← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K222966

# Cirq Arm System (2.0);Alignment System Spine (K222966)

_Brainlab AG · OLO · Oct 27, 2022 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K222966

## Device Facts

- **Applicant:** Brainlab AG
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** Oct 27, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

As an accessory to the Spine & Trauma Navigation, the Alignment System Spine is intended to surgeon to achieve a pre-defined screw with surgical instruments during the surgical procedure. It is used for spinal screw placement procedures.

## Device Story

Cirq Arm System 2.0 is a passive, semi-rigid mechatronic arm for holding/positioning surgical instruments during spinal surgery. Device attaches to operating table side rails; used in hospital operating rooms. Surgeon manually positions arm by releasing joint brakes via integrated buttons; brakes are fail-safe (close upon power loss). System provides stable support for surgical instruments to reach pre-defined trajectories. Device includes structural components, modules, and software to guide users. Compared to predecessor (K202320), system features increased payload (3.0 kg vs 1.9 kg), increased length (1272 mm vs 1200 mm), and improved stiffness via harmonic drive gears in joints 1-4. Sterility maintained via compatible drapes. Output is physical positioning/holding of instruments, aiding surgeon in accurate screw placement.

## Clinical Evidence

Bench testing only. No clinical data presented. Verification included software integration/unit testing, mechanical subsystem review (CAD, calculations, bench testing for holding force, stiffness, safety factors, lifecycle), electrical subsystem testing, and biological safety evaluation per ISO 10993-1:2018.

## Technological Characteristics

Passive mechatronic arm; 7-axis design. Axes 1-4 utilize harmonic drive gears and spring-applied brakes; axes 5-7 utilize permanent magnetic brakes. Connectivity: IGS platforms and Alignment Software Spine. Power: External base supply. Dimensions: 1272 mm length. Payload: 3.0 kg. Sterilization: Sterile drape. Software: Alignment Software Spine 2.0. Biocompatibility: ISO 10993-1:2018.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Cirq Robotic Alignment Module - Spine ([K202320](/device/K202320.md))

## Reference Devices

- Alignment Software Spine 2.0 ([K221618](/device/K221618.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

October 27, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Brainlab AG Sadwini Suresh QM Consultant Regulatory Affairs Olof-Palme-Str. 9 Munich, Bavaria 81829 Germany

Re: K222966

Trade/Device Name: Cirq Arm System (2.0); Alignment System Spine Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: September 27, 2022 Received: September 27, 2022

Dear Sadwini Suresh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-

mdr-how-report-medical-device-problems.

542 of the Act); 21 CFR 1000-1050.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K222966

Device Name

CIRQ Arm System (2.0); Alignment System Spine

Indications for Use (Describe)

As an accessory to the Spine & Trauma Navigation, the Alignment System Spine is intended to surgeon to achieve a pre-defined screw with surgical instruments during the surgical procedure. It is used for spinal screw placement procedures.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized brain icon on the left and the word "BRAINLAB" in capital letters on the right. The color of the logo is pink.

# 510(k) Summary

September 27, 2022

| General Information        |                                                                                                                             |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Manufacturer               | Brainlab AG; Olof-Palme Str.9; 81829, Munich, Germany                                                                       |
| Establishment Registration | 8043933                                                                                                                     |
| Trade Names                | Cirq Arm System (2.0) Alignment System Spine                                                                                |
| Classification Name        | Orthopedic Stereotaxic Instrument                                                                                           |
| Product Code               | OLO                                                                                                                         |
| Regulation Number          | 882.4560                                                                                                                    |
| Regulatory Class           | II                                                                                                                          |
| Panel                      | Orthopedic                                                                                                                  |
| Predicate Device           | Cirq Robotic Alignment Module - Spine (K202320)                                                                             |
| Contact Information        |                                                                                                                             |
| Primary Contact            | Sadwini Suresh<br>QM Consultant<br>Regulatory Affairs<br>Phone: +49 89 99 15 68 0<br>Email: regulatory.affairs@brainlab.com |
| Alternate Contact          | Chiara Cunico<br>Senior Manager Regulatory Affairs<br>Phone: +49 89 99 15 68 0<br>Email: chiara.cunico@brainlab.com         |

#### 1. Indication for Use

As an accessory to the Spine & Trauma Navigation, the Alignment System Spine is intended to support the surgeon to achieve a pre-defined screw with surgical instruments during the surgical procedure. It is used for spinal screw placement procedures.

## 2. Device Description

The Cirq Arm System serves as a holding and positioning system to support the surgeon in reaching a pre-defined screw trajectory with surgical instruments. It consists of a passive, semi-rigid mechatronic arm system with structural components, several modules, instrument holding and clamping mechanisms and software to quide the user.

The Cirq Arm System 2.0 is a passive mechatronic device for holding and positioning surgical instruments; it is evolved from its predecessor, the Surgical Base System 1.4 (K202320).

The Cirq Arm System 2.0 is attached to the side rail of an operating table and is intended to be used in a hospital environment, specifically in rooms which are appropriate for surgical interventions (e.g. operating rooms). The base provides external power and communication with the IGS platforms and Alignment Software Spine (if applicable, ie. if used within the active confiquration).

The Cirq Arm System 2.0 can be manually positioned by releasing the brakes in the joints, using the integrated buttons. The brakes in the joints open when powered (a voltage is

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized brain icon on the left and the word "BRAINLAB" in capital letters on the right. The color of the logo is pink.

supplied), and close without current. Therefore, the brakes close in case of a power failure. The brake state is visualized to the user by LED rings showing different colors for the two different brake states (open/closed) separately for each joint. The electronics and firmware are designed to be ready to operate meaning that the arm can be booted and used quickly after connecting it to the power supply.

The sterility of the device during the surgical intervention is ensured with a sterile drape, which is compatible to the Cirq Arm System 2.0 and the attached application module. The specific workflow and operating principles differ based on the attached module.

## 3. Substantial Equivalence

| Topic/<br>Feature     | Predicate Device (K202320)                                                                                                                                                                                                                                                                                | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Comments                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Component<br>overview | Hardware:<br>• Surgical Base System 1.4<br>(Art. No. 56000A)<br>• Cirq Robotic Alignment<br>Module (Art. No. 56100)<br>• Cirq Robotic Disposable<br>Kinematic Unit (Art. No.<br>56102)<br>• Cirq Robotic and<br>Alignment Instruments<br>(various)<br>Software:<br>Cirq Robotic Alignment<br>Software 1.0 | Hardware:<br>• Cirq Arm System 2.0 (Art.<br>No. 56500)<br>• Cirq Robotic Alignment<br>Module (Art. No. 56100)<br>• Cirq Robotic Disposable<br>Kinematic Unit (Art. No.<br>56102)<br>• Cirq Robotic and<br>Alignment Instruments<br>(various)<br>• Cirq Instrument Holder<br>Module (Art. No. 56200A)<br>• Cirq Instrument Holder<br>Spinal Drilling (Art. No.<br>56202)<br>• Accessory Package Spine<br>Drilling for Cirq Passive<br>(various)<br>Software:<br>• Alignment Software Spine<br>2.0 | The main changes relative to<br>the predicate is iteration of the<br>alignment software (Cirq<br>Robotic Alignment Software 1.0<br>to Alignment Software Spine<br>2.0, cleared in K221618), the<br>iteration of the arm hardware<br>(Surgical Base System 1.4 to<br>Cirq Arm System 2.0), and the<br>inclusion of additional<br>hardware. The operating<br>principle is maintained relative<br>to the predicate, including with<br>the additional hardware. |
| Arm weight            | Total weight of the arm is<br>11.5 kg.                                                                                                                                                                                                                                                                    | Total weight of the arm is 16<br>kg.                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Increased weight compared to<br>predicate device.                                                                                                                                                                                                                                                                                                                                                                                                           |
| Arm<br>dimensions     | The overall length of the arm<br>in its extended position is<br>approximately 1200 mm.                                                                                                                                                                                                                    | The overall length of the arm<br>in its extended position is<br>approximately 1272 mm.                                                                                                                                                                                                                                                                                                                                                                                                           | The reach of the arm has been<br>increased.                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Maximum<br>payload    | The defined maximum<br>payload of the device is 1.9<br>kg. The holding force is<br>supported by a technical<br>function called Brake Boost.                                                                                                                                                               | The defined maximum<br>payload of the device is 3.0<br>kg. This is based on the gear<br>driven joints in joints 1-4,<br>replacing the Brake Boost<br>function.                                                                                                                                                                                                                                                                                                                                   | Increasing the maximum<br>payload presents no substantial<br>differences regarding basic<br>functionality in comparison to<br>predicate device.                                                                                                                                                                                                                                                                                                             |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink brain icon on the left and the word "BRAINLAB" in pink capital letters on the right. The brain icon is abstract and appears to be made up of interconnected shapes. 

| Topic/<br>Feature         | Predicate Device (K202320)                                                                                                                                                                                                                                         | Subject Device                                                                                                                                                                                                                                                                            | Comments                  |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Brake and<br>joint design | The device features<br>permanent magnetic brakes<br>in each of the 7 axes, which<br>ensure the holding force of<br>the positioning arm in the<br>closed state, and enable<br>spatial positioning during<br>surgical procedure. A certain<br>stiffness is provided. | The device features<br>permanent magnetic brakes<br>in axes 5-7 and spring-<br>applied brakes in axes 1-4.<br>Additionally, axes 1-4 consist<br>of harmonic drive gears. Due<br>to the transmission of the<br>gears, the system is stiffer<br>than the previous version of<br>the device. | Improvement in stability. |

## 4. Performance Data

#### Software Verification:

Software verification was performed, verifying the software requirements through integration tests, and unit tests. Incremental test strategies have been set up after verification of the first release candidate for changes with limited scope. In this case, an impact analysis of the modifications is performed and tests to be performed are identified and planned correspondingly. That means, not all tests have to be performed but only a subset, as some of the previous tests are not affected by the change and remain therefore valid.

Software verification verifies all specifications, including SOUP items and cybersecurity.

#### Hardware Verification:

Hardware verification was performed, verifying the mechanical and electronic requirements, compliance to standards, and verifying the biological safety of materials and surfaces which may come in contact with the user.

Mechanical verification verifies specifications related to the mechanical subsystem, through review of CAD parts and assemblies, calculations, document review, bench testing related to holding force/payload, stiffness, mechanical safety factor, and lifecycle testing.

Electrical verification verifications related to the electrical subsystem, review of documentation and schematics, and physical testing.

Biological safety was verified according to ISO 10993-1:2018. Additional verifications have been conducted for risk-relevant specifications, user manual entries, and production steps.

#### 5. Conclusion

The comparison of the Subject Device with the predicate device shows that the Cirq Arm System 2.0 has similar functionality, intended use and technological characteristics as the predicate device. Based on the comparison to the predicate and the performance testing conducted, the Subject Device is considered substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K222966](https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K222966)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
