← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K220038

# Camber Spine Navigation System (K220038)

_Camber Spine Technologies · OLO · Apr 29, 2022 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K220038

## Device Facts

- **Applicant:** Camber Spine Technologies
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** Apr 29, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

Camber Spine Navigation System instruments are intended to be used in the preparation and placement of ORTHROS Posterior Stabilization System and ORTHROS MIS Posterior Stabilization System screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtron® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MRI based model, fluoroscopy images, or digitized landmarks of the anatomy.

## Device Story

Camber Spine Navigation System consists of reusable surgical instruments (inserters, taps, probes, awls, awl taps) used to assist surgeons in spinal screw placement. Instruments function as tracked components within the Medtronic StealthStation® Navigation System. Surgeons use the system to reference rigid anatomical structures (vertebrae) against pre-operative CT/MRI models, fluoroscopy, or digitized landmarks. The system provides real-time navigation to guide instrument positioning during surgery. By enabling precise anatomical localization, the device assists in the accurate placement of stabilization screws, potentially improving surgical outcomes in open or minimally invasive spinal procedures.

## Clinical Evidence

No clinical data. Performance evaluation relied on bench testing, including one-to-one dimensional comparisons, anatomical and navigated simulated use, and accuracy/reliability testing to verify compatibility with the Medtronic StealthStation® System and NavLock Tracker.

## Technological Characteristics

Reusable instruments (inserters, taps, probes, awls, awl taps) made of medical-grade stainless steel (ASTM F899) with titanium nitride coating (SAE AMS 2444A). Designed for integration with Medtronic StealthStation® Navigation System and NavLock arrays. Non-sterile, manual surgical instruments.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Medtronic Navigated Instruments ([K153442](/device/K153442.md))
- Medtronic Navigated Instruments ([K124004](/device/K124004.md))
- Orthros™ Posterior Stabilization System and Orthros™ MIS Posterior Stabilization ([K180980](/device/K180980.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

April 29, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

Camber Spine Technologies % Christine Scifert Partner MRC Global, LLC 9085 E. Mineral Cir., Suite 110 Centennial. Colorado 80112

Re: K220038

Trade/Device Name: Camber Spine Navigation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: March 28, 2022 Received: March 30, 2022

Dear Christine Scifert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

es. You must comply with all the Act's

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220038

Device Name Camber Spine Navigation System

#### Indications for Use (Describe)

Camber Spine Navigation System instruments are intended to be used in the preparation and placement of ORTHROS Posterior Stabilization System and ORTHROS MIS Posterior Stabilization System screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtron® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MRI based model, fluoroscopy images, or digitized landmarks of the anatomy.

| Type of Use (Select one or both, as applicable)                                                                                |  |
|--------------------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> |  |
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div>  |  |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary Camber Spine Navigation System 28 April 2022

| Company:                | Camber Spine Technologies<br>501 Allendale Rd<br>King of Prussia, PA 19406                                     |
|-------------------------|----------------------------------------------------------------------------------------------------------------|
| Company Contact:        | Noel Hetrick<br>Director, Quality Assurance/Regulatory Affairs<br>(484) 427-7060<br>nhetrick@cambermedtech.com |
| Official Correspondent: | Christine Scifert – MRC Global, LLC<br>Christine.scifert@askmrcglobal.com<br>901-831-8053                      |
| Trade Name:             | Camber Spine Navigation System                                                                                 |
| Common Name:            | Orthopedic Stereotaxic Instrument                                                                              |
| Classification:         | Class II                                                                                                       |
| Regulation Number:      | 21 CFR 882.4560 (Stereotaxic Instrument)                                                                       |
| Panel:                  | Orthopedic                                                                                                     |
| Product Code:           | OLO                                                                                                            |

## Device Description:

Camber Spine Navigation System contains reusable instruments, provided non-sterile, including inserters, taps, probes, awls, and awl taps. These instruments are intended to be used with the Medtronic StealthStation® Navigation System and its associated NavLock arrays, to assist surgeons in precisely locating anatomical structures for preparation and placement of ORTHROS Posterior Stabilization System and ORTHROS MIS Posterior Stabilization System screws during spinal surgery. The instrumentation is designed for use with the Medtronic StealthStation® Navigation System hardware and software. These instruments are made of medical grade stainless steel according to the ASTM F899 and many include a titanium nitride coating per SAE AMS 2444A.

{4}------------------------------------------------

### Indications for Use:

Camber Spine Navigation System instruments are intended to be used in the preparation and placement of ORTHROS Posterior Stabilization System and ORTHROS MIS Posterior Stabilization System screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MRI based model, fluoroscopy images, or digitized landmarks of the anatomy.

#### Substantial Equivalence:

The subject Camber Spine Navigation System is substantially equivalent to the following predicate devices:

#### Primary Predicate:

- . Medtronic Sofamor Danek – Medtronic Navigated Instruments (K153442)
#### Secondary Predicates:

- . Medtronic Sofamor Danek – Medtronic Navigated Instruments (K124004)
- . Camber Spine – Orthros™ Posterior Stabilization System and Orthros™ MIS Posterior Stabilization (K180980)

Similar to the predicate navigated instruments, Camber Spine Navigation System instruments are intended to be used with the Medtronic's StealthStation® System to assist the surgeon in locating anatomical structures. Additionally, the Camber Spine Navigation System instruments and their predicate devices have similar technological characteristics, including design, dimensions, materials and technology, and they function in the same manner. Performance testing demonstrates that the Camber Spine Navigation System instruments are substantially equivalent to the predicate devices. Thus, it can be concluded that the subject does not raise new questions about safety and effectiveness.

#### Performance Testing:

The performance evaluation included a one-to-one dimensional comparison to demonstrate substantially equivalent geometry that is critical to navigation accuracy. Testing, including anatomical and navigated simulated use and accuracy and reliability testing, has been conducted to verify that the Camber Spine Navigation System instruments are appropriate for their intended use, to ensure functionality, accuracy and compatibility with the Medtronic StealthStation® System using the NavLock Tracker, and to demonstrate substantial equivalence to the predicate instruments.

#### Conclusion:

Based on the test results and the comparison to the predicate devices, the subject device is determined to be substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K220038](https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K220038)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
