K022774 · Intl. Remote Imaging Systems · LKM · Oct 21, 2002 · Hematology
Device Facts
Record ID
K022774
Device Name
IQ 200 SYSTEM
Applicant
Intl. Remote Imaging Systems
Product Code
LKM · Hematology
Decision Date
Oct 21, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5200
Device Class
Class 2
Indications for Use
The iQ 200 System is an in-vitro diagnostic device used to automate the complete urinalysis profile, including urine test strip chemistry panel and microscopic sediment analysis. Optionally, the iQ 200 Analyzer can be used as a stand-alone unit, or the results from the iQ 200 Analyzer can be combined with other urine chemistry results received from an LIS. It produces quantitative or qualitative counts of all formed sediment elements present in urine, including cells, casts, crystals, and organisms. A competent human operator can set criteria for auto-reporting and flagging specimens for review. All instrument analyte image decisions may be reviewed and overridden by a trained technologist.
Device Story
Automated urinalysis system; processes urine samples via autosampler. Transforms samples into images for automated analyte recognition; classifies formed sediment elements (cells, casts, crystals, organisms). Operates in clinical laboratory settings; managed by trained technologists. System consolidates chemistry results from companion analyzer (ARKRAY AUTION MAX AX-4280) and microscopic findings on a dedicated Computer WorkStation. Technologists review, confirm, or modify automated classifications; results released off-line. Facilitates physician health assessment and differential diagnosis by providing quantitative/qualitative sediment counts.
Clinical Evidence
Comparative performance study against 939UDx Urine Pathology System and Sysmex UF-100. Results showed iQ 200 System had greater clinical sensitivity for abnormal specimens and individual microscopic formed elements. Specificity was comparable to predicates. With sensitivity fixed at 90%, iQ 200 specificity exceeded predicates for most elements. Quantitative reporting (R Square) for RBC, WBC, squamous epithelial cells, and casts matched expert human identification and predicate performance.
Technological Characteristics
Automated sample handling via autosampler; intelligent microscopy; automated analyte classification algorithm. Consolidates chemistry data from external analyzer via Computer WorkStation. Enables off-line review and manual override of classifications. Standalone or networked configuration.
Indications for Use
Indicated for routine urinalysis in patients undergoing screening for metabolic/systemic disease, renal/urinary tract dysfunction, infection, carcinoma, or injury; monitoring therapy effectiveness; and post-surgical/post-therapeutic recovery.
Regulatory Classification
Identification
An automated cell counter is a fully-automated or semi-automated device used to count red blood cells, white blood cells, or blood platelets using a sample of the patient's peripheral blood (blood circulating in one of the body's extremities, such as the arm). These devices may also measure hemoglobin or hematocrit and may also calculate or measure one or more of the red cell indices (the erythrocyte mean corpuscular volume, the mean corpuscular hemoglobin, or the mean corpuscular hemoglobin concentration). These devices may use either an electronic particle counting method or an optical counting method.
Predicate Devices
The Yellow IRIS® Urinalysis Workstation with CHEMSTRIP® Reader
IRIS Flow Microscope
900UDx Urine Pathology System
939UDx Urine Pathology System
Sysmex UF-100
Reference Devices
ARKRAY AUTION MAX AX-4280
Related Devices
K093861 — IQ 20 SERIES ANALYZER, IQ 200 SERIES SYSTEM, NAD IQ LAMINA CRADLE · Iris International, Inc. · Feb 5, 2010
K960054 — 900UDX URINE PATHOLOGY SYSTEM · Intl. Remote Imaging Systems · Mar 22, 1996
K252580 — iQ200 Series · Beckman Coulter, Inc. · Sep 10, 2025
K083002 — SYSMEX UF-500I AUTOMATED URINE PARTICLE ANALYZER · Sysmex America, Inc. · Feb 3, 2009
{0}
OCT 21 2002
International Remote Imaging Systems, Inc.
9162 Eton Avenue, Chatsworth, CA 91311
800-776-4747 818-709-1244 FAX 818-700-9661
# 510(k) SUMMARY
510(k) Submission: K022774
iQ™ 200 System
## General Information:
Date of Submission: August 20, 2002
Trade Name of Device: iQ™ 200 System
Common Name of Device: Automated urinalysis system
Classification Name: Automated urinalysis system, 21 CFR 862.2900, Class I
Microscope, 21 CFR 862.3600, Class I
Automated cell counter (Urine particle counter), 21 CFR 864.5200, Class II
Submitter's Name: Harvey L. Kasdan, Ph.D.
Chief Scientist
Iris Diagnostics, A Division of
International Remote Imaging Systems, Inc.
9162 Eton Avenue, Chatsworth, CA 91311
## Indications for Use (Brief):
The iQ 200 System is intended for analysis of urine chemistry, color, clarity, specific gravity, and formed sediment elements, which constitutes typical routine urinalysis.
## Intended Use:
The iQ 200 System is an in-vitro diagnostic device used to automate the complete urinalysis profile, including urine test strip chemistry panel and microscopic sediment analysis. Optionally, the iQ 200 Analyzer can be used as a stand-alone unit, or the results from the iQ 200 Analyzer can be combined with other urine chemistry results received from an LIS. It produces quantitative or qualitative counts of all formed sediment elements present in urine, including cells, casts, crystals, and organisms. A competent human operator can set criteria for auto-reporting and flagging specimens for review. All instrument analyte image decisions may be reviewed and overridden by a trained technologist.
510(k) Summary, iQ™ 200 System
{1}
510(k) Summary, iQ™ 200 System
Page 2
## Substantial Equivalence to Predicate Devices:
The iQ 200 System is substantially equivalent under Section 510(k) of the Food, Drug, and Cosmetic Act to The Yellow IRIS® Urinalysis Workstation with CHEMSTRIP® Reader, the IRIS Flow Microscope, the 900UDx Urine Pathology System, the 939UDx Urine Pathology System, and the Sysmex UF-100.
## Summary of Technological Characteristics:
The iQ 200 System provides automatic sample handling for automated intelligent microscopy and automatic analyte classification for improved data reporting, presentation and management. Specimens are aspirated by an autosampler rather than poured manually. Each image is analyzed and assigned a classification by an auto analyte recognition algorithm. Using these classifications and the known observation volume, microscopic analyte concentrations may be automatically reported. If results from a specimen are not autoreported, microscopic examination results are displayed on an independent Computer WorkStation. Operators can then confirm or modify analyte classifications and release reports off-line for enhanced convenience, obviating the need to process a second aliquot for review. Chemistry results from the companion ARKRAY AUTION MAX AX-4280 are automatically consolidated by the Computer WorkStation for display and reporting.
## Performance Studies:
Microscopic analysis performance of the iQ 200 System were compared with that of 939UDx Urine Pathology System and the Sysmex UF-100.
## Conclusions Drawn From Clinical Tests:
Clinical sensitivity (fraction of abnormal specimens detected) of the iQ 200 System exceeded that of the 939UDx Urine Pathology System and Sysmex UF-100, while clinical specificity (fraction of normal specimens detected) was comparable. The iQ 200 System generally had greater sensitivity for detection of individual microscopic formed elements than either the 939UDx Urine Pathology System or the Sysmex UF-100. With sensitivity for detection of each microscopic formed element fixed at 90%, iQ 200 System specificity was greater than that of either the 939UDx Urine Pathology System or Sysmex UF-100 for most formed elements. iQ 200 System automatically reported concentrations of RBC, WBC, squamous epithelial cells and casts matched those determined from expert human identification as well as the 939UDx Urine Pathology System predicate automatically reported concentrations matched those of the human expert when measured by the square of the regression correlation coefficient, R Square. The iQ 200 System has been shown to be suitable for its intended use.
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
OCT 21 2002
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
Harvey L. Kasdan, Ph.D.
Chief Scientist
International Remote Imaging Systems, Inc.
9172 Eton Avenue
Chatsworth, California 91311-5874
Re: k022774
Trade/Device Name: iQ™ 200 System
Regulation Number: 21 CFR § 864.5200
Regulation Name: Automated Cell Counter
Regulatory Class: II
Product Code: LKM, KQO
Dated: August 20, 2002
Received: August 21, 2002
Dear Dr. Kasdan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{4}
# PREMARKET NOTIFICATION
510(k) Number (if known): K022774
Device Name: iQ 200 System
## Indications for Use
The iQ 200 System is intended for analysis of urine chemistry, specific gravity, and formed sediment elements, which constitutes typical routine urinalysis.
Urinalysis is ordered by physicians as a screening procedure for detection of possible abnormal metabolic or systemic disease, indicated by chemical composition of urine, and of potential renal or urinary tract disease or dysfunction, indicated by urine concentration and by the nature and distribution of urinary formed elements. Urine profile testing is commonly employed in the initial clinical evaluation of patients admitted for hospital care or undergoing physical examinations.
Routine urinalysis is also indicated in diagnosis of patients with possible renal or urinary tract infection, carcinoma, or other injury, as well as for monitoring the status and effectiveness of drug, radiation, or dietary therapy, and of post-surgical or post-therapeutic recovery.
The information produced by the iQ 200 System concerning the composition of patient urines is ordered at the discretion of the physician, and is part of a larger body of laboratory and other test results available to assist the physician in health assessments or differential diagnoses. Routine urinalysis findings are always subject to judgment and interpretation by physicians relative to the patient's overall clinical presentation and history.

IQ™ 200 System
August 20, 2002
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.