K220652 · Brainlab AG · OLO · Jun 3, 2022 · Neurology
Device Facts
Record ID
K220652
Device Name
Knee 3
Applicant
Brainlab AG
Product Code
OLO · Neurology
Decision Date
Jun 3, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
Knee3 is intended to be an intraoperative localization system used during orthopedic knee surgery. It links a navigated instrument, tracked by a passive marker sensor system, to virtual computer image space on an individually acquired model of the patient's anatomical axes, which are generated through acquiring multiple anatomical bony landmarks. The system is indicated for medical conditions in which the use of stereotaxic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the femur and tibia, can be identified relative to the acquired anatomical landmarks. The system aids the surgeon to plan resections, measure cutting block alignment, measure bone resections, and/or leg alignment. The device is indicated for total and uni-compartmental knee replacement.
Device Story
Knee3 is an intraoperative software navigation module for orthopedic knee surgery. It uses infrared optical tracking of instruments (reference arrays, pointers, plane tools) to register femur and tibia anatomical landmarks. The system maps these tracked instruments to a virtual computer image space. Surgeons use the system to plan resections, measure cutting block alignment, and assess bone resections or leg alignment. Operated by surgeons in an OR setting, the system provides real-time navigation measurements on an LCD touch display. By providing precise intraoperative guidance, the device assists in the accurate placement of prosthetic knee implants, potentially improving surgical outcomes.
Clinical Evidence
No new clinical trials were conducted. Clinical validation relied on literature data from predecessor software versions (predicate devices) to establish equivalence of workflows (Motion, Universal, Partial, Express). Bench testing confirmed system accuracy within ± 2° and ± 2mm in 95% of cases. Biocompatibility of instruments was verified per ISO 10993 standards.
Technological Characteristics
Infrared passive/active marker-based optical tracking. Instruments made of surgical-grade materials. System includes software module for navigation platforms with LCD capacitive touch display. Connectivity via navigation station hardware. Biocompatibility per ISO 10993-1, -5, -10, -11, -12, -18. Software is a major level of concern.
Indications for Use
Indicated for patients undergoing total or uni-compartmental knee replacement where stereotaxic surgery is appropriate and rigid anatomical structures (femur/tibia) can be referenced via anatomical landmarks.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 3, 2022
Brainlab AG Chiara Cunico Manager Regulatory Affairs Olof-Palme-Str. 9 Munich, Bavaria 81829 Germany
Re: K220652
Trade/Device Name: Knee 3 Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: March 1, 2022 Received: March 7, 2022
Dear Chiara Cunico:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220652
Device Name Knee 3
Indications for Use (Describe)
Knee3 is intended to be an intraoperative localization system used during orthopedic knee surgery. It links a navigated instrument, tracked by a passive marker sensor system, to virtual computer image space on an individually acquired model of the patient's anatomical axes, which are generated through acquiring multiple anatomical bony landmarks. The system is indicated for medical conditions in which the use of stereotaxic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the femur and tibia, can be identified relative to the acquired anatomical landmarks. The system aids the surgeon to plan resections, measure cutting block alignment, measure bone resections, and/or leg alignment. The device is indicated for total and uni-compartmental knee replacement.
Type of Use (Select one or both, as applicable)
[X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Brainlab, a medical technology company. The logo consists of a stylized brain icon on the left, followed by the word "BRAINLAB" in bold, sans-serif font. The brain icon is composed of several curved lines, creating an abstract representation of the brain's structure.
# 510(k) Summary
May 25, 2022
| General Information | |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Brainlab AG; Olof-Palme Str.9, 81829, Munich, Germany |
| Establishment Registration | 8043933 |
| Device Name | Orthopedic Stereotaxic Instrument |
| Trade Name | Knee 3 |
| Classification Name | Stereotaxic instrument |
| Product Code | OLO |
| Regulation Number | 882.4560 |
| Regulatory Class | II |
| Panel | Orthopedic |
| Predicate Devices | Primary predicate device: Brainlab Knee (Knee 2.5) K102990<br>Secondary predicate device: uni-Knee (Uni-Knee 2.0) K080678<br>Tertiary predicate device: DASH Knee (DASH Knee 1.0) (K102251) |
| Contact Information | |
|-----------------------------------|----------------------------------------|
| Primary Contact | Alternate Contact |
| Chiara Cunico | Regulatory Affairs Brainlab |
| Manager RA | Phone: +49 89 99 15 68 0 |
| Phone: +49 89 99 15 68 0 | Fax: +49 89 99 15 68 5033 |
| Email: chiara.cunico@brainlab.com | Email: regulatory.affairs@brainlab.com |
# 1. Indications for Use
Knee3 is intended to be an intraoperative localization system used during orthopedic knee surgery. It links a navigated instrument, tracked by a passive marker sensor system, to virtual computer image space on an individually acquired model of the patient's anatomical axes, which are generated through acquiring multiple anatomical bony landmarks. The system is indicated for medical conditions in which the use of stereotaxic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the femur and tibia, can be identified relative to the acquired anatomical landmarks. The system aids the surgeon to plan resections, measure cutting block alignment, measure bone resections, and/or leg alignment.
The device is indicated for total and uni-compartmental knee replacement.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Brainlab logo. The logo consists of a stylized symbol on the left and the word "BRAINLAB" in bold, sans-serif font on the right. The symbol appears to be a stylized representation of the brain.
# 2. Device Description
The Knee3 is a software application module intended to be used on compatible navigation platforms that assist in the implantation of prosthetic knee implants. It uses instruments and accessories such as reference arrays, pointers and plane tools, which are tracked by an infrared camera of the respective platform, to determine femur and tibia anatomical landmarks as well as instrument positions in relation to the registered bones. The resulting navigation measurements support the user with the placement of the cutting guides.
| Tracking<br>Principle | Infrared passive and active marker based tracking as provided by the<br>optical tracking camera unit of the navigation station. |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Navigation | The relation between the patient and the reference attached to the patient<br>is realized with a manual registration using instruments.<br>The Knee3 software receives position and orientation of tracked instrument<br>accessories and calculates the navigation measurements out of their<br>relations to each other. The measurement values are displayed within the<br>software graphical user interface on the respective platform screen. |
| Display | LCD display with capacitive dual touch technology. |
#### 3. Substantial Equivalence
| Devices | Name (version) | K-Number | Manufacturer |
|---------------------|---------------------------|----------|--------------|
| Subject Device | Knee3 (Knee 3.3) | N/A | Brainlab AG |
| Primary Predicate | Brainlab Knee (Knee 2.5) | K102990 | Brainlab AG |
| Secondary Predicate | uni-Knee (Uni-Knee 2.0) | K080678 | Brainlab AG |
| Tertiary predicate | DASH Knee (DASH Knee 1.0) | K102251 | Brainlab AG |
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Infrared optical tracking is the technological principle for both the subject and predicate devices. It is based on the use of instruments with reflective marker technology.
At a high level, the subject and predicate devices are based on the following same technological elements:
- Same localization and referencing techniques -
- Same user interface technology -
- Same remote control devices -
- Same general acquisition methods -
- Mostly the same registration steps that make up the bone coordinate systems -
- Instruments are made of surgical grade materials -
- Instruments are pre-calibrated -
The following technological differences exist between the subject and predicate devices:
- User interface visual design -
- -Software user guidance
Knee 3
#### 510(k) Summary
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "BRAINLAB" in all capital letters. To the left of the word is a stylized logo that appears to be three curved lines stacked on top of each other. The text and logo are in black and the background is white.
- Some landmark registrations -
- -Compatible platform devices and accessories
However, these differences do not raise any new questions of safety or efficacy because the performance of the subject device has thoroughly been verified for its efficacy.
# 4. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### System
The technical navigation accuracy (bench accuracy) of the overall system consisting of platform, software and instruments has been evaluated on a bench model and has shown to be within limits of ± 2° for angles and ± 2mm for distances in 95% of cases.
#### Instruments
The safety and effectiveness of the Knee3 Instruments has been verified. Technical performance has been tested on the following main aspects:
- . Instrument system integration testing: the Knee3 instruments have been proven to meet the accuracy requirements for use with Knee3.
- Mechanical stability testing: the mechanical stability of the Knee3 instruments has proven to ● be sufficient for the intended use.
- Biocompatibility: the Knee3 instruments with intended patient contact have been supported as ● biocompatible.
#### Biocompatibility testing
The biocompatibility evaluation for the Knee3 surgical instruments was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity tests acc. to ISO 10993-5 -
- Chemical analysis acc. to 10993-12 and 10993-18 for quantification and characterization of organic leachables and extractables
In addition, the following biological endpoints were evaluated based on the test results:
- -Sensitization acc. to ISO 10993-10
- Irritation acc. to ISO 10993-10 -
- -Materials-Mediated Pyrogenicity acc. to ISO 10993-11
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Brainlab. The logo consists of a stylized symbol on the left and the word "BRAINLAB" in all caps on the right. The symbol appears to be three interconnected shapes, possibly representing neurons or brainwaves. The text is in a bold, sans-serif font, giving the logo a modern and professional look.
- -Acute Systemic Toxicity acc. to ISO 10993-11
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since failures or latent flaws could prior to mitigation of hazards result in serious injury to the patient. Information provided by the software drives a decision through the user, that, if misapplied, could result in serious injury.
#### Clinical Validation:
Clinical Validation for the Subject Device was carried out to establish equivalence of the subject device workflows (Motion, Universal, Partial and Express) to the workflows contained within the identified predicate devices.
Clinical data gathered from literature was used to validate the accuracy of the predecessor software device versions (predicate devices) and respective workflows.
#### 5. Conclusion
The comparison of the Knee3 with the predicate device(s) Brainlab Knee, Uni Knee and DASH Knee shows that the Knee3 has similar functionality, intended use, technological characteristics, and typical users as the predicate devices. Verification activities ensured that the design specifications are met and that the Knee3 does not introduce new issues concerning safety and effectiveness. Hence, the Subject Device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.