← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K202184

# Practical Navigation Surgical Guidance System (PNSGS) (K202184)

_Practical Navigation, LLC · OLO · Feb 24, 2021 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K202184

## Device Facts

- **Applicant:** Practical Navigation, LLC
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** Feb 24, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The Practical Navigation Surgical Guidance System is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of a tool holder or Guide Tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous spinal procedures in reference to rigid patient anatomy and fiducials that can be identified on an O-arm scan. The Practical Navigation Surgical Guidance System is indicated for assisting the surgeon in placing pedicle screws in vertebrae in the posterior lumbar region (L1-S1). The system is designed for lumbar pedicle screw placement with the patient in the prone position and is compatible with the Integrity LineSider Pedicle Spinal System.

## Device Story

Practical Navigation Surgical Guidance System (PNSGS) is an image-guided surgical navigation system for lumbar pedicle screw placement. System inputs include 3D intraoperative O-arm images and optical tracking data from an infrared camera. The camera tracks the surgical site and a registration array fixed to the patient's iliac crest. The workstation reconstructs images, registers patient anatomy, and allows surgical path planning. A targeting platform, mounted on a mechanical support arm, is manually positioned and activated to set the trajectory for surgical instruments. The surgeon uses tracked instruments through the targeting platform's guide tube to prepare the pedicle and place screws. Used in the OR by surgeons; provides real-time visual feedback of instrument position relative to the surgical plan. Benefits include precise anatomical localization and trajectory guidance for spinal instrumentation.

## Clinical Evidence

No clinical data. Evidence consists of non-clinical system, software, and instrument verification and validation, including cadaver and simulated-use testing under clinically relevant scenarios. Compliance with ASTM F2554-10 (positional accuracy), ANSI/AAMI ES60601-1, and IEC 60601-1-2 standards was demonstrated.

## Technological Characteristics

System includes computer workstation, software, targeting platform, infrared camera, and tracked instruments. Localization via optical infrared camera. Connectivity includes O-arm image integration. Software performs image reconstruction, registration, and trajectory planning. Complies with ASTM F2554-10, ANSI/AAMI ES60601-1, and IEC 60601-1-2.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Excelsius GPS ([K171651](/device/K171651.md))

## Reference Devices

- LineSider™ Spinal System ([K203367](/device/K203367.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

February 24, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Practical Navigation, LLC % Roger White President Phiama, Inc 236 McKinley Park Lane Louisville, Colorado 80027

Re: K202184

Trade/Device Name: Practical Navigation Surgical Guidance System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: January 22, 2021 Received: January 25, 2021

Dear Roger White:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For: Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

#### 510(k) Number (if known) K202184

Device Name

Practical Navigation Surgical Guidance System

#### Indications for Use (Describe)

The Practical Navigation Surgical Guidance System is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of a tool holder or Guide Tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous spinal procedures in rigid patient anatomy and fiducials that can be identified on an O-arm scan. The Practical Navigation Surgical Guidance System is indicated for assisting the surgeon in placing pedicle in the posterior lumbar region (1-S1).

The system is designed for lumbar pedicle screw placement with the prone position and is compatible with the Integrity LineSider Pedicle Spinal System.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary Practical Navigation Surgical Guidance System

### 1. Submitter Information

| Submitter: | Practical Navigation, LLC |
|------------|---------------------------|
| Address:   | 720 E. Wiggins Street     |
|            | Superior, CO 80027        |
| Telephone: | (303) 517-0506            |
| Telefax:   | N/A                       |

- Roger N. White Contact: rwhite@phiama.com (303) 550-2451
Date Prepared: February 24, 2021

#### 2. Device Information

| Trade Name:          | Practical Navigation Surgical Guidance System |
|----------------------|-----------------------------------------------|
| Common Name:         | Stereotaxic Instrument                        |
| Classification:      | Class II per 21 CFR 882.4560                  |
| Classification Name: | Stereotaxic Instrument                        |
| Product Code:        | OLO                                           |

#### 3. Purpose of Submission

The purpose of this submission is to gain clearance for a new Image Guided System, the Practical Navigation Surgical Guidance System.

#### 4. Predicate Device Information

The Practical Navigation Surgical Guidance System described in this submission is substantially equivalent to the following predicate:

| Predicate Device | Manufacturer         | 510(k) No. |
|------------------|----------------------|------------|
| Excelsius GPS    | Globus Medical, Inc. | K171651    |

The following device is referenced in this submission:

| Reference Device         | Manufacturer             | 510(k) No. |
|--------------------------|--------------------------|------------|
| LineSider™ Spinal System | Integrity Implants, Inc. | K203367    |

{4}------------------------------------------------

### 5. Device Description

The Practical Navigation Surgical Guidance System (PNSGS) is an image guided system primarily comprised of a computer workstation, software, a trajectory system, including a targeting platform, a Camera, and various image guided instruments intended for assisting the surgeon in placing screws in the pedicles of the lumbar spine. The PNSGS system allows for registration of the patient's anatomy to O-arm images. Once the patient anatomy is registered, a surgical plan can be created and optically-tracked surgical instruments can be used to follow the surgical plan. The system is used in the following manner:

- 1. The patient is placed in the appropriate position on the OR table.
- The compact tracking Camera is rigidly affixed to the OR table using a multi-2. functional mechanical support arm in the appropriate position to track the surgical site.
- 3. The Camera is also affixed to a pin placed in the patient's iliac to provide a fixed location relative to the patient's spinal anatomy.
- 4. The Targeting Platform is affixed to the OR Table using a multi-functional mechanical support arm, ensuring that the Targeting Platform has sufficient range of motion to be placed over the surgical site.
- 5. The Registration Array is affixed to planned surgical site.
- The appropriate area of spine is scanned with an O-arm. 6.
- 7. The scans are transferred to the PNSGS workstation, which reconstructs the images and uses the registration array image to register the patient's spine relative to the Camera location.
- 8. The registration is confirmed by placing an image guided instrument with a Tracking Array at various points in the surgical field.
- 9. The surgical paths are planned on the workstation.
- 10. The Targeting Platform is gross-positioned manually close to the first surgical plan location.
- 11. The Targeting Platform is activated to set the fine location and the trajectory based on the surgical plan.
- 12. Instruments with tracking arrays are used through the tool guide of the Targeting Platform to prepare the pedicle and place a pedicle screw.

### 6. Indications for Use

The Practical Navigation Surgical Guidance System is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of a tool holder or Guide Tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous spinal procedures in reference to rigid patient anatomy and fiducials that can be identified on an O-arm scan. The Practical Navigation Surgical Guidance System is indicated for assisting the surgeon in placing pedicle screws in vertebrae in the posterior lumbar region (L1-S1).

{5}------------------------------------------------

The system is designed for lumbar pedicle screw placement with the patient in the prone position and is compatible with the Integrity LineSider Pedicle Spinal System.

## 7. Comparison of Technological Characteristics

The substantial equivalence of the Practical Navigation Surgical Guidance System to the predicate devices is shown by the similarity in intended use, indications for use, mechanism of action, mode of operation, and performance. The table below provides a comparison of technological characteristics and principles of operation of the PNSGS and the predicate system:

| Feature                                                | Practical Navigation Surgical Guidance<br>System<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Globus Medical, Inc.<br>Excelsius GPS<br>(K171651)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use:                                   | The Practical Navigation Surgical Guidance<br>System is intended for use as an aid for precisely<br>locating anatomical structures and for the spatial<br>positioning and orientation of a tool holder or<br>Guide Tube to be used by surgeons for navigating<br>and/or guiding compatible surgical instruments in<br>open or percutaneous spinal procedures in<br>reference to rigid patient anatomy and fiducials<br>that can be identified on an O-arm scan. The<br>Practical Navigation Surgical Guidance System is<br>indicated for assisting the surgeon in placing<br>pedicle screws in vertebrae in the posterior<br>lumbar region (L1-S1). | The EXCELSIUS GPSTM is intended for use as an aid<br>for precisely locating anatomical structures and for the<br>spatial positioning and orientation of an instrument<br>holder or Guide Tube to be used by surgeons for<br>navigating and/or guiding compatible surgical<br>instruments in open or percutaneous procedures<br>provided that the required fiducial markers and rigid<br>patient anatomy can be identified on CT scans or<br>fluoroscopy. The system is indicated for the placement<br>of spinal and orthopedic bone screws. |
| Principles of<br>Operation                             | - Intraoperative/<br>preoperative images<br>- Patient registration<br>- Surgical planning<br>- Real-time tracking of<br>navigated instruments<br>- Guidance of instruments                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | - Intraoperative/<br>preoperative images<br>- Patient registration<br>- Surgical planning<br>- Real-time tracking of<br>navigated instruments<br>- Guidance of instruments                                                                                                                                                                                                                                                                                                                                                                  |
| Input Images                                           | 3D Intraoperative images                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 3D Pre-operative images<br>3D Intraoperative images<br>2D Intraoperative images                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Integrated Planning<br>Software                        | PNSGS Planning and Navigation Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Excelsius™ GPS Planning and Navigation Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Save/load Planning                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Image Merging                                          | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Trajectory planning<br>parameters                      | Entry point, target point, length of the instrument,<br>diameter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Entry point, target point, length of the instrument,<br>diameter                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Localization method                                    | Optical System (infrared Camera)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Optical System (infrared Camera)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Camera system                                          | Monocular                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Stereo                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Image-guided<br>Controller                             | Yes<br>Manual macro adjustments<br>Force-controlled movement of Targeting<br>platform                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes<br>Force-controlled movement allowing robotic arm<br>positioning                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Patient Registration<br>Method                         | Registration fixture in place during 3D<br>intraoperative images                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Registration fixture during 3D intraoperative images.<br>Fluoroscopic merge to pre-op CT<br>Registration fixture in place during fluoroscopy                                                                                                                                                                                                                                                                                                                                                                                                |
| Accuracy<br>verification on<br>anatomical<br>landmarks | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Real time display of<br>instrument position            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Instrument<br>Guidance                                 | Trajectory and location set by Targeting<br>platform. Instruments are manually                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Trajectory and location set by the robotic arm.<br>Instruments are manually positioned by the                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Feature                                                | Practical Navigation Surgical Guidance<br>System<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Globus Medical, Inc.<br>Excelsius GPS<br>(K171651)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                                        | positioned by the surgeon through the<br>Guide Tube on the Targeting Platform.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | surgeon through the Guide Tube on the robotic<br>arm.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Patient fixation                                       | Tracking Camera is fixed to OR table and to<br>the patient's iliac crest.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Dynamic reference system is fixed to patient's<br>bony structure such as a long bone, iliac crest,<br>spinous process, vertebra, etc.                                                                                                                                                                                                                                                                                                                                                                                                       |

{6}------------------------------------------------

#### 8. Performance Data

Practical Navigation Surgical Guidance System was tested in accordance with the design requirements. Testing included the following:

- Non-clinical system, software, and instrument verification and validation.
- Cadaver and simulated-use testing under clinically relevant scenarios.
- ASTM F2554-10 Standard Practice for Measurement of Positional Accuracy of . Computer Assisted Surgical Systems.
- ANSI/AAMI ES60601-1:2005/(R)2012 And A1:2012Medical Electrical ● Equipment - Part 1: General Requirements for Basic Safety and Essential Performance.
- IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: General ● Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Testing.

#### 9. Conclusion

Based on the indications for use, technological characteristics, performance results, and comparison to the predicates, the Practical Navigation Surgical Guidance System has been shown to be substantially equivalent to the predicate devices identified in this submission and does not present any new issues of safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K202184](https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K202184)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
