The SURGIVISIO Device is indicated to be used during surgical procedures in which the physician would benefit from the visualization of 2D medical intraoperatively generated 3D medical imaging of anatomical structures or objects with high x-ray attenuation such as bony anatomy or metallic objects. Such procedures during which the spine, pelvis, or articulation structures are visualized The SURGIVISIO Device through its freehand navigation feature is indicated as an intraoperative guidance system to enable open or percutaneous computer-assisted surgery. It is indicated for conditions of the spine in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical bony structure can be identified relative to the intraoperatively generated 3D image of the anatomy provided by the SURGIVISIO Device. It is indicated to precisely position the Navigated Targeting Needle SPX I during general spinal procedures with a posterior approach. The SURCIVISIO Device through its robotic guidance feature is indicated for the positioning of instrument holders or tool guides to be used by surgeons to guide the Spine CoBot instruments during general spinal surgery. Guidance is based on an intra-operative plan developed with three-dimensional imaging software based on intra-operative 3D images provided by the SURGIVISIO Device. It is indicated for positioning of surgical instruments in vertebrae with a posterior approach in the thoracolumbar region. The SPX1 Instrument and Spine CoBot instruments are designed to be used with the SURGIVISIO Device.
Device Story
SURGIVISIO provides 2D/3D intraoperative imaging and stereotactic guidance for spinal surgery. System includes stereoscopic infrared camera, computer station with monitors, navigation software, and robotic arm (CoBot). Freehand navigation uses optical tracking of reflective marker spheres on instruments. Robotic guidance uses same optical tracking to position instrument holders/tool guides based on 3D intraoperative plans. Surgeon operates robotic arm via foot pedal and manual buttons for automatic alignment and cooperative positioning. Used in OR by surgeons to guide instruments during open or percutaneous spinal procedures. Output includes real-time visualization and robotic positioning of surgical tools. Benefits include precise instrument placement in vertebrae, potentially improving surgical accuracy and safety.
Clinical Evidence
No clinical tests were conducted. Substantial equivalence is supported by non-clinical bench testing, including reprocessing validation and a cadaveric study assessing the accuracy and safety of the robotic guidance feature following design modifications.
Technological Characteristics
Stereotactic robotic navigation system. Components: infrared stereoscopic camera, computer station, robotic arm (7-DOF), and specialized surgical instruments. Connectivity: networked system. Software: 3D Spine Universal/Robotic Workflow applications. Accuracy: <2mm, <2 degrees angular error. Sterilization: reprocessing validation performed for instruments.
Indications for Use
Indicated for patients with spinal conditions requiring stereotactic surgery and placement of spinal instruments, where rigid bony anatomy can be referenced to intraoperative 3D images. Used for posterior approach spinal procedures in the thoracolumbar region.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date July 27, 2023. The month is July, the day is 27, and the year is 2023. The date is written in a clear, legible font. The text is black on a white background.
Ecential Robotics Elodie Bouillet Design Process & Regulatory Affaires Manager Zone Mayencin II, Parc Equation - Bâtiment 1
Gieres, 38610 France
tance
Re: K231886
2 avenue de Vignate
Trade/Device Name: SURGIVISIO Device Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: June 27, 2023 Received: June 27, 2023
Dear Elodie Bouillet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
## Shumaya Ali -S
Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231886
Device Name SURGIVISIO Device
#### Indications for Use (Describe)
The SURGIVISIO Device is indicated to be used during surgical procedures in which the physician would benefit from the visualization of 2D medical intraoperatively generated 3D medical imaging of anatomical structures or objects with high x-ray attenuation such as bony anatomy or metallic objects. Such procedures during which the spine, pelvis, or articulation structures are visualized
The SURGIVISIO Device through its freehand navigation feature is indicated as an intraoperative guidance system to enable open or percutaneous computer-assisted surgery.
It is indicated for conditions of the spine in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical bony structure can be identified relative to the intraoperatively generated 3D image of the anatomy provided by the SURGIVISIO Device.
It is indicated to precisely position the Navigated Targeting Needle SPX I during general spinal procedures with a posterior approach.
The SURCIVISIO Device through its robotic guidance feature is indicated for the positioning of instrument holders or tool guides to be used by surgeons to guide the Spine CoBot instruments during general spinal surgery.
Guidance is based on an intra-operative plan developed with three-dimensional imaging software based on intra-operative 3D images provided by the SURGIVISIO Device.
It is indicated for positioning of surgical instruments in vertebrae with a posterior approach in the thoracolumbar region.
The SPX1 Instrument and Spine CoBot instruments are designed to be used with the SURGIVISIO Device.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> 区 Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# Submitter Information
| Submitter: | ECENTIAL ROBOTICS<br>2 avenue de Vignate<br>Zone Mayencin II, Parc Equation – Bâtiment 1<br>38610 Gières<br>France |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Elodie BOUILLET<br>Project Quality & Regulatory Affairs Manager<br>Phone : 0033 6 37 82 07 90<br>Email: elodie.bouillet@ecential-robotics.com |
Date Summary Prepared: June 27, 2022
## Device Information
| Trade or proprietary<br>name | SURGIVISIO Device |
|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common or usual name<br>Classification Name<br>Regulation Number<br>Regulatory class<br>Primary product code | Orthopedic Stereotaxic Instrument<br>Stereotaxic Instrument<br>21 CFR 882.4560<br>II<br>OLO |
| Legally marketed device<br>to which equivalence is<br>claimed | SURGIVISIO Device by ECENTIAL ROBOTICS:<br>- Primary predicate K221028<br>- Other predicates<br>o K202547<br>o K220946<br>o K220627 |
| Scope of the<br>submission | The subject device is the SURGIVISIO Device which<br>enables 2D/3D medical imaging and stereotaxic guidance.<br>The 2D/3D medical imaging feature was cleared originally<br>via 510(k) K202547 and via Special 510(k) K220946 and<br>K220627. The stereotaxic guidance including freehand<br>navigation and robotic guidance feature was cleared via<br>510(k) K221028.<br>The scope of this current submission is limited to the design<br>modification of the Spine CoBot instruments which are used<br>for the Robotic Guidance feature. |
| Device Description | The SURGIVISIO Device is a medical device that provides<br>2D/3D medical imaging and stereotaxic guidance. The |
{5}------------------------------------------------
subject device offers two stereotaxic quidance features: freehand navigation and robotic guidance.
The freehand navigation feature is based on the standard and established technique of navigation systems utilizing optical position determination technology. Like currently marketed optical tracking navigation systems, the operating principle of the freehand navigation feature is based upon the use of a stereoscopic camera emitting infrared light which can determine a 3D position of reflective marker spheres. This allows for real-time tracking of the marker spheres. The system components include a stereoscopic camera (SURGIVISIO Camera Pole), a computer platform with monitors (SURGIVISIO Station) and navigation software (3D Spine Universal Workflow software application) and instruments equipped with marker spheres to enable an exact localization in space.
The robotic quidance feature utilizes the same principle of optical position determination technology. The system components include a stereoscopic camera (SURGIVISIO Camera Pole), a computer platform with monitors (SURGIVISIO Station) and a navigation software (3D Spine Robotic Workflow software application), a robotic arm (CoBot), and instruments equipped with marker spheres to enable an exact localization in space.
Indications for use The SURGIVISIO Device is indicated to be used during surgical procedures in which the physician would benefit from the visualization of 2D medical imaging and/or intraoperatively generated 3D medical imaging of anatomical structures or objects with high x-ray attenuation such as bony anatomy or metallic objects. Such procedures include procedures during which the spine, pelvis, or articulation structures are visualized.
> The SURGIVISIO Device through its freehand navigation feature is indicated as an intraoperative guidance system to enable open or percutaneous computer-assisted surgery. It is indicated for conditions of the spine in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical bony structure can be identified relative to the intraoperatively generated 3D image of the anatomy provided by the SURGIVISIO Device. It is indicated to precisely position the Navigated Targeting Needle SPX1 during general spinal procedures with a posterior approach.
> The SURGIVISIO Device through its robotic quidance feature is indicated for the positioning of instrument holders or tool guides to be used by surgeons to guide the Spine CoBot instruments during general spinal surgery. Guidance
{6}------------------------------------------------
is based on an intra-operative plan developed with threedimensional imaging software based on intraoperative 3D images provided by the SURGIVISIO Device. It is indicated for positioning of surgical instruments in vertebrae with a posterior approach in the thoracolumbar region.
The SPX1 Instruments and Spine CoBot instruments are designed to be used with the SURGIVISIO Device.
- Technology comparison Increase of instruments diameters, modification of the shape of the CoBot, Drill Guide - Sleeve teeth, CoBot, Drill Bit tip and CoBot, Drill Bit connection to power tool. The technology is not impacted by the discussed device change.
- Biocompatibility Same as the predicate. The discussed device change, object of this Special 510(K) submission does not require to re-perform Biocompatibility testing as the raw material, manufacturing additives as well as manufacturing processes are not modified following the design change.
- Electrical Safety Testing Same as the predicate. The discussed device change, object of this Special 510(K) submission does not impact Electrical Safety.
- Performance Testing Same as the predicate. The accuracy and safety assessment (cadaveric study) were redone in order to ensure that the design modification on the Spine CoBot Instruments do not impact the accuracy and the safety of the Robotic Guidance Feature of the SURGIVISIO Device.
Software Testing Same as the predicate
- Based upon comparison of devices and performance Conclusion testing results. SURGIVISIO Device is substantially equivalent to the predicate device.
## Summary of the technological characteristics of the device compared to the predicate device
| Characteristics | Robotic guidance feature<br>of the SURGIVISIO Device<br>(Subject device) | Robotic guidance feature<br>of the SURGIVISIO Device<br>(K221028)<br>(Predicate device) | Discussion |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The SURGIVISIO Device<br>through its robotic guidance<br>feature is indicated for the | The SURGIVISIO Device<br>through its robotic guidance<br>feature is intended for the | Similar: word "intended"<br>replaced by "indicated", |
| Characteristics | Robotic guidance feature<br>of the SURGIVISIO Device<br>(Subject device) | Robotic guidance feature<br>of the SURGIVISIO Device<br>(K221028)<br>(Predicate device) | Discussion |
| | positioning of instrument<br>holders or tool guides to be<br>used by surgeons to guide<br>the Spine CoBot instruments<br>during general spinal<br>surgery.<br>Guidance is based on an<br>intra-operative plan<br>developed with three-<br>dimensional imaging<br>software based on intra-<br>operative 3D images<br>provided by the<br>SURGIVISIO Device.<br>It is indicated for positioning<br>of surgical instruments in<br>vertebrae with a posterior<br>approach in the<br>thoracolumbar region.<br>The SPX1 instruments and<br>Spine CoBot instruments<br>are designed to be used<br>with the SURGIVISIO<br>Device. | positioning of instrument<br>holders or tool guides to be<br>used by surgeons to guide<br>the Spine CoBot instruments<br>during general spinal<br>surgery.<br>Guidance is based on an<br>intra-operative plan<br>developed with three-<br>dimensional imaging<br>software based on intra-<br>operative 3D images<br>provided by the<br>SURGIVISIO Device.<br>It is indicated for positioning<br>of surgical instruments in<br>vertebrae with a posterior<br>approach in the<br>thoracolumbar region.<br>The SPX1 instruments and<br>Spine CoBot instruments<br>are intended to be used with<br>the SURGIVISIO Device. | "intended" replaced by<br>"designed" while the<br>statement of indications<br>for use is the same. Safety<br>and performance of the<br>system are not questioned. |
| Anatomical site | The SURGIVISIO Device<br>through its robotic guidance<br>feature is indicated for<br>positioning of surgical<br>instruments in vertebrae<br>with a posterior approach in<br>the thoracolumbar region. | The SURGIVISIO Device<br>through its robotic guidance<br>feature is indicated for<br>positioning of surgical<br>instruments in vertebrae<br>with a posterior approach in<br>the thoracolumbar region. | Identical |
| Patient population | The SURGIVISIO Device<br>through its robotic guidance<br>feature is indicated for a<br>population with medical<br>conditions requiring the<br>treatment of diseases with<br>the placement of spinal<br>instruments and for which<br>the use of stereotactic<br>surgery may be considered<br>to be appropriate and after<br>consideration of the<br>compatibility of patient<br>spinal anatomy with<br>dimensions of the Spine<br>CoBot Instruments and<br>dimensions of virtual<br>implants proposed by the 3D<br>Spine Robotics workflow<br>surgical plan. | The SURGIVISIO Device<br>through its robotic guidance<br>feature is indicated for a<br>population with medical<br>conditions requiring the<br>treatment of diseases with<br>the placement of spinal<br>instruments and for which<br>the use of stereotactic<br>surgery may be considered<br>to be appropriate and after<br>consideration of the<br>compatibility of patient<br>spinal anatomy with<br>dimensions of the Spine<br>CoBot Instruments and<br>dimensions of virtual<br>implants proposed by the 3D<br>Spine Robotics workflow<br>surgical plan. | Identical |
| Characteristics | Robotic guidance feature<br>of the SURGIVISIO Device<br>(Subject device) | Robotic guidance feature<br>of the SURGIVISIO Device<br>(K221028)<br>(Predicate device) | Discussion |
| General device<br>technology description | Computer-controlled<br>electromechanical arm<br>guiding neurosurgical<br>instruments | Computer-controlled<br>electromechanical arm<br>guiding neurosurgical<br>instruments | Identical |
| Global system<br>functional principal | Stereotactic robotic<br>navigation guidance of spine<br>surgical instrument based<br>on an intra-operative plan<br>developed with three-<br>dimensional imaging<br>software which is based on<br>intraoperative 3D images<br>using an optical system<br>(infrared camera) | Stereotactic robotic<br>navigation guidance of spine<br>surgical instrument based<br>on an intra-operative plan<br>developed with three-<br>dimensional imaging<br>software which is based on<br>intraoperative 3D images<br>using an optical system<br>(infrared camera) | Identical |
| Surgical work Flow | 1. Patient and device<br>installation<br>2. Intraoperative image<br>acquisition<br>3. Registration<br>4. Intraoperative planning<br>5. Spine instrument holder<br>positioning<br>6. Instrument guidance<br>7. Screw implantation | 1. Patient and device<br>installation<br>2. Intraoperative image<br>acquisition<br>3. Registration<br>4. Intraoperative planning<br>5. Spine instrument holder<br>positioning<br>6. Instrument guidance<br>7. Screw implantation | Identical |
| Graphical User<br>Interface | System-specific GUI | System-specific GUI | Identical |
| Surgeon control | - The surgeon<br>controls the motion<br>of the robotic arm<br>from one screw<br>position to the next<br>by pressing the foot<br>pedal and the<br>manual button. | - The surgeon<br>controls the motion<br>of the robotic arm<br>from one screw<br>position to the next<br>by pressing the foot<br>pedal and the<br>manual button. | Identical |
| | The surgeon carries out the<br>final positioning through the<br>instrument guide with a<br>traditional surgical<br>instrument. | The surgeon carries out the<br>final positioning through the<br>instrument guide with a<br>traditional surgical<br>instrument. | |
| Device accuracy | - Device accuracy: <2<br>mm<br>- Angular error: <2<br>degrees | - Device accuracy: <2<br>mm<br>- Angular error: <2<br>degrees | Identical |
| SYSTEM COMPONENTS DESIGN | | | |
| | | | |
| Characteristics | Robotic guidance feature<br>of the SURGIVISIO Device<br>(Subject device) | Robotic guidance feature<br>of the SURGIVISIO Device<br>(K221028)<br>(Predicate device) | Discussion |
| Overall Device Design | SURGIVISIO Device<br>Robotic guidance feature is<br>composed of the following<br>main components:<br>- Stations<br>- Dedicated software<br>- Robotic arm<br>- Camera pole<br>- Dedicated<br>instruments | SURGIVISIO Device<br>Robotic guidance feature is<br>composed of the following<br>main components:<br>- Stations<br>- Dedicated software<br>- Robotic arm<br>- Camera pole<br>- Dedicated<br>instruments | Identical |
| Stations | - Base station<br>(SURGIVISIO<br>Station): two display<br>monitors, computer<br>unit, power supply.<br>- Robotic Station<br>(CoBot): robotic<br>arm, control box for<br>the robotic arm,<br>power supply, foot<br>pedal | - Base station<br>(SURGIVISIO<br>Station): two display<br>monitors, computer<br>unit, power supply.<br>- Robotic Station<br>(CoBot): robotic<br>arm, control box for<br>the robotic arm,<br>power supply, foot<br>pedal | Identical |
| Software | - Dedicated spine<br>application software | - Dedicated spine<br>application software | Identical |
| Robotic arm | - Guides spine<br>instruments during<br>general spine<br>surgery<br>- Instruments are<br>mounted onto the<br>robot arm's flange<br>- Uses seven degrees<br>of freedom<br>architecture to guide<br>instruments<br>- Uses foot pedal<br>and/or manual<br>buttons for<br>automatic alignment<br>and cooperative<br>mode<br>- Has light indicators<br>that reflect the<br>status of the robotic<br>guidance | - Guides spine<br>instruments during<br>general spine<br>surgery<br>- Instruments are<br>mounted onto the<br>robot arm's flange<br>- Uses seven degrees<br>of freedom<br>architecture to guide<br>instruments<br>- Uses foot pedal<br>and/or manual<br>buttons for<br>automatic alignment<br>and cooperative<br>mode<br>- Has light indicators<br>that reflect the<br>status of the robotic<br>guidance | Identical |
| Optical localizer | - An infrared camera<br>detects reflective<br>markers to track the<br>position of the | - An infrared camera<br>detects reflective<br>markers to track the<br>position of the | Identical |
| Characteristics | Robotic guidance feature<br>of the SURGIVISIO Device<br>(Subject device) | Robotic guidance feature<br>of the SURGIVISIO Device<br>(K221028)<br>(Predicate device) | Discussion |
| | robotic arm and<br>instruments | robotic arm and<br>instruments | |
| Instruments | - Robotic instruments<br>(Spine CoBot<br>Instruments: CoBot,<br>Tool Holder;<br>Robotic, Tool holder<br>- Tracker; CoBot,<br>Drill Guide - Cortical<br>Tip; CoBot, Drill<br>Guide - Sleeve;<br>CoBot, Drill Bit; 3.5<br>mm Hex Head<br>Screwdriver;<br>Fixation Screws,<br>CoBot Spine<br>Instruments Tray)<br>- Navigation<br>instruments (SPX1<br>instruments) | - Robotic instruments<br>(Spine CoBot<br>Instruments: CoBot,<br>Tool Holder;<br>Robotic, Tool holder<br>- Tracker; CoBot,<br>Drill Guide - Cortical<br>Tip; CoBot, Drill<br>Guide - Sleeve;<br>CoBot, Drill Bit; 3.5<br>mm Hex Head<br>Screwdriver;<br>Fixation Screws,<br>CoBot Spine<br>Instruments Tray)<br>- Navigation<br>instruments (SPX1<br>instruments) | Similar: the design of<br>CoBot, Tool Holder,<br>CoBot, Drill Bit, CoBot,<br>Drill Guide - Sleeve,<br>and CoBot, Drill Guide<br>– Cortical Tip is<br>modified (scope of this<br>submission) however<br>the use of those<br>instruments is<br>unchanged and the<br>performance and the<br>safety of the system is<br>not impacted (see<br>section 11 - Design<br>Control Activities<br>Summary). |
{7}------------------------------------------------
K231886
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
## Performance Data
For further information about testing, standards, acceptance criteria and results, please refer to section 11 Design Control Activities Summary.
- Nonclinical tests: The following nonclinical tests were performed on the SURGIVISIO Device to demonstrate substantial equivalence of safety and efficacy with the predicate device:
- The Sterilization validation part of the reprocessing validation of the Spine CoBot Instruments has been performed.
- A cadaveric study has been performed to assess: -
- The accuracy of the robotic quidance feature O
- The safety of the robotic guidance feature o
Clinical tests: No clinical tests were conducted to demonstrate substantial equivalence.
## Conclusions drawn from Performance Data
The SURGIVISIO Device is identical in indications for use and technological characteristics as the proposed predicate device. These aspects, along with the performance testing conducted, demonstrate the substantial equivalence SURGIVISIO to the Device (K202547/K220946/K220627/K221028) and that the SURGIVISIO Device does not raise different questions of safety and effectiveness when compared to this predicate.
● ecential robotics
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.