PATHIAM SYSTEM FOR HER2/NEU IMMUNOHISTOCHEMISTRY REAGENTS & KITS
K080910 · Bioimagene, Inc. · NOT · Feb 4, 2009 · Hematology
Device Facts
Record ID
K080910
Device Name
PATHIAM SYSTEM FOR HER2/NEU IMMUNOHISTOCHEMISTRY REAGENTS & KITS
Applicant
Bioimagene, Inc.
Product Code
NOT · Hematology
Decision Date
Feb 4, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.1860
Device Class
Class 2
Indications for Use
PATHIAM-Assisted Scoring: Intended for clinical laboratory use as an accessory to the DAKO HercepTest to aid in the detection and semi-quantitative measurement of Her2/neu in formalin fixed, paraffin-embedded normal and neoplastic tissue. When used with the DAKO HercepTest, Pathiam Assisted Scoring is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN (Trastuzumab) treatment is being considered. The pathologist should verify agreement with the PATHIAM System score. HER2/neu results are indicated for use as an aid in the management, prognosis and predication of therapy outcomes of breast cancer. Note: The actual correlation of the DAKO HercepTest to Herceptin® clinical outcome has not been established.
Device Story
PATHIAM System is an automated digital image analysis tool for pathologists. Input: digitized images of immunohistochemically stained (DAKO HercepTest) breast tissue slides captured by BioImagene iScan slide scanner. Operation: pathologist selects area of interest; software performs semi-quantitative analysis of staining extent and intensity; system suggests staining score (0 to 3). Output: digital image and suggested score displayed on monitor. Usage: clinical laboratory setting; operated by qualified pathologist. The system provides quantitative data to assist reproducibility of slide interpretation; pathologist makes final assessment and reports score. Benefits: improved consistency and data recording in HER2/neu assessment.
Clinical Evidence
Bench testing only. Study compared PATHIAM-assisted scoring to manual HercepTest scoring across 176 breast tissue specimens at three sites. Overall percent agreement between PATHIAM and manual scores ranged from 77% to 88%. Inter-site reproducibility for PATHIAM scores ranged from 89% to 92%. Scanner precision (repeatability) was 97%; inter-system reproducibility was 100%. No clinical outcome data provided.
Technological Characteristics
System includes BioImagene iScan slide scanner, Intel-based PC (dual core, dual Xeon), monitor, keyboard, mouse, and PATHIAM software. Digitizes formalin-fixed, paraffin-embedded tissue slides. Software performs object identification based on intensity, shape, size, and color. No specific ASTM materials cited. Software requires human intervention; no independent interpretation.
Indications for Use
Indicated for clinical laboratory use as an accessory to DAKO HercepTest for semi-quantitative measurement of HER2/neu in formalin-fixed, paraffin-embedded breast tissue to aid in assessing breast cancer patients being considered for HERCEPTIN (Trastuzumab) therapy.
Regulatory Classification
Identification
Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.
Special Controls
(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.
K062756 — PATHIAM IMAGING SOFTWARE FOR HER2/NEU · Bioimagene, Inc. · Feb 20, 2007
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE
A. 510(k) Number:
K080910
B. Purpose for Submission:
To obtain clearance for a new device.
C. Manufacturer and Instrument Name:
BioImagene, Inc. PATHIAM System with iScan for HER2/neu
D. Type of Test or Tests Performed:
iScan instrument and image analysis software for detection of expression of HER2/neu
E. System Descriptions:
1. Device Description:
The PATHIAM™ System is an instrument (iScan) and image analysis software system designed to assist the qualified pathologist in the consistent quantitative assessment of marker expression in immunohistochemically stained histological sections digital images. The sample tissue is breast tissue prepared using the DAKO HercepTest Reagent Kit. The PATHIAM system consists of the BioImagene iScan slide scanner, computer with the PATHIAM Software, monitor, keyboard and mouse.
Hardware: The iScan slide scanning device captures digital images of formalin-fixed, paraffin-embedded tissues that are suitable for storage, viewing and visual analysis. The device includes a digital slide scanner, racks for loading glass slides, an Intel based PC, dual core, dual Xeon processor, PATHIAM Software, and a monitor.
Software: The PATHIAM Software requires competent human intervention at all steps in the analysis process. The system is designed to complement the routine workflow of a qualified pathologist screening the immunohistochemically stained histological slides with additional quantitative data to assist the reproducibility of the slide interpretation. It allows the user to select the area of interest on the breast tissue image. The user marks the area of interest for the analysis. The system software makes no independent interpretations of the data. The software will act as a tool for the user, to improve consistency and data recording. The image
{1}
produced digitally may also be used independent of the software, by allowing the pathologist to count from the digital image, rather than from the microscope.
2. Principles of Operation:
The PATHIAM System digitizes formalin-fixed, paraffin embedded normal and neoplastic tissue and provides semi-quantitative analysis of extent and intensity of stained tissue, providing the pathologist with an aid to interpretation of the level of expression of HER2/neu in breast cancer tissue. The pathologist is presented with a digital image of the tissue section and a suggested staining score (0 to 3). The pathologist then makes an assessment of the digital image and reports his/her score. Alternately, the pathologist can simply use the digitized image to perform his interpretation of the level of expression, without employing the software.
3. Modes of Operation:
Semi-automated computer-assisted interpretation
4. Specimen Identification:
Instrument operator assigned using patient surgical pathology number (SPN).
5. Specimen Sampling and Handling:
Formalin-fixed paraffin embedded normal and neoplastic tissue
6. Calibration:
Not applicable
7. Quality Control:
The PATHIAM software assesses the quality of the image to ensure it passes minimum standards before it is analyzed.
8. Software:
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X ☐ or No ☐
F. Regulatory Information:
1. Regulation section:
21 CFR 864.1860
{2}
2. Classification:
Class II
3 Product code:
NOT
4. Panel:
88 Pathology
## G. Intended Use:
1. Indication(s) for Use:
PATHIAM-Assisted Scoring: Intended for clinical laboratory use as an accessory to the DAKO HercepTest to aid in the detection and semi-quantitative measurement of HER2/neu in formalin fixed, paraffin-embedded normal and neoplastic tissue. When used with the DAKO HercepTest, PATHIAM-Assisted scoring is indicated for use as aid in the assessment of breast cancer patients for whom HERCEPTIN® (Trastuzumab) treatment is being considered. The pathologist should verify agreement with the PATHIAM score.
HER2/neu results are indicated for use as an aid in the management, prognosis and predication of therapy outcomes of breast cancer. Note: The actual correlation of the DAKO HercepTest to HERCEPTIN® clinical outcome has not been established.
2. Special Conditions for Use Statement(s):
To be used only with DAKO HercepTest.
## H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers:
K062756 PATHIAM IHC Module software for HER2/neu
2. Comparison with Predicate Device:
| Similarities | | |
| --- | --- | --- |
| Item | Device (PATHIAM System) | Predicate (PATHIAM Software) |
| Indications for Use | Same | When used with the DAKO HercepTest, it is indicated for use as an aid in the assessment |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device (PATHIAM System) | Predicate (PATHIAM Software) |
| | | of breast cancer patients for whom HERCEPTIN. |
| Specimen type | Same | Formalin-fixed, paraffin embedded specimens stained by immunohistochemistry reagent for HER2/neu. |
| Method of Cell Detection | Same | Object identification of a digitized field of view of a pathology slide, using size, shape, color and intensity as observed by a software, and by visual observation of the digitized image by a health care professional. |
| Assay Used | Same | DAKO HercepTest™ |
| Differences | | |
| --- | --- | --- |
| Item | Device (PATHIAM System) | Predicate (PATHIAM Software) |
| Intended Use | The PATHIAM System consists of the PATHIAM Software, the BioImagene iScan Slide Scanner, computer keyboard, monitor and mouse intended to detect and classify cells of clinical interest by analyzing digitized images of microscope slides based on object identification of cellular objects of particular intensity, shape, size and color. | The imaging software is intended to detect and classify cells of clinical interest by analyzing digitized images of microscope slides based on object identification of cellular objects of particular intensity, shape, size and color. The software can be used with a computer and image digitizer with features specified in the labeling. |
| Image Analysis System | Histologic observation by a pathologist through the BioImagene iScan slide scanner. | Histologic observation by a pathologist through a specified microscope/digital camera combination or slide scanner. |
| Hardware components | PATHIAM Software, BioImagene iScan slide scanner, computer and monitor. | Computer, either microscope with digitizing camera or slide scanner, keyboard, mouse, high resolution color monitor and hard drive for storage. |
| Software components | Version PATHIAM 1.1F accepts JPEG 2000 image files from iScan. | Version PATHIAM 1.0F |
{4}
I. Special Control/Guidance Document Referenced (if applicable):
Not applicable
J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy:
Comparison with Manual HercepTest Method:
Values for staining intensity were obtained from a review of PATHIAM values by trained pathologists at three sites, who viewed both the digital images and the score provided by the software, and then selected an appropriate tissue score (0 to 3). Tissue samples for the study were all procured for a single site. The same pathologist read the same slides manually using the DAKO HercepTest package insert. The manual assessment took place seven days before the experiments with the PATHIAM System were completed at the sites.
Tables 1-3: Concordance Between the PATHIAM System and Manual Scores of HercepTest® stained Breast Tissue
Table 1. Site 1. Manual vs. PATHIAM
| Site 1 | Manual 0-1+ | Manual 2+ | Manual 3+ |
| --- | --- | --- | --- |
| PATHIAM 0-1+ | 71 | 17 | 4 |
| PATHIAM 2+ | 0 | 25 | 19 |
| PATHIAM 3+ | 0 | 1 | 39 |
Percent Agreement =135/176 X 100 = 77%
Overall % agreement (95% EXACT CI): 77% (70% - 83%)
Table 2. Site 2. Manual vs. PATHIAM
| Site 2 | Manual 0-1+ | Manual 2+ | Manual 3+ |
| --- | --- | --- | --- |
| PATHIAM 0-1+ | 80 | 4 | 0 |
| PATHIAM 2+ | 13 | 37 | 0 |
| PATHIAM 3+ | 0 | 16 | 26 |
Percent Agreement =143/176 X 100 = 81%
Overall % agreement (95% EXACT CI): 81% (75% - 87%)
{5}
Table 3. Site 3. Manual vs. PATHIAM
| Site 3 | Manual 0-1+ | Manual 2+ | Manual 3+ |
| --- | --- | --- | --- |
| PATHIAM 0-1+ | 86 | 7 | 0 |
| PATHIAM 2+ | 3 | 28 | 9 |
| PATHIAM 3+ | 0 | 2 | 41 |
Percent Agreement =155/176 X 100 = 88%
Overall % agreement (95% EXACT CI): 88% (82% - 92%)
Tables 4-6 – Comparison Manual Scoring between Sites
Table 4. Site 1 vs. 2. Manual vs. Manual
| Site 1 | Manual 0-1+ | Manual 2+ | Manual 3+ |
| --- | --- | --- | --- |
| Manual 0-1+ | 70 | 1 | 0 |
| Manual 2+ | 21 | 22 | 0 |
| Manual 3+ | 2 | 34 | 26 |
Percent Agreement = 118/176 X 100 = 67%
Overall % agreement (95% EXACT CI): 67% (60% - 74%)
Table 5. Site 2 vs. 3. Manual vs. Manual
| Site 2 | Manual 0-1+ | Manual 2+ | Manual 3+ |
| --- | --- | --- | --- |
| Manual 0-1+ | 86 | 7 | 0 |
| Manual 2+ | 3 | 30 | 24 |
| Manual 3+ | 0 | 0 | 26 |
Percent Agreement = 142/176 X 100 = 81%
Overall % agreement (95% EXACT CI): 81% (74% - 86%)
Table 6. Site 3 vs. 1. Manual vs. Manual
| Site 3 | Manual 0-1+ | Manual 2+ | Manual 3+ |
| --- | --- | --- | --- |
| Manual 0-1+ | 69 | 20 | 0 |
| Manual 2+ | 2 | 22 | 13 |
| Manual 3+ | 0 | 1 | 49 |
Percent Agreement =140 /176 X 100 = 80%
Overall % agreement (95% EXACT CI): 80% (73% - 85%)
{6}
Tables 7-9 – Comparison PATHIAM Scoring between Sites
Table 7. Site 1 vs. 2. PATHIAM vs. PATHIAM
| Site 1 vs. 2 | PATHIAM 0-1+ | PATHIAM 2+ | PATHIAM 3+ |
| --- | --- | --- | --- |
| PATHIAM 0-1+ | 83 | 9 | 0 |
| PATHIAM 2+ | 1 | 40 | 3 |
| PATHIAM 3+ | 0 | 1 | 39 |
Percent Agreement =162/176 X 100 = 92%
Overall % agreement (95% EXACT CI): 92% (87% - 96%)
Table 8. Site 2 vs. 3. PATHIAM vs. PATHIAM
| Site 2 vs. 3 | PATHIAM 0-1+ | PATHIAM 2+ | PATHIAM 3+ |
| --- | --- | --- | --- |
| PATHIAM 0-1+ | 82 | 2 | 0 |
| PATHIAM 2+ | 11 | 35 | 4 |
| PATHIAM 3+ | 0 | 3 | 39 |
Percent Agreement =156/176 X 100 = 89%
Overall % agreement (95% EXACT CI): 89% (83% - 93%)
Table 9. Site 3 vs. 1. PATHIAM vs. PATHIAM
| Site 3 vs. 1 | PATHIAM 0-1+ | PATHIAM 2+ | PATHIAM 3+ |
| --- | --- | --- | --- |
| PATHIAM 0-1+ | 88 | 5 | 0 |
| PATHIAM 2+ | 4 | 34 | 2 |
| PATHIAM 3+ | 0 | 5 | 38 |
Percent Agreement =160/176 X 100 = 91%
Overall % agreement (95% EXACT CI): 91% (86% - 95%)
b. Precision/Reproducibility:
In the reproducibility study between pathologists and the PATHIAM System the PATHIAM System was tested by analyzing images of the same set of 176 stained tissue specimens by three pathologists at three sites. Pathologists recorded their estimation of the score from the score provided by the PATHIAM System plus their review of the digital images provided by the software. Concordance for the PATHIAM System values between labs ranged from 89% to 92%.
BioImagene iScan Slide Scanner Reproducibility
1. iScan Slide Scanner Precision
Eight samples with manual scores of 0, 1+, 2+ and 3+ were scanned 5 times on the iScan slide scanner. The percent agreement was calculated to be 97% (39/40).
{7}
8
2. Inter-run/Inter System Reproducibility
Eight samples with manual scores of 0, 1+, 2+ and 3+ were scanned 5 times on 3 different iScan slide scanners. The agreement between the PATHIAM System scores for different scans is 100% and for different iScan slide scanners is 100%.
c. Linearity:
Not applicable
d. Carryover:
Not applicable
e. Interfering Substances:
Not applicable
2. Other Supportive Instrument Performance Data Not Covered Above:
None
K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.