← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K173372

# ExactechGPS Total Shoulder Application (K173372)

_Blue Ortho · OLO · Nov 24, 2017 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K173372

## Device Facts

- **Applicant:** Blue Ortho
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** Nov 24, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The ExactechGPS is intended for use during preoperative planning and during stereotaxic surgery in locating anatomical structures and aligning the endoprosthesis with the anatomical structures provided that the required anatomical landmarks can be identified on the patient's preoperative CT scan. The ExactechGPS Total Shoulder Application is specifically indicated for Total Shoulder Arthroplasty using the Equinoxe system to aid the surgeon in locating anatomical structures and aligning the glenoid component with the anatomical structures.

## Device Story

Image-guided surgery/navigation system for total shoulder arthroplasty; assists surgeons in glenoid preparation and implant alignment. Inputs: patient CT-scan data. Processing: preoperative planning via virtual 3D bone model reconstruction; allows evaluation of implant size, type, and positioning. Modification: adds semi-automatic segmentation process for virtual bone reconstruction (planning only, not for navigation). Output: visual guidance for surgical alignment. Used in clinical settings by orthopedic surgeons. Benefits: improved accuracy in implant positioning and anatomical alignment.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, including software verification to ensure design outputs meet requirements, system accuracy verification comparing system outputs to CT-scan data, and software validation to ensure conformance to user needs.

## Technological Characteristics

Image-guided navigation system; preoperative planning software. Hardware platform unchanged. Software includes semi-automatic segmentation for 3D bone reconstruction. Connectivity: requires CT-scan data input. Sterilization: N/A (software-based).

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- ExactechGPS® Total Shoulder Application ([K162567](/device/K162567.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym followed by the words "U.S. Food & Drug Administration" on the right. The Department of Health and Human Services logo is a stylized human figure, while the FDA text is in blue.

November 24, 2017

Blue Ortho Anthony Boyer President and CEO 6 Allee de Bethleem Gieres, 38610 France

Re: K173372

Trade/Device Name: ExactechGPS Total Shoulder Application Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, LLZ Dated: October 26, 2017 Received: October 27, 2017

Dear Anthony Boyer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K173372

Device Name

ExactechGPS® Total Shoulder Application

Indications for Use (Describe)

The ExactechGPS is intended for use during preoperative planning and during stereotaxic surgeon in locating anatomical structures and aligning the endoprosthesis with the anatomical structures provided that the required anatomical landmarks can be identified on the patient's preoperative CT scan.

The ExactechGPS Total Shoulder Application is specifically indicated for Total Shoulder Arthroplasty using the Equinoxe system to aid the surgeon in locating anatomical structures and aligning the glenoid component with the anatomical structures.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for BlueOrtho, a company specializing in augmented orthopedics. The logo features a stylized blue "C" shape on the left, followed by the text "BlueOrtho" in a similar blue color. Below "BlueOrtho" is the text "AUGMENTED ORTHOPAEDICS" in a smaller font size.

6 Allée de Bethléem 38610 Gières France

# ExactechGPS® Total Shoulder Application SPECIAL 510(k) - 510(k) Summary of Safety and Effectiveness

# Submission date:

October 26, 2017

# Sponsor:

BLUE ORTHO 6 Allée de Bethléem 38610 Gières France Phone: +33 (0) 4 58 00 35 25

# Contact:

Anthony BOYER Phone: +33 (0) 4 76 54 95 46 Cell: +33 (0)6 30 39 50 66 anthony.boyer@blue-ortho.com

# US Local Agent

Exactech, Inc. 2320 NW 66th Ct. Gainesville, FL. 32653 Phone: 352-377-1140

# Trade or Proprietary or Model Name(s):

ExactechGPS® Total Shoulder Application

# Common Name:

Surgical navigation system and preoperative planning tool

# Classification Number 21 CFR 882.4560

# Classification Name: Orthopedic Stereotaxic Instrument, Picture Archiving & Communications System

# Classification Class II

# Product Code OLO (21 CFR 882.4560), LLZ (21CFR 892.2050)

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for BlueOrtho, a company specializing in augmented orthopedics. The logo features a stylized blue "C" shape on the left, followed by the text "BlueOrtho" in a clean, sans-serif font. Below "BlueOrtho" is the tagline "AUGMENTED ORTHOPAEDICS" in smaller, lighter font.

6 Allée de Bethléem 38610 Gières France

# Information on Devices to which Substantial Equivalence is Claimed:

#### Primary Predicate

| 510(k) Number | Trade or Proprietary Model Name         | Manufacturer |
|---------------|-----------------------------------------|--------------|
| #K162567      | ExactechGPS® Total Shoulder Application | Blue Ortho   |

# Indications for Use:

The ExactechGPS is intended for use during preoperative planning and during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprosthesis with the anatomical structures provided that the required anatomical landmarks can be identified on the patient's preoperative CT scan.

The ExactechGPS Total Shoulder Application is specifically indicated for Total Shoulder Arthroplasty using the Equinoxe system to aid the surgeon in locating anatomical structures and aligning the glenoid component with the anatomical structures.

# Device Description:

This submission proposes modifications to the ExactechGPS Total Shoulder Application cleared per #K162567.

The ExactechGPS Total Shoulder Application is an Image Guided Surgery, or Navigation, system designed to guide surgeons during the preparation of the glenoid as part of a total shoulder arthroplasty procedure. The ExactechGPS Total Shoulder Application also offers a preoperative planning feature that enables surgeons to plan a surgical intervention by evaluating implant size, type, and positioning using reconstructed patient bone models in a virtual environment. The ExactechGPS Total Shoulder Application requires patient CT-scan data to undergo segmentation prior to being imported into the software, as part of reconstructing the bone model, for both navigation and planning.

At initial release, the segmentation step was manually conducted by Blue Ortho. This submission proposes the addition of a semi-automatic segmentation option, where surgeons can elect to not send the CT exams to Blue Ortho for segmentation but employ a semiautomatic segmentation process to virtually reconstruct the bony anatomy of the patient in three dimensions. The use of semi-automatic segmentation to virtually recreate patient bone is available for planning only, and not for use during a navigated surgical procedure.

# Testing:

This submission includes or references the following non-clinical testing:

- Software verification testing to ensure all design outputs meet all specified requirements
- System accuracy verification via comparison of system outputs to CT-scan data collected during clinical use
- Software validation to ensure software specifications conform to user needs and intended uses

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for BlueOrtho, a company specializing in augmented orthopedics. The logo features a stylized letter "C" in blue, followed by the company name "BlueOrtho" in a similar blue color. Below the company name, the words "AUGMENTED ORTHOPAEDICS" are written in smaller font, indicating the company's focus. The logo is clean and professional, conveying a sense of innovation and expertise in the field of orthopedics.

Testing demonstrates the modified ExactechGPS Total Shoulder Application is substantially equivalent to the cited predicate.

# Summary of Technological Characteristics

The rationale for substantial equivalence is based on consideration of the following characteristics:

- Intended use: the modifications do not affect device intended use. -
- -General design features and dimensions: the modifications do not affect general device features and dimensions.
- -Basic fundamental scientific technology: the modifications do not change the device computer language or other basic fundamental technologies.
- -Performance specifications: the modifications do not affect device accuracy and / or performance.
- -Hardware platform: no changes to the hardware platform or system accessories are proposed by this submission.

# Substantial Equivalence Conclusion:

A comparison of specific features included in this submission demonstrates the proposed ExactechGPS Total Shoulder Application is substantially equivalent to the cited predicate. The devices share:

- Identical Indications for Use statements
- Identical intended use
- Identical technology for stereotaxic surgery
- Similar, equivalent technology for preoperative planning

The information provided by Blue Ortho in this application confirms the proposed ExactechGPS Total Shoulder Application is substantially equivalent to the predicate cleared per #K162567.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K173372](https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K173372)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
