ORTHODOC PREOPERATIVE PLANNER MODEL 500
K960685 · Integrated Surgical Systems, Inc. · HWT · Jan 17, 1997 · Orthopedic
Device Facts
| Record ID | K960685 |
| Device Name | ORTHODOC PREOPERATIVE PLANNER MODEL 500 |
| Applicant | Integrated Surgical Systems, Inc. |
| Product Code | HWT · Orthopedic |
| Decision Date | Jan 17, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.4800 |
| Device Class | Class 1 |
| Attributes | Software as a Medical Device |
Intended Use
The ORTHODOC Preoperative Planner is a software device indicated for the planning of primary total hip arthroplasties on a personal computer. The software is intended to import computed tomography data, provide a constructed image of the data, and use computerized files, representing implant patterns of femoral components, to overlay the constructed image to aid surgeons in their selection and positioning of the femoral components.
Device Story
Software-only preoperative planning tool; runs on Pentium-based PC. Inputs: CT scan data; CAD-based geometric implant surface model files. Operation: imports CT data; constructs image; allows surgeon to superimpose/manipulate scaled 2D graphic images of femoral stem components over CT femur images using mouse pointer. Output: visual overlay of implant patterns on patient anatomy. Usage: clinical setting; operated by surgeons. Benefit: aids selection of optimal stem size, position, neck length, and head diameter to achieve surgical goals. Device does not operate other medical devices or act directly on patients.
Clinical Evidence
No pre-clinical or clinical performance data generated.
Technological Characteristics
Software-only; runs on Pentium-based PC. Utilizes CAD geometric data to generate implant surface model files. Interface: mouse-driven manipulation of 2D graphic overlays on CT-derived images. No hardware components; no direct patient contact.
Indications for Use
Indicated for planning primary total hip arthroplasties in patients requiring femoral component selection and positioning.
Regulatory Classification
Identification
A template for clinical use is a device that consists of a pattern or guide intended for medical purposes, such as selecting or positioning orthopedic implants or guiding the marking of tissue before cutting.
Predicate Devices
- Depuy AML templates
- Howmedica Osteolock HA templates
Related Devices
- K031590 — SECTRA ORTHOPEDIC PACKAGE · Sectra-Imtec AB · Oct 2, 2003
- K251863 — CustomSurg OrthoPlanner · Customsurg AG · Jul 7, 2025
- K042543 — IPLAN! DIGITAL TEMPLATING · Brainlab AG · May 17, 2005
- K032401 — ORTHOVIEW · Omsbar, Ltd. · Aug 14, 2003
- K142944 — ArthroPlan Digital Templating Software · Arthromeda, Inc. · Dec 16, 2014
Submission Summary (Full Text)
{0}
JAN 7 1997
Integrated Surgical Systems, Inc.
ORTHODOC Preoperative Planning Software
510(k) Summary
K960685
Submitter:
Integrated Surgical Systems, Inc.
829 W. Stadium Lane
Sacramento, CA 95834
(916) 646-3487
Trade name:
ORTHODOCR Model 500 Preoperative Planning Software
Common name:
Templating System
Classification name:
Template for clinical use (21 CFR 888.4800)
Indications for use:
The ORTHODOC Preoperative Planner is a software device indicated for the planning of primary total hip arthroplasties on a personal computer. The software is intended to import computed tomography data, provide a constructed image of the data, and use computerized files, representing implant patterns of femoral components, to overlay the constructed image to aid surgeons in their selection and positioning of the femoral components.
Legally marketed predicate:
All femoral component templates supplied by femoral component manufacturers. For example, Depuy AML templates and Howmedica Osteolock HA templates.
Device description:
This software is designed to run on a PentiumR-based personal computer. The device is a software-only product and is not intended to operate any other medical device or act directly on a patient. The application stores scaled graphic images of femoral stem components and provides a means for surgeons to superimpose these patterns on a CT image of the patient's femur. The implant patterns can be moved over the femur images to select what in the surgeon's judgement is the best stem size, position, neck length, and head diameter to achieve the physician's surgical goal.
The library of implant patterns, called surface model files, is developed from computer-aided-design (CAD) geometric data supplied by the implant's manufacturer. The implant surface model file used by ORTHODOC contains implant identification information as well as geometric data which defines the shape and dimensions of the implant surface.
12
{1}
Premarket Notification
Integrated Surgical Systems, Inc.
ORTHODOC Preoperative Planning Software
## Technological characteristics:
The implant patterns of the predicates are scaled computer-generated images that are plotted, photographed, and contact printed to maintain accurate scale. These are held over the x-ray and moved by hand to position them to assist the surgeon in planning the stem's placement.
The ORTHODOC software contains a library of computer-generated femoral component images which can be called-up and superimposed over a CT image of the patients femur. The images are moved by hand, employing a mouse pointer, to position them to assist the surgeon in planning the stem's placement.
## Performance data:
There has been no pre-clinical or clinical performance data generated upon which to make a decision of substantial equivalence.