← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K223350

# Remi Robotic Navigation System (K223350)

_Fusion Robotics, LLC · OLO · Mar 13, 2023 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K223350

## Device Facts

- **Applicant:** Fusion Robotics, LLC
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** Mar 13, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The Remi Robotic Navigation System is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of a tool holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous spinal procedures in reference to rigid patient anatomy and fiducials that can be identified on a 3D imaging scan or 2D fluoroscopic images. The Remi Robotic Navigation System is indicated for assisting the surgeon in placing pedicle screws in the posterior lumbar region (L1-S1). The system is designed for lumbar pedicle screw placement with the prone position and is compatible with the Accelus LineSider Spinal System.

## Device Story

Remi Robotic Navigation System is an image-guided surgical system for lumbar pedicle screw placement. Inputs include 3D intraoperative images or 2D fluoroscopic images (e.g., GE OEC 9900 Elite). System components: computer workstation, software, trajectory system with targeting platform, optical camera, and tracking instruments. Workflow: camera tracks surgical site and patient anatomy (via iliac pin); workstation reconstructs images and registers patient anatomy; surgeon plans trajectories; targeting platform automatically adjusts to planned trajectory; surgeon guides instruments through tool holder. Used in OR by surgeons. Output: real-time navigation and trajectory guidance for pedicle screw placement. Benefits: precise anatomical localization and instrument positioning. Software includes distortion correction for 2D fluoroscopic images.

## Clinical Evidence

Bench testing only. Testing included navigation accuracy verification, system accuracy validation, ASTM F2554 accuracy testing, and software unit/integration tests. Results demonstrate that 2D fluoroscopic image integration is equivalent to 3D imaging performance.

## Technological Characteristics

Optical-based navigation system; infrared camera; computer workstation; targeting platform with force-controlled movement; compatible with 3D and 2D fluoroscopic imaging; software-based image registration and distortion correction; manual gross positioning with automatic fine-tuning trajectory adjustment.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Remi Robotic Navigation System ([K223070](/device/K223070.md))
- EXCELSIUS GPS ([K171651](/device/K171651.md))

## Reference Devices

- GE OEC 9900 Elite 9" Image Intensifier ([K122234](/device/K122234.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date March 13, 2023. The text is written in a clear, sans-serif font. The date is presented in a standard format, with the month, day, and year clearly indicated. The text is horizontally aligned and evenly spaced.

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Fusion Robotics, LLC % Sarah Braun Senior Regulatory Affairs Specialist Integrity Implants Inc. dba Accelus 354 Hiatt Drive Palm Beach Gardens, Florida 33418

Re: K223350

Trade/Device Name: Remi Robotic Navigation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: November 22, 2022 Received: November 23, 2022

Dear Sarah Braun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jesse Muir -S

For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K223350

Device Name Remi Robotic Navigation System

#### Indications for Use (Describe)

The Remi Robotic Navigation System is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of a tool holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous spinal procedures in reference to rigid patient anatomy and fiducials that can be identified on a 3D imaging scan or 2D fluoroscopic images. The Remi Robotic Navigation System is indicated for assisting the surgeon in placing pedicle screws in the posterior lumbar region (L1-S1). The system is designed for lumbar pedicle screw placement with the prone position and is compatible with the Accelus LineSider Spinal System.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary

| 510(k) Owner             | Fusion Robotics, LLC<br>168 Centennial Parkway, Unit 170<br>Louisville, CO 80027 USA                     |
|--------------------------|----------------------------------------------------------------------------------------------------------|
| Contact Person           | Sarah Braun<br>Senior Regulatory Affairs Specialist<br>Tel: 423-838-4454<br>Email: sbraun@accelusinc.com |
| Date Prepared            | 11/22/2022                                                                                               |
| Classification Reference | 21 CFR 882.4560                                                                                          |
| Product Code             | OLO                                                                                                      |
| Common/Usual Name        | Stereotaxic Instrument                                                                                   |
| Trade/Proprietary Name   | Remi Robotic Navigation System                                                                           |
| Predicate Device(s)      | Remi Robotic Navigation System (K223070)<br>EXCELSIUS GPS (K171651)                                      |

The Remi Robotic Navigation System (Remi) is an image guided system primarily comprised of a computer workstation, software, a trajectory system, including a targeting platform, a camera, and various image guided instruments intended for assisting the surgeon in placing screws in the pedicles of the lumbar spine. The system operates in a similar manner to other optical-based image y systems:

- 1. The patient is placed in the appropriate position on the OR table.
- 2. The compact tracking Camera is rigidly affixed to the OR table using a multi-functional mechanical support arm in the appropriate position to track the surgical site.
- 3. The Camera is also affixed to a pin placed in the patient's iliac to provide a fixed location relative to the patient's spinal anatomy.
- 4. The Targeting Platform is affixed to the OR Table using a multi-functional mechanical support arm, ensuring that the Targeting Platform has sufficient range of motion to be placed over the surgical site.

{4}------------------------------------------------

- 5. If a 3D imaging system is being used, the Registration Array is affixed to the Targeting Platform and positioned over the planned surgical site.
- 6. If 2D fluoroscopic system is being used, the C-Arm Grid Plate, 9900 9" is installed onto the fluoroscopic imaging system.
- 7. The appropriate area of spine (L1-S1) is Imaged with a validated imaging system.
- 8. The images are transferred to the Remi system workstation, which reconstructs the images and uses the registration array images) or the C-Arm Grid Plate Tracker image (for 2D fluoroscopic Images) to register the patient's spine relative to the Camera location.
- 9. The registration is confirmed by placing an image guided instrument with an Instrument Tracker at various points in the surgical field.
- 10. The surgical paths are then planned. If the procedure is using 3D images, this is done on the workstation. If the procedure is using 2D fluoroscopic images, the Planning Probe and workstation are utilized to capture the desired position for target pedicle screw placement.
- 11. The Targeting Platform is gross positioned manually close to the first surgical plan location.
- 12. The Targeting Platform is activated to set the fine location and the trajectory based on the surgical plan.
- 13. Instruments with tracking arrays can now be used through the tool guide of the Targeting Platform to prepare the pedicle and place a pedicle screw.

## Intended Use/Indications for Use

The Remi Robotic Navigation System is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of a tool holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous spinal procedures in reference to rigid patient anatomy and fiducials that can be identified on a 3D imaging scan or 2D fluoroscopic images. The Remi Robotic Navigation System is indicated for assisting the surgeon in placing pedicle screws in the posterior lumbar region (L1-S1). The system is designed for lumbar pedicle screw placement with the prone position and is compatible with the Accelus LineSider Spinal System.

### Substantial Equivalence

The proposed Remi Robotic Navigation System (Remi) allows the use of a 2D fluoroscopic Imaging system. The system validated for this submission is the GE OEC 9900 Elite 9" Image Intensifier (K122234). The indications for use has been updated to add 2D fluoroscopic images for use with the Remi system. Four instruments were added: Long Reference Arm (PN1075), Planning Probe (PN1115), C-

{5}------------------------------------------------

Arm Grid Plate, 9900 9" (PN1138), C-Arm Grid Plate Tracker (PN1139). The software was updated to support the use of 2D fluoroscopic images. Changes include compatibility with the new instruments and an algorithm which corrects the distortion of the 2D fluoroscopic image.

### Performance Testing - Bench

The following tests were performed to support the substantial equivalence of the subject Remi Robotic Navigation System (Remi) to its predicates:

- Navigation Accuracy Verification
- . System Accuracy Validation
- Software System Test
- ASTM F2554 Accuracy Test
- Software Unit and Integration Tests

Testing was done to demonstrate that the updated requirement for this change was met and to ensure the risk profile of Remi was maintained. The testing shows that the use of the 2D fluoroscopic images with the Remi system is equivalent to the use of the validated 3D imaging systems.

| Devices                                                | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Primary Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Secondary Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                        | Remi Robotic Navigation<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Remi Robotic Navigation<br>System [K223070]                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | EXCELSIUS GPS<br>[K171651]                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Indications for<br>Use                                 | The Remi Robotic Navigation<br>System is intended for use as<br>an aid for precisely locating<br>anatomical structures and for<br>the spatial positioning and<br>orientation of a tool holder or<br>guide tube to be used by<br>surgeons for navigating<br>and/or guiding compatible<br>surgical instruments in open<br>or percutaneous spinal<br>procedures in reference to<br>rigid patient anatomy and<br>fiducials that can be identified<br>on a 3D imaging scan or 2D<br>fluoroscopic images. The Remi<br>Robotic Navigation System is<br>indicated for assisting the<br>surgeon in placing pedicle<br>screws in vertebrae in the<br>posterior lumbar region (L1-<br>S1). The system is designed<br>for lumbar pedicle screw<br>placement with the patient in | The Remi Robotic Navigation<br>System is intended for use as an<br>aid for precisely locating<br>anatomical structures and for<br>the spatial positioning and<br>orientation of a tool holder or<br>Guide Tube to be used by<br>surgeons for navigating and/or<br>guiding compatible surgical<br>instruments in open or<br>percutaneous spinal procedures<br>in reference to rigid patient<br>anatomy and fiducials that can<br>be identified on a 3D Imaging<br>scan. The Remi Robotic<br>Navigation System is indicated<br>for assisting the surgeon in<br>placing pedicle screws in<br>vertebrae in the posterior<br>lumbar region (L1-S1). The<br>system is designed for lumbar<br>pedicle screw placement with<br>the patient in the prone position | The EXCELSIUS GPS™ is intended<br>for use as an aid for precisely<br>locating<br>anatomical structures and for<br>the spatial positioning and<br>orientation of an<br>instrument holder or guide tube<br>to be used by surgeons for<br>navigating and/or<br>guiding compatible surgical<br>instruments in open or<br>percutaneous procedures<br>provided that the required<br>fiducial markers and rigid<br>patient anatomy can be<br>identified on CT scans or<br>fluoroscopy. The system is<br>indicated for the placement<br>of spinal and orthopedic bone<br>screws. |
|                                                        | the prone position and is compatible with the Accelus<br>LineSider Spinal System.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | and is compatible with the Accelus LineSider Spinal System.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | and is compatible with the Accelus LineSider Spinal System.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Product Code                                           | OLO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | OLO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | OLO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Principles of<br>Operation                             | Same as Predicates.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Intraoperative/preoperative images Patient registration Surgical planning Real-time tracking of navigated instruments Guidance of instruments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Intraoperative/preoperative images Patient registration Surgical planning Real-time tracking of navigated instruments Guidance of instruments                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Input Images                                           | 3D Intraoperative images  Medtronic O-arm GE OEC 3D Ziehm-Vision RFD 3D Stryker Airo TruCT 2D Fluoroscopic Images GE OEC 9900 Elite 9" Image Intensifier (K122234)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 3D Intraoperative images  Medtronic O-arm GE OEC 3D Ziehm-Vision RFD 3D Stryker Airo TruCT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 3D pre-operative images<br>3D intra-operative images<br>2D intra-operative images                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Trajectory<br>planning<br>parameters                   | Same as Primary Predicate.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Entry point, target point, length of the instrument, diameter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Localization<br>method                                 | Same as Primary Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Optical System (infrared Camera)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Camera system                                          | Same as Primary Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Monocular                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Controller                                             | Same as Primary Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Manual macro adjustments<br>Force-controlled movement of Targeting platform                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Patient<br>Registration<br>Method                      | Same as Primary Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Registration fixture in place during 3D<br>intraoperative images                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Intra-op CT: Registration fixture<br>Pre-op CT: Fluoroscopic to pre-op CT merge<br>Fluoroscopy: Registration fixture                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Accuracy<br>verification on<br>anatomical<br>landmarks | Same as Predicates.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Real time<br>display of<br>instrument<br>position      | Same as Predicates.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Instrument<br>Guidance                                 | Same as Primary Predicate.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Trajectory and location set by<br>Targeting platform. Instruments<br>are manually positioned by the<br>surgeon through the guide tube<br>on the Targeting Platform.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes, instruments are used<br>through the guide tube on<br>the robotic arm or are<br>manually positioned by the<br>surgeon.                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Patient fixation                                       | Same as Primary Predicate.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Tracking Camera is fixed to OR<br>table and the patient's iliac<br>crest.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Reference is fixed to<br>patient's bony structure<br>such as a long bone, iliac<br>crest, spinous process,<br>vertebra, etc. for tracking<br>system                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Positioning<br>accuracy<br>(bench                      | Same as Primary Predicate.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | $0.74 \pm 0.36$ mm (worst case)<br>95% CI: 1.46mm (worst case)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Robot collision<br>avoidance/det<br>ection             | Same as Primary Predicate.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Manual movement of Trajectory<br>Platform to gross location. Small<br>fine tuning of Trajectory<br>Platform location is automatic<br>but is currently limited to cease<br>when platform encounters a<br>force greater than 9lbs.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |

#### Substantial equivalence analysis for Remi

{6}------------------------------------------------

{7}------------------------------------------------

### Conclusions

The subject device, Remi Robotic Navigation System, described in this submission shares a majority of the same technological characteristics as the primary predicate device, Remi Robotic Navigation System (K223070). The primary difference between the subject device and the primary predicate is the addition of validated 2D fluoroscopic imaging systems and the subsequent modification of the wording of the indications for use. Like the subject device, the secondary predicate, EXCELSIUS GPS (K171651) has the capability to rely on 2D intra-operative exam as the Input Images for the stereotaxic navigation of pedicle screw placement.

The verification and validation testing demonstrated that the characteristics of the subject Remi device are substantially equivalent to the predicate device. The subject device continues

{8}------------------------------------------------

to meet design requirements, is as safe and effective as the predicate device, and performs according to its intended use. The information presented in this 510(k) premarket notification demonstrates that the subject device is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K223350](https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K223350)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
