Intellijoint® Navigation System

K191507 · Intellijoint Surgical, Inc. · OLO · Oct 1, 2019 · Neurology

Device Facts

Record IDK191507
Device NameIntellijoint® Navigation System
ApplicantIntellijoint Surgical, Inc.
Product CodeOLO · Neurology
Decision DateOct 1, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

The Intellijoint® Navigation System is a computer-controlled, optical localizer intended to provide intra-operative measurements to a surgeon to aid in selection and positioning of orthopedic implant system components, where a reference to a rigid anatomical structure can be identified. The Intellijoint® Navigation System is indicated for patients undergoing orthopedic surgery, and where the use of stereotactic surgery is considered safe and effective. Intellijoint HIP® aids the surgeon in performing intra-operative measurements including measurements of limb position, joint center-of-rotation, and implant component positioning. Intellijoint KNEE™ aids the surgeon in performing intra-operative measurements including alignment of instrumentation and bony cuts for implant component positioning. Example orthopedic surgical procedures include, but are not limited to: Total Hip Arthroplasty; Minimally Invasive Hip Arthroplasty; Revision Hip Arthroplasty; Total Knee Arthroplasty

Device Story

Imageless optical navigation system for orthopedic surgery; utilizes infrared camera, camera drape, optical tracking spheres, and computer workstation. Surgeon mounts bone trackers to femur or tibia; system digitizes anatomical landmarks to provide real-time intra-operative measurements. Outputs include varus/valgus angle, flexion/extension angle, and resection depth for TKA; limb position, joint center-of-rotation, and implant component positioning for hip procedures. Used in operating room by surgeons to guide implant selection and positioning; improves surgical accuracy via real-time feedback.

Clinical Evidence

Bench testing only. Included benchtop accuracy verification per ASTM F2554-10, software functional/unit testing, and mechanical instrumentation performance tests. Validation included simulated use on anatomical bone models and cadaver labs to assess clinical accuracy and workflow performance; all requirements met.

Technological Characteristics

Imageless optical navigation system; monocular infrared camera; optical tracking spheres; computer workstation. Connectivity: standalone. Sterilization: not specified. Software: application-specific algorithms for orthopedic measurements. Standards: ASTM F2554-10 for positional accuracy.

Indications for Use

Indicated for patients undergoing orthopedic surgery where stereotactic surgery is considered safe and effective, including Total Hip Arthroplasty, Minimally Invasive Hip Arthroplasty, Revision Hip Arthroplasty, and Total Knee Arthroplasty.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ October 1, 2019 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below. Intellijoint Surgical Inc. Breanne Cuddington Regulatory Affairs Associate 60 Bathurst Drive Unit 6 Waterloo, N2V 2A9 CA Re: K191507 Trade/Device Name: Intellijoint® Navigation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: August 30, 2019 Received: September 3, 2019 Dear Breanne Cuddington: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For, Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K191507 Device Name Intellijoint® Navigation System #### Indications for Use (Describe) The Intellijoint® Navigation System is a computer-controlled, optical localizer intended to provide intra-operative measurements to a surgeon to aid in selection and positioning of orthopedic implant system components, where a reference to a rigid anatomical structure can be identified. The Intellijoint Navigation System is indicated for patients undergoing orthopedic surgery, and where the use of stereotactic surgery is considered safe and effective. Intellijoint HIP aids the surgeon in performing intra-operative measurements of limb position, joint center-of-rotation, and implant component positioning. Intellioint KNEE aids the surgeon in performing intra-operative measurements including alignmentation and bony cuts for implant component positioning. Example orthopedic surgical procedures include, but are not limited to: - Total Hip Arthroplasty - Minimally Invasive Hip Arthroplasty - Revision Hip Arthroplasty - Total Knee Arthroplasty Type of Use (Select one or both, as applicable) | <span style="font-size:10pt">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10pt">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/3 description: The image shows the logo for "Intellijoint Surgical". The logo consists of a geometric shape on the left and the company name on the right. The geometric shape is made up of three triangles, with the top triangle being dark blue, the middle triangle being light blue, and the bottom triangle being gray. The company name is written in gray, with the word "Intellijoint" on top and the word "Surgical" on the bottom. # 510(k) Summary # 1. Submitter Information | Submitter: | Intellijoint Surgical Inc. | |----------------|-----------------------------------------------------------------| | Address: | 809 Wellington St. N, Unit 2<br>Kitchener, ON<br>Canada N2H 5L6 | | Telephone: | (519) 342-3178 | | Fax: | (226) 317-0471 | | Contact: | Breanne Cuddington | | Date Prepared: | 30 August 2019 | # 2. Device Information | Trade Name: | Intellijoint® Navigation System | |----------------------|-----------------------------------| | Common Name: | Orthopedic Stereotaxic Instrument | | Classification: | Class II per 21 CFR 882.4560 | | Classification Name: | Orthopedic Stereotaxic Instrument | | Product Code: | OLO | # 3. Purpose of Submission The purpose of this submission is to gain clearance for updates to the indications for use for the Intellijoint HIP Generation 2B System, which has been previously cleared in K171525. The updated product will be referred to as the Intellijoint® Navigation System throughout the submission. # 4. Predicate Device Information The Intellijoint® Navigation System described in this substantially equivalent to the following predicate devices: ## Primary Predicate: | Device Name | Manufacturer | 510(k) No. | |--------------------|-----------------|------------| | KneeAlign 2 System | OrthAlign, Inc. | K103829 | ## Additional Predicate: | Device Name | Manufacturer | 510(k) No. | |---------------------------------------|----------------------------|------------| | Intellijoint HIP Generation 2B System | Intellijoint Surgical Inc. | K171525 | {4}------------------------------------------------ tellijoint # 5. Device Description The Intellijoint® Navigation System (IINS) is an imageless optical navigation system intended for use in orthopedic surgery. The IINS consists of an infrared Camera, Camera Drape, optical tracking Spheres, a computer Workstation and associated hardware. This submission provides the addition of Total Knee Arthroplasty (TKA) to the indications for use (cleared in K171525), allowing real-time measurements of varus/valgus angle, flexion/extension angle, and resection depth for distal femur and proximal tibia cuts during TKA procedures. The modifications include the addition of TKA-specific mechanical instruments and software. The updated product will be referred to as the Intellijoint® Navigation System throughout this submission. # 6. Indications for Use The Intellijoint® Navigation System is a computer-controlled, optical localizer intended to provide intra-operative measurements to a surgeon to aid in selection and positioning of orthopedic implant system components, where a reference to a rigid anatomical structure can be identified. The Intellijoint® Navigation System is indicated for patients undergoing orthopedic surgery, and where the use of stereotactic surgery is considered safe and effective. Intellijoint HIP® aids the surgeon in performing intra-operative measurements including measurements of limb position, joint center-of-rotation, and implant component positioning. Intellijoint KNEE™ aids the surgeon in performing intra-operative measurements including alignment of instrumentation and bony cuts for implant component positioning. Example orthopedic surgical procedures include, but are not limited to: - Total Hip Arthroplasty । - Minimally Invasive Hip Arthroplasty । - Revision Hip Arthroplasty - - -Total Knee Arthroplasty ## 7. Comparison of Technological Characteristics The substantial equivalence of the Intellijoint® Navigation System compared to the predicate devices is based on the following technological characteristics: | Property | Intellijoint® Navigation System | Primary Predicate<br>KneeAlign® 2 System | Additional Predicate<br>Intellijoint HIP® Generation 2B System | |------------------------|---------------------------------|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------| | 510(k) No. | K191507 | K103829 | K171525 | | Tracking method | Optical | Inertial | Optical | | Imaging Modality | Imageless | Imageless | Imageless | | Camera System | Monocular | N/A | Monocular | | Registration<br>Method | HIP<br>Same as K171525 | HIP<br>N/A | HIP<br>i. Camera rigidly fixed to pelvis.<br>ii. Digitization of Anterior Pelvic Plane and/or frontal / sagittal planes. | {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for "IntelliJoint Surgical". The logo consists of a geometric shape on the left, resembling a stylized joint, with three overlapping triangles in different shades of blue and gray. To the right of the geometric shape is the word "IntelliJoint" in a sans-serif font, with the word "Surgical" in a smaller font size underneath. K191507 | Property | Intellijoint®<br>Navigation System | Primary Predicate | Additional Predicate | |----------|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|-------------------------------------------| | | | KneeAlign® 2 System | Intellijoint HIP®<br>Generation 2B System | | | KNEE | KNEE | KNEE | | | i. Bone Tracker mounted<br>rigidly to Femur or<br>Tibia<br>ii. Digitization of Femoral<br>or Tibial Landmarks | i. Navigation unit mounted<br>rigidly to Femur or Tibia<br>ii. Digitization of Femoral<br>or Tibial landmarks | N/A | # 8. Performance Data This submission provides the addition of TKA to the indications for use cleared in 510(k) K171525 and adds the Intellijoint KNEE™ Software Application to the Intellijoint® Navigation System. Other modifications include the addition of TKA specific mechanical instruments to the Intellijoint® Navigation System. The following tests were performed to support the addition of Intellijoint KNEE™ and to demonstrate the substantial equivalence of the Intellijoint® Navigation System to its predicate devices: | Test | Summary | Result | |-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------| | Verification | | | | Benchtop Accuracy | Verified clinical accuracy requirements using<br>calibrated benchtop test fixtures. | All accuracy<br>requirements were<br>met. | | Tracking System Accuracy and Robustness | Accuracy of Intellijoint KNEE was verified<br>according to the methodology in ASTM F2554-<br>10 - Standard Practice for Measurement of<br>Positional Accuracy of Computer Assisted<br>Surgical Systems. Testing simulated normal<br>conditions, and a variety of worst-case use<br>scenarios and realistic tracking disturbances. | All accuracy<br>specifications and<br>robustness<br>requirements were<br>met. | | Software Functional and Unit Tests | Verified that the Intellijoint KNEE Software<br>Application satisfies functional requirements<br>and performs as intended. Algorithms and<br>measurement calculations were also verified in<br>these tests. | Software satisfied<br>all requirements<br>and specifications. | | Mechanical Instrumentation Functional and Performance Tests | Provides confirmation that mechanical<br>instrumentation satisfies functional and<br>performance requirements. | All functional and<br>performance<br>requirements met. | {6}------------------------------------------------ Image /page/6/Picture/3 description: The image shows the logo for "Intellijoint Surgical". The logo consists of a blue and gray triangle shape on the left, followed by the word "intellijoint" in gray. Below "intellijoint" is the word "surgical" in a smaller, lighter gray font. | Validation | | | |--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------| | Anatomical Phantom<br>Simulated Use and<br>Clinical Accuracy | Simulated use testing was performed on bone<br>models by orthopedic surgeons in a simulated<br>TKA procedure following a typical workflow.<br>This test validated that Intellijoint KNEE<br>satisfies user needs, intended use and clinical<br>accuracy requirements. Accuracy was assessed<br>by comparing simulated use measurements with<br>ground truth values. | All user needs and<br>clinical accuracy<br>requirements were<br>met. | | Cadaver Simulated Use | Simulated use testing was also performed in<br>cadaver labs. These tests validated that<br>Intellijoint KNEE satisfies clinical use<br>requirements and performs as intended when:<br>- Operated by a surgeon<br>- Used on human specimens<br>- Used in a realistic operating room<br>environment | All clinical use<br>requirements were<br>met. | The testing demonstrated that the Intellijoint® Navigation System is substantially equivalent to the legally marketed predicate devices for its intended use. # 9. Conclusion Based on the indications for use, technological characteristics, performance testing, and comparison to the predicate devices, the Intellijoint® Navigation System has been shown to be substantially equivalent to the legally marketed predicate devices identified in this submission and does not present any changes to safety or effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...