K223649 · Silony Medical GmbH · OLO · Dec 22, 2022 · Neurology
Device Facts
Record ID
K223649
Device Name
VERTICALE® Navigation Instruments
Applicant
Silony Medical GmbH
Product Code
OLO · Neurology
Decision Date
Dec 22, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
Silony Medical Navigation Instruments are intended to be used during the preparation and placement of VERTICALE® pedicle screws during spinal surgery to assist the surgeon in precise locating anatomical structures in spinal procedures. These instruments are designed for use with stereotactic navigation system Medtronic StealthStation®, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
Device Story
VERTICALE® Navigation Instruments are reusable surgical tools (awls, probes, taps, drivers, inserts) used in spinal surgery; designed for use with Medtronic StealthStation® navigation system. Instruments interface with navigation hardware to assist surgeons in precise localization of anatomical structures and placement of VERTICALE® pedicle screws. Surgeon operates instruments manually while navigation system tracks position relative to patient anatomy (CT/MR/fluoroscopy/digitized landmarks). Output is visual guidance on navigation display; assists surgeon in accurate implant placement; benefits patient through increased surgical precision. Used in OR setting.
Clinical Evidence
Bench testing only. Navigational accuracy verified per ASTM F2554-18. Technical evaluation demonstrated compatibility and functionality of new instruments by comparing against previously cleared VERTICALE® Navigation Instruments (K212007) and Medtronic Navigated Instruments (K161210). Results from K212007 verification are transferable to new components.
Technological Characteristics
Reusable surgical instruments (awls, probes, taps, drivers, inserts) made of stainless steel per ASTM F899. Designed for use with Medtronic StealthStation® stereotactic navigation system. Connectivity is physical/mechanical interface with navigation hardware. Sterilization method not specified, but instruments are reusable. Accuracy testing performed per ASTM F2554-18.
Indications for Use
Indicated for use during spinal surgery for preparation and placement of VERTICALE® pedicle screws to assist surgeons in precise localization of anatomical structures. Designed for use with Medtronic StealthStation® stereotactic navigation systems in patients where stereotactic surgery is appropriate and anatomical structures can be referenced via CT, MR, fluoroscopy, or digitized landmarks.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
K212028 — Dark Star Navigation Instrument System · Republic Spine, LLC · Sep 27, 2021
K230861 — VADER® Pedicle System Navigated Instruments · Icotec AG · Jun 27, 2023
K180796 — Spineology Navigation Instruments · Spineology, Inc. · May 2, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
December 22, 2022
Silony Medical GmbH Lisa Schwarz Regulatory Affairs Manager Leinfelder Straße 60 Leinfelden-Echterdingen, Baden-Württemberg D-70771 Germany
Re: K223649
Trade/Device Name: VERTICALE® Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: December 6, 2022 Received: December 6, 2022
Dear Lisa Schwarz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jesse Muir -S
For: Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223649
Device Name VERTICALE® Navigation Instruments
#### Indications for Use (Describe)
Silony Medical Navigation Instruments are intended to be used during the preparation and placement of VERTICALE® pedicle screws during spinal surgery to assist the surgeon in precise locating anatomical structures. These instruments are designed for use with stereotactic navigation system Medtronic StealthStation®, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5.510(K) SUMMARY
| Submitter's Name: | Silony Medical GmbH |
|----------------------------|---------------------------------------------------------------------------------------|
| Submitter's Address: | Leinfelder Straße 60<br>D-70771 Leinfelden-Echterdingen, Germany |
| Submitter's Telephone: | +49 (0) 711-782 525 0 |
| Contact Person: | Lisa Schwarz<br>Silony Medical GmbH<br>+49711 78252515<br>LSchwarz@Silony-Medical.com |
| Date Summary was Prepared: | December 6th, 2022 |
| Trade or Proprietary Name: | VERTICALE® Navigation Instruments |
| Common or Usual Name: | Orthopedic Stereotaxic Instrument |
| Classification Name: | Stereotaxic instrument |
| Classification: | Class II per 21 CFR 882.4560 |
| Product Code: | OLO |
| Classification Panel: | Division of Orthopedic Devices |
## DEVICE DESCRIPTION:
Silony Medical's VERTICALE® navigation instruments are reusable surgical instruments for use with the Medtronic® StealthStation® Navigation System is designed to assist surgeons in the precise localization of anatomical structures, preparation and placement of pedicle screw implants during spinal procedures.
The VERTICALE® Navigation Instruments include awls, probes, taps, drivers and inserts. The VERTICALE® Navigation Instruments are to be used with the VERTICALE® Posterior Spinal Fixation System cleared under K171421.
All instruments are made of stainless steel per ASTM F899. The VERTICALE Navigation Instruments are not compatible with implants from other manufacturers and are designed for use only with Medtronic StealthStation Navigation System hardware and software.
## INTENDED USE / INDICATIONS FOR USE
Silony Medical Navigation Instruments are intended to be used during the preparation and placement of VERTICALE® pedicle screws during spinal surgery to assist the surgeon in precise locating anatomical structures in spinal procedures. These instruments are designed for use with stereotactic navigation system Medtronic StealthStation®, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
{4}------------------------------------------------
# SUBSTANTIAL EQUIVALENCE SUMMARY
The technological characteristics of the subject VERTICALE Navigation Instruments components remain the same as, or similar to, the predicate VERTICALE Navigation Instruments (K212007) in regards to intended use, indications for use, design, manufacturing methods and fundamental technology. The purpose of this submission is to seek clearance for four additional VERTICALE Navigation instruments for spinal surgery.
| | | Table 5 Predicate Devices | |
|--|--|---------------------------|--|
|--|--|---------------------------|--|
| 510k Number | Trade or Proprietary or Model Name | Manufacturer | Predicate Type |
|----------------------------------------------------|---------------------------------------------|------------------------------------|---------------------|
| K212007 | VERTICALE® Navigation Instruments | Silony Medical GmbH | Primary |
| Predicate Device from the previous 510(k); K212007 | | | |
| K161210<br>(PRIMARY)/K143628/<br>K140454/K143375 | Medtronic Navigated Instruments | Medtronic Sofamore Danek USA, Inc. | Reference (K212007) |
| K171421 | VERTICALE® Posterior Spinal Fixation System | Silony Medical GmbH | Reference (K212007) |
## PERFORMANCE DATA
The VERTICALE Navigation Instruments have been tested per ASTM F2554-18. "Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems".
These performance data are also applicable for the new instruments.
The additional instruments have been developed to complement Silony Medical's already approved navigable instruments (VERTICALE® Navigation Instruments; K212007) and to fully cover the intended use of the VERTICALE® Posterior Spinal Fixation System ((K171421); Section 11).
By means of a technical evaluation, the compatibility and functionality of the additional navigable instrument for spine surgery was demonstrated. Data from the already cleared navigable instruments (VERTICALE® Navigation Instruments; K212007 and Medtronic Navieated Instruments; K161210) served as the basis for the technical evaluation. The technical evaluation relates to the navigational accuracy of the instruments in terms of position verification.
As result of the verification, relevant interfaces and functionalities between the existing and the new instruments are (essentially) equivalent to the existing instruments. Therefore, the results of the verification tests (K212007) are directly transferable to the new additional instruments and repetition of these tests is not necessary.
{5}------------------------------------------------
### RISK MANAGEMENT
Risk Assessment of the VERTICALE Navigation Instruments was performed in accordance with ISO 14971:2019, Medical Devices - Application of Risk Management to Medical Devices. The purpose of this analysis was to identify potential new hazards and failure modes introduced by the new devices. This risk analysis resulted in the inclusion of additional references in the labeling (surgical technique).
The previous verifications and risk controls for the predicate VERTICALE Navigation Instruments (K212007) remain valid and acceptable as the controls used are unchanged by the introduction of the new instrument components presented in this submission.
The overall residual risk is acceptable and the medical benefits continue to outweigh the overall residual risk.
The Special 510(k) Program issued by the FDA on September 13, 2019 is appropriate for this submission as the proposed change is being submitted by the legal manufacturer and a risk analysis format supports substantial equivalence
### SUBSTANTIAL EQUIVALENCE CONCLUSION
The new VERTICALE Navigation Instruments are substantially equivalent to the predicate system as a spinal fixation device in regards to intended use, indications for use, fundamental technology including design, materials, manufacturing methods, sterility, and general operational principles. Furthermore, accuracy testing and other supporting information sufficiently demonstrate the substantial equivalence of the subject components to the navigated instrument system which has been cleared for stereotactic guidance during orthopedic surgery procedures.
Based on this information, the subject device does not raise any new issues regarding the safety or efficacy when compared to its predicates (K212007).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.