Navigation Enabled Instruments

K183381 · Stryker · OLO · May 17, 2019 · Neurology

Device Facts

Record IDK183381
Device NameNavigation Enabled Instruments
ApplicantStryker
Product CodeOLO · Neurology
Decision DateMay 17, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Stryker Navigation Enabled Instruments are intended to be used in the preparation and placement of Stryker screws (including XIA 3 (including Serrato), XIA 4.5 (including XIA CT), ES2, Mantis Redux, and XIA 2) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are specifically designed for use with the Medironic System, which is indicated for anv medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model. fluoroscopy images, or digitized landmarks of the anatomy. Stryker Navigation Enabled Screwdrivers are also compatible with the Medtronic IPC POWEREASE System.

Device Story

Navigation Enabled Instruments consist of manual taps and drivers used during spinal surgery; designed for integration with Medtronic StealthStation navigation system. Instruments track anatomical structures relative to CT/MR models, fluoroscopy, or digitized landmarks; assist surgeons in precise screw placement (XIA 3, XIA 4.5, ES2, Mantis, XIA 2). Drivers compatible with Medtronic IPC POWEREASE power system. Used in OR by surgeons to improve accuracy of instrumentation; provides visual guidance via navigation system interface; facilitates minimally invasive and open surgical workflows.

Clinical Evidence

Bench testing only. Design validation included registration and accuracy testing to confirm compatibility with Medtronic StealthStation and performance equivalence to predicate instruments.

Technological Characteristics

Nonsterile, reusable manual taps and drivers. Materials meet applicable biocompatibility standards. Designed for use with Medtronic StealthStation and IPC POWEREASE systems. Mechanical instrumentation; no integrated software or electronic sensing components.

Indications for Use

Indicated for patients undergoing spinal surgery requiring screw placement where stereotactic navigation is appropriate; compatible with open or minimally invasive procedures.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue. May 17, 2019 Stryker Megan Callanan Regulatory Affairs Associate 2 Pearl Court Allendale, New Jersey 07401 Re: K183381 Trade/Device Name: Navigation Enabled Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: May 3, 2019 Received: May 6, 2019 Dear Megan Callanan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for CAPT Raquel Peat, PhD, MPH, USPHS Director Office of Health Technology 6 Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K183381 Device Name Navigation Enabled Instruments ### Indications for Use (Describe) Stryker Navigation Enabled Instruments are intended to be used in the preparation and placement of Stryker screws (including XIA 3 (including Serrato), XIA 4.5 (including XIA CT), ES2, Mantis Redux, and XIA 2) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are specifically designed for use with the Medironic System, which is indicated for anv medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model. fluoroscopy images, or digitized landmarks of the anatomy. Stryker Navigation Enabled Screwdrivers are also compatible with the Medtronic IPC POWEREASE System. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. {3}------------------------------------------------ # 510(k) Summary | 510(k) Summary: Navigation Enabled | | |----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Stryker Spine<br>2 Pearl Court<br>Allendale, NJ 07401 | | Contact Person: | Name: Megan Callanan<br>Phone: (201)845-4900<br>Email: megan.callanan1@stryker.com | | Date Prepared: | 12/04/18 | | Trade Name: | Navigation Enabled Instruments | | Common Name: | Navigation Instruments | | Proposed Class: | Class II | | Classification Name: | Orthopedic / Orthopedic Stereotaxic Instruments | | Regulation Number: | 21 CFR 882.4560 | | Product Code: | OLO | | Predicate Devices: | Primary Predicate: Globus Navigation Instruments (K153203)<br>Additional Predicates: Medtronic Navigated Instruments (K143628,<br>K143375, K140454, K180690) | | Device Description: | Navigation Enabled Instruments are nonsterile, reusable instruments,<br>including taps which are operated manually and drivers that can be manually<br>operated or operated with the Medtronic IPC POWEREASE System. These<br>instruments are intended to be used with the Medtronic StealthStation<br>system. | | Indications for use: | Stryker Spine Navigation Enabled Instruments are intended to be used in the<br>preparation and placement of Stryker screws (including XIA 3 (including<br>Serrato), XIA 4.5 (including XIA CT), ES2, Mantis, Mantis Redux, and XIA 2)<br>during spinal surgery to assist the surgeon in precisely locating anatomical<br>structures in either open or minimally invasive procedures. These<br>instruments are specifically designed for use with the Medtronic<br>StealthStation® System, which is indicated for any medical condition in<br>which the use of stereotactic surgery may be appropriate, and where<br>reference to a rigid anatomical structure, such as a skull, a long bone, or<br>vertebra, can be identified relative to a CT or MR based model, fluoroscopy<br>images, or digitized landmarks of the anatomy. Stryker Spine Navigation<br>Enabled Screwdrivers are also compatible with the Medtronic IPC<br>POWEREASE System. | | Summary of the<br>Technological<br>Characteristics | Navigation Enabled Instruments have the same technological characteristics<br>as the predicate devices including design, material composition, intended<br>use, function, and range of sizes. Biocompatibility of patient-contacting | | 510(k) Summary: Navigation Enabled | | | | materials was demonstrated by using materials that meet applicable standards or are used in 510(k) cleared devices. | | Summary of the<br>Performance Data | Design validation testing, including registration, and accuracy, was conducted to ensure the Navigation Enabled Instruments are acceptable for their intended use, to ensure functionality and compatibility with the Medtronic StealthStation® using the NavLock tracker, and to demonstrate substantial equivalence to the predicate instruments. Registration testing was performed to ensure that the instruments can be registered to the StealthStation®.<br>Accuracy testing was completed for comparison to the predicate instruments. | | Conclusion | Stryker Spine Navigation Enabled Instruments are substantially equivalent to the noted predicate devices as they have the same intended use, technological characteristics, safety and efficacy profile, and performance specifications.<br>The content of this submission supports the determination of substantial equivalence. | {4}------------------------------------------------
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...