← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K192495

# Black Diamond Navigation Instruments (K192495)

_Osseus Fusion Systems · OLO · Jan 17, 2020 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K192495

## Device Facts

- **Applicant:** Osseus Fusion Systems
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** Jan 17, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The Black Diamond Navigation Instruments are intended to be used during the preparation and placement of Black Diamond pedicle screws during spinal surgery to aid the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The Black Diamond Navigation Instruments are specifically designed for use with the Medtronic Stealth Station System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

## Device Story

Black Diamond Navigation Instruments consist of manual, reusable surgical tools including probes, bone taps, and inserters; used in spinal surgery for pedicle screw placement. Instruments interface with Medtronic StealthStation System (S7 v2.1.0) to provide real-time, multi-plane 3D/2D image guidance. Surgeon operates instruments manually to locate anatomical structures; navigation system tracks instrument position relative to patient anatomy (CT/MR/fluoroscopy). Output provides visual confirmation of hardware placement to surgeon; aids in precise screw positioning; potentially improves surgical accuracy and patient outcomes in open or minimally invasive procedures.

## Clinical Evidence

No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing, including accuracy, compatibility, and performance testing.

## Technological Characteristics

Non-sterile, reusable manual instruments (probes, taps, inserters). Materials: Stainless steel per ASTM F899. Operates via manual manipulation in conjunction with Medtronic StealthStation System for image-guided navigation. No internal energy source; connectivity provided by integration with external navigation platform.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Navigated CD Horizon Solera Screwdrivers and Taps, Medtronic Sofamor Danek, USA Inc. ([K140454](/device/K140454.md))
- Universal Navigation Instruments for EXPEDIUM® and VIPER® MIS Spine Systems, DePuy Spine, Inc ([K140927](/device/K140927.md))
- SeaSpine Navigation System, SeaSpine Orthopedics Corporation ([K172517](/device/K172517.md))

## Reference Devices

- Black Diamond Pedicle Screw System, Osseus Fusion Systems ([K131810](/device/K131810.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 17, 2020

Osseus Fusion Systems % J. D. Webb Official Correspondent The OrthoMedix Group, Inc. 4313 W. 3800 S. West Haven, Utah 84401

## Re: K192495

Trade/Device Name: Black Diamond Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: September 5, 2019 Received: September 11, 2019

## Dear J. Webb:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K192495

Device Name Black Diamond Navigation Instruments

#### Indications for Use (Describe)

The Black Diamond Navigation Instruments are intended to be used during the preparation and placement of Black Diamond pedicle screws during spinal surgery to aid the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The Black Diamond Navigation Instruments are specifically designed for use with the Medtronic Stealth Station System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

Type of Use (Select one or both, as applicable)

| <span>         ☑ Prescription Use (Part 21 CFR 801 Subpart D)       </span> |
|-----------------------------------------------------------------------------|
| <span>         ☐ Over-The-Counter Use (21 CFR 801 Subpart C)       </span>  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary: Black Diamond Navigation Instruments

| Date Prepared                                                                  | January 9, 2020                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By                                                                   | Osseus Fusion Systems, LLC<br>2703 W. Mockingbird Ln., Ste. #102<br>Dallas, TX 75204                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Primary Contact                                                                | J.D. Webb<br>4313 W. 3800 S<br>West Haven, UT 84401<br>512-590-5810 Tele<br>e-mail: jdwebb@orthomedix.net                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Trade Name                                                                     | Black Diamond Navigation Instruments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Common Name                                                                    | Orthopedic Stereotaxic instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Classification Name                                                            | Stereotaxic Instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Class                                                                          | ll                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Product Code                                                                   | OLO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| CFR Section                                                                    | 21 CFR section 882.4560                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Panel                                                                   | Orthopedic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Primary Predicate<br>Device                                                    | Navigated CD Horizon Solera Screwdrivers and Taps, Medtronic Sofamor Danek, USA<br>Inc. (K140454)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Secondary Predicate<br>Devices                                                 | Universal Navigation Instruments for EXPEDIUM® and VIPER® MIS Spine Systems,<br>DePuy Spine, Inc (K140927)<br>SeaSpine Navigation System, SeaSpine Orthopedics Corporation (K172517)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Reference Predicate<br>Devices                                                 | Black Diamond Pedicle Screw System, Osseus Fusion Systems (K131810)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device Description                                                             | Black Diamond Navigation Instruments are non-sterile, reusable instruments; including<br>probes, bone taps, and inserters that are operated manually. These instruments are<br>intended to be used within the context and limitations of the indications for use for Osseus<br>Fusion System's FDA-cleared Black Diamond system and the Medtronic Synergy<br>Experience StealthStation System S7 (v2.1.0). Use of these navigation systems provides<br>the surgeon access to real-time, multi-plane 3D images (and 2D images) providing<br>confirmation of hardware placement.                                                                                                                                              |
| Materials                                                                      | Stainless steel per ASTM F899                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Substantial<br>Equivalence Claimed<br>to Predicate Devices                     | The Black Diamond Navigation Instruments are substantially equivalent to the<br>predicate devices in terms of intended use, design, materials used, mechanical safety and<br>performances.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Indications for Use                                                            | The Black Diamond Navigation Instruments are intended to be used during the preparation<br>and placement of Black Diamond pedicle screws during spinal surgery to aid the surgeon<br>in precisely locating anatomical structures in either open or minimally invasive procedures.<br>The Black Diamond Navigation Instruments are specifically designed for use with the<br>Medtronic Stealth Station System, which is indicated for any medical condition in which<br>the use of stereotactic surgery may be appropriate, and where reference to a rigid<br>anatomical structure, such as a vertebra, can be identified relative to a CT or MR based<br>model, fluoroscopy images, or digitized landmarks for the anatomy. |
| Summary of the<br>technological<br>characteristics<br>compared to<br>predicate | When compared to the predicate devices, Black Diamond Navigation Instruments have<br>the same intended use and similar technological characteristics, including:<br>Design Materials of Construction Function/Performance Fundamental Scientific Technology Principle of Operation                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Non-clinical Test<br>Summary                                                   | Nonclinical testing was performed to show that the subject Black Diamond Navigation<br>Instruments are substantially equivalent to the predicate device. The following testing was<br>performed:<br>Accuracy testing Compatibility testing Performance testing              The results of these evaluations indicate that the Black Diamond Navigation Instruments<br>are equivalent to predicate devices.                                                                                                                                                                                                                                                                                                                 |
| Clinical Test<br>Summary                                                       | No clinical studies were performed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Conclusions: Non-<br>clinical and Clinical                                     | Osseus Fusion Systems considers the Navigation Instruments to be equivalent to the<br>predicate devices listed above. This conclusion is based upon the devices' similarities in<br>principles of operation, technology, materials, and indications for use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

In accordance with 21 CFR 807.92 of the Federal Code of Regulations

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K192495](https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K192495)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
