← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K243419

# Altus Spine Navigation System (K243419)

_Altus Partners, LLC · OLO · May 8, 2025 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K243419

## Device Facts

- **Applicant:** Altus Partners, LLC
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** May 8, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Altus Spine Navigation System reusable instruments are indicated to be used during the preparation and placement of Altus Spine Valencia Pedicle Screw System and Altus Spine Monaco Pedicle Screw System screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The Altus Spine Navigation System reusable instruments are specifically designed for use with the Medtronic StealthStation® System S8, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy. • For a comprehensive summary of indications, the IFU for the specific system of use must be directly referenced: Altus Spine Valencia Pedicle Screw System / Altus Spine Monaco Pedicle Screw System. The Altus Spine Navigation System is intended to pair with either the Altus Spine Valencia Pedicle Screw System or Altus Spine Monaco Pedicle Screw system to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The Altus Spine Navigation system may be used for noncervical pedicle fixation via posterior percutaneous approach with MIS instrumentation.

## Device Story

Reusable surgical instruments (screwdrivers, taps) used with Medtronic StealthStation S8 navigation system; assists surgeons in locating anatomical structures during spinal surgery; facilitates preparation and placement of Altus Spine Valencia and Monaco pedicle screws; used in open or minimally invasive procedures; instruments feature NavLock tracker connection; provides real-time navigation feedback to surgeon; improves precision of screw placement; benefits patient through accurate implant positioning and stabilization of spinal segments.

## Clinical Evidence

Bench testing only. Side-by-side comparison with primary predicate device performed using Medtronic StealthStation S8 system per ASTM F2554. Testing evaluated recognition, location, dimensional accuracy, and critical length. Results demonstrate substantial equivalence in performance.

## Technological Characteristics

Stainless steel (ASTM F899); reusable; non-sterile; includes screwdrivers and taps; features NavLock tracker connection; designed for use with Medtronic StealthStation S8 (V2.1.0); mechanical navigation-compatible instrumentation.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Navigated CD Horizon® Solera® Screwdrivers and Taps ([K170679](/device/K170679.md))
- Altus Spine Monaco Pedicle Screw System ([K230766](/device/K230766.md))
- Altus Spine Valencia Pedicle Screw System ([K181281](/device/K181281.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

May 8, 2025

Altus Partners LLC.
Rand Baker
QA/RA Manager
1340 Enterprise Drive
West Chester, Pennsylvania 19380

Re: K243419
Trade/Device Name: Altus Spine Navigation System
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic instrument
Regulatory Class: Class II
Product Code: OLO
Dated: April 4, 2025
Received: April 4, 2025

Dear Rand Baker:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K243419 - Rand Baker
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K243419 - Rand Baker
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tejen D. Soni -S

For

Shumaya Ali, M.P.H.

Assistant Director

DHT6C: Division of Restorative,

Repair, and Trauma Devices

OHT6: Office of Orthopedic Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K243419  |   |
|  Device Name Altus Spine Navigation System  |   |
|  Indications for Use (Describe) The Altus Spine Navigation System reusable instruments are indicated to be used during the preparation and placement of Altus Spine Valencia Pedicle Screw System and Altus Spine Monaco Pedicle Screw System screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The Altus Spine Navigation System reusable instruments are specifically designed for use with the Medtronic StealthStation® System S8, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy. • For a comprehensive summary of indications, the IFU for the specific system of use must be directly referenced: Altus Spine Valencia Pedicle Screw System / Altus Spine Monaco Pedicle Screw System. The Altus Spine Navigation System is intended to pair with either the Altus Spine Valencia Pedicle Screw System or Altus Spine Monaco Pedicle Screw system to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The Altus Spine Navigation system may be used for noncervical pedicle fixation via posterior percutaneous approach with MIS instrumentation.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

K243419

# 510(k) Summary

**SUBMITTER:** Altus Partners, LLC
1340 Enterprise Drive
West Chester, PA 19380
Phone: 610-356-6148
Fax: 610-300-3049

**CONTACT PERSON:** Rand Baker
QA/RA Manager
rbaker@altus-spine.com

**DATE PREPARED:** May 8, 2025

**TRADE OR PROPRIETARY NAME:** Altus Spine Navigation System

**PRIMARY PREDICATE DEVICES:** Medtronic Sofamor Danek, USA Inc., Navigated CD Horizon® Solera® Screwdrivers and Taps (K170679)

**ADDITIONAL PREDICATE DEVICES:**
Altus Spine Monaco Pedicle Screw System (K230766)
Altus Spine Valencia Pedicle Screw System (K181281)

**CLASSIFICATION &amp; REGULATION #:** Class II per 21 CFR §888.4560

**PRODUCT CODES:** OLO

## DEVICE DESCRIPTION:

The Altus Spine Navigation System reusable instruments are surgical instruments for use with the Medtronic StealthStation® Navigation System to assist surgeons in precisely locating anatomical structures in either open, minimally invasive, or percutaneous procedures for preparation and placement of pedicle screw system implants. The Altus Spine Navigation System includes the following instruments dedicated to screw placement: Screwdrivers and Taps. The Altus Spine Navigation System is to be used with the following Altus Spine Systems:

- Altus Spine Valencia Pedicle Screw System
- Altus Spine Monaco Pedicle Screw System

All instruments are made of stainless steel per ASTM F899. All instruments are reusable instruments provided non sterile. The Altus Spine Navigation System instruments are not compatible with implants from other manufacturers. The Altus Spine Navigation System are designed for use only with Medtronic StealthStation® System S8 (V2.1.0) and the Medtronic NavLock® Tracker System.

## INDICATIONS FOR USE:

The Altus Spine Navigation System reusable instruments are indicated to be used during the preparation and placement of Altus Spine Valencia Pedicle Screw System and Altus Spine Monaco Pedicle Screw System screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The Altus Spine Navigation System reusable instruments are specifically designed for use with the Medtronic StealthStation® System S8, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

Page 1 of 2

{5}

- For a comprehensive summary of indications, the IFU for the specific system of use must be directly referenced: Altus Spine Valencia Pedicle Screw System / Altus Spine Monaco Pedicle Screw System.

The Altus Spine Navigation System is intended to pair with either the Altus Spine Valencia Pedicle Screw System or Altus Spine Monaco Pedicle Screw system to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

The Altus Spine Navigation system may be used for noncervical pedicle fixation via posterior percutaneous approach with MIS instrumentation.

## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS:

The Altus Spine Navigation System has the same or similar technological characteristics as the predicate instruments including design (instrument functional length and NavLock® connection feature), intended use, material composition, function, range of sizes, and sterility. The subject system shares the same principle of operation as the predicates. The subject instruments differ from the primary predicate in tip geometries and dimensions in order to accommodate the spinal screws; however, the differences are minor and do not raise any new concerns of safety or effectiveness with respect to the predicate.

## PERFORMANCE DATA:

The Altus Spine Navigation System was tested side-by-side with the primary predicate device and passed recognition and location testing with the Medtronic StealthStation® System S8 navigation system via ASTM F2554. The subject device and predicate device were compared for dimensional accuracy and the critical length. The results demonstrate that the subject device is substantially equivalent to the predicate device in use and performance.

## SUBSTANTIAL EQUIVALENCE CONCLUSION:

The overall technology characteristics and mechanical performance data lead to the conclusion that the Altus Spine Navigation System is substantially equivalent to the predicate device.

K243419

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K243419](https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K243419)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
