K130021 · Philips Medical Systems Nederland B.V. · OEO · Sep 19, 2013 · Hematology
Device Facts
Record ID
K130021
Device Name
PHILIPS HERCEPTEST DIGITAL SCORE
Applicant
Philips Medical Systems Nederland B.V.
Product Code
OEO · Hematology
Decision Date
Sep 19, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.1860
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K130021 · Sep 19, 2013
PHILIPS HERCEPTEST DIGITAL SCORE
Philips Medical Systems Nederland B.V.
Tissue bank of de-identified human breast tissue specimens
Retrospective clinical specimens were used to evaluate the performance of the Philips HER2/neu IHC Digital Manual Read compared to traditional optical microscopy (manual read) and to assess intra- and inter-pathologist precision.
Retrospective tissue bank; Method comparison; Pathologist precision; Digital pathology
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Method comparison study; Retrospective method comparison study; Follow-up/Duration: Not applicable
Formalin-fixed, paraffin-embedded breast tissue specimens; Sample Size: 200 specimens; Number of Sites: 2
Percent Agreement (PA) of HER2/neu scoring (0, 1+, 2+, 3+)
Indications for Use
The Philips HER2/neu IHC Digital Manual Read is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape. The Philips HER2/neu IHC Digital Manual Read is based on the Philips Digital Pathology Solution platform, which is an automated digital slide creation, management, viewing and analysis system. The Philips HER2/neu IHC Digital Manual Read is intended for use as an accessory to the Dako HercepTest™ to aid in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded neoplastic tissue immunohistochemically stained for HER-2 receptors on a computer monitor. When used with the Dako HercepTest™, it is indicated for use as an aid in the assessment of breast cancer patients from whom HERCEPTIN® (Trastuzumab), PERJETA® (Pertuzumab) or KADCYLA® (Ado-Trastuzumab Emtansine) treatment is being considered. Note: The actual correlation of the Dako HercepTest™ to Herceptin®, Perjeta®, or Kadcyla®, clinical outcome has not been established. Note: The Philips HER2/neu IHC Digital Manual Read is for evaluation of digital images of immunohistochemically stained slides that would otherwise be appropriate for manual visualization by conventional microscopy. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for Dako HercepTest™ to assure the validity of the scores obtained using Philips HER2/neu IHC Digital Manual Read.
Device Story
Philips HER2/neu IHC Digital Manual Read is a digital pathology application for viewing and scoring immunohistochemically stained breast tissue slides. System components include Philips Ultra Fast Scanner (UFS) for digitizing slides into whole slide images (WSI), Philips Image Management System (IMS) for storage/management, and IMS Web Viewer for pathologist review on a Barco MDCC 2121 monitor. Scanner captures high-resolution images, decodes barcodes, and performs automatic focus, compression, and stitching. Pathologist reviews digital images on a computer monitor to perform manual scoring of HER2/neu expression, replacing traditional optical microscopy. Device facilitates clinical decision-making regarding HER2-targeted therapies (Trastuzumab, Pertuzumab, Ado-Trastuzumab Emtansine). System supports LIS integration via HL7. Benefits include digital workflow efficiency, remote viewing, and standardized image management.
Clinical Evidence
Method comparison study (n=200) compared manual digital scoring vs. manual optical microscopy across three pathologists. Results showed high percent agreement for HER2 scoring categories. Intra-pathologist precision (n=8 slides, 5 reads) showed 92.50% agreement for digital vs 87.50% for optical. Inter-pathologist precision (n=200) showed 84.78% agreement for digital vs 88.04% for optical. Instrument precision (inter-system/inter-day) showed 88.24% to 92.98% agreement. Studies confirm digital manual read performance is comparable to traditional optical methods.
Technological Characteristics
System includes automated slide scanner, server, and review workstation with dedicated Barco MDCC 2121 monitor. Imaging features include automatic focus, compression, and image stitching. Connectivity via HL7 for LIS integration. Software performs digital image management and viewing; no automated image analysis algorithms are applied (manual read only). Line-powered electrical source.
Indications for Use
Indicated for use as an accessory to Dako HercepTest™ to aid in detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded neoplastic breast tissue. Intended for assessment of breast cancer patients being considered for HERCEPTIN® (Trastuzumab), PERJETA® (Pertuzumab), or KADCYLA® (Ado-Trastuzumab Emtansine) therapy. For use by qualified pathologists.
Regulatory Classification
Identification
Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.
Special Controls
(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.
Predicate Devices
Virtual Slide System, ScanScope® XT System (K111914)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k130021
B. Purpose for Submission:
New device
C. Manufacturer and Instrument Name:
Philips Digital Pathology
Philips Her2/neu IHC Digital Manual Read
D. Type of Test or Tests Performed:
Manual interpretation of digital images for immunohistochemically (IHC) stained HER2 slides
E. System Descriptions:
1. Device Description:
The Philips Her2/neu IHC Digital Manual Read is a digital score application used with the Philips Digital Pathology Solution (DPS) platform that includes a Philips Ultra Fast Scanner (UFS) Philips Image Management System (IMS) and computer monitor. The Philips Digital Pathology Solution is an automated digital slide creation, management, sharing, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape.
The Philips UFS digitizes glass slides at high resolution and generates images. First the Philips UFS takes snapshot images of the entire glass slide as well as the glass slide label and decodes the barcode. Based on the macro image of the slide, the scanner determines which region on the slide will be scanned. All images together with information about the decoded barcode are sent to the Philips Image Management System (Philips IMS).
The Phillips IMS runs on commercially available server and workstation IT hardware. The server stores and manages the digital slide images and digital slide metadata. The server supports interoperability with other information
{1}
systems such as the laboratory information systems (LIS) via a HL7 interface. The IMS Web Viewer software provides the User Interface for the pathologist to view and read on the digital slides.
2. Principles of Operation:
The Philips Her2/neu IHC Digital Manual Read is intended to provide digital images to the pathologist to perform semi-quantitative interpretation of immunohistochemistry HER2/neu stained breast cancer specimens. Formalin-fixed, paraffin embedded breast cancer specimens are stained with the Dako HercepTest™ according to the package insert. Slides are then scanned at a resolution of 0.25 um/pixel (equivalent to typical 40X objective) using the Philips UFS and stored digitally on the Philips IMS. The pathologist manually reads and interprets the digital image using the Philips IMS Web Viewer without use of image analysis software. The HER2 score is calculated by the pathologist based on the percentages of 0, 1+, 2+, and 3+ cells according to the HER2 scoring scheme in the Dako HercepTest product insert.
3. Modes of Operation:
Automatic scanning with batch and random access modes
4. Specimen Identification:
Specimens are identified by slide label. The slide label includes a barcode and a readable number. The Philips UFS takes an image of the label, the whole slide image and composite high resolution images. The barcode on the slide is read by the software and used to identify the slide in the database. If the barcode is not available or cannot be read, an image is classified as 'Action Required' and placed in a dedicated folder in the Philips IMS with the same name. If an image is placed in that folder, a user can manually enter the SlideID and retry communication with LIS. The label image is by default visible in the IMS web viewer but it is the user's preference to show or hide the label image.
5. Specimen Sampling and Handling:
The Philips UFS has the capacity to automatically scan 300 slides per run (15 glass slide racks with up to 20 slides per rack). The user may change the priority of the slide scanning sequence using the touch screen display. The display also shows the scanning status of each rack.
2
{2}
6. Calibration:
Calibrations, both manual and automated, are performed during production, system installation, operation and maintenance over the lifetime of the product.
Calibration at installation: A full calibration of the Philips UFS is performed during production at the factory using a service application. The same service application is used by service engineers to re-calibrate the Philips UFS after system installation. It includes calibration of: color (using a color target and transformation matrix), homogeneity of the illumination, alignment of mechatronics elements of the Philips UFS (e.g. stage, slide gripper), and alignment of positions of all cameras (snapshot, autofocus and line-scan-color cameras (Red, Green and Blue)).
Calibration during operation: During normal operation, the Philips UFS automatically recalibrates itself periodically for the following elements: stages, focus system, line scan cameras, and homogeneity of the illumination. While scanning, the Philips UFS continuously monitors the quality of each lane and re-scans a lane when the quality is insufficient. A user can manually start calibration process via user interface of the Philips UFS. At least every six months, the service engineer should use the same service application used during production and installation to re-calibrate the Philips UFS.
7. Quality Control:
The accuracy of the system depends on the laboratory following the quality control instructions for the Dako HercepTest™.
8. Software:
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X ☐ or No ☐
F. Regulatory Information:
1. Regulation section:
21 CFR §864.1860, Immunohistochemistry reagents and kits
2. Classification:
Class II
{3}
3. Product code:
OEO (Microscope, Automated, Digital Image, Manual Interpretation)
4. Panel:
Pathology 88
G. Intended Use:
1. Indication(s) for Use:
The Philips HER2/neu IHC Digital Manual Read is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape. The Philips HER2/neu IHC Digital Manual Read is based on the Philips Digital Pathology Solution platform, which is an automated digital slide creation, management, viewing and analysis system.
The Philips HER2/neu IHC Digital Manual Read is intended for use as an accessory to the Dako HercepTest™ to aid in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded neoplastic tissue immunohistochemically stained for HER-2 receptors on a computer monitor. When used with the Dako HercepTest™, it is indicated for use as an aid in the assessment of breast cancer patients from whom HERCEPTIN® (Trastuzumab) treatment is being considered. Note: The actual correlation of the Dako HercepTest™ to Herceptin® clinical outcome has not been established.
Note: The Philips HER2/neu IHC Digital Manual Read is for evaluation of digital images of immunohistochemically stained slides that would otherwise be appropriate for manual visualization by conventional microscopy. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for Dako HercepTest™ to assure the validity of the scores obtained using Philips HER2/neu IHC Digital Manual Read.
2. Special Conditions for Use Statement(s):
For Prescription Use Only.
H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers:
{4}
Olympus Virtual Slide System, VS800 System, k111914
2. Comparison with Predicate Device:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Semi-quantitative interpretation of HER2 | Same |
| Assay | Dako HercepTest™ | Same |
| Specimen Type | Formalin-fixed paraffin embedded human breast cancer | Same |
| Method of Interpretation | Manual | Manual |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Camera | 3 Cameras (RGB) | CCD camera |
| Image acquisition | Line scanning | Tile stitching |
| Device Components | Automated scanner and software, monitor and computer | Scanner, computer, color monitor, keyboard, scanner software and digital pathology information management software |
I. Special Control/Guidance Document Referenced (if applicable):
Guidance for Off-the-Shelf Software Use in Medical Devices (Sept. 1999)
Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 2005)
Cybersecurity for Networked Medical Devices Containing Off-the-Shelf Software Jan. 2005)
Format for Traditional and Abbreviated 510(k)s (Aug 2005)
General Principles of Software Validation; Final Guidance for Industry and Staff (Jan 2002)
Medical Device Use –Safety: Incorporating Human Factors Engineering into Risk Management (July 2000)
Guidance for In Vitro Diagnostic (IVD) Device Studies – Frequently Asked Questions (June 25, 2010)
{5}
J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy:
A method comparison study was conducted to compare traditional optical microscopy ("Manual Optical") and manual reading of digital slides on a computer monitor ("Manual Digital") with breast cancer specimens that were stained with Dako HercepTest™.
A total of two hundred (200) formalin-fixed, paraffin-embedded breast tissue specimens from a tissue bank of de-identified human specimens were selected for inclusion in the study. The slides were prescreened by a pathologist to evaluate the quality of the tissue, quality of the staining and to provide a score. Slides that passed the prescreening were randomly selected to fulfill a roughly equal distribution of HercepTest™ scores in four categories (0, 1+, 2+, 3+).
The pathologists in the study were trained in the use of the investigational device according to the labeling. The slides were scanned at three different scanners and three pathologists from two sites participated in the study. The order of the methods was randomized over the pathologists and the washout period between reads was at least 7 days.
Tables 1-3 show the 4x4 tables for the agreements between the manual microscopy reads and manual digital reads using column-wise Percent Agreement (PA; i.e., percent of manual digital reads that agree with a given manual optical read) with an Exact 95% Confidence Interval (CI) for each of the pathologists. The statistical analysis is provided for a trichotomous agreement of the HER2 scores combining 0 and 1+, and leaving 2+ and 3+ uncombined. For example, there were 93 (38 + 55) slides with manual optical read either 0 or 1+ and 85 (37 + 1 + 1 + 46) of these were called 0 or 1+ by manual digital read, resulting in a PA of 91.39% (85/93).
Table 1: Manual Optical Reads vs. Manual Digital Reads 4x4 Tables for Site 1-Pathologist 1
| Pathologist 1 | Manual Optical Read | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | 0 | 1+ | 2+ | 3+ | TOTAL |
| Manual Digital Read | 0 | 37 | 1 | 0 | 0 | 38 |
| | 1+ | 1 | 46 | 8 | 0 | 55 |
| | 2+ | 0 | 8 | 33 | 1 | 42 |
| | 3+ | 0 | 0 | 5 | 44 | 49 |
| | TOTAL | 38 | 55 | 46 | 45 | 199 |
{6}
| Score | PA | Exact 95% CI |
| --- | --- | --- |
| 0, 1+ | 91.40% | 83.75%, 96.21% |
| 2+ | 71.74% | 56.54%, 84.01% |
| 3+ | 97.78% | 88.23%, 99.94% |
Table 2: Manual Optical Reads vs. Manual Digital Reads 4x4 Tables for Site 1-Pathologist 2
| Pathologist 2 | Manual Optical Read | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | 0 | 1+ | 2+ | 3+ | TOTAL |
| Manual Digital Read | 0 | 23 | 0 | 0 | 0 | 23 |
| | 1+ | 22 | 21 | 1 | 0 | 44 |
| | 2+ | 2 | 28 | 45 | 0 | 75 |
| | 3+ | 0 | 0 | 5 | 43 | 48 |
| | TOTAL | 47 | 49 | 51 | 43 | 190 |
| Score | PA | Exact 95% CI |
| --- | --- | --- |
| 0, 1+ | 68.75% | 58.48%, 77.82% |
| 2+ | 88.24% | 76.13%, 95.56% |
| 3+ | 100.0% | 91.78%, 100.0% |
Table 3: Manual Optical Reads vs. Manual Digital Reads 4x4 Tables for Site 2-Pathologist 3
| Pathologist 3 | Manual Optical Read | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | 0 | 1+ | 2+ | 3+ | TOTAL |
| Manual Digital Read | 0 | 26 | 0 | 0 | 0 | 26 |
| | 1+ | 6 | 77 | 2 | 0 | 85 |
| | 2+ | 0 | 13 | 28 | 1 | 42 |
| | 3+ | 0 | 1 | 10 | 32 | 43 |
| | TOTAL | 32 | 91 | 40 | 33 | 199 |
| Score | PA | Exact 95% CI |
| --- | --- | --- |
| 0, 1+ | 88.62% | 81.64%, 93.64% |
| 2+ | 70.00% | 53.47%, 83.44% |
| 3+ | 96.97% | 84.24%, 99.92% |
b. Precision/Reproducibility:
Intra-Pathologist
{7}
A slide set of eight (8) slides, consisting of two slides in each of the categories $(0, 1+, 2+, 3+)$ , was used in the intra-pathologist study. Twelve additional "wild card" slides were added to minimize recall bias. These 12 slides were not used in the analyses. Each slide was read 5 times for a total of 40 reads by both the optical and digital method. For the intra-pathologist study outliers are defined as scores that are different from the median values of the scores provided by the pathologist over 5 runs of the method. For manual optical reads intra-pathologist agreement was $87.5\%$ (35/40). For manual digital reads intra-pathologist agreement was $92.5\%$ (37/40).
# Inter-Pathologist Precision
The inter-pathologist study used data collected from all sites for each of the 3 pathologists in the method comparison study $(N = 191)$ . For manual digital reads using binary percent agreement $(0, 1+$ vs. $2+, 3+)$ inter-pathologist agreement was $80.45\%$ with Exact $95\%$ CI $(76.27\%, 84.64\%)$ . For manual optical reads inter-pathologist agreement was $77.66\%$ with $95\%$ Exact CI $(73.12\%, 82.20\%)$ .
# Inter-Day/Intra-System and Inter-System Precision
A study was conducted at one clinical site to assess the inter-day/intra-instrument and inter-instrument precision for manual digital reads.
A subset of forty (40) slides from the method comparison study was used. The slides were selected to provide an equal distribution of HER2 scores in the four scoring categories $(0, 1+, 2+, 3+)$ . For the Inter-Day/Intra-System study the slide set was scanned on three different days on the same device and the images were scored by one pathologist. For the Inter-System study the slides set was scanned one time on three devices and the images were scored by a second pathologist. To prevent recall bias, slides were randomized between reads and a minimum washout period of one week was required between reading sessions. Additionally, 32 wild card images, not included in the analysis were utilized to confound the reads.
The statistical analysis is presented across all slides, with pair-wise comparisons of all reads by the pathologist, using Percent Agreement (PA) with a $95\%$ Confidence Interval (CI). The statistical analysis is provided for a trichotomous categorization of the HER2 scores combined 0 and $1+$ and leaving $2+$ and $3+$ uncombined.
| | PA | 95% CI |
| --- | --- | --- |
| Manual Digital Reads - Inter-Day/Intra-System | 92.98% | (86.76, 96.40) |
| Manual Digital Reads - Inter-System | 88.24% | (80.55, 93.14) |
{8}
c. Linearity:
Not applicable.
d. Carryover:
Not applicable.
e. Interfering Substances:
Not applicable.
# 2. Other Supportive Instrument Performance Data Not Covered Above:
The Philips UFS automatically recalibrates itself periodically for the following elements: stages, focus system, line scan cameras, and homogeneity of the illumination. Calibration is preceded by a measurement of system parameters which is performed in a certain interval. The interval is defined based on the number of slides scanned, temperature or time. If the parameters are found to have drifted outside their prescribed range, a corresponding calibration is performed to bring them back within the range. Then the parameters are measured again to verify their status. Intervals for auto-calibration are presented in Table 6.
Table 6: UFS Auto-calibration intervals for focus system, line scan cameras (LNCRs) and illumination homogeneity
| Auto-Calibration | | |
| --- | --- | --- |
| Stage to LNCR Calibration | Startup 200 slides 4 hours 2°C | 25 slides 1 hour |
| Focus System Calibration | Startup 200 slides 4 hours 2°C | 25 slides 1 hours |
| LNCR Homogeneity | Startup 1 hour 0.5°C | 1 hour |
| SNCR Background Calibration | Startup 4 hours 2°C | 1 slide 1 hours |
# K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
{9}
L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
10
{10}
11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.