NYU Langone Genome PACT (Genome Profiling of Actionable Cancer Targets)

K202304 · Nyu Langone Medical Center · PZM · Jul 14, 2021 · Pathology

Device Facts

Record IDK202304
Device NameNYU Langone Genome PACT (Genome Profiling of Actionable Cancer Targets)
ApplicantNyu Langone Medical Center
Product CodePZM · Pathology
Decision DateJul 14, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.6080
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The NYU Langone Genome PACT assay is a qualitative in vitro diagnostic test that uses targeted next generation sequencing of formalin-fixed paraffin-embedded (FFPE) tumor tissue matched with normal specimens with solid malignant neoplasms to detect tumor gene alterations in a 607-gene panel. The test is intended to provide information on somatic mutations and small insertions and deletions) for use by qualified health care professionals in accordance with professional guidelines, and is not conclusive or prescriptive for labeled use of any specific therapeutic product. NYU Langone Genome PACT is a single-site assay performed at NYU Langone Health.

Device Story

NYU Langone Genome PACT is a targeted next-generation sequencing (NGS) assay; analyzes FFPE tumor tissue matched with normal specimens; detects somatic mutations and small insertions/deletions in 607 genes. Performed as a single-site assay at NYU Langone Health. Input: tumor/normal tissue samples; Output: genomic alteration report. Used by oncologists/pathologists to inform clinical decision-making regarding cancer treatment; provides actionable genomic information; not prescriptive for specific therapies.

Technological Characteristics

Targeted next-generation sequencing (NGS) assay; 607-gene panel; utilizes FFPE tumor tissue and matched normal specimens; qualitative detection of somatic mutations and small insertions/deletions.

Indications for Use

Indicated for patients with solid malignant neoplasms. Qualitative detection of somatic mutations and small insertions/deletions in a 607-gene panel using FFPE tumor tissue matched with normal specimens. For use by qualified healthcare professionals; not conclusive or prescriptive for specific therapeutic products.

Regulatory Classification

Identification

A next generation sequencing (NGS) based tumor profiling test is a qualitative in vitro diagnostic test intended for NGS analysis of tissue specimens from malignant solid neoplasms to detect somatic mutations in a broad panel of targeted genes to aid in the management of previously diagnosed cancer patients by qualified health care professionals.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information: (i) A detailed description of all somatic mutations that are intended to be detected by the test and that are adequately supported in accordance with paragraph (b)(1)(v) of this section and reported in the test results in accordance with paragraph (b)(2)(iv) of this section, including: (A) A listing of mutations that are cancer mutations with evidence of clinical significance. (B) As appropriate, a listing of mutations that are cancer mutations with potential clinical significance. (ii) The indications for use must specify the following: (A) The test is indicated for previously diagnosed cancer patients. (B) The intended specimen type(s) and matrix ( *e.g.,* formalin-fixed, paraffin-embedded tumor tissue).(C) The mutation types ( *e.g.,* single nucleotide variant, insertion, deletion, copy number variation or gene rearrangement) for which validation data has been provided.(D) The name of the testing facility or facilities, as applicable. (iii) A detailed device description including the following: (A) A description of the test in terms of genomic coverage, as follows: ( *1* ) Tabulated summary of all mutations reported, grouped according to gene and target region within each gene, along with the specific cDNA and amino acid positions for each mutation.( *2* ) A description of any within-gene targeted regions that cannot be reported and the data behind such conclusion.(B) Specifications for specimen requirements including any specimen collection devices and preservatives, specimen volume, minimum tumor content, specimen handling, DNA extraction, and criteria for DNA quality and quantity metrics that are prerequisite to performing the assay. (C) A detailed description of all test components, reagents, instrumentation, and software required. Detailed documentation of the device software including but not limited to, software applications and hardware-based devices that incorporate software. (D) A detailed description of the methodology and protocols for each step of the test, including description of the quality metrics, thresholds, and filters at each step of the test that are implemented for final result reporting and a description of the metrics for run-failures, specimen-failures, invalids, as applicable. (E) A list of links provided by the device to the user or accessed by the device for internal or external information ( *e.g.,* decision rules or databases) supporting clinical significance of test results for the panel or its elements in accordance with paragraphs (b)(1)(v) and (b)(2)(vi) of this section.(F) A description of internal and external controls that are recommended or provided and control procedures. The description must identify those control elements that are incorporated into the testing procedure. (iv) Information demonstrating analytical validity of the device according to analytical performance characteristics, evaluated either specifically for each gene/mutation or, when clinically and practically justified, using a representative approach based on other mutations of the same type, including: (A) Data that adequately supports the intended specimen type ( *e.g.,* formalin-fixed, paraffin-embedded tumor tissue), specimen handling protocol, and nucleic acid purification for specific tumor types or for a pan-tumor claim.(B) A summary of the empirical evidence obtained to demonstrate how the analytical quality metrics and thresholds were optimized. (C) Device precision data using clinical samples to adequately evaluate intra-run, inter-run, and total variability. The samples must cover all mutation types tested (both positive and negative samples) and include samples near the limit of detection of the device. Precision must be assessed by agreement within replicates on the assay final result for each representative mutation, as applicable, and also supported by sequencing quality metrics for targeted regions across the panel. (D) Description of the protocols and/or data adequately demonstrating the interchangeability of reagent lots and multiplexing barcodes. (E) A description of the nucleic acid assay input concentration range and the evidence to adequately support the range. (F) A description of the data adequately supporting the limit of detection of the device. (G) A description of the data to adequately support device accuracy using clinical specimens representing the intended specimen type and range of tumor types, as applicable. ( *1* ) Clinical specimens tested to support device accuracy must adequately represent the list of cancer mutations with evidence of clinical significance to be detected by the device.( *2* ) For mutations that are designated as cancer mutations with evidence of clinical significance and that are based on evidence established in the intended specimen type (*e.g.,* tumor tissues) but for a different analyte type (*e.g.,* protein, RNA) and/or a measurement (*e.g.,* incorporating a score or copy number) and/or with an alternative technology (*e.g.,* IHC, RT-qPCR, FISH), evidence of accuracy must include clinically adequate concordance between results for the mutation and the medically established biomarker test (*e.g.,* evidence generated from an appropriately sized method comparison study using clinical specimens from the target population).( *3* ) For qualitative DNA mutations not described in paragraph (b)(1)(iv)(G)(*2* ) of this section, accuracy studies must include both mutation-positive and wild-type results.(H) Adequate device stability information. (v) Information that adequately supports the clinical significance of the panel must include: (A) Criteria established on what types and levels of evidence will clinically validate a mutation as a cancer mutation with evidence of clinical significance versus a cancer mutation with potential clinical significance. (B) For representative mutations of those designated as cancer mutations with evidence of clinical significance, a description of the clinical evidence associated with such mutations, such as clinical evidence presented in professional guidelines, as appropriate, with method comparison performance data as described in paragraph (b)(1)(iv)(G) of this section. (C) For all other mutations designated as cancer mutations with potential clinical significance, a description of the rationale for reporting. (2) The 21 CFR 809.10 compliant labeling and any product information and test report generated, must include the following, as applicable: (i) The intended use statement must specify the following: (A) The test is indicated for previously diagnosed cancer patients. (B) The intended specimen type(s) and matrix ( *e.g.,* formalin-fixed, paraffin-embedded tumor tissue).(C) The mutation types ( *e.g.,* single nucleotide variant, insertion, deletion, copy number variation or gene rearrangement) for which validation data has been provided.(D) The name of the testing facility or facilities, as applicable. (ii) A description of the device and summary of the results of the performance studies performed in accordance with paragraphs (b)(1)(iii), (b)(1)(iv), and (b)(1)(v) of this section. (iii) A description of applicable test limitations, including, for device specific mutations validated with method comparison data to a medically established test in the same intended specimen type, appropriate description of the level of evidence and/or the differences between next generation sequencing results and results from the medically established test ( *e.g.,* as described in professional guidelines).(iv) A listing of all somatic mutations that are intended to be detected by the device and that are reported in the test results under the following two categories or equivalent designations, as appropriate: “cancer mutations panel with evidence of clinical significance” or “cancer mutations panel with potential clinical significance.” (v) For mutations reported under the category of “cancer mutations panel with potential clinical significance,” a limiting statement that states “For the mutations listed in [cancer mutations panel with potential clinical significance or equivalent designation], the clinical significance has not been demonstrated [with adequate clinical evidence ( *e.g.,* by professional guidelines) in accordance with paragraph (b)(1)(v) of this section] or with this test.”(vi) For mutations under the category of “cancer mutations panel with evidence of clinical significance,” or equivalent designation, link(s) for physicians to access internal or external information concerning decision rules or conclusions about the level of evidence for clinical significance that is associated with the marker in accordance with paragraph (b)(1)(v) of this section.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. NYU Langone Medical Center % Derek Symula Director, Division of Laboratory Quality Certification New York State Department of Health Wadsworth Center, Empire State Plaza P.O. Box 509 Albany, New York 12237 Re: K202304 Trade/Device Name: NYU Langone Genome PACT (Genome Profiling of Actionable Cancer Targets) Regulation Number: 21 CFR 866.6080 Regulation Name: Next generation sequencing based tumor profiling test Regulatory Class: Class II Product Code: PZM Dated: July 31, 2020 Received: August 14, 2020 Dear Derek Symula: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Zivana Tezak, PhD Acting Chief Division of Molecular Genetics and Pathology OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202304 #### Device Name NYU Langone Genome PACT (Genome Profiling of Actionable Cancer Targets) #### Indications for Use (Describe) The NYU Langone Genome PACT assay is a qualitative in vitro diagnostic test that uses targeted next generation sequencing of formalin-fixed paraffin-embedded (FFPE) tumor tissue matched with normal specimens with solid malignant neoplasms to detect tumor gene alterations in a 607-gene panel. The test is intended to provide information on somatic mutations and small insertions and deletions) for use by qualified health care professionals in accordance with professional guidelines, and is not conclusive or prescriptive for labeled use of any specific therapeutic product. NYU Langone Genome PACT is a single-site assay performed at NYU Langone Health. Type of Use (Select one or both, as applicable) | <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |----------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...