← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K223553

# Spine Planning (2.0), Elements Spine Planning, Elements Planning Spine (K223553)

_Brainlab AG · OLO · Aug 2, 2023 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K223553

## Device Facts

- **Applicant:** Brainlab AG
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** Aug 2, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** AI/ML, Software as a Medical Device

## Indications for Use

Spine Planning is intended for pre- and intraoperative planning of open and minimally invasive spinal procedures. It displays digital patient images (CT, Cone Beam CT, MR, X-ray) and allows measurement and planning of spinal implants like screws and rods.

## Device Story

Spine Planning 2.0 is a software tool for pre- and intraoperative planning of spinal surgeries. It accepts 2D X-ray images and 3D datasets (CT, MR, Cone Beam CT) as input. The software uses CNN-based AI/ML algorithms to detect anatomical landmarks on 2D images for vertebrae labeling and to detect vertebrae on Digitally Reconstructed Radiographs (DRR) for atlas registration. It provides automated proposals for spinal parameters, measurements, and screw/rod placement. Used in office or operating room settings by clinicians, the software allows manual review, adjustment, and approval of all automated proposals. The output is displayed to the surgeon to assist in surgical planning and implant positioning. By providing automated anatomical labeling and implant trajectory suggestions, the device aims to improve planning accuracy and efficiency for spinal procedures.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, including software verification (unit/integration tests), AI/ML landmark detection assessment (quantifying object detection and vertebra level assignment), screw proposal algorithm evaluation (comparing new algorithm to predicate), and summative usability testing. Usability testing confirmed the device can be used by the intended user group without critical use-related problems.

## Technological Characteristics

Software-based planning tool. Utilizes CNN-based AI/ML algorithms for landmark detection and vertebra labeling. Supports CT, MR, Cone Beam CT, and X-ray modalities. Features include automated vertebra labeling, spinal parameter measurement, and screw/rod placement proposals. Operates on standard clinical computing hardware. Software verification covers SOUP and cybersecurity.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Spine Planning 1.0 ([K212245](/device/K212245.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 2, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Brainlab AG Sadwini Suresh OM Consultant Olof-Palme-Str.9 Munich, BY 81829 Germany

Re: K223553

Trade/Device Name: Spine Planning 2.0 Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, LLZ Dated: June 30, 2023 Received: June 30, 2023

Dear Sadwini Suresh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali-S

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K223553

Device Name Spine Planning 2.0

Indications for Use (Describe)

Spine Planning is intended for pre- and intraoperative planning of open and minimally invasive spinal procedures. It displays digital patient images (CT, Cone Beam CT, MR, X-ray) and allows measurement and planning of spinal implants like screws and rods.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "BRAINLAB" in pink, with a stylized pink symbol to the left of the word. The symbol appears to be three curved lines stacked on top of each other. The word "BRAINLAB" is written in all capital letters and is in a sans-serif font. The overall image has a clean and modern look.

# K223553 510(k) Summary

June 30, 2023

| General Information        |                                                                                                                                                                                                                                                                                                                           |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer               | Brainlab AG; Olof-Palme Str.9; 81829, Munich, Germany                                                                                                                                                                                                                                                                     |
| Establishment Registration | 8043933                                                                                                                                                                                                                                                                                                                   |
| Trade Names                | Spine Planning 2.0<br>• Spine Planning<br>• Elements Spine Planning<br>• Elements Planning Spine                                                                                                                                                                                                                          |
| Classification Name        | Orthopedic Stereotaxic Instrument                                                                                                                                                                                                                                                                                         |
| Product Code               | OLO                                                                                                                                                                                                                                                                                                                       |
| Regulation Number          | 882.4560                                                                                                                                                                                                                                                                                                                  |
| Regulatory Class           | II                                                                                                                                                                                                                                                                                                                        |
| Panel                      | Orthopedic                                                                                                                                                                                                                                                                                                                |
| Predicate Device           | K212245<br>Spine and Trauma Navigation System, Spine & Trauma 3D<br>Navigation, Instrument Selection, Fluoro 3D, Registration Software<br>Fluoro 3D, Registration Software Paired Point, Registration Software<br>Spine Surface Matching, Spine Planning, Elements Screw Planning<br>Spine, Elements Spine Screw Planning |
| Contact Information        |                                                                                                                                                                                                                                                                                                                           |
| Primary Contact            | Sadwini Suresh<br>QM Consultant<br>Regulatory Affairs<br>Phone: +49 89 99 15 68 0<br>Email: regulatory.affairs@brainlab.com                                                                                                                                                                                               |
| Alternate Contact          | Chiara Cunico<br>Senior Manager Regulatory Affairs<br>Phone: +49 89 99 15 68 0<br>Email: chiara.cunico@brainlab.com                                                                                                                                                                                                       |

#### INDICATIONS FOR USE 1

Spine Planning is intended for pre- and intraoperative planning of open and minimally invasive spinal procedures. It displays digital patient images (CT, Cone Beam CT, MR, X-ray) and allows measurement and planning of spinal implants like screws and rods.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink symbol on the left, resembling a brain or a network, followed by the word "BRAINLAB" in pink, block letters. The overall design is clean and modern, suggesting a focus on technology and innovation in the field of neuroscience or medical technology.

#### DEVICE DESCRIPTION 2

The Spine Planning software allows the user to plan spinal surgery pre-operatively or intra-operatively. The software is able to display 2D X-Ray images and 3D datasets (e.g. CT or MR scans). The software consists of features for automated labelling of vertebrae and proposals for screw and rod implants, proposals for measurement of spinal parameters.

The device can be used in combination with spinal navigation software during surgery, where preplanned or intra-operatively created information can be displayed, or solely as a pre-operative tool to prepare the surgery.

AI/ML algorithms are used in Spine Planning for

- . Detection of landmarks on 2D images for vertebrae labeling and measurement and
- . Vertebra detection on Digitally Reconstructed Radiograph (DRR) images of 3D datasets for atlas reqistration (labeling of the vertebra).

The AI/ML algorithm is a Convolutional Network (CNN) developed using a Supervised Learning approach. The algorithm was developed using a controlled internal process that defines from the inspection of input data to the training and verification of the algorithm.

#### SUBSTANTIAL EQUIVALENCE ന

The Subject Device has similar intended use and technological features as the predicate devices. An overview of the similarities and differences can be found in the tables below:

| Device                     | Name               |
|----------------------------|--------------------|
| Subject Device             | Spine Planning 2.0 |
| Predicate device (K212245) | Spine Planning 1.0 |

| Topic/ Feature                  | Predicate Device (Spine<br>Planning 1.0 (K212245))                                                                                                                                                                           | Subject Device (Spine<br>Planning 2.0)                                                                                                                                                                                                                 | Comment                                                                                                                                                                                                                                                       |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use             | Spine Planning is intended for pre- and intraoperative planning of open and minimal invasive spinal procedures. It displays digital bio imaging and allows measurement and planning of spinal implants like screws and rods. | Spine Planning is intended for pre- and intraoperative planning of open and minimally invasive spinal procedures. It displays digital patient images (CT, Cone Beam CT, MR, X-ray) and allows measurement and                                          | No change in the indications. Image modalities that are supported have been included.                                                                                                                                                                         |
|                                 |                                                                                                                                                                                                                              | planning of spinal implants<br>like screws and rods.                                                                                                                                                                                                   |                                                                                                                                                                                                                                                               |
| Intended use<br>environment     | The planning part of the use case<br>shall be done in an office or in the<br>operating room.                                                                                                                                 | The device can be used in an<br>office / team room or in the<br>operating room.                                                                                                                                                                        | Identical                                                                                                                                                                                                                                                     |
| Supported Imaging<br>Modalities | - Computed tomography (CT)                                                                                                                                                                                                   | - CT<br>- Magnetic resonance (MR)<br>- Cone Beam CT (XT)<br>- X-ray                                                                                                                                                                                    | Additional modalities<br>effect measurement<br>or screw planning<br>features                                                                                                                                                                                  |
| Vertebra Labeling               | Automatic detection of thoracic<br>and lumbar vertebrae. User<br>approval of calculated automatic<br>labeling. No functionality to<br>change the calculated labeling.                                                        | Automatic detection of<br>cervical, thoracic, lumbar,<br>and sacral vertebrae.<br>Ability to adjust automatic<br>labeling for correction if<br>needed. Automatic labeling<br>can be discarded and<br>performed manually. User<br>approval of labeling. | New features:<br>- Addition of cervical<br>and sacral spinal<br>regions in subject<br>device.<br>- Ability to adjust<br>and correct labeling.<br>- Possible to do<br>performed manual<br>labeling.<br>Functionality tested<br>during verification<br>testing. |
| Anatomical<br>Variations        | N/A                                                                                                                                                                                                                          | Adjustment of anatomical<br>chain to account for<br>anatomic variations.<br>Depending on patient<br>anatomy, it is possible to<br>exclude T12 and/or L5 and<br>include T13 and/or L6.<br>To be added manually by<br>user.                              | New feature.                                                                                                                                                                                                                                                  |
| Spinal Parameter<br>Proposals   | N/A                                                                                                                                                                                                                          | Able to add spinal<br>parameters on X-rays based<br>on AI/ML generated<br>landmarks. To be manually<br>reviewed and approved.                                                                                                                          | New feature.                                                                                                                                                                                                                                                  |
|                                 |                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                               |
| Manual<br>Measurements          | N/A                                                                                                                                                                                                                          | Where AI/ML landmarks are<br>unavailable all spinal<br>parameters can be added<br>manually, reviewed, and<br>approved.                                                                                                                                 | New feature.                                                                                                                                                                                                                                                  |
| Automatic Screw<br>Proposals    | Automatic screw proposals<br>provided for thoracic and lumbar<br>pedicle screws.                                                                                                                                             | Manual placement and<br>adjustment of angles and<br>open angles on X-rays.                                                                                                                                                                             | Similar function to<br>predicate. New<br>screw proposal<br>algorithm with<br>additional<br>functionality used in<br>subject device.                                                                                                                           |
| Screw Planning                  | Automatic screw proposals<br>provided for cervical (lateral<br>mass or pedicle), thoracic<br>(pedicle), lumbar (pedicle),<br>sacral (pedicle and alar iliac),<br>and iliac regions.                                          | Adjustment of screw length,<br>diameter, and position.                                                                                                                                                                                                 | Adjustment of<br>screws, creation of<br>manual screws and<br>screw approval is<br>similar to predicate.                                                                                                                                                       |
|                                 | Adjustment of screw length,<br>diameter, and position.                                                                                                                                                                       | Creation of manual screws.                                                                                                                                                                                                                             | New features.<br>Possible to change<br>type of screw<br>between lateral<br>mass and pedicle<br>screw types<br>(cervical region) or<br>pedicle and alar iliac<br>screw types (Sacral<br>region)                                                                |
|                                 | Creation of manual screws.                                                                                                                                                                                                   | Addition of automatic screw<br>proposals.                                                                                                                                                                                                              |                                                                                                                                                                                                                                                               |
|                                 | Addition of automatic screw<br>proposals.                                                                                                                                                                                    | Change of screw type<br>(cervical and sacral regions<br>only)                                                                                                                                                                                          |                                                                                                                                                                                                                                                               |
|                                 | User approval of all screws<br>required.                                                                                                                                                                                     | User approval of all screws<br>required.                                                                                                                                                                                                               |                                                                                                                                                                                                                                                               |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "BRAINLAB" in all capital letters. To the left of the word is a pink abstract symbol. The word and symbol are in a bright pink color. 

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Brainlab. The logo consists of a stylized pink caduceus symbol on the left, followed by the word "BRAINLAB" in pink, block letters. The caduceus symbol is a medical emblem featuring a staff with two snakes coiled around it.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink symbol on the left, resembling a medical symbol or abstract figure. To the right of the symbol, the word "BRAINLAB" is written in capital letters, also in pink. The overall design is clean and modern.

#### 4 PERFORMANCE DATA

## Software Verification:

Software verification was performed, verifying the software requirements through integration tests, and unit tests. Incremental test strateqies have been set up after verification of the first release candidate for changes with limited scope. In this case, an impact analysis of the modifications is performed and tests to be performed are identified and planned correspondingly. That means, not all tests have to be performed but only a subset, as some of the previous tests are not affected by the change and remain therefore valid.

Software verification verifies all specifications, including SOUP items and cybersecurity.

For newly added components, integration tests were carried out, in addition to the individual component verification.

## Al/ML Detected X-Ray Landmarks Assessment:

This test was conducted to assess the AI/ML detected landmarks on X-rays which are used by the software for automatic measurement proposals. This was done by quantifying the object detection, quality of vertebra level assignment, the quality of landmark predictions, and the performance of the observer view direction for 2D X-rays from the Universal Atlas Transfer Performer 6.0.

### Screw Proposal Alqorithm Evaluation:

The purpose of this testing was to compare the newly proposed screw proposal algorithm to the predicate and back-up algorithms. Thoracic and lumbar pedicle screw proposals generated by the new algorithm were found to be similar to thoracic and lumbar pedicle screw proposals generated by the predicate algorithm.

### Usability Evaluation:

Summative usability testing was planned and performed in order to validate that the Spine Planning 2.0 can be used by the intended user group. The summative usability testing identified no critical use related problems.

#### 5 CONCLUSION

The performed verification and validation activities established that the set requirements were met and that the device performs as intended.

The Subject Devices' comparison with the Predicate Device establishes that they have similar functionality, intended use and technological characteristics. Therefore, we consider that the Subject Device can be considered substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K223553](https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K223553)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
