APERIO EPATHOLOGY EIHC IVD SYSTEM

K141109 · Leica Biosystems Imaging, Inc. · NOT · Jul 29, 2014 · Hematology

Device Facts

Record IDK141109
Device NameAPERIO EPATHOLOGY EIHC IVD SYSTEM
ApplicantLeica Biosystems Imaging, Inc.
Product CodeNOT · Hematology
Decision DateJul 29, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.1860
Device ClassClass 2

Intended Use

The Aperio ePathology eIHC IVD System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape. The IHC HER2 Image Analysis application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffinembedded neoplastic tissue. The IHC HER2 Image Analysis application is intended for use as an accessory to the Dako HercepTest™ to aid in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded neoplastic tissue. When used with the Dako HercepTest™, it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered. The IHC ER Image Analysis application is intended for use as an aid to the pathologist in the detection and quantitative measurement of ER (Estrogen Receptor) in formalin-fixed paraffin-embedded neoplastic tissue. The IHC PR Image Analysis application is intended for use as an aid to the pathologist in the detection and quantitation measurement of PR (Progesterone Receptor) in formalin-fixed, paraffin-embedded neoplastic tissue. It is indicated for use as an aid in the management, prognosis, and prediction of therapy outcomes of breast cancer.

Device Story

System automates digital slide creation, management, viewing, and analysis for pathology. Components include digital microscope slide scanner (AT Turbo or CS2 models), computer, monitor, and eSlideManager software. Operator loads glass slides into carrier; scanner uses linear-array technique with 20x objective to generate high-resolution, seamless digital images. Software stores images in Digital Slide Repository (DSR) with SQL database metadata. Pathologists view images on monitors and use integrated image analysis applications to detect and quantify HER2, ER, and PR protein expression in breast tissue. System acts as adjunctive tool to assist pathologist reproducibility; results aid in breast cancer management and therapy prediction. Used in clinical laboratory settings by technicians and pathologists.

Clinical Evidence

No clinical tests submitted. Evidence consists of bench-level system verification, validation, and performance testing. Performance testing evaluated concordance of HER2, ER, and PR scores between updated ScanScope models and the predicate device using formalin-fixed, paraffin-embedded breast tissue slides. HER2 results analyzed via trichotomous scoring (0/1+, 2+, 3+); ER/PR results analyzed via dichotomous (positive/negative) analysis at 1% and 10% thresholds. Acceptance criteria based on percent agreement with exact 95% lower confidence intervals were met.

Technological Characteristics

System includes automated digital microscope slide scanner, computer, color monitor, and keyboard. Employs linear-array scanning with 20x objective. Energy source: 100vAC–240vAC, 50Hz/60Hz. Connectivity: Networked via eSlideManager server (SQL database, RAID storage). Software performs colorimetric and morphometric pattern recognition for cell detection. Sterilization: Not applicable (device is diagnostic imaging system).

Indications for Use

Indicated for use as an aid to pathologists in the display, detection, counting, and classification of tissues and cells in formalin-fixed, paraffin-embedded neoplastic breast tissue. Specifically used for semi-quantitative measurement of HER2/neu and quantitative measurement of Estrogen Receptor (ER) and Progesterone Receptor (PR) to aid in the management, prognosis, and prediction of therapy outcomes for breast cancer patients.

Regulatory Classification

Identification

Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.

Special Controls

(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 29, 2014 Leica Biosystems Imaging, Inc. Christine Kishi Quality Assurance Specialist 1360 Park Center Dr. Vista, CA 92081 Re: k141109 Trade/Device Name: Aperio ePathology eIHC IVD System, AT Turbo and CS2 models Regulation Number: 21 CFR 864.1860 Regulation Name: Immunohistochemistry reagents and kits Regulatory Class: Class II Product Code: NOT, NQN Dated: April 29, 2014 Received: April 30, 2014 Dear Ms. Kishi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Reena Philip -S Reena Philip, Ph.D. Director Division of Molecular Genetics and Pathology Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/0 description: The image features the logo for Leica Biosystems. The word "Leica" is written in a stylized, red font with a distinctive curve and flourish. Below "Leica", the word "BIOSYSTEMS" is written in a bold, black sans-serif font, centered beneath the Leica logo. ## Indication for Use 510(k) Number (if known): _K141109 Device Name: Aperio ePathology eIHC IVD System Indication for Use: The Aperio ePathology eIHC IVD System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape. The IHC HER2 Image Analysis application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffinembedded neoplastic tissue. The IHC HER2 Image Analysis application is intended for use as an accessory to the Dako HercepTest™ to aid in the detection and semi-quantitative measurement of Her2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded neoplastic tissue. When used with the Dako HercepTest™, it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN®(Trastuzumab) treatment is being considered. Note: The IHC HER2 Image Analysis application is an adjunctive computer-assisted methodology to assist the reproducibility of a qualified pathologist in the acquisition and measurement of images from microscope slides of breast cancer specimens stained for the presence of HER-2 receptor protein. The IHC HER2 Image Analysis application is intended to be used on images viewed on a computer monitor. The accuracy of the test result depends upon the quality of the immunohistochemical staining. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for the Dako HercepTest™ to assure the validity of the IHC HER2 Image Analysis application assisted HER-2/neu score. The actual correlation of the Dako HercepTest™ to Herceptin® clinical outcome has not been established. Prescription Use _ X And/Or Over the Counter Use (21 CFR Part 801 Subpart D) (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) {3}------------------------------------------------ #### 510(k) Number (if known): K141109 Device Name: Aperio ePathology eIHC IVD System Indication for Use: The Aperio ePathology eIHC IVD System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape. The IHC ER Image Analysis application is intended for use as an aid to the pathologist in the detection and quantitative measurement of ER (Estrogen Receptor) in formalin-fixed paraffin-embedded neoplastic tissue. The IHC PR Image Analysis application is intended for use as an aid to the pathologist in the detection and quantitation measurement of PR (Progesterone Receptor) in formalin-fixed, paraffin-embedded neoplastic tissue. lt is indicated for use as an aid in the management, prognosis, and prediction of therapy outcomes of breast cancer. Note: The IHC ER and PR Image Analysis applications are an adjunctive computer-assisted methodology to assist the reproducibility of a qualified pathologist in the acquisition and measurement of images from microscope slides of breast cancer specimens stained for the presence of estrogen and progesterone receptor proteins. The IHC ER and PR Image Analysis applications are intended to be used on images viewed on a computer monitor. The accuracy of the test result depends upon the quality of the immunohistochemical staining. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for the ER and PR reagent/kit used to assure the validity of the IHC ER and PR Image Analysis application assisted scores. Prescription Use X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use_ (21 CFR Part 801 Subpart C) #### (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Leica Biosystems. The word "Leica" is written in a stylized, red font. Below "Leica" is the word "BIOSYSTEMS" written in a simple, black, sans-serif font. The logo is set against a white background. #### Bundled 510(k) Summary Leica Biosystems Imaging, Inc. (Aperio ePathology eIHC IVD System) #### 21 CFR 807.92(a) (1): Submitter's name and address: Leica Biosystems Imaging, Inc. 1360 Park Center Dr. Vista, CA 92081 Submitter's telephone and fax numbers: Phone: (760) 539-1194 Fax: Contact person: Christine Kishi Quality Assurance Specialist 1360 Park Center Dr. Vista, CA 92081 christine.kishi@leicabiosystems.com Date this 510(k) summary was prepared: April 28, 2014 #### Bundled Submission Overview 21 CFR 807.92(a)(2): Device trade name and common name, and classification | Bundled Devices | | |-------------------------------------------------------------------------------|---------------------------------------------| | Trade Name | Aperio ePathology eIHC IVD System | | Common Name | Digital pathology and image analysis system | | Regulation Number | 21 CFR 864.1860 | | Regulation Name | Immunohistochemistry reagents and kits | | Regulatory Class | II | | Product Codes | NOT, NQN | | Bundled Indications | | | Legally Marketed predicate device to which substantial equivalence is claimed | | | Predicate Device | ScanScope XT | | Manufacturer | Aperio Technologies | | Predicate Device k# | K071128 K073677 | 21 CFR 807.92 requirements are specified individually below by cleared predicate device K #. {5}------------------------------------------------ ## K071128 (HER2 Image Analysis): 21 CFR 807.92(a)(2): Trade Name of Device: Aperio ePathology eIHC IVD System, AT Turbo and CS2 models. Regulatory Section: 21 CFR 864.1860 Immunohistochemistry reagents and kits Classification: Class II Product Code: NOT (microscope, automated, image analysis, operator intervention) ## 21 CFR 807.92(a)(3): Legally marketed predicate device to which substantial equivalence is claimed: Predicate Device: ScanScope® System XT Manufacturer: Aperio Technologies, Inc. Predicate Device k#: K071128 ## 21 CFR 807.92(a)(4): Description of the device that is the subject of this premarket notification: System: The Aperio ePathology eIHC IVD System ("System") is an automated digital slide creation, management, viewing and analysis system. The Aperio ePathology eIHC IVD System components consist of an automated digital microscope slide scanner, computer, color monitor, keyboard and digital pathology information management software. The system capabilities include digitizing microscope slides at high resolution, storing and managing the resulting digital slide images, retrieving and displaying digital slides, including support for remote access over wide-area networks, providing tools for annotating digital slides and entering and editing data associated with digital slides, and tools for image analysis of digital slides. Image analysis capabilities include the ability to quantify characteristics useful to Pathologists, such as measuring and scoring immunohistochemical stains applied to histology specimens, including the Dako HerceptTest™. The Dako HerceptTest™ reveals the presence of proteins such as Human Epidermal growth factor Receptor 2 (HER2). These results are indicated for use as an aid in the management, prognosis and prediction of therapy outcomes of breast cancer. Hardware Operation: The ScanScope digital slide scanner creates high resolution, color digital slide images of entire glass slides in a matter of minutes. High numeric aperture 20x objectives, as found on conventional microscopes, are used to produce high-quality images. The ScanScope digital slide scanner employs a linear-array scanning technique that generates digital slide images that have no tiling artifacts and that are free from optical aberrations along the scanning axis. Software Operation: The eSlideManager software (formally named Spectrum) is a full-featured digital pathology information management system. The software runs on a server computer called {6}------------------------------------------------ a Digital Slide Repository (DSR), which stores digital slide images on disk storage such as a RAID array, and which hosts an SQL database that contains digital slide metadata. eSlideManager includes a web application and services which encapsulate database and digital slide image access for other computers. The eSlideManager server supports the capability of running a variety of digital slide image analysis algorithms on digital slides, and storing the results of the analysis into the database. ImageScope also includes support for locally or remotely connected image workstation computers, which run digital slide viewing and analysis software provided as part of eSlideManager. Overview of System Operation: The laboratory technician or operator loads glass microscope slides into either a specially designed slide carrier with a capacity of up to 400 slides for the ScanScope AT Turbo or a 5 slide slide-tray for the ScanScope CS2. The scanning process begins when the operator starts the ScanScope® scanner and finishes when the scanner has completed scanning of all loaded slides. As each glass slide is processed, the system automatically stores individual "striped" images of the tissue contained on the glass slide and integrates the striped images into a single digital slide image, which represents a histological reconstruction of the entire tissue section. After scanning is completed, the operator is able to view and perform certain analytical tests on the digital slides. #### 21 CFR 807.92(a)(5): Intended use and labeled indications for use: The Aperio ePathology eIHC IVD System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape. The IHC HER2 Image Analysis application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin embedded neoplastic tissue. The IHC HER2 Image Analysis application is intended for use as an accessory to the Dako HercepTest™ to aid in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded neoplastic tissue. When used with the Dako HercepTest™, it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered. Note: The IHC HER2 Image Analysis application is an adjunctive computer-assisted methodology to assist the reproducibility of a qualified pathologist in the acquisition and measurement of images from microscope slides of breast cancer specimens stained for the presence of HER-2 receptor protein. The IHC HER2 Image Analysis application is intended to be used on images viewed on a computer monitor. The accuracy of the test result depends upon the quality of the immunohistochemical staining. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for the Dako HercepTest™ to assure the validity of the IHC HER2 Image Analysis application assisted HER-2/neu score. The actual correlation of the Dako HercepTest™ to Herceptin® clinical outcome has not been established. {7}------------------------------------------------ ## 21 CFR 807.92(a)(6): Technological characteristics: The design, construction, energy source and other characteristics of the ScanScope System candidate device are considered to be substantially equivalent to the relevant features of the predicate device. A summary of the technological characteristics of the ScanScope System candidate device in comparison to the predicate device follows: Method of cell detection. The method of cell detection is by colorimetric pattern recognition by microscopic examination of prepared cells by size, shape, hue and intensity as observed by a computer-automated, microscopic digital slide scanner system and/or by visual observation by a health care professional. System Components. The system components comprising the Aperio ePathology eIHC IVD System candidate device are substantially equivalent to those in the predicate device; i.e., a computer-automated digital microscope slide scanner, computer, color monitor, and keyboard. Energy Source. The electrical service is 100vAC – 240vAC, 50Hz/60 Hz, 2 amp, which is similar to the predicate device electrical service requirements. ## 21 CFR 807.92(b): 510(k) summaries for those premarket submissions in which determination of substantial equivalence is also based on an assessment of performance data shall contain the following information: ## 21 CFR 807.92(b)(1): Brief discussion of nonclinical tests submitted, referenced or relied on in this premarket notification: The testing conducted to support this premarket notification includes - 1) System verification and validation testing and 2) System level performance testing using clinical samples. Each of these will be briefly summarized below: #### 1) System Verification and Validation Testing Verification was conducted to ensure the specified Design Inputs met the Design Outputs as specified in the Marketing and/or Product Requirement Specifications and Design Documents. Verification testing includes performance and physical testing, environmental testing, interface compatibility, user interface, worst case testing and negative testing as applicable. Design validation was conducted following successful completion of Verification to ensure the design met user needs. Validation is conducted through running a Beta program and/or User Acceptance Testing (UAT) as applicable. {8}------------------------------------------------ ## 2) System Performance Testing The system level performance testing for the Aperio ePathology eIHC IVD System was conducted through a series of internal studies. Each evaluated the accuracy of the HER2 image analysis algorithm on the updated ScanScope instruments. Paraffin embedded breast tissue slides prepared with the appropriate Dako HER2 IHC test kit was used. The slides represented the range of HER2 scores (0, 1+, 2+ and 3+) that are observed clinically, and were scanned on the ScanScope instruments. The HER2 scores obtained from each of the updated ScanScope systems were evaluated for concordance with the scores obtained from the predicate device. These results were evaluated in a trichotomous (0 and1+, 2+, 3+) analysis. The analysis acceptance criteria was percent agreement using an exact 95% lower confidence interval. Based on these tests and analyses, the updated ScanScope models passed the acceptance criteria for substantial equivalence to the original ScanScope XT. ## 21 CFR 807.92(b)(2): Brief discussion of clinical tests submitted, referenced or relied on in this premarket notification: There are no clinical tests submitted, referenced or relied on in this submission. {9}------------------------------------------------ ## K073677 (ER/PR Image Analysis): | Trade Name of Device: | Aperio ePathology eIHC IVD System, AT Turbo and CS2 models. | |-----------------------|------------------------------------------------------------------------------------------------------------------------------------------| | Regulatory Section: | 21 CFR 864.1860 Immunohistochemistry reagents and kits | | Classification: | Class II | | Product Code: | NQN (Microscope, Automated, Image Analysis,<br>immunohistochemistry, Operator Intervention, Nuclear Intensity<br>and Percent Positivity) | ## 21 CFR 807.92(a)(3): Legally marketed predicate device to which substantial equivalence is claimed: | Predicate Device: | ScanScope® System XT | |----------------------|--------------------------| | Manufacturer: | Aperio Technologies, Inc | | Predicate Device k#: | K073677 | ## 21 CFR 807.92(a)(4): Description of the device that is the subject of this premarket notification: System: The Aperio ePathology eIHC IVD System ("System") is an automated digital slide creation, management, viewing and analysis system. The Aperio ePathology eIHC IVD System components consist of an automated digital microscope slide scanner, computer, color monitor, keyboard and digital pathology information management software. The system capabilities include digitizing microscope slides at high resolution, storing and managing the resulting digital slide images, retrieving and displaying digital slides, including support for remote access over wide-area networks, providing tools for annotating digital slides and entering and editing data associated with digital slides, and tools for image analysis of digital slides. Image analysis capabilities include the ability to quantify characteristics useful to Pathologists, such as measuring and scoring immunohistochemical stains applied to histology specimens, including Dako ER/PR, which reveal the presence of ER (Estrogen Receptor) protein and PR (Progesterone Receptor) protein expression.These results are indicated for use as an aid in the management, prognosis and prediction of therapy outcomes of breast cancer. Hardware Operation: The ScanScope digital slide scanner creates seamless true color digital slide images of entire glass slides in a matter of minutes. A high numeric aperture 20x, as found on conventional microscopes, is used to produce high-quality images. The ScanScope digital slide scanner employs a linear-array scanning technique that generates images free from optical aberrations along the scanning axis. The result is digital slide images that have no tiling artifacts and are seamless. {10}------------------------------------------------ Software Operation: The eSlideManager software (formally named Spectrum) is a full-featured digital pathology management system. The software runs on a server computer called a Digital Slide Repository (DSR), which stores digital slide images on disk storage such as a RAID array, and which hosts an SQL database that contains digital slide metadata. eSlideManager includes a web application and services which encapsulate database and digital slide image access for other computers. The eSlideManager server supports the capability of running a variety of image analysis algorithms on digital slides, and storing the results of analysis into the database. ImageScope includes support for locally or remotely connected image workstation computers, which run digital slide viewing and analysis software provided as part of eSlideManager. Overview of System Operation: The laboratory technician or operator loads glass microscope slides into either a specially designed slide carrier with a capacity of up to 400 slides for the ScanScope AT Turbo or a 5 slide slide-tray for the ScanScope CS2. The scanning process begins when the operator starts the ScanScope® scanner and finishes when the scanner has completed scanning of all loaded slides. As each glass slide is processed, the system automatically stores individual "striped" images of the tissue contained on the glass slide and integrates the striped images into a single digital slide image, which represents a histological reconstruction of the entire tissue section. After scanning is completed, the operator is able to view and perform certain analytical tests on the digital slides. #### 21 CFR 807.92(a)(5): Intended use and labeled indications for use: The Aperio ePathology eIHC IVD System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape. The IHC ER Image Analysis application is intended for use as an aid to the pathologist in the detection and quantitative measurement of ER (Estrogen Receptor) in formalin-fixed paraffinembedded neoplastic tissue. The IHC PR Image Analysis application is intended for use as an aid to the pathologist in the detection and quantitation measurement of PR (Progesterone Receptor) in formalin-fixed, paraffin-embedded neoplastic tissue. It is indicated for use as an aid in the management, prognosis, and prediction of therapy outcomes of breast cancer. Note: The IHC ER and PR Image Analysis applications are an adjunctive computer-assisted methodology to assist the reproducibility of a qualified pathologist in the acquisition and measurement of images from microscope slides of breast cancer specimens stained for the presence of estrogen and progesterone receptor proteins. The IHC ER and PR Image Analysis applications are intended to be used on images viewed on a computer monitor. The accuracy of the test result depends upon the quality of the immunohistochemical staining. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls {11}------------------------------------------------ as specified in the instructions for the ER and PR reagent/kit used to assure the validity of the IHC ER and PR Image Analysis application assisted scores. ## 21 CFR 807.92(a)(6): Technological characteristics: The design, construction, energy source and other characteristics of the Aperio ePathology eIHC IVD System candidate device are considered to be substantially equivalent to the relevant features of the predicate device. A summary of the technological characteristics of the ScanScope System candidate device in comparison to the predicate device follows: Method of cell detection. The method of cell detection is by colorimetric and morphometric pattern recognition by microscopic examination of prepared cells by size, shape, color, and intensity as observed by a computer-automated, microscopic digital slide scanner system and/or by visual observation by a health care professional. System Components. The system components comprising the ScanScope System candidate device are substantially equivalent to those in the predicate device; i.e., a computer-automated digital microscope slide scanner, computer, color monitor, and keyboard. Energy Source. The electrical service is 100vAC - 240vAC, 50Hz/60 Hz, 2 amps which is similar to the predicate device electrical service requirements. ## 21 CFR 807.92(b): 510(k) summaries for those premarket submissions in which determination of substantial equivalence is also based on an assessment of performance data shall contain the following information: ## 21 CFR 807.92(b)(1): Brief discussion of non-clinical tests submitted, referenced or relied on in this premarket notification: The nonclinical testing conducted to support this premarket notification includes - 1) System verification and validation testing and 2) System level performance testing using clinical samples. Each of these will be briefly summarized below: ## 1) System Verification and Validation Testing Verification was conducted to ensure the specified Design Inputs met the Design Outputs as specified in the Marketing and/or Product Requirement Specifications and Design Documents. Verification testing includes performance and physical testing, environmental testing, interface compatibility, user interface, worst case testing and negative testing as applicable. Design validation was conducted following successful completion of Verification to ensure the design met user needs. Validation is conducted through running a Beta program and/or User Acceptance Testing (UAT) as applicable. ## 2) System Performance Testing {12}------------------------------------------------ The system level performance testing for the ScanScope System was conducted through a series of internal studies. Each evaluated the accuracy of the ER and PR image analysis algorithms on the updated ScanScope instruments. Paraffin embedded breast tissue slides prepared with the appropriate Dako ER and PR IHC test kits were used. The slides represented the range of ER and PR scores that are observed clinically, and were scanned on the ScanScope instruments. The ER and PR scores obtained from each of the updated ScanScope systems were evaluated for concordance with the scores obtained from the predicate device. These results were evaluated in a dichotomous (Positive vs Negative) analysis at two percent positive thresholds (1% and 10%). The analysis acceptance criteria was percent agreement using an exact 95% lower confidence interval. Based on these tests and analyses, the updated ScanScope models passed the acceptance criteria for substantial equivalence to the original ScanScope XT. ### 21 CFR 807.92(b)(2): Brief discussion of clinical tests submitted, referenced or relied on in this premarket notification: There are no clinical tests submitted, referenced or relied on in this submission. ....End of 510(k) Summary....
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...