← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K240211

# VELYS Robotic-Assisted Solution (K240211)

_Depuy Ireland UC · OLO · Jun 5, 2024 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K240211

## Device Facts

- **Applicant:** Depuy Ireland UC
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** Jun 5, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The VELYS Robotic-Assisted Solution is intended for stereotaxic surgery to aid the surgeon in identifying the relative position and orientation of anatomical structures, planning the position of the femoral and tibial implant components intraoperatively, and preparing the bones during unicompartmental knee arthroplasty. When indicated for unicompartmental knee arthroplasty, the VELYS Robotic-Assisted Solution is indicated for use with the SIGMA HP Partial Knee Implants and INTUITION Instruments for SIGMA HP Partial Knee and their cleared indications for use.

## Device Story

Image-free robotic-assisted surgery system for unicompartmental knee arthroplasty (UKA). Uses optical tracking to acquire intraoperative anatomical landmarks; surgeon plans femoral/tibial implant positions based on preferred technique. System includes robotic arm mounted on OR bed rail; provides real-time compensation for patient leg movement during bone resection. Surgeon manually manipulates saw handpiece within planned resection planes; system constrains handpiece orientation. Operated by surgeon in OR via touchscreen and footswitch. Stores case reports for surgeon review. Benefits include precise bone preparation and alignment during knee arthroplasty.

## Clinical Evidence

Bench testing only. No clinical data required. Testing included sagittal border guard evaluation, dynamic compensation performance, cadaveric resection accuracy and soft tissue study, summative usability evaluation, and design validation.

## Technological Characteristics

Image-free robotic-assisted system; optical tracking; robotic arm mounted on OR bed rail; touchscreen/footswitch interface. Software-controlled bone resection guidance. Complies with 21 CFR Part 820 design controls.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Mako Partial Knee Application ([K172301](/device/K172301.md))
- VELYS Robotic-Assisted Solution ([K233227](/device/K233227.md))

## Reference Devices

- VELYS Robotic-Assisted Solution with TKA

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 5, 2024

DePuy Ireland UC Camille Chua Regulatory Affairs Manager Loughbeg Ringaskiddy Ireland

Re: K240211

Trade/Device Name: VELYS Robotic-Assisted Solution Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OLO Dated: May 6, 2024 Received: May 6, 2024

Dear Camille Chua:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely. Shumaya Ali -S Shumaya Ali, M.P.H. Assistant Director

DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

Submission Number (if known)

K240211

Device Name

VELYS Robotic-Assisted Solution

#### Indications for Use (Describe)

The VELYS Robotic-Assisted Solution is intended for stereotaxic surgery to aid the surgeon in identifying the relative position and orientation of anatomical structures, planning the position of the femoral and tibial implant components intraoperatively, and preparing the bones during unicompartmental knee arthroplasty.

When indicated for unicompartmental knee arthroplasty, the VELYS Robotic-Assisted Solution is indicated for use with the SIGMA HP Partial Knee Implants and INTUITION Instruments for SIGMA HP Partial Knee and their cleared indications for use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) #: K240211

510(k) Summary

Prepared on: 2024-06-05

### Contact Details

### 21 CFR 807.92(a)(1)

| Applicant Name                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                          | DePuy Ireland UC                  |                              |                     |              |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|-----------------------------------|------------------------------|---------------------|--------------|
| Applicant Address                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                          | Loughbeg Ringaskiddy Ireland      |                              |                     |              |
| Applicant Contact Telephone                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                          | +1 949-795-6191                   |                              |                     |              |
| Applicant Contact                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                          | Ms. Camille Chua                  |                              |                     |              |
| Applicant Contact Email                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                          | vchua1@its.jnj.com                |                              |                     |              |
| Correspondent Name                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                          |                                   | DePuy Ireland UC             |                     |              |
| Correspondent Address                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                          |                                   | Loughbeg Ringaskiddy Ireland |                     |              |
| Correspondent Contact Telephone                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                          |                                   | +1 949-795-6191              |                     |              |
| Correspondent Contact                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                          |                                   | Ms. Camille Chua             |                     |              |
| Correspondent Contact Email                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                          |                                   | vchua1@its.jnj.com           |                     |              |
| Device Name                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                          |                                   |                              | 21 CFR 807.92(a)(2) |              |
| Device Trade Name                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                          | VELYS Robotic-Assisted Solution   |                              |                     |              |
| Common Name                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                          | Stereotaxic instrument            |                              |                     |              |
| Classification Name                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                          | Orthopedic Stereotaxic Instrument |                              |                     |              |
| Regulation Number                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                          | 882.4560                          |                              |                     |              |
| Product Code(s)                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                          | OLO                               |                              |                     |              |
| Legally Marketed Predicate Devices<br>21 CFR 807.92(a)(3)                                                                                                                                                                                                                                                                                                                                                                                              |                                                          |                                   |                              |                     |              |
| Predicate #                                                                                                                                                                                                                                                                                                                                                                                                                                            | Predicate Trade Name (Primary Predicate is listed first) |                                   |                              |                     | Product Code |
| K172301                                                                                                                                                                                                                                                                                                                                                                                                                                                | Mako Partial Knee Application                            |                                   |                              |                     | OLO          |
| K233227                                                                                                                                                                                                                                                                                                                                                                                                                                                | VELYS Robotic-Assisted Solution                          |                                   |                              |                     | OLO          |
| Device Description Summary                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                          |                                   |                              | 21 CFR 807.92(a)(4) |              |
| The VELYS™ Robotic-Assisted Solution is an image-free robotic-assisted surgery system. It is intended for stereotaxic surgery to aid the<br>surgeon in identifying the relative position and orientation of anatomical structures and landmarks, planning the position of the femoral<br>and tibial implant components intraoperatively, and preparing the bones during total knee arthroplasty (TKA) and unicompartmental<br>knee arthroplasty (UKA). |                                                          |                                   |                              |                     |              |

The image-free system uses a dedicated optical tracking device to acquire anatomical landmarks intra-operatively. These landmarks are then used to plan the femoral and tibial implant locations based on the surgeon's preferred surgical technique and placement preferences. Following the planning step, the VELYS™ Robotic-Assisted Solution helps the surgeon to execute the bone preparation

{4}------------------------------------------------

#### K240211 510(k) #:

according to the plan.

The system includes a Robotic-Assisted Device that constrion and orientation of the saw handpiece and blade inside each resection plane on the patient's femur and tibia. The surgeon actuates and manipulates the saw handpiece, within the planned resection plane, to execute the bone resection. This is analogous to using manual instruments in TKA or UKA. If the patient's leg moves during the resection, the Robotic-Assisted Device compensates for such movement in real-time.

The Robotic-Assisted Device is assembled with a Robotic-Assisted Device arm, mounted on the Operating Room (OR) bed rail, for a minimal footprint

The VELYS® Robotic-Assisted Solution incorporates several software subsystems, including applications responsible for general operation of the system and clinical applications dedicated to the surgery workflow.

The users interact with the clinical applications via a touchscreen and footswitch to navigate through the surgery steps. Case Reports including key surgical procedure information are stored on the system and can be retrieved by the surgeon for future use. Cases Reports including PHI are only available to the surgeon who performed the procedure.

# Intended Use/Indications for Use

The VELYS Robotic-Assisted Solution is intended for stereotaxic surgery to aid the surgeon in identifying the relative position and orientation of anatomical structures, planning the femoral and tibial implant components intraoperatively, and preparing the bones during unicompartmental knee arthroplasty.

When indicated for unicompartmental knee arthroplasty, the VELYS Robotic-Assisted Solution is indicated for use with the SIGMA HP Partial Knee Implants and INTUITION Instruments for SIGMA HP Partial Knee and their cleared indications for use.

# Indications for Use Comparison

The subject device has the same indications for use as the predicate device.

# Technological Comparison

The subject device and the predicate device have similar technological characteristics with respect to principles of operation, system design, and performance. The subject and predict device achieve the same outcomes, however the main design differences of the subject device include: image-free planning and patient anatomy assessment, robotic-assisted bone resection, and surgeon control. Performance testing on the subject device and comparison to the reference device, VELYS™ Robotic-Assisted Solution with TKA, support that there are no new questions of safety or effectiveness.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Bench testing was performed on the subject device, VELYS™ Robotic-Assisted Solution for Unicompartmental Knee Arthroplasty (UKA), in accordance with the product risk analysis and product requirements, in line with FDA 21 CFR Part 820, and to demonstrate substantial equivalence with the predicate device. Where available, FDA-consensus standards and/or guidance documents were used. New bench testing specific to the expanded indication to include UKA and leveraged testing from the VELYS™ Robetic-Assisted Solution Total Knee Arthroplasty (TKA) reference device were performed for assessing device performance.

Performance testing activity included

- · Sagittal Border Guard Evaluation
- · Dynamic Compensation Performance Test
- · Cadaveric Resection Accuracy and Soft Tissue Study
- · Summative Usability Evaluation
- · Design Validation

No clinical data was necessary to support the determination of substantial equivalence

Both the subject device and predicate device have intended use, same indications for use, and similar technological characteristics. Where there are technological differences, performance testing was conducted on the subject device and VELYS™ Robotic-Assisted Solution with TKA was utilized as a reference device to show that no new questions of safety or effectiveness were raised; therefore DePuy Synthes concludes that the subject and predicate devices are substantially equivalent.

# 21 CFR 807.92(a)(5)

### 21 CFR 807.92(a)(6)

## 21 CFR 807.92(a)(5)

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K240211](https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K240211)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
