VYSIS EGR1 FISH PROBE KIT - SC (SPECIMEN CHARACTERIZATION)
Applicant
Abbott Molecular, Inc.
Product Code
PDO · Hematology
Decision Date
Jul 29, 2013
Decision
DENG
Submission Type
Post-NSE
Regulation
21 CFR 864.1870
Device Class
Class 2
Intended Use
The Vysis EGR1 FISH Probe Kit - SC detects the LSI EGR1 probe target on chromosome 5q in bone marrow specimens. The Vysis EGR1 FISH Probe Kit -SC assay results characterize bone marrow specimens from patients with acute myeloid leukemia or myelodysplastic syndrome. The assay results are intended to be interpreted by a qualified pathologist or cytogeneticist. This device is not intended for high-risk uses such as selecting therapy, predicting therapeutic response or disease screening. The use of this product for diagnosis, monitoring or risk assessment has not been established.
Device Story
Vysis EGR1 FISH Probe Kit - SC uses fluorescence in situ hybridization (FISH) to detect EGR1 gene (5q31) and control probe (5p15.2) in bone marrow specimens. Kit includes SpectrumOrange-labeled LSI EGR1 probe, SpectrumGreen-labeled LSI D5S23/D5S721 probe, hybridization buffer, DAPI II counterstain, and NP-40. Procedure involves attaching bone marrow cells to slides, DNA denaturation, hybridization, washing, and DAPI counterstaining. Performed in clinical laboratories by trained personnel. Results viewed via fluorescence microscope; qualified pathologist or cytogeneticist enumerates orange/green signals to determine copy number and identify chromosomal alterations. Expected pattern 2R2G; 5q alteration indicated by 1R2G. Provides characterization of bone marrow specimens to assist clinical evaluation.
Clinical Evidence
No prospective clinical trial conducted. Clinical validity supported by three peer-reviewed literature references (Sun et al. 2010, Galvan et al. 2010, Vance et al. 2007) using the device on bone marrow specimens from AML/MDS patients. Analytical performance: Precision/reproducibility met acceptance criteria across 3 sites/5 days. Analytical sensitivity 99.6% (95% CI: 99.4-99.7) based on 5000 nuclei. Analytical specificity 100% (95% CI: 98-100) based on 200 metaphase loci. Reference limit established at 6% (12 1R2G patterns per 200 nuclei).
Technological Characteristics
FISH DNA probe technology. Components: SpectrumOrange-labeled LSI EGR1 probe (~209 kb), SpectrumGreen-labeled LSI D5S23/D5S721 probe (~561 kb), hybridization buffer, DAPI II counterstain, NP-40, 20X SSC salt. Requires fluorescence microscope with appropriate excitation/emission filters. Manual interpretation. No software/algorithm.
Indications for Use
Indicated for characterization of bone marrow specimens from patients with acute myeloid leukemia (AML) or myelodysplastic syndrome (MDS). Not for therapy selection, therapeutic response prediction, disease screening, diagnosis, monitoring, or risk assessment.
Regulatory Classification
Identification
An early growth response 1 (EGR1) gene fluorescence in-situ hybridization (FISH) test system for specimen characterization is a device intended to detect the EGR1 probe target on chromosome 5q in bone marrow specimens from patients with acute myeloid leukemia (AML) or myelodysplastic syndrome (MDS). The assay results are intended to be interpreted only by a qualified pathologist or cytogeneticist. These devices do not include automated systems that directly report results without review and interpretation by a qualified pathologist or cytogeneticist. These devices also do not include any device intended for use to select patient therapy, predict patient response to therapy, or to screen for disease as well as any device with a claim for a particular diagnosis, prognosis, monitoring, or risk assessment.
Special Controls
EGR1 Gene FISH test system for specimen characterization must comply with the following special controls: 1) Premarket notification submissions must include the following information: i) A detailed description of all probes included in the kit ii) Purpose of each probe iii) Probe molecular specificity iv) Probe specificity v) Probe limits vi) Probe sensitivity vii) Specification of required ancillary reagents, instrumentation and equipment viii) Specification of the specimen collection, processing, storage and slide preparation methods ix) Specification of the assay procedure x) Specification of control elements that are incorporated into the recommended testing procedures xi) Specification of risk mitigation elements: description of all additional procedures. methods, and practices incorporated into the directions for use that mitigate risks associated with testing xii) Specification of the criteria for test result interpretation and reporting xiii) Device analytical sensitivity data xiv) Device analytical specificity data xv) Device reference limit data xvi) Device precision/reproducibility data xvii) Device stability data to include: A) Real-time Stability B) Freeze-Thaw Stability C) Transport and Temperature Stability D) Post-Hybridization Signal Stability E) Photostability of Probe xviii) Documentation that demonstrates the clinical validity of the device. The documentation must include data from clinical studies, a minimum of two peer-reviewed published literature references using the specific device seeking marketing clearance, or both. Documentation for peer-reviewed published literature references cited must include the following elements: A) Documentation that the sponsor's probe was used in the literature reference B) Number & type of specimens C) Target population studied. Target population must include the intended use population D) Upper reference limit F) Range of positive probe results 2) 21 CFR 809.10(b)(12) compliant labeling must include a statement summarizing the data identified in subparagraphs (1)(xiii)-(xviii) and a description of the studies supporting the information, including the pre-specified acceptance criteria for these performance studies, justification for the prespecified acceptance criteria, and whether the prespecified acceptance criteria were met. 3) 21 CFR 809.10 compliant labeling must include: i) A warning that reads "The assay results are intended to be interpreted only by a qualified pathologist or cytogeneticist." ii) A warning that reads "This device is not for high-risk uses such as selecting therapy, predicting therapeutic response or disease screening." iii) A warning that reads "The use of this device for diagnosis, monitoring or risk assessment has not been established."
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must also include the following information:
(i) A detailed description of all probes included in the kit;
(ii) Purpose of each probe;
(iii) Probe molecular specificity;
(iv) Probe specificity;
(v) Probe limits;
(vi) Probe sensitivity;
(vii) Specification of required ancillary reagents, instrumentation, and equipment;
(viii) Specification of the specimen collection, processing, storage and slide preparation methods;
(ix) Specification of the assay procedure;
(x) Specification of control elements that are incorporated into the recommended testing procedures;
(xi) Specification of risk mitigation elements: Description of all additional procedures, methods, and practices incorporated into the directions for use that mitigate risks associated with testing;
(xii) Specification of the criteria for test result interpretation and reporting;
(xiii) Device analytical sensitivity data;
(xiv) Device analytical specificity data;
(xv) Device reference limit data;
(xvi) Device precision/reproducibility data;
(xvii) Device stability data to include:
(A) Real-time stability,
(B) Freeze-thaw stability,
(C) Transport and temperature stability,
(D) Post-hybridization signal stability,
(E) Photostability of probe, and
(xviii) Documentation that demonstrates the clinical validity of the device. The documentation must include data from clinical studies, a minimum of two peer-reviewed published literature references using the specific device seeking marketing clearance, or both. Documentation for the clinical studies and peer-reviewed published literature references cited must include the following elements:
(A) Documentation that the sponsor's probe was used in the literature reference,
(B) Number and type of specimens,
(C) Target population studied,
(D) Upper reference limit, and
(E) Range of positive probe results.
(2) Your § 809.10(b)(12) of this chapter compliant labeling must include a statement summarizing the data identified in paragraphs (b)(1)(xiii) through (xviii) of this section and a description of the studies supporting the information, including the pre-specified acceptance criteria for these performance studies, justification for the pre-specified acceptance criteria, and whether the pre-specified acceptance criteria were met.
(3) Your § 809.10 of this chapter compliant labeling must include:
(i) A warning that reads “The assay results are intended to be interpreted only by a qualified pathologist or cytogeneticist.”
(ii) A warning that reads “This device is not for high-risk uses such as selecting therapy, predicting therapeutic response or disease screening.”
(iii) A warning that reads “The use of this device for diagnosis, monitoring or risk assessment has not been established.”
Related Devices
K091960 — VYSIS EGR1 FISH PROBE KIT (VYSIS LSI EGR1 SPECTRUM/ORANGE/D5S23, D5S721, SPECTRUM GREEN PROBE), MODEL 4N37-020 · Abbott Molecular, Inc. · Aug 29, 2011
K131508 — VYSIS D7S486/CEP 7 FISH PROBE KIT · Abbott Molecular, Inc. · Sep 13, 2013
DEN170070 — MLL (KMT2A) Breakapart FISH Probe Kit; AML1 (RUNX1) Breakapart FISH Probe Kit, P53 (TP53) Deletion FISH Probe Kit; EVI1 (MECOM) Breakapart FISH Probe Kit, Del(20q) Deletion FISH Probe Kit; AML1/ETO (RUNX1/RUNXIT1)) Translocation, Dual Fusi, CBFB (CBFB) /MYH11 Translocation, Dual Fusion FISH Probe Kit,Del(5q)Deletion FISH Probe Kit; Del(7q)Deletion FISH Probe Kit · Cytocell, Ltd. · Dec 21, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
# EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR VYSIS EGR1 FISH PROBE KIT - SC (SPECIMEN CHARACTERIZATION)
#### Decision Summary
| A. 510(k) Number | K123951 |
|-------------------------------|--------------------------------------------------------------------|
| B. Purpose of the Submission: | Clearance of new assay |
| C. Measurand: | LSI EGR1 probe target on chromosome 5q in<br>bone marrow specimens |
| D. Type of Test: | Flourescence in-situ hybridization (FISH) |
| E. Applicant: | Abbott Molecular Inc. |
| F. Device Name: | Vysis EGR1 FISH Probe Kit - SC (Specimen<br>Characterization) |
#### G. Regulatory Information:
FDA identifies this type of device as:
An early growth response 1 (EGR1) gene fluorescence in-situ hybridization (FISH) test system for specimen characterization is a device intended to detect the EGR1 probe target on chromosome 5q in bone marrow specimens from patients with acute myeloid leukemia (AML) or myelodysplastic syndrome (MDS). The assay results are intended to be interpreted only by a qualified pathologist or cytogeneticist. These devices do not include automated systems that directly report results without review and interpretation by a qualified pathologist or cytogeneticist. These devices also do not include any device intended for use to select patient therapy, predict patient response to therapy or to screen for disease as well as any device with a claim for a particular diagnosis, prognosis, monitoring or risk assessment.
| 1. New Regulation Number: | 21 CFR 864.1870 |
|---------------------------|----------------------------------------|
| 2. Classification: | Class II (special controls) |
| 3. Product code: | PDO |
| 4. Panel: | Hematology and Pathology Devices Panel |
# H. Intended Use:
{1}------------------------------------------------
#### 1. Intended use(s):
The Vysis EGR1 FISH Probe Kit - SC detects the LSI EGR1 probe target on chromosome 5q in bone marrow specimens. The Vysis EGR1 FISH Probe Kit -SC assay results characterize bone marrow specimens from patients with acute myeloid leukemia or myelodysplastic syndrome. The assay results are intended to be interpreted by a qualified pathologist or cytogeneticist. This device is not intended for high-risk uses such as selecting therapy, predicting therapeutic response or disease screening. The use of this product for diagnosis, monitoring or risk assessment has not been established.
### 2. Indication(s) for use:
Same as intended use
- 3. Special conditions for use statement(s):
Warnings included in the labeling are as follows:
- The assay results are intended to be interpreted only by a qualified pathologist or cytogeneticist.
- · The Vysis EGR1 FISH Probe Kit SC is not for high-risk uses such as selecting therapy, predicting therapeutic response or disease screening.
- · The use of this product for diagnosis, monitoring or risk assessment has not been established.
- · Caution: Federal law restricts this device to sale by or on the order of a physician or other practitioner licensed by the law of the State in which he practices, to use or order the use of the device.
- 4. Special instrument requirements:
Fluorescence microscope equipped with appropriate excitation and emission filters
# I. Device Description:
The Vysis EGR1 FISH Probe Kit - Specimen Characterization uses fluorescence in situ hybridization (FISH) DNA probe technology to detect probe target LSI EGR1 (containing early growth response 1 gene; location chromosome 5g31). The LSI D5S23, D5S721 probe (location chromosome 5p15.2) serves as a control.
The Vysis EGR1 FISH Probe Kit – SC (List No. 04N37-001) consists of the following components which are sufficient to process 20 assays:
- 1. Vysis LSI EGR1 SpectrumOrange/D5S23, D5S721 Spectrum Green Probes
- 2. Vysis LSI/WCP Hybridization Buffer
- 3. DAPI II Counterstain
- 4. NP-40
{2}------------------------------------------------
#### 5. 20X SSC Salt
Items 2 through 5 above are general purposes reagents.
### DNA Probes: Vysis LSI EGR1 SpectrumOrange/ D5S23, D5S721 SpectrumGreen.
- The SpectrumOrange-labeled LSI EGR1 probe, approximately 209 kb in a. length (chr5:137654208-137862738; February 2009 Assembly; UCSC Human Genome Browser) is located at 5g31 and contains the complete EGR1 gene.
- b. The SpectrumGreen-labeled LSI D5S23, D5S721 probe, approximately 561 kb in length (chr5:9397109-9958407; February 2009 Assembly; UCSC Human Genome Browser) is located at 5p15.2.
### J. Substantial Equivalence Information:
- 1. Predicate device name(s):
No predicate device exists for this intended use
- 2. Predicate 510(k) number(s):
Not applicable
- 3. Comparison with predicate:
Not Applicable
# K. Standard/Guidance Document Referenced (if applicable):
Not Applicable
# L. Test Principle:
Alterations in chromosome 5 are common aberrations in myelodysplastic syndromes (MDS) and acute myeloid leukemia (AML). Chromosome band 5q31 contains the early growth response 1 gene (EGR1). The Vysis locus specific identifier (LSI) EGR1 SpectrumOrange Probe and D5S23, D5S721 SpectrumGreen Probe are components of the Vysis EGR1 FISH Probe Kit, and are used to detect the target EGR 1 using FISH DNA probe technology. The results of this assay are used to characterize bone marrow specimens.
Bone marrow cells from AML patients attached to microscope slides using standard cytogenetic procedures are used for this assay. The resulting cellular DNA is denatured to single-stranded form and subsequently allowed to hybridize with the LSI EGR1 and LSI D5S23, D5S721 probes. Following hybridization, the unbound probe is removed by a series of washes, and the nuclei are counterstained with DAPI, a DNA-specific stain that fluoresces blue. Hybridization of the LSI EGR1 and LSI D5S23, D5S721 probes is viewed using a fluorescence microscope equipped with
{3}------------------------------------------------
appropriate excitation and emission filters, allowing visualization of the orange and green fluorescent signals.
The expected signal pattern in a cell with two copies of the EGR1 gene covered by the probes using the Vysis EGR1 FISH Probe Kit is 2 orange and 2 green signals (2R2G). In a cell with the 5q alteration, one orange signal (LSI EGR1) and two green signals (LSI D5S23, D5S721) will be expected. Enumeration of the orange LSI EGR 1 and green LSI D5S23, D5S721 signals provide a mechanism for determining absolute copy number of the probe targets and the presence of chromosomal alterations of interest.
# M. Performance Characteristics (if/when applicable):
The sponsor provided the following information to support the analytical performance of the device:
- 1. Analytical performance:
- a. Precision/Reproducibility: Repeatability and reproducibility of the Vysis EGR1 FISH Probe Kit were tested as shown in the table below:
{4}------------------------------------------------
| Study | Study Protocol | Conclusion |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Precision:<br>Intra-Day and<br>Inter-Day | • 6 Bone marrow specimens<br>- 2 high positive (HP1 & 2)<br>(69.3%, 49.9% positivity)<br>- 2 low positive (LP1 & 2)<br>(12% positivity)<br>- 2 negative (Neg1 & 2)<br><br>• 3 test sites<br>• 5 different testing days<br>• 1 lot of probe<br><br>• Red and green signal<br>patterns of 200 nuclei<br>evaluated by 2<br>technologists (each<br>technologist evaluated 100<br>nuclei per panel member)<br><br>• Pre-specified acceptance criteria:<br>Agreement with the expected result<br>of greater than or equal to 90% for<br>the high positive specimen<br>category for each site and 90% for<br>the negative specimen category<br>across all sites with no more than 3<br>discordant results occurring at one<br>site | Acceptance criteria met |
{5}------------------------------------------------
| Study | Study Protocol | Conclusion |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| <b>Reproducibility:</b><br>Inter-Site | • Testing configuration same<br>as above<br>• In addition, 2 replicates of<br>each panel member were<br>tested<br>• Pre-specified acceptance<br>criteria: Same as above | Acceptance criteria met |
| Lot to Lot<br><b>Reproducibility</b> | • 6 Bone marrow specimens<br>- 2 high positive (HP1 & 2)<br>(69.3%, 49.9% positivity)<br>- 2 low positive (LP1 & 2)<br>(12% positivity)<br>- 2 negative (Neg1 & 2)<br>• 3 lots of probe<br>• 4 replicates<br>• 1 site<br>• 2 operators per site<br>• Pre-specified acceptance<br>criteria: Same as above | Acceptance criteria met |
# Between Site Analysis of Variance
| Sample Type | N | Mean %a | SDb |
|-----------------|----|---------|------|
| High Positive 1 | 30 | 70.0 | 5.44 |
| High Positive 2 | 30 | 47.6 | 0.74 |
| Low Positive 1 | 30 | 18.1 | 1.03 |
| Low Positive 2 | 30 | 14.9 | 0.00 |
| Negative 1 | 30 | 0.7 | 0.99 |
| Negative 2 | 30 | 0.9 | 1.5 |
ª Percentage of cells with 1R2G signal pattern
bSD = standard deviation
| Sample Type | N | Mean %a | Within-Day<br>Component<br>SDb | Between-Day<br>Component<br>SD |
|-----------------|----|---------|--------------------------------|--------------------------------|
| High Positive 1 | 30 | 70.0 | 3.28 | 4.01 |
| High Positive 2 | 30 | 47.6 | 5.56 | 0.00 |
| Low Positive 1 | 30 | 18.1 | 3.00 | 3.82 |
| Low Positive 2 | 30 | 14.9 | 3.25 | 1.54 |
| Negative 1 | 30 | 0.7 | 0.71 | 0.00 |
| Negative 2 | 30 | 0.9 | 0.66 | 1.42 |
#### Within-day/Between-day Analysis of Variance
ª Percentage of cells with 1R2G signal pattern
bSD = standard deviation
{6}------------------------------------------------
| Sample Type | N | Mean %a | SDb |
|-----------------|----|---------|------|
| High Positive 1 | 12 | 66.2 | 0.00 |
| High Positive 2 | 12 | 47.4 | 3.04 |
| Low Positive 1 | 12 | 12.7 | 0.00 |
| Low Positive 2 | 12 | 12.3 | 1.12 |
| Negative 1 | 12 | 0.0 | 0.00 |
| Negative 2 | 12 | 0.1 | 0.20 |
Lot-to-Lot Analysis of Variance
ª Percentage of cells with 1R2G signal pattern
bSD = standard deviation
- b. Linearity/assay reportable range:
Not applicable
- c. Traceability, Stability, Expected values (controls, calibrators, or methods):
#### Kit stability
The kit stability was assessed as detailed in the table below:
| Study | StudyProtocol | Conclusion |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Real-Time<br>Stability | • Attributes evaluated: signal<br>intensity, target<br>background, cross-hybridization,<br>specificity, overall readability<br>• 3 lots of the device<br>• 3 specimens<br>• The Vysis EGRI FISH Probe Kit<br>dating is determined by the<br>component with the shortest<br>expiration dating.<br>• Pre-specified acceptance criteria:<br>Acceptable quality of all attributes<br>for all samples tested | Acceptance criteria met for<br>12 month stability |
| In-Use Freeze-<br>Thaw Stability | • A series of 20 freeze-thaw cycles<br>was performed on the probes,<br>hybridization buffer and DAPI II<br>counterstain.<br>• 1 lot of the device<br>• 3 specimens<br>• Attributes evaluated: Same<br>as above<br>• Pre-specified acceptance criteria:<br>Same as above | Acceptance criteria met<br>throughout and at the end<br>of 20 freeze-thaw cycles |
| Transport and<br>Temperature<br>Extreme | • Device components were removed<br>from -20°C and cycled for 48 hours<br>on dry ice, 72 hours at 25°C, 72 | Acceptance criteria met |
{7}------------------------------------------------
| Stability | hours at 40°C, and -20°C for 24<br>hours prior to testing<br>• 1 lot of the device<br>• 3 specimens<br>• Attributes evaluated: Same as<br>above<br>• Pre-specified acceptance criteria:<br>Same as above | |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Post-<br>hybridization<br>Signal Stability | • A single slide from each of 3 bone<br>marrow specimens was tested at<br>baseline and then stored at -20°C ±<br>10°C protected from light<br>• The samples were tested at three<br>time points -Day 7, Day 14, and<br>Day 25<br>• 1 lot of the device<br>• Attributes evaluated: Same<br>as above<br>• Pre-specified acceptance criteria:<br>Same as above | Acceptance criteria met for<br>Post-hybridization stability<br>of 3 weeks |
| Probe<br>Photostability | • The device was exposed<br>continuously under white<br>fluorescent light (to mimic<br>standard laboratory conditions) at<br>15-30°C for different time lengths<br>(0 hours, 3 hours, 8 hours, 24 hours<br>and 48 hours)<br>• 1 lot of the device<br>• Attributes evaluated: Same as<br>above<br>• Pre-specified acceptance criteria:<br>Same as above | Acceptance criteria met for<br>Photo-stability of 48 hours |
# Controls:
The Vysis EGR1 FISH Probe Kit also contains the LSI D5S23, D5S721 probe (location chromosome 5p15.2) which serves as a control. Observation of at least 1R signal and 1G signal, assures that the assay conditions are adequate to allow both the orange and green labeled probes to bind properly to their respective targets.
- d. Detection limit:
The analytical sensitivity of the Vysis LSI EGRI SpectrumOrange/D5S23, D5S721SpectrumGreen probes was established using interphase nuclei prepared from 25 bone marrow specimens that were either karyotypically normal or 5pl5 and 5q31 deletion-free. The orange and green signal patterns of nuclei for these 25 specimens were evaluated by two technologists. Each
{8}------------------------------------------------
technologist evaluated 100 nuclei per specimen for a total of 200 nuclei per specimen and 5000 scoreable nuclei from normal specimens. The analytical sensitivity was calculated as the percentage of scoreable interphase nuclei with the expected 2 red/2 green signal pattern.
| Probe | Total Number<br>of Nuclei<br>Scored | Number of Nuclei<br>With Expected<br>Signal Pattern | Analytical Sensitivity<br>(95% Confidence Interval) |
|--------------------------------|-------------------------------------|-----------------------------------------------------|-----------------------------------------------------|
| Vysis<br>EGR1/D5S23,<br>D5S721 | 5000 | 4979 | 99.6% (99.4, 99.7) |
### e. Analytical specificity:
Analytical specificity is defined as the percentage of signals that hybridize to the correct locus and no other location. The analytical specificity of the Vysis LSI EGR1 SpectrumOrange/D5S23, D5S721 SpectrumGreen probes for their respective chromosome target loci was established using metaphase chromosomes prepared from peripheral blood cultures of five karyotypically normal males that were pooled prior to dropping on microscope slides. The hybridization location of each FISH signal on chromosomes of 100 consecutive metaphase nuclei was evaluated by one technologist for a total of 200 target loci.
For each probe and sample, the number of metaphase chromosome FISH signals hybridized to the correct locus and the number of metaphase chromosome FISH signals hybridized to the incorrect locus were enumerated.
The analytical specificity of each probe was calculated as the number of metaphase chromosome FISH signals hybridized to the correct locus divided by the total number of metaphase chromosome FISH signals hybridized and multiplied by 100 to give a percentage. The analytical specificity of the Vysis LSI EGR1 SpectrumOrange/D5S23, D5S721 SpectrumGreen Probes was 100%.
| Probe | Target | Total No. of<br>Metaphase<br>Chromosome<br>Hybridized | No. of<br>Metaphase<br>Chromosome<br>Hybridized to<br>Correct Locus | Analytical Specificity<br>(95% Confidence<br>Intervals) |
|---------------|--------|-------------------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------------|
| D5S23, D5S721 | 5p15.2 | 200 | 200 | 100%<br>(98, 100) |
| EGR 1 | 5q31 | 200 | 200 | 100%<br>(98, 100) |
f. Assay cut-off:
Same as reference range (see #5 below)
{9}------------------------------------------------
### 2. Comparison studies:
- a. Method comparison with predicate device:
Not applicable
- b. Matrix comparison:
Not applicable
- 3. Clinical studies:
- a. Clinical Sensitivity:
Not applicable
- b. Clinical specificity:
Not applicable
- c. Other clinical supportive data (when a. and b. are not applicable):
The sponsor provided three peer-reviewed published papers to support the clinical validity of the device in characterizing bone marrow specimens from patients with AML. All three papers used the Abbott Vysis EGR1 FISH Probe Kit. The information is listed in the table below:
{10}------------------------------------------------
| Conditions | Data Source 1<br>Sun et. al | Data Source 2<br>Galvan et. al | Data Source 3<br>Vance et. al |
|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------|----------------------------------------------------------------------|---------------------------------------------------------------------|
| Was the specific<br>device under<br>review in the<br>submission used in<br>the study? | Yes | Yes | Yes |
| Was the specimen<br>type in the study<br>representative of<br>the claimed<br>specimen type(s) | Yes | Yes | Yes |
| Target population<br>(disease status) | Known or<br>suspected del(5q)<br>having MDS or<br>AML | MDS, therapy-<br>related MDS (t-MDS) and AML<br>patients | AML |
| Upper reference<br>limit (percentage<br>and per 200 nuclei) | 6% or 12 1R2G<br>patterns per 200<br>scoreable<br>interphase nuclei | 6% or 12 1R2G<br>patterns per 200<br>scoreable<br>interphase nuclei | 6% or 12 1R2G<br>patterns per 200<br>scoreable<br>interphase nuclei |
| Total Number of<br>specimens tested<br>for each claimed<br>type | 320 bone marrow<br>specimens | 28 bone marrow<br>specimens | 181 bone marrow<br>specimens* |
| Number of<br>specimens with a<br>positive probe<br>result [5q- (1R2G)] | 66 | 23<br>MDS - 6<br>t-MDS - 3<br>t-MDS&MM – 1<br>AML - 13 | 8<br>AML - 8 |
| Range of positive<br>probe results | Not available | MDS: 35 - 81.5%<br>t-MDS: 25 - 75%<br>t-MDS&MM: 76%<br>AML: 20 - 99% | 45 - 91 % |
MDS – Myelodysplastic syndrome
AML – Acute myeloid leukemia
t-MDS – Therapy related MDS
MM – Multiple Myeloma
* Line data unpublished, but this data was provided in the 510(k) submission
Sun Y, Cook JR. Fluorescence in situ hybridization for del(5q) in myelodysplasia/acute myeloid lukemia: Comparison of EGR1 vs. CSF1R probes and diagnostic yield over metaphase cytogenetics alone. Leuk Res 2010; 34: 340-343.
Galvan AB, Mallo M, Arenillas L, et al. Does monosomy 5 really exist in myelodysplastic syndromes and acute myeloid leukemia? Leuk Res 2010; 34: 1242– 1245.
Vance GH, Haesook K, Hicks GA, et al. Utility of interphase FISH to stratify patients into cytogenetic
{11}------------------------------------------------
risk categories at diagnosis of AML in an Eastern Cooperative Oncology Group (ECOG) clinical trial (E1900). Leuk Res 2007; 31:605-609.
- 4. Clinical cut-off:
Not applicable
- 5. Expected values/Reference range:
The upper reference limit is defined as the maximum quantity of scoreable interphase nuclei with an altered signal pattern from either karyotypically normal specimens or 5p15.2 and 5g31 deletion-free specimens. The upper reference limit is expressed in terms of a percentage or the actual number of atypical nuclear FISH signal pattern per the standard number of nuclei tested.
The upper reference limit for this assay is 6% or 12 1R2G patterns per 200 scoreable interphase nuclei. Published literature was used to establish the reference range for this assay.
In order to validate the 6% upper reference limit of the Vysis EGR 1 FISH Probe Kit, the assay was performed on interphase nuclei from 25 bone marrow specimens from either karyotypically normal specimens or 5p15.2 and 5q31 deletion-free specimens. The signal patterns of 200 nuclei were evaluated by counting the number of orange and green signals. Each of two technologists evaluated 100 nuclei per specimen. Among the 25 normal specimens, none produced 1R2G signals at or above the 6% upper reference limit.
# N. Other Supportive Instrument Characteristics Data Not Covered In The "Performance Characteristics" Section above:
None
# O. Proposed Labeling:
Labeling satisfies the requirements of 21 CFR 809.10, 21 CFR 801.109, including an appropriate prescription statement as required by 21 CFR 801.109(b), and the special controls for this type of device.
| Identified Potential Risk | Required Mitigations |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| False negative result | 1) Premarket notification submissions must<br>also include the following information:<br>i) A detailed description of all probes<br>included in the kit<br>ii) Purpose of each probe<br>iii) Probe molecular specificity<br>iv) Probe specificity<br>v) Probe limits |
| vi) | Probe sensitivity |
| vii) | Specification of required ancillary<br>reagents, instrumentation and<br>equipment |
| viii) | Specification of the specimen<br>collection, processing, storage and<br>slide preparation methods |
| ix) | Specification of the assay procedure |
| x) | Specification of control elements<br>that are incorporated into the<br>recommended testing procedures |
| xi) | Specification of risk mitigation<br>elements: description of all<br>additional procedures, methods, and<br>practices incorporated into the<br>directions for use that mitigate risks<br>associated with testing |
| xii) | Specification of the criteria for test<br>result interpretation and reporting |
| xiii) | Device analytical sensitivity data |
| xiv) | Device analytical specificity data |
| xv) | Device reference limit data |
| xvi) | Device precision/reproducibility<br>data |
| xvii) | Device stability data to include:<br>A) Real-time Stability<br>B) Freeze-Thaw Stability<br>C) Transport and Temperature<br>Stability<br>D)Post-Hybridization Signal<br>Stability<br>E) Photostability of Probe |
| xviii) | Documentation that demonstrates the<br>clinical validity of the device. The<br>documentation must include data<br>from clinical studies, a minimum of<br>two peer-reviewed published<br>literature references using the<br>specific device seeking marketing<br>clearance, or both. Documentation<br>for the clinical studies and peer-<br>reviewed published literature<br>references cited must include the<br>following elements:<br>A) Documentation that the |
| | literature reference<br>B) Number & type of specimens<br>C) Target population studied<br>D) Upper reference limit<br>E) Range of positive probe results |
| | 2) Your 21 CFR 809.10(b)(12) compliant labeling must include a statement summarizing the data identified in subparagraphs (1)(xiii)-(xviii) and a description of the studies supporting the information, including the pre-specified acceptance criteria for these performance studies, justification for the pre-specified acceptance criteria, and whether the pre-specified acceptance criteria were met. |
| | 3) Your 809.10 compliant labeling must include:<br>i) A warning that reads “The assay results are intended to be interpreted only by a qualified pathologist or cytogeneticist.”<br>ii) A warning that reads “This device is not for high-risk uses such as selecting therapy, predicting therapeutic response or disease screening.”<br>iii) A warning that reads “The use of this device for diagnosis, monitoring or risk assessment has not been established.” |
| False Positive Result | 1) Premarket notification submissions must also include the following information:<br>i) A detailed description of all probes included in the kit<br>ii) Purpose of each probe<br>iii) Probe molecular specificity<br>iv) Probe specificity<br>v) Probe limits<br>vi) Probe sensitivity<br>vii) Specification of required ancillary reagents, instrumentation and equipment<br>viii) Specification of the specimen collection, processing, storage and slide preparation methods |
| ix) | Specification of the assay procedure |
| x) | Specification of control elements that are incorporated into the recommended testing procedures |
| xi) | Specification of risk mitigation elements: description of all additional procedures, methods, and practices incorporated into the directions for use that mitigate risks associated with testing |
| xii) | Specification of the criteria for test result interpretation and reporting |
| xiii) | Device analytical sensitivity data |
| xiv) | Device analytical specificity data |
| xv) | Device reference limit data |
| xvi) | Device precision/reproducibility data |
| xvii) | Device stability data to include:<br>A) Real-time Stability<br>B) Freeze-Thaw Stability<br>C) Transport and Temperature Stability<br>D) Post-Hybridization Signal Stability<br>E) Photostability of Probe |
| xviii) | Documentation that demonstrates the clinical validity of the device. The documentation must include data from clinical studies, a minimum of two peer-reviewed published literature references using the specific device seeking marketing clearance, or both. Documentation for the clinical studies and peer-reviewed published literature references cited must include the following elements:<br>A) Documentation that the sponsor's probe was used in the literature reference<br>B) Number & type of specimens<br>C) Target population studied<br>D) Upper reference limit<br>E) Range of positive probe results |
| 2) | Your 21 CFR 809.10(b)(12) compliant labeling must include a statement summarizing the data identified in |
| | description of the studies supporting the information, including the pre-specified acceptance criteria for these performance studies, justification for the pre-specified acceptance criteria, and whether the pre-specified acceptance criteria were met. |
| | 3) Your 809.10 compliant labeling must include: |
| i) | A warning that reads “The assay results are intended to be interpreted only by a qualified pathologist or cytogeneticist.” |
| ii) | A warning that reads “This device is not for high-risk uses such as selecting therapy, predicting therapeutic response or disease screening.” |
| iii) | A warning that reads “The use of this device for diagnosis, monitoring or risk assessment has not been established.” |
# P Risks to Health and Mitigation Measures:
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
# Q. Benefit/Risk Analysis:
| Summary of the<br>Benefits | Summary of<br>the Risks | Summary of Other<br>Factors | Conclusions |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AML and MDS<br>patients may<br>potentially<br>benefit in the<br>intended use<br>population by use<br>of the device to<br>characterize bone<br>marrow<br>specimens with<br>assay results | Erroneous<br>device results<br>could adversely<br>influence<br>expectation of<br>more or less<br>favorable<br>clinical course<br>of AML or<br>MDS patients<br>due to false | In addition to cited<br>clinical studies using<br>the device probes,<br>analytical evaluation<br>and labeling supports<br>the intended use. De<br>novo regulatory<br>approach leverages<br>device use by a<br>qualified pathologist or<br>cytogeneticist in the | Yes. Based on<br>the supporting<br>clinical studies<br>for the<br>diagnostic<br>device along<br>with review of<br>the analytical<br>performance and<br>labeling, the<br>probable |
| interpreted by a<br>qualified<br>pathologist or<br>cytogeneticist. | negative or<br>false positive<br>results. | context of<br>histopathological<br>evaluation (e.g.,<br>immunohistochemistry). | benefits<br>outweigh the<br>probable risks. |
# R. Conclusion:
{16}------------------------------------------------
The request for Evaluation of Automatic Class III Designation for this device is granted. The device is classified as Class II under regulation 21 CFR 864.1870 with special controls. The device is classified under the following:
| Product Code: | PDO |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Type: | Early growth response 1 (EGR1) gene fluorescence in-situ<br>hybridization (FISH) test system for specimen<br>characterization |
| Class: | II (special controls) |
| Regulation: | 21 CFR 864.1870 |
| | (a) IDENTIFICATION: An early growth response 1<br>(EGR1) gene fluorescence in-situ hybridization (FISH) test<br>system for specimen characterization is a device intended<br>to detect the EGR1 probe target on chromosome 5q in bone<br>marrow specimens from patients with acute myeloid<br>leukemia (AML) or myelodysplastic syndrome (MDS).<br>The assay results are intended to be interpreted only by a<br>qualified pathologist or cytogeneticist. These devices do<br>not include automated systems that directly report results<br>without review and interpretation by a qualified pathologist<br>or cytogeneticist. These devices also do not include any<br>device intended for use to select patient therapy, predict<br>patient response to therapy or to screen for disease as well<br>as any device with a claim for a particular diagnosis,<br>prognosis, monitoring or risk assessment. |
| | (b) CLASSIFICATION: Class II (special controls). EGR1<br>Gene FISH test system for specimen characterization must<br>comply with the following special controls:<br><br>1) Premarket notification submissions must include the<br>following information:<br>i) A detailed description of all probes included in the kit<br>ii) Purpose of each probe<br>iii) Probe molecular specificity<br>iv) Probe specificity<br>v) Probe limits<br>vi) Probe sensitivity<br>vii) Specification of required ancillary reagents,<br>instrumentation and equipment |
{17}------------------------------------------------
- viii) Specification of the specimen collection, processing, storage and slide preparation methods
- Specification of the assay procedure ix)
- Specification of control elements that are x) incorporated into the recommended testing procedures
- xi) Specification of risk mitigation elements: description of all additional procedures. methods, and practices incorporated into the directions for use that mitigate risks associated with testing
- xii) Specification of the criteria for test result interpretation and reporting
- xiii) Device analytical sensitivity data
- xiv) Device analytical specificity data
- XV) Device reference limit data
- Device precision/reproducibility data xvi)
- Device stability data to include: xvii)
- A) Real-time Stability
- B) Freeze-Thaw Stability
- C) Transport and Temperature Stability
- D) Post-Hybridization Signal Stability
- E) Photostability of Probe
- xviii) Documentation that demonstrates the clinical validity of the device. The documentation must include data from clinical studies, a minimum of two peer-reviewed published literature references using the specific device seeking marketing clearance, or both. Documentation for peer-reviewed published literature references cited must include the following elements:
- A) Documentation that the sponsor's probe was used in the literature reference
- B) Number & type of specimens
- C) Target population studied. Target population must include the intended use population
- D) Upper reference limit
- F) Range of positive probe results
- 2) 21 CFR 809.10(b)(12) compliant labeling must include a statement summarizing the data identified in subparagraphs (1)(xiii)-(xviii) and a description of the studies supporting the information, including
{18}------------------------------------------------
the pre-specified acceptance criteria for these performance studies, justification for the prespecified acceptance criteria, and whether the prespecified acceptance criteria were met.
- 3) 21 CFR 809.10 compliant labeling must include:
- i) A warning that reads "The assay results are intended to be interpreted only by a qualified pathologist or cytogeneticist."
- ii) A warning that reads "This device is not for high-risk uses such as selecting therapy, predicting therapeutic response or disease screening."
- iii) A warning that reads "The use of this device for diagnosis, monitoring or risk assessment has not been established."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.