← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K240625

# CORUS Navigation Access System (K240625)

_Providence Medical Technology, Inc. · OLO · May 3, 2024 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K240625

## Device Facts

- **Applicant:** Providence Medical Technology, Inc.
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** May 3, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The CORUS™ Navigation Access System for use with the CORUS™ Spinal System is intended to be used during spinal surgery to assist the surgeon in locating and preparing facet joints in either open, or minimally invasive procedures. The CORUSTM Navigation Access System is specifically designed for use with the Medtronic StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

## Device Story

Manually operated, disposable instrument set; includes Navigated Access Chisel, Guide Tube, and Trephine Decorticator. Used in spinal surgery to assist surgeons in locating and preparing facet joints. Designed for integration with Medtronic StealthStation™ System; utilizes stereotactic navigation to track instrument position relative to patient anatomy (CT/MR/fluoroscopy). Provides physical site preparation; navigation output assists surgeon in precise instrument placement. Used in OR by surgeons. Benefits include improved precision in facet joint preparation during open or minimally invasive spinal procedures.

## Clinical Evidence

Bench testing only. Evaluation included engineering analysis, geometric comparison to predicate devices, and validation of navigation compatibility with the Medtronic StealthStation™ System. No clinical data presented.

## Technological Characteristics

Materials: Stainless steel. Components: Navigated Access Chisel, Guide Tube, Trephine Decorticator. Energy: Manual operation. Connectivity: Compatible with Medtronic StealthStation™ System for navigation. Sterilization: Not specified. Form factor: Disposable surgical instruments.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Navigated Anterolateral Disc Prep Instruments ([K211441](/device/K211441.md))
- CORUS™ Spinal System-X ([K212636](/device/K212636.md))
- Navigated Disc Preparation Instruments ([K150231](/device/K150231.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below.

May 3, 2024

Providence Medical Technology, Inc. % Hannah Taggart, MS Regulatory Associate Empirical Technologies 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K240625

Trade/Device Name: CORUSTM Navigation Access System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: March 5, 2024 Received: March 6, 2024

Dear Ms. Taggart:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Shumaya Ali -S

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K240625

Device Name CORUSTM Navigation Access System

Indications for Use (Describe)

The CORUS™ Navigation Access System for use with the CORUS™ Spinal System is intended to be used during spinal surgery to assist the surgeon in locating and preparing facet joints in either open, or minimally invasive procedures. The CORUSTM Navigation Access System is specifically designed for use with the Medtronic StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

Type of Use (Select one or both, as applicable) � Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection including suggestions for reducing this burden to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Providence Medical Technology. The logo consists of an orange symbol on the left and the words "PROVIDENCE" and "MEDICAL TECHNOLOGY" on the right. The word "PROVIDENCE" is in gray, and the words "MEDICAL TECHNOLOGY" are in orange. There is a trademark symbol after the word "PROVIDENCE".

| 510(k) Owner:                | Providence Medical Technology, Inc.<br>4234 Hacienda Dr. Suite 150, Pleasanton, California 94588<br>T: 415-923-9376; F: 415-923-9377 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact Person:      | Edward Liou, Providence Medical Technology, Inc.<br>415-754-8593; ed@providencemt.com                                                |
| Submission Correspondent:    | Hannah Taggart, MS, Empirical Technologies<br>719-457-1152; htaggart@empiricaltech.com                                               |
| Date Summary Prepared:       | April 23, 2024                                                                                                                       |
| Trade Name:                  | CORUST™ Navigation Access System                                                                                                     |
| Common Name:                 | Stereotaxic Instruments                                                                                                              |
| Device Classification Name:  | Orthopedic Stereotaxic Instruments                                                                                                   |
| Classification & Regulation: | Class II per 21 CFR §882.4560                                                                                                        |
| Product Code:                | OLO                                                                                                                                  |
| Panel:                       | Orthopedic - Stereotaxic, Trauma, and Restorative Devices (DHT6C)                                                                    |

## PREMARKET NOTIFICATION 510(K) SUMMARY

### DEVICE DESCRIPTION:

The CORUS™ Navigation Access System is a manually operated disposable instrument set to be used with the Medtronic StealthStation™ System to assist the surgeon in precise site preparation during open or minimally invasive spinal surgery. The CORUS™ Navigation Access System includes the Navigated Access Chisel, Guide Tube, and Trephine Decorticator. The instruments are manufactured from stainless steel.

### INDICATIONS FOR USE

The CORUS™ Navigation Access System for use with the CORUS™ Spinal System is intended to be used during spinal surgery to assist the surgeon in locating and preparing facet joints in either open, or minimally invasive procedures. The CORUSTM Navigation Access System is specifically designed for use with the Medtronic StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a long bone, or vertebra can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

### TECHNOLOGICAL CHARACTERISTICS

The subject has similar indications for use, materials, sterilization method, technological characteristics, geometry, and compatibilities with Medtronic StealthStation™ System as the predicates devices and the minor differences do not raise any new issues of safety and effectiveness.

| Predicate Devices |  |
|-------------------|--|
|-------------------|--|

| 510k #  | Trade Name                                    | Manufacturer                      | Product<br>Code | Predicate<br>Type |
|---------|-----------------------------------------------|-----------------------------------|-----------------|-------------------|
| K211441 | Navigated Anterolateral Disc Prep Instruments | Medtronic Sofamor Danek USA, Inc. | OLO             | Primary           |
| K212636 | CORUS™ Spinal System-X                        | Providence Medical Technology     | HRX             | Additional        |
| K150231 | Navigated Disc Preparation Instruments        | Medtronic Sofamor Danek, USA Inc. | OLO             | Additional        |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Providence Medical Technology. The logo consists of an orange symbol on the left and the words "PROVIDENCE" in gray and "MEDICAL TECHNOLOGY" in orange on the right. The symbol is a stylized letter "P" made up of three curved lines.

### PERFORMANCE DATA

The CORUS™ Navigation Access System has been evaluated though an engineering analysis and geometric comparison to the predicate device. A validation was also conducted to demonstrate navigation compatibility with the Medtronic StealthStation™ System. The results show that the subject device is substantially equivalent to the cleared predicates.

### CONCLUSION

The overall technological characteristics, engineering analysis, geometric comparison, and navigation compatibility validation lead to the conclusion that the CORUS™ Navigation Access System is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K240625](https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K240625)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
