The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons during open surgical procedures with appropriate bone preparation. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure that can be identified on CT derived patient images for pedicle screw cannulation of the thoracic to sacrum vertebrae.
Device Story
Paradigm System is a stereotaxic image guidance system for spinal surgery. Inputs: preoperative CT scans; real-time camera imaging of surgical site via Prism Sensor Array. Operation: system segments CT data to generate 3D models of vertebrae; intraoperative registration matches live anatomy topography to 3D model; creates common coordinate space; overlays virtual information onto live anatomy for guidance. Output: visual display on external monitor showing instrument position/orientation relative to patient anatomy. Used in OR by surgeons. Benefits: precise instrument placement during pedicle screw cannulation; improved spatial positioning.
Clinical Evidence
Bench testing only. Evidence includes non-clinical hardware, software, and instrumentation verification tests, and non-clinical design validation conducted in a cadaveric model (cadaveric simulated workflow study). Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60825-1, IEC 62366-1, ISO 10993-1, ISO 14971, ISO 17665-1, AAMI/ANSI ST79, and ASTM F2554-18.
Technological Characteristics
Stereotaxic image-guided navigation system. Components: cart, arm, Prism Sensor Array, external monitor, surgical instruments. Energy: 120V. Sterilization: Steam (SAL 1x10^-6). Connectivity: Preoperative CT data upload to console. Software: 3D model generation via segmentation of CT data; intraoperative registration of live anatomy. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60825-1, IEC 62366-1, ISO 10993-1, ISO 14971, ISO 17665-1, AAMI/ANSI ST79, ASTM F2554-18.
Indications for Use
Indicated for patients undergoing posterior approach spine surgery requiring pedicle screw cannulation of thoracic to sacrum vertebrae, where rigid anatomical structures are identifiable on CT images.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is the Department of Health & Human Services logo. The FDA acronym is in a blue square, and the full name of the agency, "U.S. Food & Drug Administration," is in blue text to the right of the square.
April 21, 2023
Proprio, Inc. Shannon Eubanks Vice President, HW Engineering and Regulatory 111 W John Street Suite 308 Seattle, Washington 98119
Re: K222291
Trade/Device Name: Paradigm System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: March 20, 2023 Received: March 22, 2023
Dear Shannon Eubanks:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali-S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K22291
Device Name Paradigm System
#### Indications for Use (Describe)
The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons during open surgical procedures with appropriate bone preparation. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure that can be identified on CT derived patient images for pedicle screw cannulation of the thoracic to sacrum vertebrae.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt;"></span> | <span style="font-size:10pt;"><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:10pt;"></span> | <span style="font-size:10pt;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K222291
# 2.0 510(k) Summary/Statement
## Table 1: 510(k) Summary
| Submitter: | Proprio, Inc |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Shannon Eubanks<br>VP of HW Engineering and Regulatory<br>Phone: (425) 802-6063<br>E-mail: seubanks@propriovision.com |
| Trade Name: | Paradigm System |
| Common Name: | Orthopedic stereotaxic instrument |
| Classification: | Class II |
| Product Code: | OLO |
| Regulation | 21 CFR 882.4650 |
| Predicate Device(s): | The subject device is equivalent to the following device:<br>Envision 3DTM: Image Guidance System (K162375) |
| Device Description: | The Paradigm System is a stereotaxic image guidance system intended<br>for the spatial positioning and orientation of spinal surgical instruments<br>used by surgeons. The Paradigm System fuses high-resolution, real-time<br>camera imaging of the surgical site (such as the spine) with medical<br>imagery (such as CT) and surgical navigation data, such as<br>predetermined instrument trajectories for precise instrument placement.<br>The Paradigm System encompasses non-sterile operating room<br>equipment components as well as sterilizable, reusable instruments, and<br>off-the-shelf sterile-packaged, single-use accessories.<br>The Paradigm System is used with an external monitor. |
| Indication for Use: | The Paradigm System is a stereotaxic image guidance system intended for<br>the spatial positioning and orientation of spinal surgical instruments used<br>by surgeons during open surgical procedures with appropriate bone<br>preparation. The device is indicated for posterior approach spine surgery<br>where reference to a rigid anatomical structure that can be identified on<br>CT derived patient images for pedicle screw cannulation of the thoracic to<br>sacrum vertebrae. |
{5}------------------------------------------------
| | Paradigm System<br>(Subject Device) | Envision 3DTM: Image<br>Guidance System<br>(Predicate Device) | Comment |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Device Overview | | | |
| 510(k) Number | To be determined | K162375 | |
| Decision Date | | December 29, 2016 | |
| Manufacturer | Proprio, Inc | 7D Surgical, Inc | |
| Classification | Class II | Class II | Same |
| Product Code | OLO | OLO | Same |
| Regulation | 21 CFR 882.4560 | 21 CFR 882.4560 | Same |
| Medical Specialty | Neurology | Neurology | Same |
| Indications for<br>Use | The Paradigm System is a stereotaxic<br>image guidance system intended for the<br>spatial positioning and orientation of<br>spinal surgical instruments used by<br>surgeons during open surgical procedures<br>with appropriate bone preparation. The<br>device is indicated for posterior approach<br>spine surgery where reference to a rigid<br>anatomical structure that can be<br>identified on CT derived patient images<br>for pedicle screw cannulation of the<br>thoracic to sacrum vertebrae. | The Envision 3DTM: Image<br>Guidance System is a stereotaxic<br>image guidance system intended<br>for the spatial positioning and<br>orientation of neurosurgical<br>instruments used by surgeons.<br>The system is also intended to be<br>used as the primary surgical<br>luminaire during image guided<br>surgery. The device is indicated<br>for posterior approach spine<br>surgery where reference to a<br>rigid anatomical structure can be<br>identified. | Subject device is<br>not the primary<br>luminaire. |
{6}------------------------------------------------
| | Paradigm System<br>(Subject Device) | Envision 3DTM: Image<br>Guidance System<br>(Predicate Device) | Comment |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Principle of<br>operation | Paradigm System fuses high-<br>resolution, real-time camera imaging<br>of the surgical site (such as the spine)<br>with medical imagery (such as CT)<br>and surgical navigation data, such as<br>predetermined instrument trajectories<br>for precise instrument placement.<br><br>The Paradigm System enables surgical<br>navigation of instruments relative to<br>the patient spinal anatomy by<br>combining preoperative imaging<br>segmentation and intraoperative<br>registration of the patient's live<br>anatomy. Preoperative CT data is<br>uploaded to the console and the system<br>software generates a 3D model of the<br>patient's operative vertebrae. The 3D<br>model is used for preoperative<br>planning such as implant placements<br>or instrument trajectories. During the<br>procedure, once the patient's anatomy<br>is exposed, the Prism Sensor Array<br>acquires imaging to match the<br>topography of the live anatomy to the<br>3D model. This intraoperative<br>registration of live anatomy to the 3D<br>model creates a common coordinate<br>space between the preoperative CT<br>data and the patient anatomy. This<br>enables virtual information to be<br>overlaid on the live anatomy during<br>the procedure for intraoperative<br>guidance. | The Envision 3D system<br>provides image registration<br>between preoperative scan data<br>and data captured<br>intraoperatively from the<br>Envision 3DTM integrated<br>structured light scanner<br>and/or user selected points. The<br>system provides guidance data<br>by displaying the<br>locations of wireless optically<br>tracked Envision 3DTM Spinal<br>Instruments (examples<br>include pedicle probe and awl)<br>relative to the patient. Position<br>and orientation data of<br>tracked Envision 3DTM Spinal<br>Instruments are linked to the<br>preoperative scan data using<br>the Envision 3DTM workstation. | Subject device<br>uses light field<br>where predicate<br>uses structured<br>light scanner.<br>Subject device<br>segments each<br>vertebra where the<br>predicate as if it<br>were a single<br>anatomical body. |
| Technical<br>Comparison | Stereotaxic image guided surgical<br>navigation system during spine | Stereotaxic image guided<br>surgical navigation system | Subject device has<br>option for |
{7}------------------------------------------------
| | Paradigm System<br>(Subject Device) | Envision 3D™: Image<br>Guidance System<br>(Predicate Device) | Comment |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| | surgery. Preoperative CT data is<br>segmented and a 3D model is<br>generated. Preoperative planning can<br>be performed. Intraoperative image<br>capture using reflective markers on<br>spinal instruments and registers the<br>patient's anatomy to the preoperative<br>3D model for intraoperative guidance<br>during the procedure. | during spine surgery.<br>Preoperative CT data is loaded<br>onto the cart. Intraoperative<br>registration of the patient's<br>anatomy with the preloaded CT<br>data using 3D structured light<br>images. Reflective markers on<br>spinal instruments aid the<br>surgeon in viewing the position<br>and orientation of instruments<br>relative to registered pre-<br>operative image data while<br>performing the surgical<br>procedure. | preoperative<br>planning. |
| Components | Cart, Arm, Sensor Array, External<br>Monitor, Surgical Instruments,<br>Software | Cart, Arm, Head, tracked<br>surgical Envision 3D™ Spinal<br>Instruments, Software | Same. |
| Energy Source | 120V | 120V | Same |
| Sterilization<br>Method -<br>Surgical<br>Instruments | Steam | Steam | Same |
| Minimum SAL | $1 x 10^{-6}$ | $1 x 10^{-6}$ | Same |
| Required<br>Accessories | Reflective Markers (Northern Digital<br>K033621) | Reflective Markers | Same |
{8}------------------------------------------------
| Functional and Safety<br>Testing: | Verification and Validation activities have been conducted to provide<br>assurance that the device meets the performance requirements under the<br>indications for use conditions. Proprio performed the following testing<br>to ensure the safety and effectiveness of the Paradigm System device:<br>Non-Clinical Hardware, Software, and Instrumentation Verification Tests Non-Clinical Design Validation conducted in Cadaveric Model Cadaveric Simulated Workflow Study Assessing Usability Compliance Conformity Assessments |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The following standards were used in testing:<br>IEC 60601-1:2015 Medical electrical equipment. General requirements for basic safety and essential performance IEC 60601-1-2:2014 Medical Electrical Equipment – Part 1-2, General Requirements for Basic Safety and Essential Performance - Collateral Standard Electromagnetic Compatibility IEC 60601-1-6:2010+AMD1:2013 Medical Electrical Equipment - Part 1-6, General Requirements for Basic Safety and Essential Performance - Usability IEC 60825-1:2014 Safety of laser products - Part 1: Equipment classification and requirements IEC 62366-1: 2015 Ed1Medical Devices – Part 1: Application of Usability Engineering To Medical Devices ISO 10993-1:2003 Biological evaluation of medical devices: Part 1: Evaluation and testing ISO 14971:2019 Medical devices – Application of risk management to medical devices ISO 17665-1:2006 Sterilization of health care products – Moist heat - Part:1 Requirements for the development, validation and routine control of a sterilization process for medical device AAMI / ANSI ST79:2017 Guide to Steam Sterilization ASTM F2554-18 Standard Practice For Measurement Of Positional Accuracy Of Computer Assisted Surgical Systems |
| Conclusion: | The Paradigm System intended use, indications for use, and<br>fundamental scientific technology is similar to the predicate device.<br>Performance, safety and usability testing demonstrate that the<br>differences between the subject device and the predicate device do not<br>raise new risks of safety and effectiveness. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.