted interference signal. The differential amplifier operates by subtracting the reference signal from the EMG / EEG electrode signal. This leaves only the small EMG / EEG signal, which goes on to be amplified. Unwanted noise is thus eliminated. The cannula of the Disposable Concentric Needle Electrode, located at the tip of the needle, acts as its reference electrode. Its active core acts as its active electrode. The active core is embedded in an insulating material to electrically insulate the two electrodes from each other. Both the subject devices and the predicate devices have the following advantages: - Precision: Allows for targeted muscle activity recording. Technomed Europe Wiebachstraat 25a 6466 NG Kerkrade The Netherlands T +31 (43) 408 6868 infotechnomed@welcony.com www.technomedmedical.com Chamber of Commerce Reg.nr. 14049781 IBAN nr.: NL80ABNA0441485987 Vat-nr.: NL803379936B01 {8} Technomed media taking you further. Step by step. # Disposable Concentric Needle Electrode (DCNE) 510(k) - Safety: Disposable design reduces the risk of infection. - Minimal Invasiveness: The needle is small and sharp, causing minimal discomfort. In conclusion, the principles of operation for the subject device and the predicate device are identical. ## G. Non-Clinical and/or Clinical Test Summary &amp; Conclusions Performance testing of the Disposable Concentric Needle Electrode (DCNE) demonstrates that the device performs as intended for and is therefore substantially equivalent to the primary predicate device. The following nonclinical testing within this submission is relied upon for a determination of substantial equivalence: ## Visual inspection The visual inspection was carried out using a microscope (20x magnification) to examine the integrity of subject devices, ensuring that the needle is not damaged, and the needle tip is free of burr and is not blunt. Additionally, the alignment of the indicator of the hub, the color of the hub, and the recording area of the needle tip was verified. Furthermore, the back side connector was inspected to be clean and not damaged. The straightness of the needle, the length and diameter of the needle were also verified. ## Puncture test The puncture test was executed on needles using the tensile testing machine. The needle was fixed in the upper compartment of the tensile testing machine, and the PU foil (simulating the skin) is placed in the lower compartment, on a holder that features a 10mm diameter circle area where the PU foil is exposed. The puncture test is executed at a constant velocity established at 50mm/min. The tested needle was inserted into the PU foil for a puncture depth of 15mm. During this test, the puncture force and the friction force values were examined and reported. The puncture force represents the force at the end of the cutting phase, while the friction force represents the force exerted by sliding the cannula through the PU foil. ## Impedance The electrical test was performed on needles, where the LCR meter is used to measure the capacitance (C, [nF]), and the resistance (R, [kΩ]) of the subject devices. The tested needle was placed in a container with physiological solution (0.9% NaCl). The measured capacitance and resistance values were examined and reported. Technomed Europe Wiebachstraat 25a 6466 NG Kerkrade The Netherlands T +31 (43) 408 6868 infotechnomed@welcony.com www.technomedmedical.com Chamber of Commerce Reg.nr. 14049781 IBAN nr.: NL80ABNA0441485987 Vat-nr.: NL803379936B01 {9} Technomed media medical accessories Taking you further. Step by step. # Disposable Concentric Needle Electrode (DCNE) 510(k) ## Push test needle to connector This test measured the connection between the hub and the connector by doing a mechanical push test using the tensile testing machine. The pushing force was measured and recorded. ## Pull test needle to connector This test measured the connection between the needle and the connector by doing a mechanical pull test using the tensile testing machine. The pull force was measured and recorded. ## Biocompatibility Biological tests were conducted to demonstrate the subject devices are biologically safe. According to the FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation medical devices - Part 1: Evaluation and testing within a risk management process", the test conducted including - Physical and chemical characterization - Cytotoxicity testing - Toxicological evaluation - Sensitization testing - Intracutaneous reactivity testing - Acute systemic toxicity testing Clinical Testing - Not Applicable ## H. Conclusion Based on the above testing and analysis of performance data – it can be concluded that the subject device Disposable Concentric Needle Electrode is substantially equivalent to the primary predicate device cleared under K990015. Refer to Attachment 3.08 Bench Testing for the details of the bench testing. Technomed Europe Wiebachstraat 25a 6466 NG Kerkrade The Netherlands T +31 (43) 408 6868 infotechnomed@welcony.com www.technomedmedical.com Chamber of Commerce Reg.nr. 14049781 IBAN nr.: NL80ABNA0441485987 Vat-nr.: NL803379936B01 {10} Technomed medical accessories Taking you further. Step by step. Disposable Concentric Needle Electrode (DCNE) 510(k) # I. Predicate Performance Comparison Table / Reference Device Table 1: Substantial Equivalence / reference Summary | Criteria | Disposable Concentric Needle Electrode (Subject Device in this submission) | Technomed Europe diagnostic needle electrodes (Primary Predicate) | DISPOSABLE HYPODERMIC EMG NEEDLE ELECTRODE (Reference Device) | Disposable Monopolar Needle Electrode (Reference Device) | Evaluation of Differences | | --- | --- | --- | --- | --- | --- | | Manufacturer | Technomed Europe | Technomed Europe | Technomed Europe | Technomed Europe | N/A | | Device Name | Disposable Concentric Needle Electrode | Disposable Concentric Needle Electrode | Disposable Hypodermic Needle Electrode | Disposable Monopolar Needle Electrode | N/A | | 510(k) number | / | K990015 | K062437 | K130136 | N/A | | Product code | IKT and GXZ | IKT and GXZ | IKT | IKT and GXZ | Identical to the primary predicate device | | Intended use / Indications for use | Disposable Concentric Needle Electrodes are intended for use with recording, monitoring equipment for the recording of biopotential signals including electroencephalography (EEG), electromyography (EMG) and nerve potential signal. | Technomed Europe diagnostic needle electrodes are intended to be inserted in the subdermal, muscle or nerve tissue to sense bio-electric, EMG or EEG signals distally, and are intended to be proximal connected to electromyography / electroencephalogram recording equipment. | Disposable Hypodermic Needle Electrodes are intended for use in muscle stimulation and monitoring, motor unit action potential recording, and drug delivery. The only drug that may be used with this device is Botox Type A./ The Disposable Hypodermic Needle Electrodes are used for EMG recording and stimulation in patients undergoing clinical neurophysiological testing combined with delivery of injectable drugs (Botox Type A). | Disposable Monopolar Needle Electrodes are intended for use with recording and monitoring equipment for the recording of biopotential signals including electroencephalography (EEG), electromyography (EMG) and nerve potential signals, and are intended for stimulation and recording with stimulation / recording equipment for electromyography (EMG) and nerve potential signals. | Substantial equivalent to the predicate device, as described in current submission | | Regulation name | Diagnostic electromyograph needle electrode | Diagnostic electromyograph needle electrode | Electrode, Needle, Diagnostic Electromyograph | Electrode, Needle | Identical to the primary predicate device | | Regulation number | 21 CFR 890.1385 | 21 CFR 890.1385 | 21 CFR 890.1385 | 21 CFR 882.1350 | Identical to the primary predicate device | | Target population | The intended patients benefiting from the use of these devices are any patients identified by the medical experts to benefit from such diagnostic procedure. | The intended patients benefiting from the use of these devices are any patients identified by the medical experts to benefit from such diagnostic procedure. | The intended patients benefiting from the use of these devices are any patients identified by the medical experts to benefit from such diagnostic procedures | The intended patients benefiting from the use of these devices are any patients identified by the medical experts to benefit from such diagnostic procedures | Identical to the primary predicate device | Page 7(11) Technomed Europe Wiebachstraat 25a 6486 NG Karlvrede The Netherlands T +31 (43) 408 6868 info@technomed@waksmy.com www.technomedmedical.com Chamber of Commerce Reg.nr. 14049781 IBAN nr.: NL80A8NA0441485987 Vak nr.: NL803379936B01 {11} Technomed medical accessories Taking you further. Step by step. Disposable Concentric Needle Electrode (DCNE) 510(k) | Criteria | Disposable Concentric Needle Electrode (Subject Device in this submission) | Technomed Europe diagnostic needle electrodes (Primary Predicate) | DISPOSABLE HYPODERMIC EMG NEEDLE ELECTRODE (Reference Device) | Disposable Monopolar Needle Electrode (Reference Device) | Evaluation of Differences | | --- | --- | --- | --- | --- | --- | | Where used | The Disposable Concentric Needle Electrodes are used in highly controlled environments, according to facility specific procedures aligned with local, national and/or international guidance | The Disposable Concentric Needle Electrodes are used in highly controlled environments, according to facility specific procedures aligned with local, national and/or international guidance | The Disposable Hypodermic Needle Electrodes are used in highly controlled environments, according to facility specific procedures aligned with local, national and/or international guidance | The Disposable Monopolar Needle Electrodes are used in highly controlled environments, according to facility-specific procedures aligned with local, national and / or international guidance | Identical to the primary predicate device | | Dimensions | | | | | | | Hub diameter | 8 mm | 8 mm | 7.78 mm (luer lock) | 8 mm | Identical to the primary predicate device | | Hub length | 20 mm | 20 mm | 20 mm | 20 mm | Identical to the primary predicate device | | Needle shape | Straight | Straight | Straight | Straight | Identical to the primary predicate device | | Needle surface area | Recording area 0.03 or 0.07 mm² depending on needle size | Recording area 0.03 or 0.09 mm² depending on needle size | Recording/ Stimulation area 0.06 mm² 0.10 mm² 0.12 mm² 0.15 mm² 0.26 mm² | Recording/ Stimulation area ø 0.35mm/ ø 0.45mm 0.48mm²/ 0.78mm² | Substantial equivalent to the predicate device, as described in current submission. Non-clinical testing did not raise new or different questions relating to safety and effectiveness | | Needle length (Height of needle windings coming out of the hub) | 25mm 37mm 50mm 75mm | 25mm 37mm 50mm 75mm | 25mm 37mm 50mm 75mm | 25mm 37mm 50mm 75mm | Identical to the primary predicate device | | Needle diameter | 0.30mm 0.40mm 0.45mm 0.60mm | 0.30mm 0.40mm 0.45mm 0.60mm | 0.30mm 0.40mm 0.45mm 0.50mm 0.70mm | 0.35mm 0.45mm | Identical to the primary predicate device | | Tip geometry / angle (defining sharpness) | Trocar/ 15° | Trocar/ 12° | Trocar (pencil)/ 9° | Trocar | Substantial equivalent to the predicate device, as described in current submission. Non-clinical testing did not raise new or different questions relating to safety and effectiveness | | Materials of Construction | | | | | | Page 8(11) Technomed Europe Wiebachstraat 25a 6400 NG Karlsruhe The Netherlands T +31 (43) 408 6868 info@technomed@welcony.com www.technomedmedical.com Chamber of Commerce Reg.nr. 14049781 IBAN nr.: NL80A8NA0441485987 Vak nr.: NL803379936B01 {12} Technomed medical accessories Taking you further. Step by step. Disposable Concentric Needle Electrode (DCNE) 510(k) | Criteria | Disposable Concentric Needle Electrode (Subject Device in this submission) | Technomed Europe diagnostic needle electrodes (Primary Predicate) | DISPOSABLE HYPODERMIC EMG NEEDLE ELECTRODE (Reference Device) | Disposable Monopolar Needle Electrode (Reference Device) | Evaluation of Differences | | --- | --- | --- | --- | --- | --- | | Hub to needle connection | Crimping + Glue | Solder | Brass (crimp connection lead wire -needle) | Solder | Equivalent connection with the reference device Substantial equivalent to the predicate device, as described in current submission, Non-clinical testing did not raise new or different questions relating to safety and effectiveness | | Core wire | Stainless Steel AISI304 | Tungsten | N/A | N/A | Substantial equivalent to the predicate device, as described in current submission. Non-clinical testing did not raise new or different questions relating to safety and effectiveness | | Cannular | Stainless Steel AISI304 | Stainless Steel AISI304 | Stainless Steel AISI304 | Stainless Steel AISI304 | Identical to the primary predicate device | | Connector cage | Tin plated brass | Gold plated brass | N/A (lead wire) | Tin plated brass | Substantial equivalent to the predicate device, as described in current submission. Non-clinical testing did not raise new or different questions relating to safety and effectiveness | | Hub | Polyamide PA6 | Polyamide PA6 | Polyamide, food grade | Polyamide PA6 | Identical to the primary predicate device | | Protecting tube | Low-density polyethylene | Low-density polyethylene | polyethylene | Low-density polyethylene | Identical to the primary predicate device | | Isolation core wire | Polyesterimide | Polyesterimide | N/A | N/A | Modification, as described in current submission. Non-clinical testing did not raise new or different questions relating to safety and effectiveness | | Lubrication | Non curable Silicone | Non curable Silicone | N/A | Silicone mixture 20% MF 360 | Identical to the primary predicate device | Page 9(11) Technomed Europe Wiebachstraat 25a 6486 NG Karlvrede The Netherlands T +31 (43) 408 6868 info@technomed@welcony.com www.technomedmedical.com Chamber of Commerce Reg.nr. 14049781 IBAN nr.: NL80A8NA0441485987 Vak nr.: NL803379936B01 {13} Technomed medical accessories Taking you further. Step by step. Disposable Concentric Needle Electrode (DCNE) 510(k) | Criteria | Disposable Concentric Needle Electrode (Subject Device in this submission) | Technomed Europe diagnostic needle electrodes (Primary Predicate) | DISPOSABLE HYPODERMIC EMG NEEDLE ELECTRODE (Reference Device) | Disposable Monopolar Needle Electrode (Reference Device) | Evaluation of Differences | | --- | --- | --- | --- | --- | --- | | Performance | | | | | | | Impedance | Resistance ≤ 20 kΩ Capacitance ≥ 3.5 nF | Resistance ≤ 20 kΩ Capacitance ≥ 3.5 nF | Impedance < 100 kΩ | Impedance < 200 kΩ | Identical to the primary predicate device | | Tensile strength needle to hub | 0.30mm: > 11N 0.40mm: > 22N 0.45mm: > 22N 0.60mm: > 34N | 0.30mm: > 11N 0.40mm: > 22N 0.45mm: > 22N 0.60mm: > 34N | 0.30mm: > 11N 0.40mm: > 22N 0.45mm: > 22N 0.50mm: > 22N 0.70mm: > 40N | 0.30mm: > 11N 0.45mm: > 22N | Identical to the primary predicate device | | Connector cage to Hub Compression | > 3N | > 3N | Tensile strength > 25N (cable to connector) | > 3N | Identical to the primary predicate device | | Energy used and/or delivered | No energy used, no energy delivered, only conduction of electrical signals | No energy used, no energy delivered, only conduction of electrical signals | Used conduction of electrical signals (recording & stimulation) | Used conduction of electrical signals (recording & stimulation) | Identical to the primary predicate device | | Packaging | | | | | | | Single use | Yes | Yes | Yes | Yes | Identical to the primary predicate device and reference devices | | Supplied as sterile | Yes | Yes | Yes | Yes | Identical to primary predicate and reference devices | | Sterilization method | Ethylene oxide | Ethylene oxide | Ethylene oxide | Ethylene oxide | Identical to primary predicate and reference devices | | Sterility assurance level (SAL) | 10^{-6} | 10^{-6} | 10^{-6} | 10^{-6} | Identical to primary predicate and reference devices | | Shelf life | Three years | Three years | Three years | Three years | Identical to primary predicate and reference device | | Sterile barrier / container (transparent - film) | Film PET/PE | Film PET/PE | Film PET/PE | Film PET/PE | Identical to primary predicate and reference devices | | Sterile barrier / container (paper) | Medical paper | Medical paper | Medical paper | Medical paper | Identical to primary predicate and reference devices | | Pouch size | 50 x 200 mm (with thumbhole) | 50 x 200 mm (with or without thumbhole) | 100 x 200 (with or without thumbhole) | 50 x 200 mm (with thumbhole) | Identical to the primary predicate device | Page 10(11) Technomed Europe Wiebachstraat 25a 6486 NG Karlvade The Netherlands T +31 (43) 408 6868 info@technomed@welcony.com www.technomedmedical.com Chamber of Commerce Reg.nr. 14049781 IBAN nr.: NL80A8NA0441485987 Vak nr.: NL803379936B01 {14} Technomed medical accessories Taking you further. Step by step. Disposable Concentric Needle Electrode (DCNE) 510(k) | Criteria | Disposable Concentric Needle Electrode (Subject Device in this submission) | Technomed Europe diagnostic needle electrodes (Primary Predicate) | DISPOSABLE HYPODERMIC EMG NEEDLE ELECTRODE (Reference Device) | Disposable Monopolar Needle Electrode (Reference Device) | Evaluation of Differences | | --- | --- | --- | --- | --- | --- | | Packaging Shelf box | 200 x 140 x 50mm | 200 x 140 x 50 mm | 200 x 140 x 50 mm | 200 x 140 x 50mm | Identical to primary predicate and reference devices | | Number of items per Shelf box | 25 | 25 | 10 | 25 | Identical to the primary predicate device | | Packaging Shipping carton | 48 shelf boxes per Shipping carton | 48 shelf boxes per Shipping carton | 48 shelf boxes per Shipping carton | 48 shelf boxes per Shipping carton | Identical to primary predicate and reference devices | Technomed Europe Wiebachstraat 25a 6486 NG Karlvrede The Netherlands T +31 (43) 408 6868 infotechnomed@welcony.com www.technomedmedical.com Chamber of Commerce Reg.nr. 14049781 IBAN nr.: NL80A8NA0441485987 Vak nr.: NL803378636B01 Page 11(11) 11
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...