← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K190881

# Spinal Elements Navigated Instruments (K190881)

_Spinal Elements, Inc. · OLO · Dec 19, 2019 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K190881

## Device Facts

- **Applicant:** Spinal Elements, Inc.
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** Dec 19, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Intended Use

Spinal Elements Navigated Instruments are intended to be used during the preparation and placements screws (Overwatch®, Savannah® and Mercury® Fixation Systems) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

## Device Story

Navigated instruments (stainless steel) designed for use with Medtronic StealthStation® System; assist surgeons in precise anatomical localization during spinal screw placement (Overwatch®, Savannah®, Mercury® systems). Instruments function as part of a computer-assisted surgical navigation workflow; surgeon uses instruments in conjunction with CT/MR models, fluoroscopy, or digitized landmarks to guide screw placement in open or minimally invasive spinal procedures. Output is visual guidance on navigation system display; aids surgeon in real-time positioning; improves accuracy of screw placement; reduces risk of malpositioning.

## Clinical Evidence

Bench testing only. Performance validated per ASTM F2554 (Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems) to demonstrate positional accuracy equivalent to predicate devices.

## Technological Characteristics

Stainless steel instruments; reusable; non-sterile. Designed for integration with Medtronic StealthStation® navigation system. Principle of operation: computer-assisted surgical navigation using rigid anatomical reference. Tested per ASTM F2554.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Medtronic Navigated Horizon screwdriver/Taps ([K140454](/device/K140454.md))
- [K143375](/device/K143375.md)
- [K143628](/device/K143628.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

December 19, 2019

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

Spinal Elements Inc. Julie Lamothe Vice President Regulatory Affairs 3115 Melrose Dr. Suite 200 Carlsbad, California 92010

Re: K190881

Trade/Device Name: Navigated Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: November 18, 2019 Received: November 20, 2019

Dear Julie Lamothe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190881

#### Device Name Spinal Elements Navigated Instruments

#### Indications for Use (Describe)

Spinal Elements Navigated Instruments are intended to be used during the preparation and placements screws (Overwatch®, Savannah® and Mercury® Fixation Systems) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

| Type of Use (Select one or both, as applicable) |                                                                                                                                                                     |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span></span>                                   | <span></span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary Navigated Instruments

# 510(k) Number: K190881

| Manufacturer Identification<br>Submitted by:                                                                      | Spinal Elements, Inc.<br>3115 Melrose Dr., Suite 200<br>Carlsbad, CA 92010<br>760-607-0121                                                                                                         |
|-------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Information:                                                                                              | Julie Lamothe<br>Vice President, Regulatory Affairs/Quality Assurance<br>Spinal Elements, Inc.<br>3115 Melrose Dr., Suite 200<br>Carlsbad, CA 92010<br>760-607-1816<br>jlamothe@spinalelements.com |
| Date Prepared:                                                                                                    | December 16, 2019                                                                                                                                                                                  |
| Proprietary Device<br>Common Name<br>Classification Name<br>Regulation Number<br>Regulatory Class<br>Product Code | Spinal Elements Navigated Instruments<br>Stereotaxic Instrument<br>Orthopedic Stereotaxic Instrument<br>21 CFR 882.4560<br>Class II<br>OLO                                                         |

# Device Description

Spinal Elements Navigated instruments are non-sterile, reusable instruments designed to function with the the the Medtronic StealthStation® System. Spinal v Elements Navigated instruments are for use with Spinal Elements pedicle screw systems, specifically, Overwatch®, Mercury® and Savannah® pedicle screw systems. Instruments are manufactured from stainless steel.

# Indications for Use

Spinal Elements Navigated Instruments are intended to be used during the preparation and placement of Spinal Elements screws (Overwatch®, Savannah® and Mercury® Fixation Systems) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

{4}------------------------------------------------

### Substantial Equivalence

The subject devices are substantially equivalent to the primary predicate devices cleared in K140454 and the additional predicates cleared in K143375 and K143628.

### Technological Characteristics

The subject device has equivalent technological characteristics to its predicates presented below through comparison in areas including labeling/indications for use, general design features, function, material, manufacturing process and principle of operation:

- Medtronic Navigated Horizon screwdriver/Taps. K140454 Primary .
### Performance Data

Spinal Elements Navigated Instruments have been tested per ASTM F2554 "Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems". The results of this non-clinical testing show that performance of the Spinal Elements Navigated instruments is sufficient for its intended use and is substantially equivalent to legally marketed devices.

# Conclusion

Based on the indications for use, technological characteristics, and comparison to predicate devices, the subject device has been shown to be substantially equivalent to the aforementioned predicate devices cleared by FDA for commercial distribution in the United States.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K190881](https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K190881)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
