P950002 · Zimmer Spine, Inc. · MAX · Sep 20, 1996 · Orthopedic
Device Facts
Record ID
P950002
Device Name
BAK INTERBODY FUSION SYSTEM
Applicant
Zimmer Spine, Inc.
Product Code
MAX · Orthopedic
Decision Date
Sep 20, 1996
Decision
APRL
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic, Expedited Review
Intended Use
The BAK™ Interbody Fusion System is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). BAK™ devices are to be implanted via an open anterior or posterior approach. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
Device Story
BAK™ Interbody Fusion System is a hollow, threaded titanium alloy cylinder used for spinal fusion. Device implanted via open anterior or posterior surgical approach by surgeons. Input: autogenous bone graft packed into device. Principle: device provides structural support and stabilization to intervertebral space, promoting bony ingrowth through transverse holes. Output: stabilized spinal segment. Clinical use: treats discogenic back pain; increases neuroforaminal volume; reduces stenosis. Benefits: provides mechanical stability comparable to rigid instrumentation; facilitates fusion; improves patient pain, function, and muscle strength.
Clinical Evidence
PMA clinical study (IDE G900193) of 947 patients (54% male, mean age 41.5). Primary endpoints: fusion, pain, function, muscle strength. 24-month success rate 72% overall. Fusion rate 91% at 24 months. Pain improvement 85% at 24 months. Function improvement 94% at 24 months. Muscle strength maintenance 94% at 24 months. Compared to literature controls (ALIF/PLIF); BAK™ fusion rates were generally superior or comparable. Complication rates within range of literature controls.
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1, including those with up to Grade I spondylolisthesis or retrolisthesis. Contraindicated in patients with active infection at the operative site.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Reference Devices
Bagby Basket
Related Devices
P970015 — INTER FIX THREADED FUSION DEVICE · Sofamor Danek · May 14, 1999
K192248 — Cortina [MAX] Lumbar Cage System · Neurostructures, Inc. · Nov 25, 2019
K201671 — A-Link Z · Acuity Surgical Devices, LLC · May 4, 2021
K193418 — SeaSpine Skipjack System · SeaSpine Orthopedics Corporation · Mar 13, 2020
Submission Summary (Full Text)
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS
I. GENERAL INFORMATION
DEVICE GENERIC NAME: Intervertebral Body Fusion Device
DEVICE TRADE NAME: BAK™ Interbody Fusion System with Instrumentation
APPLICANT'S NAME: Spine-Tech, Inc.
7375 Bush Lake Road
Minneapolis, MN 55439-2029
PREMARKET APPROVAL (PMA) APPLICATION NUMBER: P950002
DATE OF PANEL RECOMMENDATION: May 23, 1996
DATE OF NOTICE OF APPROVAL TO THE APPLICANT: September 20, 1996
II. INDICATIONS FOR USE
The BAK™ Interbody Fusion System is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). BAK™ devices are to be implanted via an open anterior or posterior approach.
DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
III. DEVICE DESCRIPTION
The BAK™ device is a hollow, threaded cylinder available in five sizes. The sizes (diameter x length) are: 13mm x 20mm; 15mm x 20mm; 15mm x 24mm; 17mm x 24mm; and 17mm x 28mm. Each device has 10° modified square threads covering the entire outer surface of the implant and the first 5mm of one end of each device is tapered to allow for easier insertion into a pre-tapped intervertebral cavity. Each device has two large through-holes which are placed cephalad and caudad and multiple small transverse holes to enhance bony ingrowth. The BAK™ device may be used with an optional endcap which is available in corresponding diameters of 13mm, 15mm, and 17mm.
The BAK™ device is manufactured from titanium 6Al-4V (extra low interstitial) alloy which conforms to American Society Testing and Materials (ASTM) F136-92. The endcaps are manufactured from ultra-high molecular weight polyethylene (UHMWPe) which conforms to ASTM F648-84. The BAK™ device and endcap are provided sterile.
{1}
The BAK™ device and endcaps are implanted using a defined set of instruments. While some of the instruments are specific to either the anterior or posterior surgical approach, many of the instruments are common between the two surgical approaches. The instruments used for the anterior surgical approach include the following: anterior alignment guide; 8mm drill; distraction plug inserter; distraction plugs; guide pin; slap hammer; drill tube sheath; alignment guide handle; short series spacer; T-handle; starter reamer; final reamer; drill tube sleeve; guide pin; pituitary rongeur; drill tube sheath; trail implant; bone tap; implant driver; and endcap inserter. The instruments used for the posterior surgical approach include the following: starter alignment guide; alignment guide handle; 8mm drill; 8mm drill tube; distraction plug inserter; distraction plugs; drill tube guide; drill tube; guide pin; posterior drill tube sheath; slap hammer; short series spacer; T-handle; starter reamer; guide pin; final reamer; pituitary rongeur; trail implant; bone tap; implant driver; and endcap inserter. All instruments are manufactured from stainless steel which conforms to ASTM F899-94. All instruments are provided nonsterile and must be sterilized prior to use or reuse.
## IV. CONTRAINDICATIONS
BAK™ implants should not be implanted in patients with an active infection at the operative site.
## V. WARNINGS AND PRECAUTIONS
There are no known warnings associated with the BAK™ Interbody Fusion System.
The surgeon should only implant the BAK™ device after adequate training and familiarity with the surgical technique manual.
Safety and effectiveness have not been established in patients with the following conditions: gross obesity; three or more levels to be fused; symptomatic cardiac disease; pregnancy; previous fusion attempt at the involved level(s); spondylolisthesis or retrolisthesis of Grade II or greater; systemic or terminal illness; significant loss of quantity or quality of vertebral bone stock usually due to osteoporosis or osteomalacia; conditions requiring steroid use; or active drug abuse.
Two devices should be implanted at each surgical level whenever possible. One device may be used if patient anatomy or surgical exposure does not allow for placement of two devices.
The BAK™ implant and endcap are supplied sterile and should be handled in a manner to avoid contamination. In the event of damage to the sterile packaging or inadvertent contamination, implants may be steam sterilized using a gravity cycle of 270°F for 3 minutes. The endcaps should not be re-sterilized if contaminated.
No implant or endcap should be re-used if it has come into contact with human tissue or bodily fluid.
PMA P950002: Summary of Safety and Effectiveness
Page 2
{2}
Instruments for implantation of the BAK™ device are provided non-sterile and must be sterilized prior to use. They may be sterilized using gravity steam at 270°F for 10 minutes or 250°F for 15 minutes.
## VI. ALTERNATIVE PRACTICES AND PROCEDURES
Nonoperative alternative treatments may include, but are not limited to, physical therapy, medications, braces, chiropractic care, or exercise programs. In addition, there are alternative spinal fusion techniques. These include, but are not limited to, posterior lumbar interbody fusion (PLIF) procedures without instrumentation, anterior lumbar interbody fusion (ALIF) procedures without instrumentation, combined anterior and posterolateral (360°) fusion procedures, anterior/anterolateral spinal systems (e.g., plate and screw systems), or posterior spinal systems (e.g., hook and rod systems).
## VII. POTENTIAL ADVERSE EFFECTS
From the investigational device exemption (IDE) G900193, a total of 947 patients were evaluated for adverse events with the BAK™ Interbody Fusion System. The adverse events (complications) were stratified by those that did not require surgical intervention and those that did. Further, because not all of the patients entered into the study had reached their 24 month postoperative time point, a time course distribution of the complications was provided. The intervals are as follows: operative (day of surgery); post-operative (first day post-op up to 2 months); 3 month (2 to 5 months); 6 month (5 to 9 months); 12 month (9 to 19 months); and 24 month (19 to 30 months). The overall rate was determined by adding the rates from the operative time point to 24 months. The rates shown in Table 1 below are the patient rates (i.e., number of patients with a particular complication divided by the total number of patients with available data at a given time point).
Table 1 - Complication Rates
| | Op.
N=947 | Post-Op
N=947 | 3 mo
N=847 | 6 mo
N=770 | 12 mo
N=661 | 24 mo
N=283 | Overall
Rate |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Complications not requiring surgical intervention | | | | | | | |
| dura related | 3.1% | 0% | 0% | 0% | 0.2%* | 0% | 3.3% |
| neurologic | 1.9% | 0.2% | 0.4% | 0.4%* | 0% | 0% | 2.9% |
| infection | 0% | 2.1% | 0.2% | 0% | 0.1%* | 0% | 2.4% |
| implant migration | 0% | 0.3% | 0.9% | 0.4% | 0% | 0% | 1.6% |
| ileus | 0.4% | 1.1% | 0% | 0% | 0% | 0% | 1.5% |
| vessel damage,
bleeding | 1.5% | 0% | 0% | 0% | 0% | 0% | 1.5% |
PMA P950002: Summary of Safety and Effectiveness
Page 3
{3}
| atelectasis, pneumonia | 0% | 1.2% | 0.1% | 0% | 0% | 0% | 1.3% |
| --- | --- | --- | --- | --- | --- | --- | --- |
| hematoma, seroma | 0.1% | 1.0% | 0.1% | 0.1% | 0% | 0% | 1.3% |
| retrograde ejaculation | 0.2% | 0.1% | 0.7% | 0.3% | 0% | 0% | 1.3% |
| other** | 0.1% | 0.7% | 0% | 0.1% | 0.1% | 0.3% | 1.3% |
| urologic, swollen testicle, prostatic, epididymitis | 0.2% | 0.6% | 0.1% | 0% | 0% | 0% | 0.9% |
| wound dehiscence, incisional hernia | 0% | 0.4% | 0.1% | 0.1% | 0.1% | 0% | 0.7% |
| thrombophlebitis, embolism | 0% | 0.5% | 0.2% | 0% | 0% | 0% | 0.7% |
| leg pain | 0.1% | 0.1% | 0.1% | 0% | 0% | 0.3%* | 0.6% |
| Complications requiring surgical intervention*** | | | | | | | |
| additional stabilization | 0% | 0.2% | 0.3% | 0.6% | 1.5% | 2.1% | 4.7% |
| additional level fusion | 0% | 0% | 0.1% | 0.1% | 0.8% | 0.7% | 1.7% |
| implant migration | 0% | 0.6% | 0.5% | 0.1% | 0% | 0% | 1.2% |
| leg pain | 0% | 0.4% | 0.1% | 0.1% | 0.3% | 0.3% | 1.2% |
| dura related | 0.5% | 0% | 0.1% | 0% | 0% | 0% | 0.6% |
| implant reposition | 0% | 0.3% | 0% | 0.1% | 0% | 0% | 0.4% |
| other decompression | 0% | 0.1% | 0% | 0.1% | 0% | 0% | 0.2% |
| anterior ligament penetration | 0.1% | 0% | 0% | 0% | 0% | 0% | 0.1% |
| fractured sacrum | 0% | 0.1% | 0% | 0% | 0% | 0% | 0.1% |
*due to secondary surgical intervention
**other includes: anemia, colitis, gastro-intestinal (GI) bleeding, undisplaced sacral fracture, unrelated sacro-iliac joint infection, chronic subdural hematoma, umbilical hernia (unrelated to surgery), loose implant perioperatively, occipital infarction, possible pre-existing disc space infection
PMA P950002: Summary of Safety and Effectiveness
Page 4
{4}
**includes 9 revisions, 7 removals, 27 reoperations, and 26 supplemental fixations (see definitions below)
A *revision* is a procedure which adjusts or in any way modifies the original implant configuration (e.g., adjusting position of original configuration, removal with replacement of component). A *removal* is a procedure which removes one or more components of the original implant configuration without replacement of any components. A *reoperation* is a procedure which involves any surgical procedure at the involved level(s) which does not remove, modify, or add any components. A *supplemental fixation* is a procedure in which additional instrumentation not approved as part of the protocol is placed. This may include supplemental placement of a rod/screw system or a plate/screw system.
## VIII. MARKETING HISTORY
The BAK™ Interbody Fusion System has been marketed in approximately 20 international countries. It has not been withdrawn from marketing for any reason relating to its safety or effectiveness.
## IX. SUMMARY OF PRECLINICAL STUDIES
### A. Nonclinical Testing
Nonclinical tests were conducted to characterize the mechanical properties of the BAK™ Interbody Fusion System. The smallest size BAK™ device (13mm x 20mm) was used for this testing because it represents the worst case device (thinnest internal web and thinnest cylindrical wall thickness) of the five sizes.
1. Ultimate Compressive Strength Testing
A single BAK™ device was loaded statically in compression until implant failure or until 30,600 N (6876 lbs), the failure load of the jig, was reached. Five (5) devices sustained a static load of 30,600 N (6876 lbs) without failure. None of the five specimens were crushed or exhibited any cracking under microscopic examination following the testing. The device’s compressive strength greatly exceeds the compressive strength of bone which is estimated to be approximately 1500 N (337 lbs).
2. Fatigue Testing
A single BAK™ implant was loaded cyclically in compression at loads which ranged from 880 N (198 lbs) to 9600 N (2157 lbs) at 15 Hz. None of five (5) devices failed after 5 million cycles. Five million cycles typically represents the number of loading cycles a device might experience within two years. This assumes moderate loading and the device’s goal of stabilizing until fusion occurs within those two years. Because the device’s compressive fatigue strength exceeds the static compressive strength of bone by at least a factor of six, the
PMA P950002: Summary of Safety and Effectiveness
Page 5
{5}
BAK™ device can be expected to withstand anticipated physiologic fatigue loads.
3. Polyethylene Endcap Pushout Test
This test was conducted to determine the static force required to dislodge the endcap from a BAK™ device. After the endcap was snap-fit to the BAK™ device, an axial load was applied through the cavity of the BAK™ device to the endcap. The average push-out load determined for five (5) samples was $111 \pm 53$ N ($25 \pm 12$ lbs). Based on the expected minimal loading on the endcap, the endcap should not become dislodged from the BAK™ device.
B. Stability Testing
Three tests were performed to determine the effect of the device on the flexion/extension (F/E) and lateral stability of the spine after implantation of either one or two BAK™ devices. Bovine, porcine, and primate spines were used in these tests.
1. The first test was conducted to determine the effect of implanting one and two BAK™ devices on spinal stability, relative to the intact spine. One BAK™ device was implanted in a bovine spine (n=7) and two BAK™ devices were implanted in a porcine spine (n=2) and the flexion/extension and lateral stiffness of the spines were measured. Results with one device implanted showed that the spinal segments had an increased stiffness of 71% for flexion/extension and 77% for lateral bending relative to the same tests in spinal segments without implants. Results with two devices implanted showed that the spinal segments had an increased stiffness of 81% for flexion/extension and 477% for lateral bending relative to the same tests in spinal segments without implants. Therefore, the implantation of either one or two devices increased the initial spinal stability.
2. The second test was designed to compare the initial biomechanical stability of the following constructs in a calf spine model: 1) BAK™ implant; 2) PLIF with iliac crest bone graft alone; 3) PLIF with rigid posterior instrumentation; 4) BAK™ implant with rigid posterior instrumentation; and 5) rigid posterior instrumentation (pedicle screws). Non-destructive testing was performed on eight (8) lumbar (L3-L4) calf spines of similar age and size. Each specimen was tested for stiffness in flexion/extension, torsion, and axial loading using the intact spine as a baseline and the five constructs above.
The BAK™ system, used alone, doubled the stiffness of the intact spine in terms of flexion/extension loading (2.6 versus 1.3 Nm/deg) and torsional loading (14 versus 7.5 Nm/deg) and did not migrate during testing. This BAK™ construct was stiffer than the PLIF with iliac crest bone graft alone (which had a flexion/extension stiffness of 1 Nm/deg and a torsional stiffness of 5 Nm/deg) and was similar in stiffness to the PLIF with rigid posterior instrumentation (which had a flexion/extension stiffness of 2.9 Nm/deg and a torsional stiffness of 12.5 Nm/deg). The BAK™ increased the initial stiffness of a calf spine motion
PMA P950002: Summary of Safety and Effectiveness
Page 6
{6}
segment relative to the intact spine and compares favorably to the other stabilization methods with respect to initial spinal stability.
3. The third test assessed the mechanical flexibility of the motion segment both before and after instrumentation using the BAK™ device in a Chagma baboon model. Testing was performed on six (6) Chagma baboon spines at the L4-L5 motion segment. Each specimen was tested for stiffness in flexion and extension in its intact condition and with a single BAK™ implant. The six (6) instrumented lumbar segments showed a trend for increased stiffness in flexion (p<0.07). In the four comparisons made in extension, the instrumented segments were stiffer (p<0.05). The BAK™ device increased the flexion/extension stiffness of the spine relative to the intact spine.
### C. Foraminal Volume Study
The objective of this study was to quantitatively assess changes in the size of the neuroforamen after anterior distraction using the BAK™ Interbody Fusion System in degenerated cadaveric lumbar spines. Nine (9) fresh frozen degenerative cadaver lumbar spines were obtained for the study. Anteroposterior and lateral x-rays were taken for each specimen with neuroforaminal stenosis at L4-L5 and L5-S1. The lateral x-rays were used to measure the anterior and posterior disc heights before and after insertion of the BAK™ implants. Measurements of the volumes and areas were conducted before and after anterior application of the BAK™ implants using three techniques: blunt probe; silicone mold; and computerized tomography. All data were normalized to the pre-implantation measurements.
After BAK™ implantation, the anterior disc heights increased 35.2 ± 19.2% at L4-L5 and 28.4 ± 12.0% at L5-S1 (p<0.001); the posterior disc heights increased 37.1 ± 20.9% at L4-L5 and 45.1 ± 28.0% at L5-S1 (p<0.001). The neuroforamen diameters measured with the blunt probe increased 13.3 ± 4.1% at L4-L5 and 12.3 ± 4.3% at L5-S1 after the BAK™ implantation. The areas measured by CT increased 29.0 ± 18.6% at L4-L5 and 33.8 ± 22.2% at L5-S1 after BAK™ implantation. The volume of each neuroforamen was calculated from the mass of the silicone mold and material density. The volumes increased 22.9 ± 8.3% at L4-L5 and 21.5 ± 11.5% at L5-S1. These results indicate that the BAK™ implant can increase the neuroforaminal volume and possibly reduce neuroforaminal stenosis.
### X. SUMMARY OF ANIMAL STUDIES
#### A. Horse Studies
Studies involving the Bagby Basket, a smooth, stainless steel cylinder with transverse holes, as an interbody fusion device were conducted.
1. The first study involved the use of the Bagby Basket in stabilizing equine cervical vertebra. The goal of this study was to compare commercially available xenograft
PMA P950002: Summary of Safety and Effectiveness
Page 7
{7}
with the Bagby Basket filled with local autograph bone for equine interbody fusion. 16 horses were assigned into one of two treatment groups, eight (8) received a dowel of bovine xenograft and eight (8) received the Bagby Basket. At six months postoperatively, the horses were sacrificed and the spines evaluated.
Mobility studies produced an index value for evaluation of both treatment groups. The index value of the implant was representative of substantially less perceptible motion than the index of the grossly fibrous bovine xenografts. Implantation of the bovine xenograft and the implant resulted in different forms of vertebral fusion. The use of the implant with autogenous bone graft usually resulted in an osseous union and provided a superior fusion. Gross examination of the fusion sites on cut section revealed several differences. The sites in which bovine xenografts had been implanted were extremely pale and frequently contained gross evidence of fibrous connective tissue dispersed through the graft sites. There was minimal evidence of fibrous connective tissue dispersed through the Bagby Basket implantation sites. The autograft within the implant was more consistent with the color and texture of normal cancellous bone.
2. A second study was conducted to evaluate the clinical effectiveness of the Bagby Basket in decompressing caudal cervical spinal cord compression in the horse. In this test, eight (8) male horses with caudal cervical spinal cord compression were implanted with a single bone graft-filled Bagby Basket in an effort to fuse the intervertebral space at the level of compression. The horses were evaluated five months postoperatively by neurologic evaluation, radiology, myelography, CSF analysis, serology, virology and bacteriology. Neurologic exams were repeated 10 months postoperatively. This test showed that all of the implants appeared to have been incorporated in osseous arthrodesis and that this was effective in decompressing all horses by five months post-op. All other results (CSF analysis, serology, virology, and bacteriology) were normal, both preoperatively and postoperatively.
## B. Primate Study
This study was undertaken to evaluate the safety and effectiveness of the BAK™ device when used in the Chagma baboon. The effectiveness in producing a stable lumbar interbody construct was measured by radiologic, biomechanics, and histologic methods.
Comparing the BAK™ device and fresh frozen allograft in this experimental model, it was found that: 1) the Chagma baboon is an adequate model in evaluating lumbar interbody fusion techniques; 2) radiologic assessment showed that disc height loss and increased local kyphosis occurred over time in both groups, although the allografts stabilized earlier (6 weeks) compared to the BAK™ device (24 weeks); 3) biomechanical evaluation revealed no statistically significant differences between the two implant groups; 4) histologic findings confirmed bone ingrowth in the BAK™ group; and 5) the BAK™ Interbody Fusion System was found to be an adequate fusion technique compared to fresh-frozen allograft in this experimental model and justified further human clinical assessment.
PMA P950002: Summary of Safety and Effectiveness
Page 8
{8}
# XI. SUMMARY OF CLINICAL INVESTIGATIONS
A clinical study of the BAK™ Interbody Fusion System was conducted in accordance with approved IDE G900193.
## A. Objective
The objective of the study was to determine the safety and effectiveness of the BAK™ Interbody Fusion System in stabilizing and fusing the affected vertebrae when compared to literature controls.
## B. Inclusion and Exclusion Criteria
The inclusion criteria were males and females between the ages of 21 and 65 with a diagnosis of DDD at 1 or 2 contiguous levels from L2-S1. These patients may have had disc herniation and/or no greater than Grade I spondylolisthesis. Additionally, these patients were to have had six months of unresponsive, nonoperative treatment. Note that based on Panel input, the definition for DDD was refined to that reflected in Section II above, Indications for Use.
The exclusion criteria were as follows: patients with active infection; osteoporosis or osteomalacia; a medical condition that interfered with the postoperative management program; circulatory problems; symptomatic cardiac disease; active malignancy; gross obesity; Grade II or greater spondylolisthesis at involved spine levels; pregnancy; DDD affecting three or more spine motion segments; and the presence of more three of the following psycho-social factors: alcoholism; drug dependence; recent or pending divorce; high level of job dissatisfaction; pending litigation; depression; multiple unsuccessful surgeries; smoking one or more packs per day; or a Waddel score of three or more.
Patients meeting the inclusion criteria were stratified into four treatment groups by the surgical approach (i.e., open anterior or open posterior) and the number of involved levels (i.e., one or two).
- Anterior surgical approach involving one disc level (A1)
- Anterior surgical approach involving two disc levels (A2)
- Posterior surgical approach involving one disc level (P1)
- Posterior surgical approach involving two disc levels (P2)
For both surgical approaches, autogenous bone graft was packed into the BAK™ devices after implantation. If the disc space had sufficient room after devices placement, autogenous bone graft could be placed around them.
## C. Patient Population and Demographics
The BAK™ study population is comprised of 54% (512/947) males and 46% (435/947) females. The mean age at the time of study enrollment was 41.5 years with a range of
PMA P950002: Summary of Safety and Effectiveness
{9}
20 to 73 years. 57% (544/947) of the patients were enrolled in the study with compensation related injuries. 36% (342/947) of the patients had prior back surgery. The average duration of back symptoms prior to enrollment in this study was 5.5 years.
All 947 patients enrolled in the BAK™ study had a diagnosis of DDD, 43% (408/947) with disc herniation and 57% (539/947) without herniation. Grade I spondylolisthesis or retrolisthesis (reverse spondylolisthesis) was present in 12% (112/947) of the population.
A total of 1317 levels were implanted in 947 patients. The distribution of the levels by patient were: <1% (5/947) at L2-L3; 5% (49/947) at L3-L4; 56% (530/947) at L4-L5; 38% (356/947) at L5-S1; <1% (4/947) at L5-L6; <1% (1/947) at L6-S1; and <1% (2/947) at L6-S1. Of the 947 patients, 62% (591/947) had one-level fusions and 38% (356/947) had two-level fusions. There were 55 single cages and 1262 pairs of cages implanted for a total of 2579 cages implanted.
## D. Evaluation Schedule
Patients were evaluated preoperatively, immediately postoperatively (i.e., hospital discharge), at 3 months, 6 months, 12 months, 24 months, and biennially thereafter until the last patient had his/her two-year evaluation. Radiographic studies were conducted at 12 and 24 months postoperatively.
## E. Patient Accountability
A total of 947 patients were enrolled at 19 investigational sites in the United States by 44 investigators. Of these 947 patients, there were 32% (305/947) in the A1 subgroup, 30% (286/947) in the A2 subgroup, 29% (272/947) in the P1 subgroup, and 9% (84/947) in the P2 subgroup. As of November, 1995, a total of 307 patients had reached his/her two-year postoperative time point. Follow-up evaluations, which included an assessment of fusion, pain, function and muscle strength, were performed on 283 of these 307 patients (92%). Complete follow-up evaluations (i.e., measurement of each of the four major outcome parameters) were performed on 254 of these 283 patients (90%). Partial follow-up evaluations were available for 29 patients (10%).
## F. Study Design and Analyses
### 1. Literature Study Control
Literature controls were employed in this study. Outcomes of patients implanted with BAK™ devices via an anterior approach were compared to outcomes of patients who received ALIFs. Outcome of patients implanted with BAK™ devices via a posterior approach were compared to outcomes of patients who received PLIFs. Literature references were deemed acceptable as controls if at least 50% of patients had back pain due to DDD; this group may or may not have had spondylolisthesis of Grade I or less. This differs from the BAK™ group in which all patients had back pain due to DDD and had no greater than Grade I
PMA P950002: Summary of Safety and Effectiveness
Page 10
{10}
spondylolisthesis.
The literature controls used in this study had many differences relative to the BAK™ population with respect to the indication for use, the method by which DDD was assessed, the number of levels fused, the age of the patients, the types of outcome criteria assessed, the method of outcome assessment, the definitions for successful outcome, the duration and nature of follow-up, the incidence of previous back surgery at the same level, and whether the patients were affected by more than two psychological/behavioral risk factors (e.g., alcoholism, drug abuse).
Use of a literature control group was common at the time of the submission of this study, although it is now recognized as less desirable than a randomized, concurrent control study. The advantages of using a randomized, concurrent control reflects the disadvantages of literature controls. In general, in a randomized, concurrent control study, potential bias is eliminated or at least reduced, unknown or known baseline factors tend to be balanced between the two groups, the statistical properties of hypothetical tests are improved, time trends are controlled because of concurrency, and the results tend to be more successfully convincing.
## 2. Data Pooling
Pooling the data between investigational sites and other stratified groups were justified based on a statistical analyses using Pearson chi-squared tests or Fisher's exact tests.
## G. Effectiveness Analyses
The effectiveness variables included an assessment of fusion at the involved level(s), pain, function, and muscle strength; neurological information was captured in the safety assessment. In some cases, only partial data were available (i.e., not all of the four outcome measures were obtained for all patients at all follow-up points). In these cases, all available outcomes for fusion, pain, function, and muscle strength were summarized in these analyses. Therefore, the number of patients included in the assessment of the four outcomes varies slightly due to missing data. Because not all of the patients had reached his/her two-year postoperative time point, the effectiveness analyses involved both the 12 month and 24 month time points for comparison purposes.
## H. Effectiveness Analysis - Fusion
Successful fusion was defined as less than 5° of motion on a flexion/extension series of x-rays at the involved level(s). In cases where two levels were implanted, both levels must have been fused in order for that patient to be considered fused. An independent radiologist reviewed the films. The successful fusion rates are provided in Table 2 below for each study subgroup and the overall population.
PMA P950002: Summary of Safety and Effectiveness
Page 11
{11}
Table 2 - Successful Fusion Rates at 12 and 24 Months
| Study Subgroups | 12 Month Rate | 24 Month Rate |
| --- | --- | --- |
| Anterior, 1-level | 92% (211/229) | 98% (106/108) |
| Anterior, 2-levels | 79% (118/150) | 80% (56/70) |
| Posterior, 1-level | 87% (156/180) | 94% (68/72) |
| Posterior, 2-levels | 75% (45/60) | 71% (12/17) |
| All Study Subgroups | 86% (530/619) | 91% (242/267) |
## I. Effectiveness Analysis - Pain
Pain was measured on a 6-point scale where 1 = no recurrent episode of low back pain and able to perform all previous activities, 2 = occasional recurrence of low back pain, 3 = mild pain, 4 = moderate pain, 5 = marked pain, and 6 = disabling pain.
The distribution of pain scores preoperatively and at 24 months is shown in Table 3 below. One patient was missing the preoperative pain score.
Table 3 - Distribution of Pain Scores
| Pain Level | Preoperative Rate | 24 Month Rate |
| --- | --- | --- |
| None (1) | 0% | 19% (54/280) |
| Episodic (2) | <1% (4/946) | 18% (49/280) |
| Mild (3) | <1% (6/946) | 27% (75/280) |
| Moderate (4) | 22% (210/946) | 24% (68/280) |
| Marked (5) | 54% (506/946) | 9% (25/280) |
| Disabled (6) | 23% (220/946) | 3% (9/280) |
All patients experiencing an improvement by at least one level in the pain score relative to their preoperative score were considered to have a successful result in terms of the pain outcome measure. Table 4 below shows the successful pain rates for each study subgroup and the overall population.
PMA P950002: Summary of Safety and Effectiveness
Page 12
{12}
Table 4 - Successful Pain Rates at 12 and 24 Months
| Study Subgroups | 12 Month Rate | 24 Month Rate |
| --- | --- | --- |
| Anterior, 1-level | 84% (202/240) | 84% (93/111) |
| Anterior, 2-levels | 88% (136/155) | 86% (62/72) |
| Posterior, 1-level | 81% (158/196) | 87% (66/76) |
| Posterior, 2-levels | 77% (48/62) | 81% (17/21) |
| All Study Subgroups | 83% (544/653) | 85% (238/280) |
It is important to distinguish between patients with a successful pain outcome and the amount of pain experienced by patients after implantation with the BAK™ device. A successful outcome did not necessarily mean that a patient experienced no pain; instead, it means that there was at least one level of improvement.
J. **Effectiveness Analysis - Function**
Function was measured on a 26-point scale which ranged from 7 (best) to 32 (worst) points. This function scale measures the patient’s ability to perform activities of daily living (i.e., standing, sitting, walking, squatting, ability to put on shoes and socks), level of recreational activity, and level of employment.
The average preoperative function score was $20.9 \pm 3.7$ (range 9-32) for 946 patients; a preoperative function score was missing for one patient. The average function score at 24 months was $14.5 \pm 4.5$ for 280 patients.
All patients maintaining or experiencing an improvement by at least one point in the function score relative to their preoperative score, were considered to have a successful result in terms of the function outcome measure. Table 5 shows the successful function rates for each study subgroup and the overall population.
Table 5 - Successful Function Rates at 12 and 24 Months
| Study Subgroups | 12 Month Rate | 24 Month Rate |
| --- | --- | --- |
| Anterior, 1-level | 93% (224/240) | 95% (105/111) |
| Anterior, 2-levels | 89% (138/155) | 94% (68/72) |
| Posterior, 1-level | 89% (174/196) | 92% (70/76) |
| Posterior, 2-levels | 89% (55/62) | 95% (20/21) |
| All Study Subgroups | 91% (591/653) | 94% (263/280) |
PMA P950002: Summary of Safety and Effectiveness
Page 13
{13}
# K. Effectiveness Analysis - Muscle Strength
Muscle strength was evaluated bilaterally at four sites: quadriceps, dorsiflexion-inversion, great toe extension, and plantarflexion. Each of the sites was measured on a 3-point scale ranging from 1 (normal) to 3 (marked decrease). The majority of patients (82%) demonstrated normal muscle strength preoperatively.
Maintenance or improvement in muscle strength for all eight sites evaluated was required to be considered a success. Table 6 shows the successful muscle strength rates for each study subgroup and the overall population.
Table 6 - Successful Muscle Strength Rates at 12 and 24 Months
| Study Subgroups | 12 Month Rate | 24 Month Rate |
| --- | --- | --- |
| Anterior, 1-level | 92% (221/240) | 94% (102/109) |
| Anterior, 2-levels | 92% (135/147) | 94% (64/68) |
| Posterior, 1-level | 95% (176/185) | 94% (67/71) |
| Posterior, 2-levels | 92% (55/60) | 100% (20/20) |
| All Study Subgroups | 93% (587/629) | 94% (253/268) |
# L. Safety Analysis
Safety analyses included all patients regardless of the completeness of their follow-up data or length of follow-up. Safety was assessed through physical examinations, x-rays, and by questioning of all patients enrolled in the study. For a summary of the safety data, please see Table 1 in Section VII above, Potential Adverse Effects.
# M. Study Success / Statistical Differences
To be considered an overall study success, the patient must have met each of the following four criteria: 1) fusion of the involved level(s); 2) improvement in pain; 3) maintenance or improvement in function; and 4) maintenance or improvement in muscle strength. The overall success rates at 12 and 24 months for each subgroup and the overall patient population are provided in Table 7 below. Note that the number of patients with data available differs slightly for each based on the study follow-up.
PMA P950002: Summary of Safety and Effectiveness
Page 14
{14}
Table 7 - Study Success Rates at 12 and 24 Months
| Study Subgroups | 12 Month Rate | 24 Month Rate |
| --- | --- | --- |
| Anterior, 1-level | 73% (163/223) | 81% (85/105) |
| Anterior, 2-levels | 64% (90/140) | 59% (39/66) |
| Posterior, 1-level | 64% (108/168) | 78% (52/67) |
| Posterior, 2-levels | 52% (30/58) | 50% (8/16) |
| All Study Subgroups | 66% (391/589) | 72% (184/254) |
The clinical utility of the BAK™ procedure also was evaluated by the proportion of patients that returned to work or were able to work. At the time of the initial BAK™ procedure, 36% of all patients were working full-time or part-time. The proportion of patients working was decreased to 30% at three months follow-up and then progressively increased to 44%, 54%, and 62% at 6, 12, and 24 month follow-up visits, respectively.
Because of the many differences between the literature control groups and the BAK™ group, the longitudinal analyses performed on the BAK™ patient population was extremely important in the assessment of the safety and effectiveness of the BAK™ device. The longitudinal analyses (using Generalized Estimating Equation (GEE) model) showed that the outcomes of fusion, pain, function, and muscle strength did not worsen over time for all study subgroups combined. Specifically, for the total patient population, the rate of fusion increased with time, the amount of pain decreased with time, and the patient's ability to function increased with time.
From the longitudinal analyses of these clinical data, the following statistical differences were observed among the study subgroups up to or at the two-year time point:
- The two year postoperative data indicates that the likelihood of needing additional supplemental fixation increases over time in patients who were not fused or showed no improvement in pain.
- For both the anterior and posterior approaches, patients with one level fusion had lower overall complication rates than patients with two level fusions.
- Patients implanted with the BAK™ device from the posterior approach had higher rates of intra-operative complications and early postoperative surgical interventions than patients implanted from the anterior approach.
- Patients implanted from the anterior approach had a higher overall rate of early postoperative complications than patients implanted from the posterior approach.
PMA P950002: Summary of Safety and Effectiveness
Page 15
{15}
- For both the anterior and posterior approaches, patient who had higher preoperative baseline pain scores showed better pain improvement throughout the study than patients with lower baseline pain scores.
## N. Comparison with Literature Controls
A total of 20 ALIF and 9 PLIF literature articles were used as controls; these are identified in Section XVI below, References. As previously discussed, these literature controls were often greatly different from the BAK™ population. However, clinical results and complication information were extracted for purposes of this comparison.
### 1. Anterior Results
As stated above, there were 20 ALIF literature control articles. The sample sizes reported in these articles ranged from 20 to 150. The data from the control articles were compared to the data from the BAK™ Interbody Fusion System for the total anterior patient population (i.e., A1 and A2) at 24 months.
The fusion rate for the BAK™ was 91% (162/178) for the anterior group. The range of fusion rates reported in the ALIF literature controls was 46% to 96%. Fusion results of the BAK™ were better than literature results in 18 of 20 articles, significantly better in 13, and none were significantly worse than the literature controls.
The definition of clinical success in the ALIF literature primarily involved an assessment of pain but occasionally included some work status, function, and narcotic use information. The clinical success rates reported in the ALIF literature ranged from 41% to 100%. Taking into consideration the same types of measurements, these literature control rates were compared to the following BAK™ clinical rates: 85% (155/183) for pain and 95% (173/183) for function. There were 22 possible assessments in the 20 ALIF literature controls. BAK™ clinical results were better than ALIF in 8 assessments, within 5% of the ALIF results in 3 assessments, and worse than the ALIF results in 11 assessments. It cannot be determined if the differences are due to true differences in clinical success, or due to differences in patient population, data collection, or interpretation of methods.
### 2. Posterior Results
As stated above, there were 9 PLIF literature control articles. The sample sizes reported in these articles ranged from 20 to 462. The data from the control articles were compared to the data from the BAK™ Interbody Fusion System for the total posterior patient population (i.e., P1 and P2) at 24 months.
The fusion rate for the BAK™ was 90% (80/89) for the posterior group. The range of fusion rates reported in literature was 74% to 96%. Fusion results for the
PMA P950002: Summary of Safety and Effectiveness
Page 16
{16}
BAK™ device were higher than 5 PLIF studies and statistically higher in 1. Although 4 of the PLIF studies reported results better than the BAK™, none of these differences were statistically significant. It is unclear if these differences are due to true differences in clinical success or due to differences in patient population, data collection or interpretation of methods.
As with the ALIF literature, the definition of clinical success in the PLIF literature primarily involved an assessment of pain but occasionally included some work status, function, and narcotic use information. The clinical success rates reported in the PLIF literature ranged from 60% to 94%. Taking into consideration the same types of measurements, these literature control rates were compared to the following BAK™ clinical rates: 86% (83/97) for pain and 93% (90/97) for function.
## 3. Comparison of Complications
The experience in this clinical investigation with the BAK™ system compares favorably with the ALIF and PLIF literature complication rates. Reported complications for the BAK™ system were within the range reported for the literature control groups.
## XII. CONCLUSIONS DRAWN FROM THE STUDIES
The nonclinical (i.e., mechanical), animal, and clinical data provide reasonable assurance of the safety and effectiveness of the BAK™ Interbody Fusion System for the treatment of degeneration disc disease (DDD), when used as indicated.
## XIII. PANEL RECOMMENDATION
The Orthopedic and Rehabilitation Devices Panel met to discuss the application on May 23, 1996. The Panel recommended that the application be approved pending submission to and approval by the Center for Devices and Radiological Health (CDRH) of: a reanalysis of the study outcomes with a revised definition of patient success; modifications to the labeling; creation of a patient information document; development of post-approval studies; continued follow-up of study patients; and completion of the IDE randomized study arm of the BAK™ Interbody Fusion System. The Panel agreed with FDA’s recommendation to define patient success as fusion of involved level(s); improvement in pain; maintenance or improvement in function; maintenance or improvement in muscle strength; and maintenance or improvement in neurological reflexes.
The Panel recommended that the labeling be modified to: (1) limit use to the treatment of patients with DDD where DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by historical and radiographic studies; (2) recommend a minimum of six months of non-operative treatment prior to surgery; (3) report the study’s success rates and trends noted in the statistical analyses; (4) require that the device be packed with autograft bone; (5) limit implantation of the device to an open
PMA P950002: Summary of Safety and Effectiveness
Page 17
{17}
approach; and (6) limit use of the device to fusions involving one level or two adjacent levels.
As stated above, the Panel also recommended that two post-approval studies be developed. The first post-approval study is to obtain continued follow-up for a subset of the patients from the IDE study to evaluate the long-term device performance and patient outcomes for a minimum of five years. The second post-approval study is to retrieve and analyze any $\mathrm{BAK}^{\mathrm{TM}}$ device that was implanted and subsequently removed. The Panel recommended that retrieved implants be analyzed metallurgically and histologically for bone quality/quantity and potential wear debris.
## XIV. CDRH DECISION
CDRH agreed with the Panel’s recommendation that the PMA be approved subject to conditions and concurred with each of the conditions recommended by the Panel except for the recommendations to require continued follow-up of study patients and completion of the IDE randomized study arm of the $\mathrm{BAK}^{\mathrm{TM}}$ Interbody Fusion System. While CDRH agreed that it would be desirable to obtain long-term follow-up on the entire study population and the results of a randomized, concurrent clinical trial of the $\mathrm{BAK}^{\mathrm{TM}}$ device, sufficient information currently exists to approve this application. The Panel’s recommendations were modified and made conditions of approval. In addition to the Panel’s recommended conditions, CDRH also required the following information: a time course distribution of the complications; revision of the labeling to incorporate all applicable changes (e.g., indications for use, clinical results, complications); inclusion of sterilization instructions for the instrumentation in the labeling; modifications to the surgical technique manuals; generation of a surgeon training program; and development of the post-approval studies with specific elements.
FDA issued a letter to Spine-Tech, Inc. on June 20, 1996, advising that its PMA was approvable subject to the conditions listed above as recommended by the Panel and required by FDA.
In amendments received by FDA on July 26, August 12, and August 23, 1996, Spine-Tech, Inc. submitted the requested information. The company reanalyzed the clinical outcomes using the revised definition of overall patient success (redefined again to be based on only fusion, pain, function, and muscle strength while capturing neurological information in the complication section), provided the time course distribution of complications, revised the labeling, described their surgeon training program, and developed two post-approval studies. The first post-approval study involves the collection of clinical and radiographic data for long term device performance and patient outcomes for an additional four years of follow-up (for a total of six years of postoperative data) on a subset of the IDE patient population; the goal is to obtain six years of postoperative data on a minimum of 100 patients per surgical approach. The second post-approval study involves the retrieval assessment of any $\mathrm{BAK}^{\mathrm{TM}}$ device that is implanted and subsequently removed.
PMA P950002: Summary of Safety and Effectiveness
Page 18
{18}
In amendments received by FDA on September 11 and September 13, 1996, Spine-Tech, Inc. submitted the required information which included revisions to the labeling and post-approval studies and agreed to the conditions cited in the approvable with conditions letter dated June 20, 1996. CDRH determined that, based on the above modifications, the applicant's response was adequate.
FDA inspections completed on August 6 and 7, 1996, determined the manufacturing facilities to be in compliance with the Good Manufacturing Practices (GMP) regulations.
CDRH issued an approval order on _______.
## XV. APPROVAL SPECIFICATIONS
Directions of Use: See labeling.
Hazards to Health from Use of the Device: See indications, contraindications, warnings, precautions, and adverse events in labeling.
Post-approval Requirements and Restrictions: See approval order.
## XVI. REFERENCES FOR CONTROLS
Blumenthal, S., et al., "The Role of Anterior Lumbar Fusion for Internal Disc Disruption," Spine, 13: 986-989, 1988.
Cheng, C., et al., "Anterior Spinal Fusion for Spondylolysis and Isthmic Spondylolisthesis," The Journal of Bone and Joint Surgery, 72: 264-267, 1989.
Chow, S., et al., "Anterior Spinal Fusion for Deranged Lumbar Intervertebral Disc: A Review of 97 Cases," Spine. 5: 452-458, 1980.
Cloward, R., "Long-Term Result of PLIF," In Lin, P. (es). Posterior Lumbar Interbody Fusion. Springfield, Ill. Charles C. Thomas. Chapter 8, 1982.
Dennis, S., et al., "Comparison of Disc Space Heights after Anterior Lumbar Interbody Fusion," Spine, 14: 876-878, 1989.
Flynn, J., et al., "Anterior Fusion of the Lumbar Spine," The Journal of Bone and Joint Surgery, 61A: 1143-1150, 1979.
Freebody, D., et al., "Anterior Transperitoneal Lumbar Fusion," The Journal of Bone and Joint Surgery, 53: 617-627, 1971.
PMA P950002: Summary of Safety and Effectiveness
Page 19
{19}
Fujimaki, A., et al., "The Results of 150 Anterior Lumbar Interbody Fusion Operations Performed by Two Surgeons in Australia," Clinical Orthopaedics and Related Research, 165: 164-167, 1982.
Gill, K. and Blumenthal, S., "Functional Results After Anterior Lumbar Fusion at L5-S1 in Patients with Normal and Abnormal MRI Scans," Spine, 17: 940-942, 1992.
Goldner, L., et al., "Anterior Disc Excision and Interbody Spinal Fusion for Chronic Low Back Pain," Orthopedic Clinics of North America, 2: 510-568, 1971.
Greenough, C., "Anterior Lumbar Fusion: A Comparison of Non-Compensation Patients with Compensation Patients," Clinical Orthopaedics and Related Research, 300: 30-37, 1994.
Kim, N. and Kim, D., "Anterior Interbody Fusion for Spondylolisthesis," Orthopedics, 14: 1069-1076, 1991.
Kim, N., et al., "A Computed Tomographic Analysis of Changes in the Spinal Canal After Anterior Lumbar Interbody Fusion," Clinical Orthopaedics and Related Research, 286: 180-191, 1993.
Knox, B. and Chapman, T., "Anterior Lumbar Interbody Fusion for Discogram Concordant Pain," Journal of Spinal Disorders. 6: 242-244, 1993.
Lee, K., "Chronic Disabling Low Back Pain Syndrome Caused by International Disc Derangements. The Results of Disc Excision and Posterior Lumbar Interbody Fusion," Spine, 20: 356-361, 1995.
Lin, P., et al., "Posterior Lumbar Interbody Fusion," Clinical Orthopaedics and Related Research, 180: 154-168, 1983.
Linson, M. and William, H., "Anterior and Combined Anteroposterior Fusion for Lumbar Disc Pain: A Preliminary Study," Spine, 16: 143-145, 1991.
Loguidice, V., et al., "Anterior Lumbar Interbody Fusion," Spine, 13: 366-369, 1988.
Ma, G., "Posterior Lumbar Interbody Fusion with Specialized Instruments," Clinical Orthopaedics and Related Research, 193: 57-63, 1985.
Mitsunaga, M., et al., "Microscopically Assisted Posterior Lumbar Interbody Fusion," Clinical Orthopaedics and Related Research, 263: 121-127, 1991.
Newman, M. and Grinstead, G., "Anterior Lumbar Interbody Fusion for Internal Disc Disruption," Spine, 17: 831-833, 1992.
PMA P950002: Summary of Safety and Effectiveness
Page 20
{20}
Prolo, D., et al., "Toward Uniformity in Evaluating Results of Lumbar Spine Operations: A Paradigm Applied to Posterior Lumbar Interbody Fusions," *Spine*, 11(6): 601-606, 1986.
Raugstad, T., et al., "Anterior Interbody Fusion of the Lumbar Spine." *Acta Orthop Scand*, 53: 561-565, 1982.
Rish, B., "A Critique of Posterior Interbody Fusion: 12 Years' Experience with 250 Patients," *Surg. Neurology*, 31: 281-289, 1989.
Schechter, N., et al., "Disc Disease Update Painful Internal Disc Derangements of the Lumbosacral Spine: Discographic Diagnosis and Treatment by Posterior Lumbar Interbody Fusion," *Orthopedics*, 14: 447-450, 1991.
Sorensen, K., "Anterior Interbody Lumbar Spine Fusion for Incapacitating Disc Degeneration and Spondylolisthesis," *Acta Orthop Scand*, 49: 267-277, 1978.
Stauffer, R. and Coventry, M., "Anterior Interbody Lumbar Spine Fusion," *Journal of Bone and Joint Surgery*, 54: 756-768, 1972.
Thomasen, E., "Intercorporal Lumbar Spondylosis," *Acta Orthop Scand*, 56: 287-293, 1985.
Verlooy, J., et al., "Failure of a Modified Posterior Lumbar Interbody Fusion Technique to Produce Adequate Pain Relief in Isthmic Spondylitic Grade 1 Spondylolisthesis Patients: A Prospective Study of 20 Patients," *Spine*, 16: 839-845, 1991.
PMA P950002: Summary of Safety and Effectiveness
Page 21
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.