K242244 · Lumea, Inc. · QKQ · Mar 14, 2025 · Pathology
Device Facts
Record ID
K242244
Device Name
Viewer+
Applicant
Lumea, Inc.
Product Code
QKQ · Pathology
Decision Date
Mar 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.3700
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Viewer+ is a software only device intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review, interpret and manage digital images of pathology slides for primary diagnosis. Viewer+ is not intended for use with frozen sections, cytology, or non-FFPE hematopathology specimens. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. Viewer+ is intended for use with Hamamatsu NanoZoomer S360MD Slide scanner and BARCO MDPC-8127 display.
Device Story
Viewer+ is a web-based software-only device for viewing/managing digitized pathology images (FFPE tissue) acquired via Hamamatsu NanoZoomer S360MD scanner. Input: digital pathology images (NDPI format) uploaded to cloud storage. Operation: pathologist accesses case worklist, views images, performs panning/zooming, measures distances/areas, and annotates images via web browser (Chrome, Edge, Firefox, Safari). Output: rendered digital slide images on BARCO MDPC-8127 display. Used in clinical pathology settings by pathologists for primary diagnosis. Device facilitates review and interpretation; pathologist retains responsibility for diagnostic decisions, including use of conventional light microscopy when necessary. Benefits: enables remote/digital review of pathology slides, streamlining workflow and diagnostic management.
Clinical Evidence
No clinical data. Bench testing only. Pixel-wise comparison of 270 image-pairs (30 slides, 3 ROIs, 3 magnifications) across 4 browsers showed 95th percentile ΔE00 < 3 (max 1.141), confirming identical reproduction to predicate. Turnaround time testing (loading, panning, zooming) met acceptance criteria (max 10s load, 5s pan/zoom). Measurement accuracy testing (length/area) met criteria (+/- 0.1mm/0.1mm²). Summative human factors testing confirmed safety and effectiveness for intended users.
Technological Characteristics
Web-based software-only device. Operates on client PC (Windows/Apple OS) via standard browsers. Interoperable with Hamamatsu NanoZoomer S360MD scanner and BARCO MDPC-8127 display. Performs image rendering, navigation, annotation, and measurement. No hardware components. Software life cycle per IEC 62304; risk management per ISO 14971.
Indications for Use
Indicated for use by pathologists to review, interpret, and manage digital images of FFPE-tissue surgical pathology slides for primary diagnosis. Not indicated for frozen sections, cytology, or non-FFPE hematopathology specimens.
Regulatory Classification
Identification
The whole slide imaging system is an automated digital slide creation, viewing, and management system intended as an aid to the pathologist to review and interpret digital images of surgical pathology slides. The system generates digital images that would otherwise be appropriate for manual visualization by conventional light microscopy.
Special Controls
A whole slide imaging system must comply with the following special controls: (1) Premarket notification submissions must include the following information: (i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system. (ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate: (A) Slide feeder; (B) Light source; (C) Imaging optics: (D)Mechanical scanner movement; (E) Digital imaging sensor; (F) Image processing software; (G)Image composition techniques; (H)Image file formats; (I) Image review manipulation software; (J) Computer environment; (K)Display system. (iii)Detailed bench testing and results at the system level, including for the following, as appropriate: (A)Color reproducibility; (B) Spatial resolution; (C) Focusing test; (D) Whole slide tissue coverage; (E) Stitching error: (F) Turnaround time. (iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate: (A)Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (e.g., main sign-out diagnosis). (D) A detailed human factors engineering process must be used to evaluate the whole slide imaging system user interface(s). (2) Labeling compliant with 21 CFR 809.10(b) must include the following: The intended use statement must include the information described in paragraph (i) (1)(i) of this section, as applicable, and a statement that reads, "It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device." (ii) A description of the technical studies and the summary of results, including those that relate to paragraph (1)(ii) and (1)(iii) of this section, as appropriate. (iii) A description of the performance studies and the summary of results, including those that relate to paragraph (1)(iv) of this section, as appropriate. (iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system.
(ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate:
(A) Slide feeder;
(B) Light source;
(C) Imaging optics;
(D) Mechanical scanner movement;
(E) Digital imaging sensor;
(F) Image processing software;
(G) Image composition techniques;
(H) Image file formats;
(I) Image review manipulation software;
(J) Computer environment; and
(K) Display system.
(iii) Detailed bench testing and results at the system level, including for the following, as appropriate:
(A) Color reproducibility;
(B) Spatial resolution;
(C) Focusing test;
(D) Whole slide tissue coverage;
(E) Stitching error; and
(F) Turnaround time.
(iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate:
(A) Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (
*e.g.,* main sign-out diagnosis).(D) A detailed human factor engineering process must be used to evaluate the whole slide imaging system user interface(s).
(2) Labeling compliant with 21 CFR 809.10(b) must include the following:
(i) The intended use statement must include the information described in paragraph (b)(1)(i) of this section, as applicable, and a statement that reads, “It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device.”
(ii) A description of the technical studies and the summary of results, including those that relate to paragraphs (b)(1)(ii) and (iii) of this section, as appropriate.
(iii) A description of the performance studies and the summary of results, including those that relate to paragraph (b)(1)(iv) of this section, as appropriate.
(iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K242244
B Applicant
Lumea, Inc.
C Proprietary and Established Names
Viewer+
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QKQ | Class II | 21 CFR 864.3700 - Whole slide imaging system | PA - Pathology |
## II Submission/Device Overview:
A Purpose for Submission:
New device
B Type of Test:
Software only device
## III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
For In Vitro Diagnostic Use
Viewer+ is a software only device intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review, interpret and manage digital images of pathology slides for primary diagnosis. Viewer+ is not intended for use with frozen sections, cytology, or non-FFPE hematopathology specimens.
It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. Viewer+ is intended for use with Hamamatsu NanoZoomer S360MD Slide scanner and BARCO MDPC-8127 display.
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
## IV Device/System Characteristics:
### A Device Description:
Viewer+, version 1.0.1, is a web-based software device that facilitates the viewing and navigating of digitized pathology images of slides prepared from FFPE-tissue specimens acquired from Hamamatsu NanoZoomer S360MD Slide scanner and viewed on BARCO MDPC-8127 display. Viewer+ renders these digitized pathology images for review, management, and navigation for pathology primary diagnosis.
Viewer+ is operated as follows:
1. Image acquisition is performed using the NanoZoomer S360MD Slide scanner according to its Instructions for Use. The operator performs quality control of the digital slides per the instructions of the NanoZoomer and lab specifications to determine if re-scans are necessary.
2. Once image acquisition is complete and the image becomes available in the scanner's database file system, a separate medical image communications software (not part of the device) automatically uploads the image and its corresponding metadata to persistent cloud storage. Image and data integrity checks are performed during the upload to ensure data accuracy.
3. The subject device enables the reading pathologist to open a patient case, view the images, and perform actions such as zooming, panning, measuring distances and areas, and annotating images as needed. After reviewing all images for a case, the pathologist will render a diagnosis.
Viewer+ operates with and is validated for use with the components specified the tables below:
K242244 - Page 2 of 9
{2}
Table 1. Interoperable Components for Use with Viewer+
| Components | Manufacturer | Model |
| --- | --- | --- |
| Scanner | Hamamatsu Photonics K.K. | NanoZoomer S360MD Slide scanner |
| Display | BARCO | MDPC-8127 |
Table 2. Computer Environment/System Requirements
| Environment | Component | Minimum Requirements |
| --- | --- | --- |
| Client PC | | |
| Hardware | Processor | • 1 CPU
• 4 cores
• 2.0GHz
• Standard consumer GPU, such as Intel, AMD, or NVIDIA with 2GB VRAM |
| | Memory | 8 GB or more |
| | Network | 100 Mbps or faster (1 Gbps recommended) |
| Software | Operating System/Browser | • Windows (10 or higher) for
○ Google Chrome (130.06723.117 or higher)
○ Microsoft Edge (132.0.2.2849.80 or higher)
○ Mozilla Firefox (132.0.1 or higher)
• Apple OS (13.7.2 Ventura or higher) for
○ Apple Safari (18.1 (20619.2.8.11.10) or higher) |
B Instrument Description Information:
1. Instrument Name:
Viewer+
2. Specimen Identification:
Viewer+ uses digital pathology images obtained from the Hamamatsu NanoZoomer S360MD Slide scanner of Hematoxylin and Eosin (H&E) stained glass slides. The reading pathologist selects a case (patient) from a worklist (within the device or external to the device) whereby the subject device fetches the associated images from the external image storage. The scanned images are identified based on the previously assigned specimen identifier such as the laboratory specimen accession number.
3. Specimen Sampling and Handling:
Specimen sampling and handling are performed upstream and independent of the use of the subject device. Specimen sampling includes biopsy or resection specimens which are processed using histology techniques. The FFPE tissue section is Hematoxylin and Eosin (H&E) stained. Digital images are then obtained from these glass slides using the Hamamatsu NanoZoomer S360MD Slide scanner.
4. Calibration:
K242244 - Page 3 of 9
{3}
Not applicable
5. Quality Control:
The scanning technician should perform quality control of all glass slides following the instructions. Upon scanning, quality control should be performed on all digital slide images following procedures defined by the scanner manufacturer prior to import into Viewer+. The pathologist should review all images to ensure that all expected slides have been imported by viewing the thumbnails and labels, and manually verifying the tissue specimen, block, and staining information is present.
V Substantial Equivalence Information:
A Predicate Device Name(s):
NanoZoomer S360MD Slide scanner system
B Predicate 510(k) Number(s):
K233027
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K242244 | K233027 |
| --- | --- | --- |
| Device Trade Name | Viewer+ | NanoZoomer S360MD Slide scanner system |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | For In Vitro Diagnostic Use
Viewer+ is a software only device intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review, interpret and manage digital images of pathology slides for primary diagnosis. Viewer+ is not intended for use with frozen sections, cytology, or non-FFPE hematopathology specimens.
It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. Viewer+ is intended for use with | The NanoZoomer S360MD Slide scanner system (“NanoZoomer System”) is an automated digital slide creation, viewing, and management system. The NanoZoomer System is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (“FFPE”) tissue. The NanoZoomer System is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.
The NanoZoomer System comprises the NanoZoomer S360MD Slide scanner, the NZViewMD Software and a compatible display that has been 510(k) cleared for use with the NanoZoomer system or a 510(k)-cleared display that has been assessed in accordance with the Predetermined Change Control Plan |
K242244 - Page 4 of 9
{4}
VI Standards/Guidance Documents Referenced:
1. Guidance for Industry “Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices”, April 20, 2016.
2. Guidance for Industry "Applying Human Factors and Usability Engineering to Medical Devices", February 3, 2016.
3. IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION, 13-79. Medical device software – Software life cycle processes.
4. ISO 14971 Third Edition 2019-12, 5-125, Medical devices – Applications of risk management to medical devices.
5. IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION, 5-129, Application of usability engineering to medical devices.
K242244 - Page 5 of 9
{5}
6. AAMI TIR 45:2012, 13-36, Guidance on the use of AGILE practices in the development of medical device software.
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
1. Precision/Reproducibility: Not applicable
2. Linearity: Not applicable
3. Analytical Specificity/Interference: Not applicable
4. Accuracy (Instrument): Not applicable
5. Carry-Over: Not applicable
### B Other Supportive Instrument Performance Characteristics Data:
Technical performance testing was conducted with the subject device, Viewer+ as specified below.
#### 1. Bench Testing – Pixelwise comparison test
Pixel-wise comparison testing to demonstrate identical image reproduction was conducted to compare WSIs reproduced by the subject device and the comparators as listed in Table 3 below. The subject device was compared to the predicate device’s image review manipulation software (IRMS, as defined in FDA guidance document, “Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices” dated April 20, 2016) using the quantitative pixel-wise comparison method. The basis for the comparison was the CIEDE2000 color difference equation, $\Delta E_{00}$. The devices were tested as operating with the intended components, including the scanner, specific file format, image management systems (subject device with the intended browsers, comparator [predicate device IRMS]) and display, as specified in the Table 3 below.
For each of the 4 configurations in Table 3 below, the device was tested with multiple slides across multiple regions of interest (ROI) at multiple magnification levels. A total of 30 H&E-stained, FFPE glass slides of normal and tumor tissues from various human anatomical organs were used in the testing. The glass slides were scanned on a Hamamatsu NanoZoomer S360MD Slide scanner to obtain 30 WSIs. For each of the 30 WSIs, 3 ROIs were selected by qualified personnel to represent various features in the tissue samples. Each ROI was captured at 3 magnification levels (10x, 20x, 40x).
K242244 - Page 6 of 9
{6}
The screenshots were captured from the intended display while viewing with the subject device and predicate device IRMS. The screenshots were cropped and registered to be pixelwise comparable. The cropped image included most of the pixels in the image except for those in the viewer-specific user interface areas.
For each configuration, two sets of images were collected: comparator (predicate device IRMS) and the subject device (Viewer+ with the intended browser). Each image set included 270 images that covered all combinations of 30 slides, 3 ROIs, and 3 magnification levels. The testing data, including the overview images of the 30 glass slides with annotations of the ROIs, registration/cropping information, and captured images, were provided in the FDA specific format.
The comparator (predicate device IRMS) image set was used as the reference to compare the subject device image set to determine whether all the 270 image-pairs were identical for each configuration. Two images are considered identical if the 95th percentile of the pixelwise differences, computed using the International Commission on Illumination (CIE) color difference metric CIEDE2000 ( $\Delta$ E00), is less than $3\Delta \mathrm{E}_{00}$ . Testing results showed that the pixelwise differences across all 270 image-pairs per configuration were less than $3\Delta \mathrm{E}_{00}$ . The maximum (max), minimum (min), and mean of the 95th percentile $\Delta$ E00 value were reported in Table 3. Testing results demonstrated that WSIs reproduced by Viewer+ are identical to images reproduced by the predicate device.
Table 3. Viewer+ Pixelwise Comparison Testing Results
| Scanner | Image File Format | Subject Device /Browser | Comparator (Predicate device IRMS) | Display | Results |
| --- | --- | --- | --- | --- | --- |
| Hamamatsu NanoZoomer S360MD Slide Scanner | NDPI | Viewer+ /Chrome | NZViewMD | BARCO MDPC-8127 display | max (95th percentile ΔE00) = 1.133 min (95th percentile ΔE00) = 0.550 mean (95th percentile ΔE00) = 0.773 |
| | | Viewer+ /Edge | NZViewMD | | max (95th percentile ΔE00) = 1.133 min (95th percentile ΔE00) = 0.550 mean (95th percentile ΔE00) = 0.773 |
| | | Viewer+ /Firefox | NZViewMD | | max (95th percentile ΔE00) = 1.141 min (95th percentile ΔE00) = 0.526 mean (95th percentile ΔE00) = 0.775 |
| | | Viewer+ /Safari | NZViewMD | | max (95th percentile ΔE00) = 1.077 min (95th percentile ΔE00) = 0.620 mean (95th percentile ΔE00) = 0.791 |
# 2. Turnaround Time
The turnaround time for image loading, panning and zooming were tested in the subject device when using the 4 supported browsers (Google Chrome, Microsoft Edge, Mozilla Firefox, and Apple Safari) over different magnifications levels and over multiple fields of view (FOVs). Test results for different scenarios met the test acceptance criteria and showed acceptable turnaround time for the intended use of the subject device, as shown in table 4 below.
K242244 - Page
{7}
Table 4. Viewer+ Turnaround Time Study Test Results
| Description | Test Method/Sample Size | Acceptance Criteria | Results (Seconds) |
| --- | --- | --- | --- |
| Loading | 4 participants/4 browsers * 2 patient cases * 3 H&E-stained whole slide images per case = total of 24 image loading time measurements | Max 10 seconds per action | Mean = 3.3 Min = 2.0 Max = 8.0 |
| Panning | 4 participants/4 browsers * 2 patient cases * 3 H&E-stained whole slide images per case * 3 pans per image = total of 72 panning time measurements | Max 5 seconds per action | Mean = 1.3 Min = 1.0 Max = 3.0 |
| Zooming | 4 participants/4 browsers * 2 patient cases * 3 H&E-stained whole slide images per case * 3 zooms per image = total of 72 zooming time measurements | Max 5 seconds per action | Mean = 1.3 Min = 1.0 Max = 3.0 |
# 3. Measurement - distance and area
Measurement accuracy testing was performed to demonstrate that Viewer+ represents length and areal measurements accurately across multiple magnification levels. Four biological glass slides were scanned on Hamamatsu NanoZoomer S360MD Slide scanner, and then measured by qualified medical professionals using the Hamamatsu NZViewMD software and Viewer+ measurement tools on the supported Barco MDPC-8127 display. A randomized set of tests is chosen to test 4 biological glass slide images, 4 supported browsers (Google Chrome, Microsoft Edge, Mozilla Firefox, and Apple Safari), 4 rotations, 4 zooms, 2 directions, and 2 types of measurements (length and areal). The intended and reported metrics were recorded for each annotation. The differences between intended and reported measurements were calculated, as summarized below in Table 5. Acceptance criteria was set at $+/-0.1\mathrm{mm}$ of pathologist annotation for length, and $+/-0.1\mathrm{mm}^2$ for area measurements. Test results showed that the acceptance criteria were met, and the subject device performed accurate measurements of length across multiple magnification settings with respect to its intended use.
Table 5. Viewer+ Measurement Results
| Description | Acceptance Criteria | Results |
| --- | --- | --- |
| Length | +/- 0.1mm of pathologist annotation | -0.02 mm < Variance < 0.0000 mm |
| Area | +/- 0.1mm² of pathologist annotation | -0.0180 mm² < Variance < 0.0600 mm² |
K242244 - Page 8 of 9
{8}
4. Human Factors (Usability) Testing
Usability testing was conducted per FDA’s Guidance on Applying Human Factors and Usability Engineering to Medical Devices (2016). A summative human factors test, designed around critical user tasks and use scenarios using multiple representative users (pathologists), was conducted. All tasks associated with reviewing and reporting results for cases including confirmation that all slides belonging to specific cases are reviewed before reporting results were included in the study. The Viewer+ device has been found to be safe and effective for intended users, users and use environments.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K242244 - Page 9 of 9
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.